Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K131383

    Validate with FDA (Live)

    Date Cleared
    2013-11-27

    (197 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X-Series System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, transmit, display and store physiological signals from patients ages 6 and older. The X-Series system requires operation by a trained technician. The X-Series System acquires, transmits, displays and stores electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), and accelerometer signals. The X-Series System only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.

    Device Description

    The X-Series System is indicated for acquiring, transmitting, displaying and storing physiological data in patients. It can be used with ambulatory patients in the home, health care facility, or clinical research environment. The X-Series system requires operation by a trained technician. The X-Series System.is comprised the X10 and X24 Headsets and accessories, Synapse Cream, X-Series Basic Software and BT receiving unit. The X-Series Basic Software is also compatible with the Models X4-E, and X4-M (K130013) when used in wireless mode. The X-Series System combines hardware, firmware and software to acquire physiological signals. It acquires physiological data through a battery powered headset worn by the patient and provides a flexible platform for applying sensors using synapse cream and acquiring signals from multiple locations on the head or body, transmitting and recording the signals and providing visual indications to ensure high quality data are obtained. Model X24 provides for acquisition of twenty channels of electroencephalography (EEG) and four optional channels connected to two sensors via a dual-lead connector. Model X10 provides for acquisition of nine channels of electroencephalography (EEG) and an optional channel connected to two sensors via a dual-lead connector. Both models measure movement and position measured via a 3-D accelerometer. The device is designed so it can be affixed by a technician and displays the signals via a wireless connection during acquisition. The X-Series Basic software monitors signal quality to ensure that the sensors are properly applied and that high quality signals are being acquired. The X-Series Basic software provides a means to: a) initiate a study and track patient information, b) acquire and wirelessly transmit signals from the device, c) visually inspect the signal quality. The acquired signals are saved in a universal data format (European Data Format Plus, EDF+) that is intended to be analyzed by a Physician using FDA cleared third party software, i.e. Persyst Software (K011397).

    AI/ML Overview

    This document describes the X-Series System, an electroencephalograph device for acquiring, transmitting, displaying, and storing physiological data. The submission focuses on demonstrating substantial equivalence to a predicate device (K130013 X4 System) through non-clinical and limited clinical testing.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document leverages comparison to a predicate device and adherence to recognized standards as its primary acceptance criteria. Performance metrics are largely presented as being equivalent to the predicate.

    Acceptance Criteria CategorySpecific Criteria (from document)Reported Device Performance (X-Series System)
    Functional EquivalenceEquivalence to predicate device (X4 System) for electrophysiological (EEG, EOG, ECG, EMG), wireless acquisition, and actigraphy. Technologies used in the same manner.EEG: Sampling Rate: 256 s/s, 0.1 Hz High Pass (hardware), 100 Hz Low Pass (hardware), Dynamic range: +/- 1000μV, Resolution 0.03 µV, Peak to peak noise: 3.7 µV (typical), 110dB Common Mode Rejection Ratio (typically), Input Impedance: 100GOhm (Equivalent to X4) Aux - EOG/EMG: Sampling rate 256 Hz, 0.1 Hz High Pass (hardware), 100 Hz Low Pass (hardware). EMG/EOG: Dynamic range: +/- 1000μV, Resolution 0.03 μV, Peak to peak noise: 3.7 μV (typical) (Equivalent to X4) Aux - ECG: Bandwidth: 0.1 to 100Hz -3dB, Notch Filters: 50, 60Hz (software), Common Mode Rejection Ratio: 110dB, Input Impedance: 100GOhm, Input Range: +/-4mV, Accuracy: 16 bits, Noise: <5μVpp (typically 4.2μVpp), Max Electrode DC Offset: +/-125mV, Dynamic range: +/- 2000μV, Resolution: 0.06 μV (Equivalent to X4) Accelerometer: Sampled 100Hz, downsampled to 10Hz; X x Y x Z @ 10 s/s, Resolution nominal 10 bit at 2g, Actual output range for -90 to 90 degrees is 12 bit. Position accuracy typically +/- 3.0 degrees, maximum +/- 5.0 degrees in the +/- 60 degrees range. Sensitivity: 10 bit / 2g, Non-linearity: +/-0.5 %FS, Cross-Axis: +/-1 %, Zero-Level: +/-0.35-0.4 mg, Sensitivity change due temperature: +/-0.01 LSB/°C (Equivalent to X4) Bench tests confirmed EEG and optional dual-lead (for ECG) analog signals were equivalent to predicate. Actigraph measures were equivalent.
    Electrical SafetyCompliance with IEC 60601-2-26:2002 (Electroencephalographs) and IEC 60601-1-11: 2010 (Home healthcare environment)All tests "passing demonstrating the compliance of the X-Series system to FDA recognized standards for electro-medical equipment."
    BiocompatibilityCytotoxicity, Sensitization, Irritation per ISO 10993 standards (for patient-contacting components).Cytotoxicity (ISO 10993-5): No reactivity (grade 0) in all cultures. Concluded "Non-cytotoxic." Sensitization (ISO 10993-10): Irritation absent from all animals, animals gained weight, no overt toxicity, no skin reaction scores. Concluded "Non-sensitizer." Irritation (ISO 10993-10): No signs of erythema or edema at observation points. Test Primary Irritation Index (PII) was 0.0. Concluded "Non-irritant."
    Cleaning & DisinfectionAcceptance criteria: ≥ log 3.0 reduction in bioburden after cleaning according to instructions, per AAMI TIR 12-94 and AAMI TIR 30: 2003."All tests passed and demonstrate that the cleaning methods are appropriate for ensuring the X-Series acquisition device is clean between uses."
    Software IntegrityVerification and validation of specifications, function, and intended use, per FDA's Guidance for Premarket Submissions for Software Contained in Medical Devices (May 2005)."The results of the verification and validation activities that have been performed demonstrate that the software meets requirements for safety, function, and intended use."
    EEG Electrode PerformanceElectrical performance acceptable, using applicable methods from AAMI/ANS1 EC12:2000 (R)2010."The results demonstrate electrical performance of the EEG electrodes used with the X-Series System is acceptable."
    Clinical Performance (Ages 6+)Ability to acquire appropriate EEG recordings in children aged 6-8, with equivalent data quality to adults.100% of EEG sessions contained appropriate EEG recordings. Data quality from children was equivalent to an equivalent adult population. Technician setup was quick, and no discomfort or agitation was reported by subjects.

