510(k) Data Aggregation

    K Number
    K242926

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-30

    (248 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XBODY Go USA and XBODY Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used only for healthy muscles and clients, not for rehabilitation purposes.

    Both models are intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Neither model is intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBODY Go USA or XBODY Pro USA training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.

    For both models, electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    The various types of muscle work that the XBODY Go USA and XBODY Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. Both models may therefore be considered a technique of muscle training.

    Device Description

    The XBody Go USA and the XBody Pro USA generate electronic muscle stimulation based on electrical muscle stimulation (EMS) technology. The devices are designed as additions to other sports and for training muscles. They are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. They are intended for use only by persons with healthy muscles, not for rehabilitation purposes.

    AI/ML Overview

    Based on the provided FDA 510(k) Clearance Letter and Submission Summary, here's a description of the acceptance criteria and the study that proves the device meets them:

    Important Note: The provided document is a 510(k) submission summary, primarily focused on establishing substantial equivalence to a predicate device. It details the similarities and minor differences between the new device and the predicate. For devices like the XBody Go USA and XBody Pro USA, which are "Powered Muscle Stimulators," the primary "performance" and "acceptance criteria" revolve around demonstrating equivalence in electrical safety, electromagnetic compatibility, and fundamental stimulation characteristics, rather than a clinical study measuring a specific physiological outcome (like muscle strength improvement). The document explicitly states, "Any differences that exist between the subject and predicate devices do not affect safety or performance." This implies that the acceptance criteria are largely met by demonstrating adherence to recognized safety and performance standards and showing that any changes do not negatively impact these.

    Therefore, the "proof" described is not a typical clinical trial demonstrating superiority or non-inferiority in effectiveness, but rather a demonstration of continued safety and equivalent performance to a device already deemed safe and effective.


    Acceptance Criteria and Device Performance Study

    The acceptance criteria for the XBody Go USA and XBody Pro USA, as implied by the 510(k) submission, are primarily centered around demonstrating essential safety and equivalent performance to their predicate devices, particularly regarding electrical muscle stimulation parameters and the introduction of new accessories (the XBody DrySuit and DryWear). Since this is a 510(k) for a powered muscle stimulator, the "performance" proved is not a clinical outcome like muscle strength gain, but rather the device's ability to safely and effectively deliver electrical impulses as intended, consistent with the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of this 510(k) (substantial equivalence for a powered muscle stimulator), the "acceptance criteria" are predominantly related to compliance with recognized standards and maintaining equivalent technical specifications and safety profiles. The "device performance" is therefore demonstrated by meeting these standards and showing that the subject device's operation is not adversely affected by the minor changes.

    Acceptance Criteria (Implied)Reported Device Performance (as demonstrated)
    Electrical Safety (IEC 60601-1, 60601-2-10)Validated through independent testing. Product meets max output voltage, current, phase charge, current density, power density, net charge, waveform shape specifications, and incorporates safety circuits (overload, no-load trip, battery monitoring, watchdog).
    Electromagnetic Compatibility (IEC 60601-1-2)Validated through independent testing. Device is compliant.
    Usability (IEC 60601-1-6, IEC 62366-1)User interface is described as "large and easily controllable" with clear stimulation parameters and channel identification. Operator training and certification are required. Updated control units maintain ease of use.
    Software Life Cycle (IEC 62304)Software differences are minor, aiming to enhance user interface without impacting safety/effectiveness. Compliance with IEC 62304 is stated.
    Biocompatibility (ISO 10993-1, -5, -10)Cotton underwear is "biocompatibility certified." New DryWear and DrySuit materials are implied to be compliant through risk management and biocompatibility standards.
    Maintaining Intended Use & Indications for UseDevices have "same in indication for use, intended use performance, design, dimensions, and materials as the predicate devices." The new DrySuit is stated to have "acceptable similarity in intended use."
    Equivalent Output SpecificationsOutput voltage, current, phase charge, current density, power density, net charge, waveform, burst mode parameters, output frequency, and pulse widths are all "Same" as the predicate.
    Equivalent Portability/Environment of Use"Portable with ease," "indoors," "home healthcare environment." Same as predicate.
    Safe Use of New Accessories (DrySuit/DryWear)Performance testing of the XBody DrySuit demonstrated "acceptable similarity in intended use." Instructions for use (e.g., wearing undergarments) ensure safe conductivity.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a "sample size" in the context of a data-driven test set for complex performance metrics like accuracy or sensitivity/specificity for an AI/ML device. For this type of device, the "test set" is more conceptual, referring to the various configurations and conditions under which the device's electrical safety, EMC, and fundamental performance characteristics were validated through bench testing.

