510(k) Data Aggregation

    K Number
    K210322

    Validate with FDA (Live)

    Date Cleared
    2021-10-21

    (259 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XO Cross is a support catheter intended to guidewire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.

    Device Description

    The XO Cross Support Catheter is identical to the predicate device in that it is a single-lumen support catheter designed to support a guidewire during access of the peripheral vasculature, allowing for exchange of guidewires during vascular access procedures, and providing a conduit for the delivery of saline solutions or diagnostic contrast agents. The shaft and tip are both radiopaque to track the location of the XO Cross Support Catheter within the vasculature, while under fluoroscopy. The distal end of the subject catheter is also coated with a hydrophilic coating to reduce surface friction. It is available in various lengths (90cm, 135cm, 15cm) and ranges in diameter from 2 Fr (0.7 mm) to 4 Fr (1.35 mm) with a tapered polymer tip at its distal end.

    AI/ML Overview

    The provided text describes the submission for a 510(k) premarket notification for the XO Cross Support Catheter, which is a revised version of a previously cleared device. The document primarily focuses on demonstrating substantial equivalence to the predicate device rather than providing a detailed study of the device's performance against specific acceptance criteria in a clinical setting.

    Here's an analysis of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a table of specific acceptance criteria with corresponding performance statistics for the device in terms of clinical outcomes or diagnostic accuracy. Instead, it details that "Verification & Validation activities were performed on the subject XO Cross Support Catheter to demonstrate substantial equivalence to the predicate device."

    The closest the document comes to "acceptance criteria" and "reported performance" is in the context of the functional/safety testing and biocompatibility testing. For these, the acceptance criteria are generally "passed" or "met/exceeded" internal requirements and ISO standards.

    Acceptance Criteria CategorySpecific Test/EvaluationReported Device Performance
    BiocompatibilityCytotoxicity (ISO 10993-1)Passed (per the same protocol as predicate)
    Sensitization (ISO 10993-1)Passed (per the same protocol as predicate)
    Irritation/Intracutaneous Toxicity (ISO 10993-1)Passed (per the same protocol as predicate)
    Acute Systemic Toxicity (ISO 10993-1)Passed (per the same protocol as predicate)
    Material Mediated Pyrogenicity (ISO 10993-1)Passed (per the same protocol as predicate)
    Hemolysis Assay (ISO 10993-1)Passed (per the same protocol as predicate)
    Complement Activation (ISO 10993-1)Passed (per the same protocol as predicate)
    LAL Pyrogenicity (ISO 10993-1)Passed (per the same protocol as predicate)
    Design VerificationRadiopacity Acceptability (Polymer Tip Extrusion)Met or exceeded Transit Scientific's in-house requirements
    Tip ID TestingMet or exceeded Transit Scientific's in-house requirements
    Tip Tensile TestingMet or exceeded Transit Scientific's in-house requirements
    Biocompatibility Testing (Hydrophilic Coating)Passed
    Catheter OD TestingMet or exceeded Transit Scientific's in-house requirements
    Coating Lubricity TestingMet or exceeded Transit Scientific's in-house requirements
    Coating Durability TestingMet or exceeded Transit Scientific's in-house requirements
    Distal Catheter TensileMet or exceeded Transit Scientific's in-house requirements
    Coating Particulate TestingMet or exceeded Transit Scientific's in-house requirements
    Simulated UseMet or exceeded Transit Scientific's in-house requirements
    Coating IntegrityMet or exceeded Transit Scientific's in-house requirements
    Coaxial TestingMet or exceeded Transit Scientific's in-house requirements
    Overall Design Verification (ISO 10555-1 requirements)All models met or exceeded requirements

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide details on a 'test set' in the context of clinical data or patient outcomes. The testing described is primarily pre-clinical (biocompatibility, engineering design verification). Therefore, information on sample size and data provenance for a clinical test set is not available.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable to the provided document. The submission is for a medical device (support catheter), not an AI/diagnostic software. Ground truth, in the sense of expert consensus on imaging or pathology, is not part of this type of submission. The "ground truth" for the engineering and biocompatibility tests would be the established scientific methods and standards (e.g., ISO 10993-1, ISO 10555-1) and internal specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. There is no clinical 'test set' involving human interpretation or adjudication for this device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a physical support catheter and not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical support catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the evaluations performed is based on:

    • International Standards: ISO 10993-1 for biocompatibility and ISO 10555-1 for catheter design verification.
    • Internal Requirements: Transit Scientific's in-house specifications for design verification tests (e.g., radiopacity, tip ID, tensile, lubricity, durability, particulate, integrity, coaxial).

    8. The sample size for the training set

    This information is not applicable. The document does not describe the development of an AI algorithm with a training set. The descriptions relate to physical device design and manufacturing.

