Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223675

    Validate with FDA (Live)

    Date Cleared
    2023-01-06

    (30 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WatchPAT ONE (WP1) device is a noninvasive home care device for use with patients suspected to have sleep related breathing disorders. The WP1 is a diagnostic aid for the detection of sleep related breathing disorders, sleep staging (Rapid Eye Movement (REM) Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WP1 generates a peripheral arterial tonometry ("PAT") Respiratory Disturbance Index ("PRDI"), Apnea- Hypopnea index ("PAHI"), Central Apnea-Hypopnea index ("PAHIc"), PAT sleep staging identification (PSTAGES) and optional snoring level and body position discrete states from an external integrated snoring and body position sensor. The WP1's PSTAGES and snoring level and body position provide supplemental information to its PRD/PAHI/PAHIc. The WP1's PSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted.

    PAHIc is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older.

    Device Description

    The subject WatchPAT ONE (WP1) has the same intended use and indications for use as the cleared predicate WP300 (K222331) and the cleared reference WP1 (K183559). The subject WP1 combines the software arrhythmia feature of the WP300 cleared predicate (K222331) into the WP1 cleared reference (K183559).

    The subject WP1 consists of: A wrist worn Control Unit, Actigraphy (in the Control Unit), Wrist Strap, Power Supply (battery), uPAT probe, Chest sensor (optional), Management and Analysis Software (SW) and mobile application.

    The subject WP1, like the predicate WP300 and reference WP1, uses the same Management and Analysis Software (SW). The arrhythmia feature is to be used for informational use only, to flaq patients suspected of having arrhythmias, thereby aiding the physician to decide if further arrhythmia investigation is needed. The device is not intended to be used as a diagnostic device for cardiac arrhythmia and it is not intended to replace traditional methods of diagnosis. The arrhythmia detection is not intended for use in life supporting or sustaining systems or monitor and alarm devices. It provides the sleep physician with additional information to that of the WP1 cleared capabilities of detecting sleep disorders, to be considered in conjunction with the physician's knowledge of patient background, clinical history, symptoms, and other diagnostic information.

    AI/ML Overview

    The document provided does not contain the specific details required to complete all sections of your request. This is largely due to the nature of a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a full clinical study report with detailed performance metrics, ground truth establishment, or multi-reader multi-case studies.

    However, I can extract the following information based on the text provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" with numerical targets or direct "reported device performance" in the context of a clinical study for the newly added arrhythmia feature. It states: "SW Design verification was performed on the subject WP1 device with the SW containing the additional arrhythmia feature. The testing and acceptance criteria are the same as those in the predicate WP300 and reference WP1 devices." This implies that the acceptance criteria for the existing functionalities of the WatchPAT ONE (WP1) were met, and the new arrhythmia feature also adhered to the criteria established for the predicate WP300. Without further information on the WP300's performance from its original 510(k) for the arrhythmia feature, specific numerical acceptance criteria and performance for the arrhythmia detection cannot be provided from this document.

    The document emphasizes substantial equivalence to the predicate device WP300 (K222331) and reference device WP1 (K183559). The "performance data" section states: "Bench testing was previously conducted on the reference WP1 (K183559) to show that the acquisition system of the WP1 and WP300 generate equivalent input signals to the analysis algorithms. Specifically, the equivalence of the PAT and the actigraphy signals which are used in the newly added arrhythmia algorithm were tested. This testing provides evidence that the WP1 signals are equivalent to the WP300 signals."

    This suggests that the performance of the arrhythmia detection algorithm in the subject WP1 is assumed to be equivalent to that of the predicate WP300 because the input signals are equivalent and the software algorithm is identical.

    2. Sample size used for the test set and the data provenance

    The document does not specify a separate test set sample size or data provenance for the new arrhythmia feature. It refers to "bench testing" that validated the equivalence of signals between WP1 and WP300. This implies that no new clinical test set was used for the arrhythmia feature's performance validation in this specific 510(k) submission, as the feature itself originated from a previously cleared device (WP300).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable for this submission, as no new clinical test set for the arrhythmia detection was performed or described. The validation focuses on signal equivalence and software transfer from the predicate.

    4. Adjudication method for the test set

    Not applicable, as no new clinical test set for the arrhythmia detection was performed or described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study is mentioned or implied for the new arrhythmia feature. The device's arrhythmia flagging is described as "informational use only" to "flag patients suspected of having arrhythmias, thereby aiding the physician to decide if further arrhythmia investigation is needed." It is explicitly stated: "The device is not intended to be used as a diagnostic device for cardiac arrhythmia and it is not intended to replace traditional methods of diagnosis." Therefore, a comparative effectiveness study with human readers improving with AI assistance is outside the scope of this particular submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The document implies a standalone algorithm performance implicitly by stating the "SW Design verification was performed on the subject WP1 device with the SW containing the additional arrhythmia feature. The testing and acceptance criteria are the same as those in the predicate WP300". The feature "flags" abnormal events for a physician's review, suggesting the algorithm operates in a standalone manner to generate these flags. However, whether a dedicated standalone performance study (beyond software design verification) was done for the arrhythmia feature (as distinct from device equivalence) is not detailed here. The focus is on the equivalence of the acquisition system and the software algorithm to the predicate device, where the arrhythmia feature was initially cleared.

    7. The type of ground truth used

    The document does not describe the ground truth used for the arrhythmia feature validation in this submission. For the sleep-related breathing disorders, the general method would typically involve comparison to Polysomnography (PSG) as the gold standard, but this document focuses on the arrhythmia feature's integration, not re-validation of the entire device. Since the arrhythmia feature was inherited from the WP300, the ground truth establishment would have been part of the WP300 clearance.

    8. The sample size for the training set

    Not specified in this document. This submission is for enabling an existing algorithm from a predicate device (WP300) onto a different hardware platform (WP1), not for developing a new algorithm. Therefore, training set information is not relevant to this specific 510(k) summary.

    9. How the ground truth for the training set was established

    Not specified in this document, as it is not a new algorithm development and relies on the ground truth establishment from the predicate device (WP300) for the arrhythmia feature.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1