    • Sample Size: Not applicable in the typical sense for a clinical data set. The validation is based on engineering tests performed on production units or prototypes.
    • Data Provenance: The data comes from "independent testing" of the XBody Actiwear G2 main unit, the control unit, the XBody Training Suit 2.1, the XBody Training Suit 3.0, and the new XBody DrySuit. This implies lab-based testing. The country of origin of the data/testing is not specified, but the manufacturer is XBody Hungary Kft. The testing would be considered prospective in the sense that it's conducted to demonstrate compliance for the submission, rather than retrospective analysis of existing clinical data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. For a powered muscle stimulator's safety and performance testing, "ground truth" is established by calibrated instruments and adherence to engineering standards (e.g., peak voltage, current, waveform shape, and safety circuit functionality). There is no "expert consensus" on an image or clinical finding that requires adjudication.

    4. Adjudication Method for the Test Set

    Not applicable. There is no human interpretation or subjective assessment of data requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI/ML diagnostic or image-based device that would typically undergo an MRMC study to assess human reader improvement with AI assistance. The device's function is direct electrical muscle stimulation, not assisting in interpretation.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This device is a physical electromedical device, not a standalone algorithm. Its performance is intrinsically linked to its hardware and software working together to deliver electrical impulses.

    7. Type of Ground Truth Used

    The "ground truth" is based on:

    • Engineering specifications and standards: Adherence to output parameters (voltage, current, frequency, pulse width) as defined by electrical engineering principles and the device's design.
    • Safety standards: Compliance with IEC 60601 series standards for electrical safety and electromagnetic compatibility.
    • Predicate device characteristics: The predicate device itself effectively serves as a "ground truth" or benchmark for acceptable performance and safety, as the submission aims to demonstrate substantial equivalence.

    8. Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that learns from a "training set" of data in the typical sense of machine learning. The device's behavior is deterministic, based on its programmed logic and hardware design.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no "training set" for this type of device. The ground truth for the device's design and functionality is rooted in established electrical engineering principles, medical device safety standards, and the performance characteristics of the legally marketed predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221200

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-31

    (280 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XBODY Go USA and XBODY Pro USA is a machine with electronic muscle stimulation based on EMS technology. The device is specifically designed as an addition to other sports and for training muscles. It must be used for only healthy muscles and clients, not for rehabilitation purposes.

    The XBODY Go USA and XBODY Pro USA is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The XBODY Go USA and XBODY Pro USA is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBODY Go USA training programs is designed for injured or ailing muscles and its use on such muscles is contraindicated.

    The XBODY Go USA and XBODY Pro USA electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber respond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    The various types of muscle work that the XBODY Go USA and XBODY Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. The XBODY Go USA and XBODY Pro USA may therefore be considered a technique of muscle training.

    Device Description

    The XBody Go USA and the XBody Pro USA generate electronic muscle stimulation based on EMS technology. The devices are designed as additions to other sports and for training muscles. They are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. They are intended for use only by persons with healthy muscles, not for rehabilitation purposes.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the XBody Go USA and XBody Pro USA, which are powered muscle stimulators. While it thoroughly compares the new devices to predicates and lists many standards and tests, it does not describe a study that proves the device meets specific acceptance criteria in the way a clinical trial or performance study would for an AI/ML medical device.

    The document primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices (XBody Newave USA and WiemsPro) through a comparison of various characteristics (technological, performance, safety standards, materials). For a device like a powered muscle stimulator, the "acceptance criteria" are generally tied to meeting recognized safety and performance standards (bench testing) and demonstrating similar intended use and technological characteristics to a legally marketed predicate, rather than a clinical effectiveness study with a specific metric like accuracy, sensitivity, or specificity for a diagnostic AI.

    Therefore, many of the specific points requested regarding AI/ML device performance studies (e.g., sample size for test set, ground truth experts, MRMC studies, standalone performance, training set details) are not applicable or are not explicitly stated in this 510(k) document because the approval pathway relies on substantial equivalence to existing devices rather than a de novo clinical performance validation against a defined ground truth.

    However, I can extract the relevant "acceptance criteria" which, in this context, are the various technical and safety parameters that need to be "similar" or "meet" established standards, and how these performances are demonstrated.

    Here is the information that can be extracted or reasonably inferred from the provided text, addressing the prompt's points where applicable, and noting when the requested information is not present or not relevant to this type of device submission:


    1. A table of acceptance criteria and the reported device performance

    The "acceptance criteria" in this context are the characteristics and performance parameters that are compared to the predicate devices to demonstrate substantial equivalence. The "reported device performance" is essentially the description of the subject device's characteristics and how they compare to the predicates.