    9. How the ground truth for the training set was established

    This information is not applicable. As there is no AI algorithm or training set discussed, this question is not relevant to the provided content.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193420

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-05-06

    (149 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XO Cross is a support catheter intended to guide and support a guidewire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.

    Device Description

    The XO Cross catheter is a single-lumen support catheter designed to support a quidewire during access of the peripheral vasculature, allowing for exchange of guidewires during vascular access procedures, and providing a conduit for the delivery of saline solutions or diagnostic contrast agents. The shaft and tip are both radiopaque to track the location of the XO Cross catheter within the vasculature, while under fluoroscopy. It is available in various lengths (90cm, 135cm, 150cm, 175cm) and ranges in diameter from 2 Fr (0.7 mm) to 4 Fr (1.35 mm) with a tapered polymer tip at its distal end.

    AI/ML Overview

    The provided text is a 510(k) summary for the XO Cross Support Catheter. It outlines the device description, indications for use, and a comparison to a predicate device. However, it does not contain detailed information about specific acceptance criteria or the study data that proves the device meets those criteria, especially in the context of device performance metrics with numerical values or statistical analyses.

    The document states that "All necessary verification steps met pre-determined acceptance criteria to confirm substantial equivalence" and "All data met pre-determined acceptance criteria" for functional tests, but it does not specify what those criteria are beyond general statements. It also doesn't describe the methodology of any studies that would demonstrate these criteria were met in a quantifiable way.

    Therefore, I cannot populate the table or answer most of the specific questions as the required information is not present in the provided text.

    Here's what can be extracted and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Stated or Implied)Reported Device Performance
    Functional/Safety Testing:
    Biocompatibility (ISO 10993-1)Met ISO 10993-1 requirements (Cytotoxicity, Irritation/Intracutaneous Toxicity, Sensitization, Systemic Injection, Material Mediated Pyrogenicity, Hemolysis Complete, Complement Activation, Thrombogenicity)
    Design Verification (ISO 10555-1)Met or exceeded Transit Scientific's in-house requirements and ISO 10555-1 requirements
    Packaging integrityValidated in conjunction with sterilization studies
    Max Infusion PressureTested for manual infusion, not for use with power injectors (No specific pressure limit given for manual infusion, implying it passed internal criteria for manual use)
    Materials (similar to predicate)Uses single Lumen metal alloy hypo-tube shaft with tapered tip and polymer jacket, different from predicate's SS braided shaft (This is a design difference, not a performance metric)
    Dimensions (similar to predicate)Various lengths (90cm, 135cm, 150cm, 175cm), diameters (2Fr to 4Fr), and distal tip profiles (1.6Fr, 1.9Fr, 3.2Fr), similar to, but not identical to, predicate (Again, a design spec, not a performance metric with specific acceptance criteria)
    Visibility (radiopaque markers)Radiopaque catheter body, 40% tungsten loaded polymer tip, different from predicate's 3 radiopaque markers (Design difference)
    Guidewire Compatibility0.014", 0.018", 0.035" (Same as predicate)
    Sheath/Introducer Compatibility4Fr (Predicate mentions 4Fr, 5Fr. Subject device is only 4Fr, but "same" is noted in the table)
    Sterility (EO)Sterile - EO (Same as predicate)

    Missing Information:

    • Specific quantitative acceptance criteria for each functional test (e.g., maximum acceptable cytotoxicity index, specific force required for kink resistance in design verification).
    • Quantitative results of the functional tests against the specific criteria.
    • Any performance testing related to guidewire support strength, catheter trackability, pushability, or other in-use functional attributes beyond basic material and design verification.

    2. Sample size used for the test set and the data provenance

    • Missing. The document does not describe any specific "test set" in terms of patient data or clinical samples that would have a sample size or provenance. The "functional/safety testing" refers to laboratory-based verification, and no sample sizes for these tests are provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This information would typically be relevant for studies involving qualitative assessments, such as image interpretation or clinical diagnosis. The described testing is primarily laboratory-based engineering and biocompatibility testing. No "ground truth" established by experts in this context is mentioned.

    4. Adjudication method for the test set

    • Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a physical medical device (support catheter), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical medical device, not an algorithm.

    7. The type of ground truth used

    • For the functional/safety testing, the "ground truth" would be the standards and specifications (e.g., ISO 10993, ISO 10555) that the device must meet, as well as the manufacturer's own internal design requirements. These are objective engineering and biological standards. There is no "expert consensus," "pathology," or "outcomes data" ground truth described for the tests mentioned.

    8. The sample size for the training set

    • Not applicable. This device is a physical medical device. There is no concept of a "training set" as there would be for AI/machine learning models.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.