Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K232175

    Validate with FDA (Live)

    Date Cleared
    2024-04-14

    (268 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatic arthritis.

    Post-traumatic loss of knee joint configuration and function.

    Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

    Revision of previous unsuccessful uni-knee replacement or other procedure.

    The system components are intended for cemented use only.

    Device Description

    The XN Knee Prosthesis System is a knee implant developed by Beijing Chunlizhengda Medical Instruments Co. Ltd. The subject system includes femoral condyle, tibial tray, tibial insert and patella. The fixation method of the system is bone cement fixation. It is mainly used in total knee replacement to repair the diseased or injured knee joint, improve the range of motion of the knee joint and reduce the pain of the patient. The XN Knee Prosthesis System is a Posterior Stabilized (PS) total knee replacement system.

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device (XN Knee Prosthesis System). It confirms that the device is substantially equivalent to a previously marketed predicate device. For devices seeking 510(k) clearance based on substantial equivalence, the "study that proves the device meets the acceptance criteria" is typically a series of non-clinical performance tests and engineering analyses that demonstrate the new device performs as safely and effectively as a legally marketed predicate device.

    It's important to understand that a 510(k) submission generally does not involve large-scale clinical trials or studies akin to those for drug approvals. Instead, the focus is on demonstrating equivalence through comparison to an existing device, primarily via bench testing, material characterization, and engineering simulations.

    Based on the provided text, here's a breakdown of the "acceptance criteria" and the "study" (non-clinical testing) that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states: "Testing and engineering analyses showed that the subject components met the pre-determined acceptance criteria identified in the Bench Testing Summary, demonstrating that the subject components are substantially equivalent compared to the predicate components."

    While explicit numerical acceptance criteria for each test are not detailed in this summary (they would be in the full submission, which is not provided), the document lists the types of performance tests conducted. The "reported device performance" is implicitly that the device passed these tests and was found substantially equivalent.

    Acceptance Criterion (Implied)Reported Device Performance
    Sufficient Component Interlocking StrengthMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable Fatigue Performance Under High FlexionMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable Tibial Insert ConstraintMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable Surface RoughnessMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable Patella Contact PressureMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable Patella ConstraintMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable Static Contact StressMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable Dynamic FatigueMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable Wear PerformanceMet acceptance criteria (demonstrated acceptable mechanical performance)
    Acceptable UHMWPE CharacterizationMet acceptance criteria (demonstrated acceptable mechanical performance)
    Mechanical performance for intended useAcceptable mechanical performance demonstrated for intended use
    Equivalence to predicate device in intended use, materials, and fundamental technologySubstantially equivalent to predicate device

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify exact sample sizes for each test. For non-clinical bench testing, sample sizes are typically determined by relevant ASTM or ISO standards for testing medical devices (e.g., often N=3 or N=5 for certain mechanical tests, depending on variability and statistical power needs). This information would be in the detailed test protocols within the full 510(k) submission.
    • Data Provenance: The testing was non-clinical (bench testing and engineering analysis) and performed by the manufacturer, Beijing Chunlizhengda Medical Instruments Co. LTD. The location of primary testing would likely be China, where the manufacturer is based. It is retrospective in the sense that the data was collected and analyzed for this specific submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    • This question is not applicable in the context of a 510(k) non-clinical bench testing submission. "Ground truth" for these tests is established by engineering specifications, accepted industry standards (e.g., ISO, ASTM), and comparison to the predicate device's known performance characteristics. There are no human "experts" establishing a "ground truth" for each individual test sample in the way a clinical trial might use radiologists for image interpretation. The expertise lies in the engineers performing and reviewing the tests, and the FDA reviewers assessing the data against regulatory standards.

    4. Adjudication Method for the Test Set

    • This question is not applicable as the "test set" consists of physical devices undergoing mechanical and material property testing, not subjective human assessments that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • This question is not applicable. The XN Knee Prosthesis System is a mechanical orthopedic implant, not an AI/software device used for diagnostic imaging or a human-in-the-loop system. The clearance is based on substantial equivalence to a predicate device through non-clinical performance data.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • This question is not applicable. As stated, this is a mechanical implant, not an algorithm.

    7. The Type of Ground Truth Used

    • For this device, the "ground truth" for substantial equivalence is primarily established through:
      • Engineering specifications and design requirements: The device must meet predefined mechanical and material properties that ensure safety and efficacy.
      • Comparison to the predicate device: The new device's performance is measured against the known and accepted performance of the legally marketed predicate device (Stryker Triathlon® Total Knee System). The standard is "substantially equivalent," meaning it performs as safely and effectively as the predicate.
      • Relevant industry standards (e.g., ISO, ASTM): These standards often define the specific tests and acceptable ranges for performance for orthopedic implants.

    8. The Sample Size for the Training Set

    • This question is not applicable. This is a mechanical device, not a machine learning or AI model that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    • This question is not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1