    Acceptance Criteria (Parameter)Predicate Performance (XBody Newave USA, WiemsPro)Reported Device Performance (XBody Go USA, XBody Pro USA)Assessment / Conclusion of Substantial Equivalence
    Intended UseTo stimulate healthy muscles to improve or facilitate muscle performance; not for rehabilitation or medical conditions.Same. Designed as an addition to other sports and for training muscles; for healthy muscles/clients only.Same.
    Powered Muscle StimulatorYesYesSame.
    Regulated VoltageYesYesSame.
    Portability/Mobile UsePortable (17 lbs.), not a mobile device; indoors, requires trained operator.Portable with ease; not a mobile device; intended environment is indoors; requires qualified and trained operator.Similar. Both are portable, not mobile.
    User InterfaceDevice with touchscreen for control; large START/STOP buttons; clear stimulation controls and parameters; channel identification with pictures; STOP button always visible.The device can be controlled using graphical windows on the touchscreen of an external control unit (Microsoft Surface Go 2 / Surface Pro 7+). Similar controls and visible STOP button.Similar. Both utilize a touchscreen control unit. Supports usability performance testing.
    Operator RequirementsTrainer must complete an XBody US EMS Trainer Course; certification data required for registration.Trainer must complete an XBody US EMS Trainer Course; certification data required for registration; only registered trainers can use.Same.
    Display10.4" resistive touchscreenGO: <10.5" LCD; PRO: <12.3" LCDDifferent/Similar. Subject devices have larger, capacitive LCD touchscreens. Considered similar as it supports usability.
    Max Output VoltagePredicate: 20.8V @500Ω; Reference: 30V @500Ω62.5V @500ΩSimilar to Predicate (higher than predicate but tested to IEC 60601-1 for safety and effectiveness). Different compared to Reference. "This particular feature has already been tested to IEC 60601-1:2005+ A1:2012+A2:2020. This difference does not raise any concern for safety and effectiveness."
    Max Output CurrentPredicate: 125mA @500Ω60mA @500ΩSimilar (lower than predicate, tested to IEC 60601-1 for safety and effectiveness). "This difference does not raise any concern for safety and effectiveness."
    Max Phase ChargePredicate: 24µC@500Ω16.64µC@500ΩSimilar (lower than predicate, tested to IEC 60601-1 for safety and effectiveness). "This difference does not raise any concern for safety and effectiveness."
    Max Current DensityPredicate: 0.61mA/cm2 @500Ω; Reference: 1.92mA/cm2 @500Ω0.65mA/cm2 @500ΩSimilar to Predicate (slightly higher but tested to IEC 60601-1 for safety and effectiveness). Different compared to Reference. "This difference does not raise any concern for safety and effectiveness."
    Max Power DensityPredicate: 7.27mW/cm2 @500Ω; Reference: 9.61mW/cm2 @500Ω3.46mW/cm2 @500ΩSimilar to Predicate (lower than predicate, tested to IEC 60601-1 for safety and effectiveness). Different compared to Reference. "This difference does not raise any concern for safety and effectiveness."
    WaveformSymmetric biphasicSymmetric biphasicSame.
    Output Channels10 individual, galvanically isolated channels for each output (Predicate); 1 output channel can shift in time to 10 outputs (Reference).6 output channels, but 12 independently regulated outputs.Similar to Predicate. Different compared to Reference (Subject devices have more independently regulated output channels).
    Output Frequency1-100Hz1-150HzDifferent. "Object devices provide better customization through a wider spectrum of output frequency. During the summative evaluation, the users evaluated the feeling of stimulation positively. The main output parameters (output current and output voltage) are lower for the subject devices than for the primary predicate and reference device. Therefore, the difference would not affect safety and effectiveness of the subject device."
    Pulse Width (Positive & Negative)50-400µsec (Predicate); 100-400µsec (Reference)50-500µsecDifferent. "Object devices provide better customization through a wider spectrum of positive/negative pulse width. During the summative evaluation, the users evaluated the feeling of stimulation positively. The main output parameters (output current and output voltage) are lower for the subject devices than for the primary predicate and reference device. This particular feature has already been tested to IEC 60601-1:2005+ A1:2012+ A2:2020. Therefore, the difference would not affect safety and effectiveness of the subject device."
    Power Source - Battery12V 10Ah LiFePO4 battery in sealed housing (Predicate); LiPo 3.7V (2.4Ah) (Reference)Li-ion 4x 3.7V (3.4Ah)Similar. All use rechargeable batteries. Subject device chosen for light weight and compact design. Bench tested for function and safety (similar to WiemsPro).
    Safety CircuitsOverload trip, No-load trip, battery voltage monitoring, hardware error detection at startup, watchdog monitoring (Predicate); Software/Firmware/Microprocessor Control, Automatic Overload Trip, Automatic Shut Off, Patient Override Control (Reference).Overload trip, No-load trip, battery voltage monitoring, hardware error detection at startup, watchdog monitoring.Same as Predicate. Different from Reference (subject devices utilize a wider range of safety features than reference).
    Connectivity (Plugs/Wires)Spiral cable connects control unit to training suit (Predicate).Tablets and Actiwear G2 wireless stimulation unit are connected wirelessly. Actiwear G2 and Training Suit connected with magnetic connectors. Internal suit cable connects to snap fasteners to which detachable electrodes are attached via waterproof connections.Different from Predicate. "Primary predicate has built-in control panel, while the subject device connects wirelessly with the control unit. The subject devices are worn by the client, therefore no spiral cable is needed to connect with the training suit." Wires are similar/compliant with safety requirements.
    ConductivityClient wears XBody cotton underwear; electrodes in cotton covers watered with tap water for conductivity.Client wears XBody cotton underwear; electrodes in cotton covers watered with tap water for conductivity; washable/disinfectable.Same.
    Programs4+10 (Predicate); 20 (Reference)XBody Go USA: 5; XBody Pro USA: 6 (including manual settings and various training programs).Same to Go USA vs. Predicate. Similar to Pro USA vs. Predicate. Different vs. Reference (XBody devices offer broader customization through manual programs).
    Treatment Duration1 min to 60 min maximum1 min to 60 min maximumSame.
    Environment of UseHome healthcare environment, according to IEC 60601-1-11.Home healthcare environment, according to IEC 60601-1-11.Same.
    Compliance StandardsIEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-10, IEC 62366, IEC 62304.All predicate standards, plus additional for wireless (ANSI/AAMI TIR 69:2017, IEEE/ANSI C63.27-2017, ANSI/AAMI TIR57:2016).Same/Different. Subject device was tested to additional standards due to wireless capability, ensuring overall safety and effectiveness.
    Training SuitPredicate TS2.0: 75% Polyamide, 25% Polyester; 10 channels; 20 electrodes; 3kg.TS2.1: Same material as TS2.0; 12 channels; 24 electrodes; 2.7-3.3kg. TS3.0: 40% Polyamide, 30% Acrylic, 25% Polyester, 5% Elastane; 12 channels; 24 electrodes; 1.8-2.4kg. Wider range of sizes and electrode shapes for both new suits.Similar. Differences in material (TS3.0) and number of channels/electrodes for improved comfort/performance, without affecting safety or primary intended use. Performance testing supported similarity. Output specs for new suits are within similar ranges or lower than predicates.
    SoftwareNot explicitly detailed beyond 'similar software platforms'.Similar software platforms with some additional programming (e.g., UPBEAT, group training) for Pro USA. Validated to IEC 62304:2006+A1:2015.Similar. Differences noted "do not affect the safety and effectiveness" and are for "user-friendly interface".

