510(k) Data Aggregation

    K Number
    K023917

    Validate with FDA (Live)

    Date Cleared
    2003-02-26

    (93 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Antineedle Stick Protector is an integral, active safety device intended to minimize accidental needle stick injuries when used to shield needles for Kawasumi products manufactured with a wing and needle assembly for solution infusion and blood sampling.

    The Small Vein Infusion Set with Antineedle Stick Protector is a single use, sterile device used for peripheral venous access for solution administration or blood collection. The Antineedle Stick Protector is an integral, active safety device intended to minimize accidental needle stick injuries when used to shield needles.

    The Winged Collection Set with Multi-Sample Luer Adapter and Antineedle Stick Protector is a single use, sterile, blood collection device used for peripheral venous access for blood sampling. The Antineedle Stick Protector is an integral, active safety device intended to minimize accidental needle stick injuries when used to shield needles.

    Device Description

    The Antineedle Stick Protector is a polymeric device designed to be used integral with winged needle sets and shields the needle when the needle with hub and wing assembly is removed from the patient

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided K023917 510(k) Summary, the acceptance criteria are implicitly tied to the performance of the predicate device and the prevention of needlestick injuries. The document does not explicitly state quantitative acceptance criteria with specific thresholds (e.g., "99% activation rate"). Instead, it focuses on achieving the same results as the predicate device regarding needle tip protection.

    Acceptance Criteria (Implied)Reported Device Performance
    Minimize accidental needlestick injuriesSuccessful simulated use study
    Shield the needle effectively after useNeedle tip is protected inside the device after use
    Substantially equivalent to predicate devices in functionAchieves the same results as predicate device (Becton Dickinson Vacutainer Brand Safety-Lok Blood Collection Set)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not specified. The document mentions "a successful simulated use study" but does not provide details on the number of units or uses tested.
    • Data Provenance: The study was conducted by Kawasumi Laboratories, Inc., located in Tokyo, Japan. The study is described as a "simulated use study," implying it was conducted in a controlled environment, likely within their facility. It is a prospective study as it was conducted specifically to demonstrate the device's performance for this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the given text. The study is described as a "simulated use study" but does not mention the involvement of experts in establishing ground truth or evaluating the results. The determination of "success" likely relied on internal assessment criteria by Kawasumi Laboratories.

    4. Adjudication Method for the Test Set

    This information is not provided in the given text. There is no mention of independent reviewers or an adjudication process for the "simulated use study."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically associated with diagnostic imaging devices or algorithms where human readers interpret medical images with or without AI assistance. The device in question is a medical device designed to prevent needlestick injuries, and such a study would not be applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable to the device described. The device is a physical product (winged needle set with an antineedle stick protector), not an algorithm or software. Its performance inherently involves human interaction during its use to activate the protector.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance would likely be a direct observation of whether the needle tip was successfully shielded or retracted after simulated use, thereby preventing accidental needlestick. The text states: "In both devices, the needle tip is protected inside the device after use." This implies the ground truth for success was the successful physical enclosure of the needle tip.

    8. The Sample Size for the Training Set

    This question is not applicable as there is no "training set" for this type of physical medical device. Training sets are relevant for machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no "training set."