Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K112199

    Validate with FDA (Live)

    Date Cleared
    2011-09-01

    (31 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The xTAG Respiratory Viral Panel (RVP) is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of multiple respiratory virus nucleic acids in nasopharyngeal swabs from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using RVP: Influenza A subtype H1, Influenza A subtype H3, Influenza B, Respiratory Syncytial Virus subtype A, Respiratory Syncytial Virus subtype B, Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus, Human Metapneumovirus, Rhinovirus, and Adenovirus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection if used in conjunction with other clinical and laboratory findings. It is recommended that specimens found to be negative for Influenza B, Respiratory Syncytial Virus subtype A and B, Parainfluenza 1, Parainfluenza 2, Parainfluenza 3 and Adenovirus, after examination using RVP be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

    Positive results do not rule out bacterial infection, or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection.

    Due to seasonal prevalence, performance characteristics for Influenza A/H1 were established primarily with retrospective specimens.

    The RVP assay cannot adequately detect Adenovirus species C, or serotypes 7a and 41. The RVP primers for detection of rhinovirus cross-react with enterovirus. A rhinovirus reactive result should be confirmed by an alternate method (e.g. cell culture).

    Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infections with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter for the xTAG® Respiratory Viral Panel (RVP), not a study report. Therefore, it does not contain the detailed acceptance criteria and study information typically found in a clinical study report.

    The provided text only includes the "Indications for Use" for the device, which describes its intended purpose and some performance considerations, but not the specific acceptance criteria or the study that demonstrates the device meets those criteria.

    Therefore, I cannot fulfill the request to provide:

    1. A table of acceptance criteria and the reported device performance
    2. Sample size used for the test set and the data provenance
    3. Number of experts used to establish the ground truth for the test set and their qualifications
    4. Adjudication method
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or its effect size
    6. If a standalone performance study was done
    7. The type of ground truth used
    8. The sample size for the training set
    9. How the ground truth for the training set was established

    The document explicitly states: "Performance characteristics for Influenza A/H1 were established primarily with retrospective specimens." and "Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary." This indicates that studies were indeed performed, but the details of those studies (acceptance criteria, sample sizes, ground truth establishment, etc.) are not present within the given text. This information would typically be found in the 510(k) submission itself, which is not provided here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103776

    Validate with FDA (Live)

    Date Cleared
    2011-07-01

    (186 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The xTAG® Respiratory Viral Panel Fast (RVP FAST) is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of multiple respiratory virus nucleic acids in nasopharyngeal swabs from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using RVP FAST: Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza B, Respiratory Syncytial Virus, Human Metapneumovirus, Rhinovirus, and Adenovirus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection if used in conjunction with other clinical and epidemiological information.

    Negative results do not preclude respiratory viral infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Positive results do not rule out bacterial infection or co-infection with other organisms. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory infection.

    Due to the genetic similarity between human Rhinovirus and Enterovirus, the RVP FAST primers for the detection of rhinovirus cross react with enterovirus. A rhinovirus reactive result should be confirmed by an alternate method (e.g. cell culture).

    Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

    Device Description

    RVP FAST is a PCR-based system for detecting the presence / absence of viral DNA / RNA in clinical specimens. The oligonucleotide primer / probe components of the RVP FAST have been designed to specifically target unique regions in the RNA / DNA of each molecular species listed in the intended use. Amplified products are then sorted and analyzed on the Luminex 100 or 200 instrument, which generates signals based on the acquisition of spectrofluorometric data. The raw signals are median fluorescence intensities (MFI) which are acquired in a Luminex Output.csv file that is subsequently analyzed by the xTAG Data Analysis Software (TDAS RVP FAST) to establish the presence of all viral types / subtypes for which a Luminex microsphere population has been dedicated.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the xTAG® RVP FAST device, based on the provided document:


    Acceptance Criteria and Device Performance

    The provided document details the performance characteristics through clinical sensitivity and specificity. While explicit quantitative "acceptance criteria" tables are not presented as separate targets, the FDA's clearance implies these performance metrics were deemed acceptable for the device's intended use. The table below summarizes the reported clinical performance.

