510(k) Data Aggregation

    K Number
    K073246

    Validate with FDA (Live)

    Device Name
    ZENIEVA
    Date Cleared
    2008-07-23

    (247 days)

    Product Code
    Regulation Number
    878.4022
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Zenieva is used to manage and relieve the burning and itching experienced with various types of dermatoses, including radiation dermatitis, atopic dermatitis, and allergic contact dermatitis. Zenieva helps relieve dry waxy skin by maintaining a moist wound and skin environment, which is beneficial to the healing process.

    Device Description

    Zenieva is a non-sterile, semi-viscous emulsion intended for topical application. It is presented for prescription (requires a physician diagnosis of disease state) use. The product is formulated as an oil-in-water emulsion containing a crosslinked polyacrylic acid polymer, natural gum, and cellulose as thickening agents. The oil composition of Zenieva is composed of glyceride, squalane, lecithin, and fatty acids.

    AI/ML Overview

    This submission describes Zenieva, a hydrogel wound dressing, and its substantial equivalence to the predicate device, MimyXTM cream (K041342). The provided text focuses on the device's composition, intended use, and non-clinical performance data, as well as the FDA's clearance letter.

    Device Acceptance Criteria and Performance

    The acceptance criteria for Zenieva are based on demonstrating substantial equivalence to the predicate device, MimyXTM cream. The study presented is a non-clinical performance evaluation, which aims to show that Zenieva performs similarly to the predicate and meets safety standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance (Zenieva)
    FormulationIdentical chemical ingredients as predicate (MimyX™)Zenieva's ingredients are identical to MimyX™. (Ingredients provided in the table: Purified water, olive oil, glycerin, pentylene glycol, palm glycerides, vegetable oil, hydrogenated lecithin, squalane, betaine, palmitamide MEA, sarcosine, acetamide MEA, hydroxyethyl cellulose, sodium carbomer, carbomer, xanthan gum)
    Intended UseIdentical intended use as predicateZenieva's intended use is identical to MimyX™: to manage and relieve burning and itching from various dermatoses (radiation, atopic, allergic contact dermatitis) and maintain a moist wound/skin environment for healing. (Claim identical to predicate)
    Physical PropertiesIdentical physical properties as predicateZenieva's physical properties are identical to MimyX™: Non-sterile white to off-white thick cream; Water-based emulsion.
    Application InstructionsIdentical application frequency as predicate3 times per day or as needed, identical to MimyX™.
    In Vitro CytotoxicityNon-toxicZenieva was shown to be non-toxic when tested as a "surface device" with limited contact (24 hours) on breached surfaces.
    Skin SensitizationNon-sensitizerZenieva was shown to be a non-sensitizer.
    Primary Skin IrritationNon-irritantZenieva was shown to be a non-irritant.
    Acute Systemic ToxicityNot applicable (topical use, not injection)Test was performed but deemed not applicable, as the product is for topical application.
    Microbial LimitsAcceptable bioburden levels throughout shelf-life (non-sterile)Microbial limit testing was performed to ensure acceptable bioburden levels throughout the shelf-life, as the product is non-sterile. (No specific numerical results provided, but stated as a conducted and met measure).

    2. Sample Size for the Test Set and Data Provenance

    The provided summary does not detail the sample sizes for the in vitro cytotoxicity, skin sensitization, and primary skin irritation tests. It states that Zenieva was tested as a "surface device" with "limited contact duration (24 hours) on breached surfaces." The data provenance is not explicitly stated as retrospective or prospective, nor does it mention the country of origin, but it is implied to be from testing conducted by Gorbec Pharmaceutical Services Inc. to support this 510(k) submission.

    3. Number of Experts and Qualifications for Ground Truth

    This submission does not involve clinical studies with human assessors establishing ground truth for performance metrics in a typical sense (e.g., image interpretation). Instead, the "ground truth" for the non-clinical tests (cytotoxicity, sensitization, irritation) would be established by the results of standardized laboratory assays, interpreted by qualified laboratory personnel and toxicologists. The specific number and qualifications of experts for interpreting these non-clinical tests are not detailed in the provided document.

    4. Adjudication Method for the Test Set

    Not applicable. The non-clinical tests (cytotoxicity, sensitization, irritation) are typically objective laboratory assays with predefined endpoints, rather than requiring expert consensus or adjudication on subjective interpretations for a "test set" in the context of diagnostic or treatment efficacy.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This document describes a 510(k) submission for a hydrogel wound dressing, not an AI-assisted diagnostic or treatment device that would typically involve a MRMC study. Therefore, no effect size for human readers with and without AI assistance is reported.

    6. Standalone Performance Study

    Yes, in a sense. The non-clinical performance tests (cytotoxicity, skin sensitization, primary skin irritation) assess the intrinsic biological safety of the device (Zenieva) in isolation, without human-in-the-loop performance influencing the assay results. The device's "performance" in this context is its safety profile.

    7. Type of Ground Truth Used

    The ground truth used for these non-clinical tests is based on standardized laboratory assay results and regulatory definitions for toxicity, sensitization, and irritation. For example, a "non-toxic" finding is based on the cellular response in a cytotoxicity assay meeting predefined criteria, rather than expert consensus on a clinical outcome or pathology.

    8. Sample Size for the Training Set

    Not applicable. This device is a medical product (hydrogel wound dressing) and not an algorithm or AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of medical device submission.