Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K070655

    Validate with FDA (Live)

    Date Cleared
    2007-04-02

    (24 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

    Device Description

    The WU'S Powered Wheelchair MAMBO 36 is an indoor / outdoor electric scooter that is battery operated. It has a base with four-wheeled with a seat, armrests, and a front basket. The movement of the scooter is controlled by the rider who uses hand controls located at the top of the steering column. The device can be disassembled for transport and is provided with an onboard battery charger.

    AI/ML Overview

    This is a 510(k) premarket notification for a Powered Wheelchair, which is a Class II medical device. The information provided describes the performance testing and comparison to a predicate device, as opposed to a study on software or AI performance. Therefore, many of the requested categories (such as sample size for test/training sets, expert qualifications, adjudication methods, MRMC studies, and standalone algorithm performance) are not applicable to this type of submission.

    Here's an analysis of the provided text in the context of your request:

    1. Table of acceptance criteria and reported device performance:

    Acceptance Criteria / Test MethodReported Device Performance (Mambo 36)
    EMC Report ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995 (Electrically powered wheelchairs, scooters, and their chargers - requirements and test methods)The device "passed" these tests. Specifically, the electronic systems use "the same suppliers" as the predicate device and "all passed by the UL certificated," assuring the "same safety level."
    Strength and fatigue tests (for mainframes)Mainframes of the two devices (Mambo 36 and Mambo 36X) "all meet the strength and fatigue tests".
    Resistance ignition test (for back upholstery materials)Back upholstery materials are the "same fabric and passed the resistance ignition test."
    Cruising range per charge20-30 miles (Improved compared to 10-15 miles for predicate device)
    Incline angle8 degrees (Improved compared to 5 degrees for predicate device)
    Weight capabilities130 kgs (Same as predicate device)
    Maximum speed6.4 km/hr (Same as predicate device)

    Notes on "Acceptance Criteria" and "Reported Performance":
    For a physical device like a powered wheelchair, the "acceptance criteria" are generally established by the referenced standards (e.g., ANSI/RESNA WC/Vol.2-1998 for performance, CISPR 11 for EMC). The document states that the new device passed these tests or met the requirements, which implies it met the acceptance criteria implicitly defined by those standards.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not applicable. This submission is for a physical device and references performance testing against established standards for a single device, not a data-driven study with a test set.
    • Data Provenance: Not applicable. The testing appears to be physical testing of the device itself, likely performed in Taiwan where the manufacturer is located.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. Ground truth as typically understood in AI/software validation (e.g., expert consensus on medical images) is not relevant for the performance testing of a powered wheelchair. The "ground truth" for a wheelchair's performance is adherence to engineering and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No expert adjudication method is mentioned or relevant for this type of device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a submission for a physical medical device, not an AI or software assistant designed to improve human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not applicable in the typical sense. The "ground truth" for this device's performance is compliance with established international and national standards for wheelchair safety and performance (e.g., ANSI/RESNA, CISPR, EN, IEC standards). This is determined through physical testing and measurement.

    8. The sample size for the training set:

    • Not applicable. This is not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. There is no training set mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K050010

    Validate with FDA (Live)

    Date Cleared
    2005-02-11

    (39 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

    Device Description

    The WU'S Powered Wheelchair MAMBO 36X is an indoor / outdoor electric scooter that is battery operated. It has a base with four-wheeled with a seat, armrests, and a front basket. The movement of the scooter is controlled by the rider who uses hand controls located at the top of the steering column. The device can be disassembled for transport and is provided with an onboard battery charger.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a powered wheelchair, the WU's POWERED WHEELCHAIR, MAMBO 36X. This document focuses on establishing substantial equivalence to a legally marketed predicate device, the TEH LIN JUPITER Powered Wheelchair TL-320 (K022697), rather than detailing acceptance criteria and performance data from a specific study designed to "prove the device meets acceptance criteria" in the way one might expect for a novel AI/software-as-a-medical-device (SaMD) product.

    Therefore, many of the requested categories (e.g., sample size for test set, number of experts, adjudication method, MRMC studies, ground truth details, training set size) are not applicable or cannot be extracted directly from this type of regulatory submission for a powered wheelchair.

