510(k) Data Aggregation

    K Number
    K172735

    Validate with FDA (Live)

    Device Name
    WR19 System
    Manufacturer
    Date Cleared
    2018-04-17

    (218 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WR19 System is intended for prescription use in a health care facility or clinical research environment to acquire, transmit, display and store primarily EEG and optional auxiliary signals for adults and children, not including newborns.

    The WR19 System requires operation by a healthcare professional familiar with EEG.

    The WR19 System acquires, transmits, displays and stores electroencephalogram (EEG), and optionally electrocardiogram (ECG), accelerometer, photic sensor, external trigger signals and video.

    Device Description

    The WR19 System is primarily intended to acquire, transmit, display and store EEG and optionally do so for auxiliary signals. WR19 headset is designed to perform Routine or Outpatient EEG using 19 dry signal electrodes and 1 dedicated dry ground/driven-right-leg (DRL) electrode, adjusted and placed to comply with the 10-20 EEG system.

    The device consists of the following components:

    • . Headset
    • Electrodes (affixed to the underside of the headset) ●
    • Charger
    • Charging cable .
    • Software ●
      • Headset firmware -
      • Client application -
      • Data center application
    AI/ML Overview

    Here is an analysis of the acceptance criteria and supporting study for the WR19 System, based on the provided document:

    Acceptance Criteria and Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a single, clear format as might be desired for a device performance study. Instead, it describes performance testing outcomes and comparisons to a predicate device. The primary performance criteria are implicitly related to the ability of the WR19 System to acquire EEG signals comparably to a legally marketed predicate device (XLTEK EMU40EX) and meet established electrical and safety standards.

    Here's an attempt to synthesize the acceptance criteria and stated performance based on the "510(k) Summary" and "PERFORMANCE DATA" sections, particularly "8. CLINICAL PERFORMANCE TESTING":

    Acceptance Criteria CategorySpecific Criteria (Implicit or Explicit)Reported Device Performance (WR19 System)
    EEG Signal QualityComparable EEG signal quality to XLTEK EMU40EX (K163163)."The subject device was found to perform at least as well as the comparator device based on predefined acceptance criteria." Evaluated via: Qualitative waveform comparisons, Likert scoring, spectral correlation, and SNR comparisons.
    Setup TimeNot explicitly stated but evaluated."Results of time to setup... were analyzed to assess device performance." (Implied comparable or better than predicate, otherwise would be noted as a deficiency).
    Artifact and Error Generation HandlingAbility to handle sources of artifact and/or error.Healthy volunteers were "asked to generate sources of artifact and/ or error," and performance was assessed. (Implied satisfactory handling, as it contributed to the overall positive performance conclusion).
    Functionality of Optional Auxiliary SignalsVerification of functionality for infrared receiver/transmitter, external optical input, photic trigger detector, ECG input, accelerometer, video capture, contact quality detection.Testing confirmed functionality.
    Electrical SafetyConformance to ANSI / AAMI 60601-1:2005 + A1:2012."The WR19 System was evaluated and the device was found to conform to ANSI / AAMI 60601-1:2005 + A1:2012."
    Electromagnetic Compatibility (EMC)Conformance to IEC 60601-1-2:2014 and wireless coexistence testing per FDA guidance."The WR19 System was evaluated and found to conform to IEC 60601-1-2:2014. In addition, wireless coexistence testing was conducted, per FDA’s Guidance Document..."

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: 7 subjects.
        • 2 cohorts: EEG patients and healthy volunteers.
        • Recording durations: 15-30 minutes for 2 subjects (EEG patients and healthy volunteers) and 2 hours for 3 healthy volunteers.
      • Data Provenance: The study was conducted "at a United States hospital." This indicates prospective data collection for this specific study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number of experts used to establish ground truth or their qualifications. The phrase "qualitative waveform comparisons" suggests expert review, but details are absent.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • The document does not describe any specific adjudication method for the test set.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This was not an MRMC comparative effectiveness study involving AI assistance for human readers. The study compared the signal quality of the WR19 System (which is an EEG acquisition device) to another EEG acquisition device. There is no mention of AI in the context of improving human reader performance.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, in essence, but not in the context of an "AI algorithm." The "standalone" performance here refers to the device's ability to acquire and produce EEG signals. The clinical performance testing ("8. CLINICAL PERFORMANCE TESTING") evaluates the device's inherent signal quality and functionality (e.g., contact quality detection feature, various auxiliary signals). This is a standalone evaluation of the device's performance characteristics.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the clinical performance study was established by comparing the WR19 System's EEG signals to those obtained from a "common EEG device cleared in K163163 - XLTEK EMU40EX." This implies a comparison to an established, legally marketed predicate device's output considered to be the standard for EEG signal quality. The evaluation metrics included "qualitative waveform comparisons, Likert scoring, and spectral correlation and SNR comparisons." This suggests expert qualitative review and quantitative signal analysis relative to the predicate.
    7. The sample size for the training set:

      • The document does not mention a training set in the context of the clinical performance study. The WR19 System is an EEG acquisition device, not a diagnostic AI algorithm that typically requires a large training dataset. The software components are described as firmware, client application, and data center application, performing functions like presenting waveforms, controlling sessions, and standard EEG transformations (filters, montages), not complex pattern recognition that would necessitate a traditional "training set."
    8. How the ground truth for the training set was established:

      • Not applicable as no training set (in the AI/machine learning sense) is described for the clinical performance study.