Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111584

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-08-25

    (79 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Xylos® Macro-Porous Surgical Mesh is intended to be used for implantation to reinforce and protect soft tissue including but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, hernias, suture-line reinforcement and other reconstructive procedures. This mesh may be used for the repair of hernia and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

    Xylos® Macro-Porous Surgical Mesh is also intended for the reinforcement of soft tissues repaired by sutures or suture anchors during tendon repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps or other tendons. Xylos® Macro-Porous Surgical Mesh is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear, and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair.

    Xylos® Macro-Porous Surgical Mesh is indicated for open and endoscopic procedures and is intended for one-time use only.

    Device Description

    Xylos® Macro-Porous Surgical Mesh is a flexible, non-resorbable implant composed of microbial-derived cellulose; there are no animal or human derived components in the device. Xylos® Macro-Porous Surgical Mesh is strong, conformable, biocompatible and includes macro-pores to facilitate tissue in-growth. The device is not subject to degradation or weakening by the action of tissue enzymes.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

    Disclaimer: This device is a surgical mesh, not an AI-powered diagnostic tool. Therefore, many of the typical questions related to AI device evaluation (such as multi-reader multi-case studies, effect size of AI assistance, standalone performance, and ground truth establishment for training/test sets in an AI context) are not applicable to this submission. The information provided focuses on the regulatory submission for a physical medical device.


    Acceptance Criteria and Device Performance for Xylos® Macro-Porous Surgical Mesh

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes performance testing for mechanical properties and biocompatibility. The acceptance criteria are implicit in the claim of "substantial equivalence" to predicate devices, meaning the Xylos® Macro-Porous Surgical Mesh must perform at least as well as, or equivalently to, the predicate devices in the measured physical properties and biocompatibility.

    Acceptance Criterion (Implicit)Reported Device Performance and Evidence
    Mechanical Equivalence to specified predicate devices"Xylos® Macro-Porous Surgical Mesh was subjected to mechanical performance tests typical for surgical mesh products (tensile strength, burst strength, and suture pull out strength). This testing demonstrated that Xylos® Macro-Porous Surgical Mesh is mechanically equivalent to the predicate devices: Pegasus' OrthADAPT Bioimplant (K071065), Ethicon's Mersilene Mesh (Pre-amendment), and Covidien's Parietex Composite Mesh (K040998)."
    Sufficient Physical Strength for Intended Use"Further, animal study results show that Xylos® Macro-Porous Surgical Mesh possesses sufficient physical strength for the intended use."
    Biocompatibility Equivalence to specified predicate device"Xylos® Macro-Porous Surgical Mesh is substantially equivalent in biocompatibility to Xylos® Securian® Tissue Reinforcement Matrix (K083823) because they have the identical material composition." "The material has demonstrated its biocompatibility via testing as defined in ANSI/AAMI/ISO 10993 Biological evaluation of medical devices."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Mechanical Testing: The document does not specify exact sample sizes for the mechanical tests (tensile strength, burst strength, suture pull-out strength). It refers to "mechanical performance tests typical for surgical mesh products."
    • Animal Study: The document does not specify the sample size for the animal study.
    • Biocompatibility Testing: The document does not specify the sample size for biocompatibility testing, but indicates it was conducted "as defined in ANSI/AAMI/ISO 10993 Biological evaluation of medical devices."
    • Data Provenance: The origin of the animals or the specific details of the testing environment (country, etc.) are not provided. The studies appear to be prospective tests conducted specifically for this device's submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This is not applicable as the submission is for a physical surgical mesh, not an AI device requiring expert interpretation of output. The "ground truth" for this device's performance is established through direct physical and biological testing against established standards and predicate devices.

    4. Adjudication Method for the Test Set:

    Not applicable for a physical device. Performance is measured objectively through standardized tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    This is not applicable. This is not an AI device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This is not applicable. This is not an AI device.

    7. The Type of Ground Truth Used:

    • Mechanical Performance: The ground truth is objective measurement against established engineering standards and comparative data from the identified predicate devices.
    • Biocompatibility: The ground truth is established through adherence to international standards for biological evaluation of medical devices (ANSI/AAMI/ISO 10993) and comparison to a predicate device with identical material composition that was previously cleared for biocompatibility.
    • Physical Strength for Intended Use: The ground truth is derived from the results of an animal study.

    8. The Sample Size for the Training Set:

    This is not applicable. This is not an AI device that requires a training set in the machine learning sense. The "training" for this device involved its design and manufacturing processes based on knowledge of similar products and materials.

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable. As stated above, this is not an AI device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1