    K Number
    K030838

    Validate with FDA (Live)

    Device Name
    WARTNER PRO
    Date Cleared
    2003-06-27

    (102 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    WARTNER PRO® is intended for the treatment of common warts and plantar warts in adults and children over the age of 4. Common warts are recognized by the rough 'cauliflower-like' appearance ci the surface. The plantar wart is recognized by its location only on the bottom of the foot, its tenderness, and the interruption of the footprint pattern. The product is intended as an OTC device, which will be marketed to professionals, allied professionals, and semi-professionals.

    Device Description

    WARTNER PRO® is a cryosurgical system for the treatment of warts. It consists of

    • A canister filled with 125 ml of a liquid mixture of the compressed gases . dimethyl ether and propane. This gas mixture does not harm the ozone layer, and has a four-year shelf life.
    • 42 Foam applicators in three different sizes (6 mm, 7 mm, 8 mm) .
    • An applicator stick/key for holding a foam applicator which is required to . dispense the liquid to the applicator, and is held by the person during treatment
    • . Information booklet
      The pressurized canister is filled with dimethyl ether and propane, which acts as a cryogen. The key is fitted with a foam applicator, which is inserted into the safety valve opening in the canister. The valve releasing the cryogen can only be actuated when the key is in place and depressed. When the key is depressed it saturates the foam applicator with the cold cryogenic gases and lowers the temperature of the applicator to approximately -50℃. The foam applicator is then placed on the wart. The freezing kills the wart and the infected skin causing the wart to fall off within a few days.
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the WARTNER PRO® device:

    The provided text is a 510(k) summary for a cryosurgical system called WARTNER PRO®. It focuses on demonstrating substantial equivalence to predicate devices rather than detailing independent clinical studies with specific performance metrics against pre-defined acceptance criteria. Therefore, many of the requested elements for a typical clinical study, such as specific acceptance criteria for efficacy or safety, effect sizes from MRMC studies, or detailed ground truth establishment, are not explicitly present.

    However, based on the information provided, we can infer some "acceptance criteria" through the lens of performance equivalence to the predicate device and the laboratory testing described.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from Predicate Equivalence and Lab Testing)Reported Device Performance
    Cryogen Composition: Identical to predicate device (dimethyl ether and propane).WARTNER PRO® uses a cryogen composed of dimethyl ether and propane. The Wartner® Wart Removal System OTC predicate device cryogen is identical.
    Applicator Material: Identical to predicate device (foam material).The material used to transfer the cold to the patient in both WARTNER PRO® and Wartner® Wart Removal System OTC use is the same foam material.
    Applicator Effectiveness Temperature: Maintains temperature of less than minus 50°C.The average temperature of the applicator surface after saturation is -56.4°C. The WARTNER PRO® foam applicator maintains a temperature of less than minus 50°C. The Wartner® Wart Removal System OTC predicate device effectiveness is identical (implies it also maintains < -50°C).
    Safety Valve Mechanism: Incorporates a safety valve that cannot be actuated unless the foam applicator and applicator stick/key are in place and depressed, identical to predicate device.The design of both WARTNER PRO® and Wartner® Wart Removal System OTC incorporate the use of a safety valve that cannot be actuated unless the foam applicator and applicator stick/key are in place and depressed.
    Intended Use: For treatment of common warts and plantar warts in adults and children over 4.WARTNER PRO® is intended for the treatment of common warts and plantar warts in adults and children over the age of 4. This expands on the primary predicate, which only covers common warts.
    Labeling Safety & Warnings: Identical to predicate device for common warts, comparable to other predicate devices for plantar warts.The safety and warning statements for WARTNER PRO® and the primary predicate device (Wartner® Wart Removal System for OTC use) are identical except for any reference to plantar warts. The WARTNER PRO® safety and warning statements for the plantar wart indication have been developed using the other labeling predicates to ensure that they are comparable.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes laboratory testing and comparison to predicate devices rather than a clinical human subject test set for efficacy.

    • Test Set Sample Size: Not applicable in the context of a clinical study. For the laboratory testing, while it mentions "Several lengths, sizes, and shapes of foam were tested," a specific quantitative sample size for this internal testing is not provided.
    • Data Provenance: The data appears to be internal laboratory testing data from Wartner Medical Products. No information is given about the country of origin of subjects/data (as it's not a clinical study) or if it's retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The filing is a 510(k) based on substantial equivalence, relying on physical and functional comparisons to predicate devices and internal lab measurements, not on clinical expert consensus for ground truth on patient outcomes.


    4. Adjudication Method for the Test Set

    This information is not applicable as there is no mention of a test set requiring adjudication in the context of a clinical study or human subject data.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done or described in this 510(k) summary. This type of study comparing human readers with and without AI assistance is not relevant to a cryosurgical device for wart removal, which is a direct treatment device, not an imaging or diagnostic aid.


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This device is a cryosurgical tool, not an algorithm or AI system. Its performance is related to its physical properties (temperature, cryogen, applicator design) and its ability to deliver the therapy as intended, not an algorithm's standalone performance.


    7. Type of Ground Truth Used

    The "ground truth" for the device's technical specifications and equivalence is established through:

    • Laboratory Measurements: For physical characteristics like applicator temperature (-56.4°C).
    • Chemical and Material Analysis: Confirming identical cryogen composition and applicator foam material to the predicate device.
    • Design Comparison: Verifying similarity in application method, safety features, and overall design to predicate devices.
    • Labeling Comparison: Ensuring comparable instructions for use, warnings, and indications, particularly by referring to existing predicate devices for the plantar wart indication.

    8. Sample Size for the Training Set

    There is no mention of a "training set" in the context of this device. A training set is typically used for machine learning models, which is not applicable here. The device's design and characteristics were developed through engineering processes and laboratory testing, rather than a data-driven training process.


    9. How the Ground Truth for the Training Set Was Established

    Since there is no training set mentioned or implied for this device, this question is not applicable.