    Here's an analysis based on the provided text, addressing the applicable points and explaining why others are not present:


    Acceptance Criteria and Study for WU's POWERED WHEELCHAIR, MAMBO 36X

    This 510(k) submission primarily focuses on demonstrating substantial equivalence to a predicate device, as opposed to a de novo approval or clinical trial for a novel device. Therefore, the "acceptance criteria" are largely based on meeting established standards and demonstrating similarity in safety and performance to the predicate device. The "study" referenced is the performance testing against these standards, which are not designed to measure metrics like sensitivity, specificity, or accuracy (which are common for AI/SaMD devices).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryCriterionReported Device Performance/Comparison to Predicate
    Intended UseSame as predicate device.The intended use between the two devices is the same: "to provide mobility to persons restricted to a seated position."
    Design/StructureMainframes are foldable, similar materials, same suspension (cross brace).Mainframes of both devices are foldable. Mainframe materials "all meet the strength and fatigue tests and they are similar for the material aspects." The suspension of the cross brace is the same.
    ComponentsSame type of armrest, same back upholstery material, same electronic systems.The two devices used the same type of armrest. Back upholstery material is the same fabric. Electronic systems (controller, batteries, charger) are from the "same suppliers" and "all passed by the UL certificated."
    Safety (Electrical/Mechanical)Complies with relevant electrical safety and wheelchair performance standards.Demonstrated through compliance with: - EMC Report ANSI / RESNA WC/Vol.2-1998: Standard for electrically powered wheelchairs, scooters, and their chargers - requirements and test methods. - CISPR 11: 1990: Industrial, scientific and medical (ISM) radio-frequency equipment – Electromagnetic disturbance characteristics – Limits and methods of measurement. - EN61000-3-2: 1995: Limits for harmonic current emissions. - IEC61000-3-3: 1995: Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems. - Electronic systems are UL certified.
    Minor Differences (Not impacting safety)Differences in overall dimensions, seat size, weight limit, speed, range, incline.The major differences existing are the overall dimension, the size of seat, weight limit, maximum speed, cruising range, and the incline degrees. The overall appearance differences are "not safety aspect" and do not affect substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable in the context of this 510(k) for a powered wheelchair. Performance testing refers to engineering and electrical safety standards applied to the device itself (e.g., a sample unit of the wheelchair), not a patient-based test set or data set in the manner of AI/SaMD.
    • Data Provenance: Not applicable. The "data" here are test results from engineering and electrical safety evaluations of the device's physical and electrical characteristics.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not applicable. Ground truth, in the AI/SaMD context, refers to a definitively correct label or outcome for a medical condition. For a powered wheelchair, "ground truth" relates to compliance with engineering standards, which is assessed through standardized testing and certification by notified bodies (e.g., UL, and compliance to standards like ANSI/RESNA).

    4. Adjudication Method for the Test Set

    • Not applicable. This concept is relevant for reconciling discrepancies in expert opinions on medical data, which is not part of this type of device submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • Not applicable. This is a powered wheelchair, not an AI-powered diagnostic or assistive tool for human readers (e.g., radiologists).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This device does not involve a standalone algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • Not applicable in the typical medical imaging/AI sense. The "ground truth" for this device is based on established engineering and electrical safety standards (e.g., ANSI/RESNA, CISPR, EN, IEC) as assessed through physical and electrical performance tests. Successful passing of these tests serves as the "truth" that the device meets the safety and performance requirements.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not employ machine learning or AI that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K030707

    Validate with FDA (Live)

    Date Cleared
    2003-04-04

    (29 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    0 - 999
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

    Device Description

    WU'S POWER WHEELCHAIR, MAMBO 3

    AI/ML Overview

    This document is a 510(k) premarket notification letter from the FDA for a power wheelchair. It is not a study report and therefore does not contain any information regarding acceptance criteria or the results of a study proving a device meets these criteria.

    The letter essentially states that the FDA has reviewed the submission for the "WU'S Power Wheelchair, MAMBO 3" and determined it is "substantially equivalent" to legally marketed predicate devices. This means it can be marketed, subject to general controls provisions.

    To directly answer your request:

    1. A table of acceptance criteria and the reported device performance: Not available in this document.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not available in this document.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not available in this document.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not available in this document.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not available in this document. This device is a power wheelchair, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not available in this document. This device is a power wheelchair.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not available in this document.
    8. The sample size for the training set: Not available in this document.
    9. How the ground truth for the training set was established: Not available in this document.

    In summary, this document is solely an FDA clearance letter and does not provide the detailed study information you are seeking.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1