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Non-clinical bench tests (EEG, ECG, Actigraph comparison): The sample size is not explicitly stated as 'n=X' but implies a sufficient number of measurements were taken from the X-Series and predicate devices to compare their performance in a jig and with an oscilloscope/signal generator.
    • Biocompatibility: The specific number of items/animals per test is mentioned in the results (e.g., "all the animals," "all test animals").
      • Cytotoxicity: Mammalian cell cultures (mouse fibroblast L929).
      • Sensitization: Albino guinea pigs.
      • Irritation: New Zealand White rabbits.
    • Clinical tests (for ages 6+):
      • Sample size: Seven children aged 6-8 and seven adults.
      • Data provenance: "A subset of data from a research study was analyzed." This suggests the data was retrospective and likely collected in a clinical research environment (Country of origin not specified, but likely USA given the FDA submission).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • Non-clinical tests: The document implies standard laboratory testing procedures and comparisons to the predicate, overseen by qualified personnel. No specific "experts" for ground truth establishment are mentioned as the tests rely on objective measurements and established standards.
    • Clinical tests (for ages 6+): The document states that the acquired EEG recordings "could be further examined in third-party software" by a Physician. While a physician would interpret the EEG, their role in establishing "ground truth" for the device's acquisition capabilities (the focus of this study) is not explicitly detailed. The acceptance criteria here are about the ability to acquire appropriate EEG recordings and data quality equivalence, which are evaluated directly from the device's output. The "ground truth" for this aspect is the objective quality of the signals themselves, compared between subject groups.

    4. Adjudication Method for the Test Set:

    • No formal adjudication method (like 2+1 or 3+1) is mentioned for any of the studies.
    • The non-clinical tests rely on objective measurement against defined specifications or predicate device performance.
    • The clinical study for children's EEG acquisition focused on the presence of "appropriate EEG recordings" and data quality equivalence, which would likely be assessed by comparing signal characteristics.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance:

    • No MRMC comparative effectiveness study was done.
    • The X-Series System is a data acquisition device; it explicitly states, "The X-Series System only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability." It is not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done:

    • Not applicable. This device is purely for data acquisition and display, to be interpreted by a physician using separate (third-party) software. There is no algorithm providing diagnostic or analytical output within the X-Series System itself.

    7. The Type of Ground Truth Used:

    • Non-clinical/Bench tests: Objective measurements (e.g., signal generator output, oscilloscope readings, defined angles for actigraph, standard test methods for biocompatibility and cleaning). The comparison to the predicate device serves as a de facto "ground truth" for performance equivalence where applicable.
    • Clinical tests (for ages 6+): The "ground truth" for this specific study was the presence of "appropriate EEG recordings" and subjectively assessed "equivalent data quality" for the acquired signals between age groups. This relies on the qualitative assessment of electrical signal characteristics typically reviewed by trained professionals.

    8. The Sample Size for the Training Set:

    • Not explicitly stated, nor is a dedicated "training set" relevant for this device. The X-Series System is a data acquisition device, not a machine learning model that requires a training set. The clinical study for ages 6+ was a validation of its acquisition capabilities, not for training an algorithm.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable, as no training set was used for a machine learning algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1