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set (Performance Testing): The document states "summative evaluation, the users evaluated the feeling of stimulation positively" and "Performance testing of the Training Suit 3.0 demonstrated acceptable similarity". It also relies heavily on compliance with various IEC standards (bench testing). No specific sample size for a "test set" of patient data (as relevant for AI/ML diagnostic devices) is mentioned. The "test set" here refers to the device and its components undergoing engineering and electrical safety tests.
    • Data Provenance: Not applicable in the context of human data. The tests are bench tests and "summative evaluations" of "users" (likely trained operators or healthy volunteers) in a controlled environment as part of the engineering validation.
    • Retrospective/Prospective: Not applicable to this type of device submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This device is a powered muscle stimulator, not a diagnostic AI/ML device that requires expert-established ground truth from medical images or clinical outcomes. The "ground truth" for this device's safety and performance is established through compliance with recognized electrical safety, EMC, usability, and biocompatibility standards, and comparison to existing predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No human interpretation or qualitative assessment of a test set requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted diagnostic device. The device itself delivers electrical stimulation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/ML diagnostic algorithm. The device's "performance" is its ability to safely and effectively deliver electrical stimulation according to its specifications, which is validated through bench testing and compliance with standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable/Implicit. For this type of device, the "ground truth" for demonstrating safety and efficacy is established by:
      • Regulatory Standards: Compliance with international consensus standards like IEC 60601 series, ISO 14971, ISO 10993, etc.
      • Predicate Device Performance: Demonstrating that the new device's technical characteristics and performance are substantially equivalent to a previously cleared device.
      • Bench Testing: Engineering measurements conducted to verify specifications and safety.
      • Usability Testing: Verification of safe and effective user interaction (mentioned as "usability performance testing" and "summative evaluation" where users "evaluated the feeling of stimulation positively").

    8. The sample size for the training set

    • Not applicable. This device does not use a "training set" in the context of AI/ML model development. Its development is based on engineering design, comparison to known safe and effective technologies, and adherence to established manufacturing and quality systems.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an AI/ML model, the concept of "ground truth for the training set" is irrelevant in this submission.