    Virus (Analyte)Reported Sensitivity (%)Lower-Bound 95% CI for Sensitivity (%)Reported Specificity (%)Lower-Bound 95% CI for Specificity (%)
    Human Influenza A94.288.895.594.0
    H196.387.398.297.2
    H396.189.097.896.8
    Human Influenza B94.384.398.898.0
    RSV90.984.397.796.6
    Rhinovirus95.684.992.590.8
    Adenovirus85.757.299.398.6
    Metapneumovirus (hMPV)97.285.598.998.2
    Adenovirus (Supplemented)97.06 (PPA)86.47N/AN/A

    Study Details

    1. Sample Size used for the test set and the data provenance:

      • Sample Size: 1191 prospectively collected clinical specimens (nasopharyngeal swabs).
      • Data Provenance: Specimens were collected during the 2007/2008 and 2008/2009 flu seasons from enrolled clinical sites. They were a mix of "left-over, fresh and frozen." An additional 34 banked, pre-selected positive clinical specimens for Adenovirus were used to supplement the prospective set. Furthermore, 77 clinical specimens confirmed positive for Novel 2009/H1N1 (swine flu) were also tested.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number or qualifications of experts directly. Ground truth was established using comparator methods such as Direct Fluorescent Antibody Test (DFA), viral culture, and well-characterized RT-PCR amplification followed by bidirectional sequencing. These methods are laboratory-based and generally performed by trained laboratory personnel, rather than explicitly "experts" in the context of clinical interpretation.
    3. Adjudication method for the test set:

      • The document does not describe an adjudication method in the sense of multiple human readers resolving discrepancies. The ground truth was established by laboratory-based comparator methods.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a qualitative nucleic acid multiplex test for direct detection of viral RNA/DNA, not an imaging or diagnostic support system for human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance characteristics (sensitivity and specificity) presented in Table 10 represent the standalone performance of the xTAG RVP FAST assay itself. The assay utilizes multiplex RT-PCR and Luminex technology, with data analyzed by the xTAG Data Analysis Software (TDAS RVP FAST) to provide a qualitative summary. There is no human interpretation of image data or similar involved in the reported performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Laboratory-based Comparator Methods:
        • For Influenza B, RSV, and Adenovirus: Direct Fluorescent Antibody Test (DFA) and/or viral culture.
        • For Influenza A subtyping, hMPV, and Rhinovirus: Well-characterized RT-PCR amplification followed by bidirectional sequencing.
        • For the additional Adenovirus samples: Culture.
        • For the Novel 2009/H1N1 samples: CDC real-time PCR test.
    7. The sample size for the training set:

      • The document does not explicitly state a "training set" in the context of machine learning or AI models. This device is a molecular diagnostic assay where primers and probes are designed based on known viral genetic sequences. The development and optimization process for such assays typically involves laboratory experimentation with known positive and negative controls and clinical samples, rather than a distinct "training set" in the AI sense for which performance metrics against a held-out test set would be reported for FDA submission.
    8. How the ground truth for the training set was established:

      • As there isn't a stated "training set" for a machine learning model, the concept of establishing ground truth for it is not directly applicable in the terms usually associated with AI/ML device submissions. The assay itself is designed based on known biological targets, and its analytical and clinical performance is validated against established laboratory reference methods and clinical outcomes.
    Ask a Question

    Ask a specific question about this device

    K Number
    K091667

    Validate with FDA (Live)

    Date Cleared
    2009-06-25

    (16 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The xTAG Respiratory Viral Panel (RVP) is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of multiple respiratory virus nucleic acids in nasopharyngeal swabs from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using RVP: Influenza A. Influenza A subtype H1, Influenza A subtype H3, Influchza B, Respiratory Syncytial Virus subtype A, Respiratory Syncytial Virus subtype B, Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus. Human Metapneumovirus, Rhinovirus, and Adenovirus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection if used in conjunction with other clinical and laboratory findings. It is recommended that specimens found to be negative for Influenza B, Respiratory Synctial Virus subtype A and B, Parainfluenza 2, Parainfluenza 3 and Adenovirus, after examination using RVP be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Positive results do not rule out bacterial infection, or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection.

