510(k) Data Aggregation

    K Number
    K211466

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-06-14

    (399 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Yomi is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. Yomi is intended for use in partially edentulous adult patients who qualify for dental implants.

    When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of Yomi Plan is to be used with the Yomi Robotic System.

    Device Description

    The subject of this submission is YomiPlan Go, a feature of the Yomi Robotic System which enables the use of the system without the uploading of a preoperative CT scan. The dynamic planning feature in K202264 requires a pre-operative CT scan for use. This submission includes a new workflow called YomiPlan Go and provides instructions on how to use this feature without the need of uploading a CT scan to the Yomi system. This planning involves placing the robotic arm drill tip to the point where an osteotomy is to be performed i.e., the surgeon performs planning with their direct visualization of the anatomy and with the use of the robotic arm to select where the osteotomy is to be performed. YomiPlan Go gives surgeons the ability to perform an osteotomy under robotic guidance at the point that the surgeon selects on the patient's anatomy. The selected point, axis, and trajectory are maintained by the robotic arm while the surgeon performs the osteotomy.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria / Performance MetricReported Device Performance (YomiPlan Go)Predicate Device Performance (NGS with YomiPlan v2.0)Freehand Surgery Performance (Varga, et al, 2020)
    PrecisionUpper 95% Probability with 95% Model Fit < 1.00 mmUpper 95% Probability with 95% Model Fit < 1.00 mmNot specified directly, but YomiPlan Go offers improved precision over freehand.
    AccuracyN/A (not a specification or advantage for YomiPlan Go)Upper 95% Probability with 95% Model Fit < 1.00 mmNot a formal metric, but indirect comparisons are made.
    Depth Error (mm)1.38 +/- 0.71Not directly provided for comparison in summary tableNot specified in this document for direct comparison
    Lateral Error (mm)2.04 +/- 1.59Not directly provided for comparison in summary tableNot specified in this document for direct comparison
    Angular Error (degs)12.91 +/- 7.563.3° (from conclusion for predicate performance)7.03° (average), 0.7–21.3° (range)
    Human Factors (user needs)Met design and performance requirements (qualitative survey)Not applicable as this is a new featureNot applicable
    SafetyNo vital anatomical structure concerns (3rd party clinical review)Not applicable as this is a new featureConsidered acceptable for healthy patients with adequate bone
    No Serious Adverse EventsConfirmed in clinical study (except for one minor event related to user error)Not applicable as this is a new featureNot applicable

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 44 implants placed in 15 patient arches.
    • Data Provenance: Prospective, multi-center, single-arm study (G210363) involving partially or fully edentulous patients. The study was conducted clinically with actual patients. The specific country of origin is not explicitly stated, but the submission is to the US FDA, implying it would be relevant to US practice.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Number of Experts:
      • 3 investigators (general dentists) who performed the procedures.
      • 1 independent 3rd party oral surgeon who performed an independent clinical assessment of all 15 cases.
    • Qualifications of Experts:
      • Investigators: General dentists.
      • Independent Assessor: 3rd party oral surgeon. Specific years of experience are not mentioned for any of the experts.

    4. Adjudication Method for the Test Set:

    • The text describes case-by-case feedback from investigators (through CRF feedback) and an independent 3rd party clinical review of all 15 cases. This suggests a form of independent review rather than a strict 2+1 or 3+1 consensus process among multiple readers for ground truth establishment. Agreement between these groups (investigators' safety feedback, 3rd party review), especially regarding "no vital anatomical structure concerns," served as a form of "ground truth" for clinical safety and acceptability.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    • No, an MRMC comparative effectiveness study was not explicitly stated to have been done in the traditional sense of evaluating human readers with and without AI assistance.
    • The study primarily focused on the standalone performance of "YomiPlan Go" relative to freehand techniques and comparison metrics with the predicate device. The "human factors" evaluation was for the new YomiPlan Go workflow itself by the operating surgeons, not a comparison of human interpretation/performance with and without AI.
    • The benefits assessment does highlight how YomiPlan Go assists surgeons (e.g., depth stop, prevention of skiving, parallelism, precision in repeated drilling), which could be interpreted as improving human performance, but this wasn't quantified through an MRMC study design measuring specific effect sizes of human improvement.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    • Yes, in part, a standalone assessment of the algorithmic component's performance (precision, depth, and lateral error) was conducted as a secondary objective of the clinical study, comparing pre-operative plans (ideal locations) to post-operative implant placement.
    • The system's precision specification (Upper 95% Probability with 95% Model Fit < 1.00 mm) is also mentioned as supported by benchtop verification testing, which would typically be a standalone performance test. However, the accuracy is explicitly stated as "N/A for YomiPlan Go" as it's not a specification for this specific mode, which relies on the surgeon's real-time visual assessment rather than a pre-loaded CT.

    7. The Type of Ground Truth Used:

    • The ground truth for the performance metrics (depth, lateral, and angular error) was established by comparing preoperative clinical plans (ideal planned implant locations on CBCT scans) with postoperative CBCT scans capturing the actual placement of implants.
    • For clinical safety and acceptability, the ground truth was based on the assessment and feedback from the investigators and an independent 3rd party oral surgeon, specifically their confirmation of "no vital anatomical structure concerns."

    8. The Sample Size for the Training Set:

    • The document does not provide information on the sample size for the training set for the Yomi or YomiPlan Go system. The focus of this submission is on the clinical validation of a new feature (YomiPlan Go) for a previously cleared device.

    9. How the Ground Truth for the Training Set Was Established:

    • The document does not provide information on how the ground truth for the training set was established. Since the details of the training set size or methodology are not given, the ground truth establishment method for it is also absent.