    Due to seasonal prevalence, performance characteristics for Influenza A/H1 were established primarily with retrospective specimens.

    The RVP assay cannot adequately detect Adenovirus species C, or serotypes 7a and 41. The RVP primers for detection of thinovirus cross-react with enterovirus. A rhinovirus reactive result should be confirmed by an alternate method (e.g. cell culture).

    Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infections with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

    Device Description

    The xTAG Respiratory Viral Panel includes the following components:

    • Multiplex PCR primer mix (without dNTPs) .
    • Multiplex target specific primer extension (TSPE) primers (includes dNTPs) .
    • Coupled bead mix .
    • 10x buffer
    • xTAG® Data Analysis Software (TDAS RVP-I) .
    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the xTAG® RVP device, based on the provided 510(k) summary:

    The 510(k) summary provided primarily focuses on establishing substantial equivalence to a predicate device and specifically details the device's ability to detect novel 2009 Influenza A/H1N1 strains, rather than establishing acceptance criteria against a predefined performance target in a typical clinical trial format.

    Key takeaway: The document highlights the device's performance in identifying the 2009 Influenza A/H1N1 strain, which was a new and important consideration at the time of this 510(k) submission. It describes how the device handles this specific strain compared to a reference method (CDC rRT-PCR assay).


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state quantitative acceptance criteria (e.g., minimum sensitivity, specificity thresholds) for the device's overall performance as would be expected in a typical clinical study report for initial market clearance. Instead, it demonstrates the device's capability to detect a newly emerging influenza strain. The "acceptance" can be inferred as the ability to effectively detect the 2009 Influenza A/H1N1 strain, often by agreement with a reference method.

    Acceptance Criteria (Implied for 2009 Influenza A/H1N1 Detection)Reported Device Performance (xTAG® RVP)
    Detection of 2009 Influenza A/H1N1 (matrix gene)Study 1 (Ginocchio & George, 2009): - Out of 141 "unsubtypeable" Flu A samples by xTAG RVP, 101 were further tested with CDC rRT-PCR. - 99 out of 101 (98.0%) identified as positive for 2009 Influenza A/H1N1 by CDC rRT-PCR (CT<37). - 2 remaining samples produced weak positive signals (CT>37) by CDC rRT-PCR and by xTAG RVP, and could not be classified as 2009 Influenza A/H1N1 positive. Study 2 (Ginocchio et al., 2009): - Out of 1265 samples positive for Influenza A by xTAG RVP, 1108 were "flu A unsubtypeable". - All 1108 (100%) "flu A unsubtypeable" samples were confirmed to be Influenza A/H1N1 with the CDC rRT-PCR assay.
    Identification of other respiratory virusesThe device identified Influenza B (2 samples) and other respiratory viruses including adenovirus, metapneumovirus, Parainfluenza 1, 2, 3, RSV, and rhinovirus (58 samples) in the Ginocchio & George (2009) study, alongside the Influenza A positives.
    Subtyping of seasonal Influenza A strainsIn the Ginocchio & George (2009) study, 60 Flu A positive samples were identified by xTAG RVP as seasonal strains (2 as H1 and 58 as H3).

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Study 1 (Ginocchio & George, 2009):
      • Initial samples tested with various methods: 1,382 nasopharyngeal swab samples.
      • Samples further tested with xTAG RVP: 375 samples (positive for Influenza A by other methods or from high-risk patients).
      • Samples identified as "unsubtypeable" Flu A by xTAG RVP: 141.
      • Test Set for 2009 H1N1 confirmation: 101 frozen residual samples out of the 141 unsubtypeable samples.
      • Data Provenance: Retrospective specimens from the 2009 Influenza A H1N1 (swine flu) outbreak in New York.
    • Study 2 (Ginocchio et al., 2009):
      • Test Set: 2,715 nasopharyngeal swab samples.
      • Data Provenance: Not explicitly stated, but context suggests it's related to 2009 Influenza A/H1N1 surveillance, likely also retrospective and from the outbreak regions.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The studies described use a highly sensitive and specific laboratory reference method, not expert human readers, to establish ground truth.

    • Ground truth was established by the CDC rRT-PCR assay for 2009 Influenza A H1N1 (swine flu). This is a highly specialized molecular diagnostic test, and the "experts" would be the trained laboratory personnel performing and interpreting this assay. Specific qualifications beyond performing the reference assay are not detailed.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was established by a laboratory reference method (CDC rRT-PCR assay), not through human reader adjudication.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The device is a diagnostic assay, and its performance is typically evaluated against laboratory reference standards or clinical outcomes, not against variations in human reader interpretation of images or signals.


    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    Yes, the studies describe the standalone performance of the xTAG RVP assay. While the assay requires laboratory personnel to run the test and interpret results from the xTAG® Data Analysis Software (TDAS RVP-I), the performance metrics reported (e.g., number of positives concordant with CDC rRT-PCR) reflect the assay's direct detection capabilities, independent of human clinical interpretation of symptoms or subsequent reader adjustments.


    7. Type of Ground Truth Used

    The primary ground truth used for confirming the detection of 2009 Influenza A/H1N1 was a laboratory reference method: the CDC rRT-PCR assay for 2009 Influenza A H1N1. This is a highly sensitive and specific molecular test, considered the gold standard for identifying this specific viral strain at the time.

    For other viruses, the general intended use suggests confirmation by cell culture for negative results of certain viruses, but the performance studies specifically for 2009 H1N1 used the CDC PCR assay as ground truth.


    8. Sample Size for the Training Set

    The document does not specify a separate training set or its sample size. The studies described are performance evaluations on clinical samples, likely acting as a test set for the already developed assay. Diagnostic molecular assays typically establish their core design and parameters through internal development and validation, and these clinical studies serve to demonstrate performance on real-world samples.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a distinct training set with its own ground truth establishment is not detailed in the provided summary. The clinical samples evaluated in the studies served as performance verification, using the CDC rRT-PCR as the ground truth comparator.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081483

    Validate with FDA (Live)

    Date Cleared
    2008-06-25

    (28 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The xTAG™ Respiratory Viral Panel (RVP) is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of multiple respiratory virus nucleic acids in nasopharyngeal swabs from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using RVP: Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza B, Respiratory Syncytial Virus subtype A, Respiratory Syncytial Virus subtype B, Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus. Human Metapneumovirus, Rhinovirus, and Adenovirus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection if used in conjunction with other clinical and laboratory findings. It is recommended that specimens found to be negative for Influenza B, Respiratory Syncytial Virus subtype A and B, Parainfluenza 2, Parainfluenza 3 and Adenovirus, after examination using RVP be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Positive results do not rule out bacterial infection, or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection.

    Device Description

    The xTAGTM Respiratory Viral Panel includes the following components:

    • Multiplex PCR primer mix (without dNTPs) .
    • Multiplex target specific primer extension (TSPE) primers (includes dNTPs) .
    • Coupled bead mix ●
    • 10x buffer .
    • xTAG™ Data Analysis Software (TDAS RVP-I) .
    AI/ML Overview

    The provided document does not contain any information about acceptance criteria or a study proving the device meets those criteria, nor does it include reported device performance.

    The document is a 510(k) summary for the xTAG™ Respiratory Viral Panel, indicating it's a re-submission or a modification of a previously cleared device (K063765). The key sections state:

    • "The xTAG™ Respiratory Viral Panel performance parameters remain unchanged."
    • "Summary of Performance Data: Not applicable."

    This strongly suggests that a new performance study was NOT conducted for this specific 510(k) submission (K081483) because the performance characteristics were deemed to be the same as the predicate device (K063765). Therefore, the information requested in your prompt is not available in the provided text.

    To answer your request, one would need to refer to the original 510(k) submission for the xTAG™ Respiratory Viral Panel (K063765) where the initial performance data would have been presented.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1