510(k) Data Aggregation

    K Number
    K053131

    Validate with FDA (Live)

    Date Cleared
    2005-12-30

    (52 days)

    Product Code
    Regulation Number
    862.1110
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indication of serum and plasma bilirubin is useful in the screening of liver function disorders or in the diagnosis of jaundice.

    Device Description

    Wako Total Bilirubin V is based on a chemical oxidation method, utilizing vanadate as an oxidating agent, shows good correlation with conventional methods, practically no interference by coexistent serum and plasma substances, and is convenient ready-to-use liquid type reagent. When a sample is mixed with the reagent containing the detergent and the vanadate, at around pH3, total bilirubin in the sample is oxidized to biliverdin. This causes the absorbance of yellow, specific to bilirubin, to decrease. Therefore, the total bilirubin concentration in the sample can be obtained by measuring the adsorbances before and after the vanadate oxidation.

    AI/ML Overview

    This submission is for a reagent, the Wako Total Bilirubin V assay, and thus the acceptance criteria and study design are related to the analytical performance of the assay rather than a diagnostic device that requires expert consensus or advanced imaging.

    Here's an analysis of the provided text in relation to your request:

    Acceptance Criteria and Device Performance

    The provided text does not explicitly state specific numerical acceptance criteria for the Wako Total Bilirubin V assay. Instead, it relies on demonstrating "substantial equivalency" to a previously marketed device (510(K) # 970985).

    The reported device performance, in terms of meeting implicit acceptance criteria, is based on the following:

    • Substantial Equivalency: The key claim is that "The safety and effectiveness of the Wako Total Bilirubin V assay is demonstrated by its substantial equivalency to our previous Total Bilirubin assay (510(K) # 970985)."
    • Correlation with Conventional Methods: The device "[Wako Total Bilirubin V] shows good correlation with conventional methods."
    • Interference by Coexistent Serum and Plasma Substances: It demonstrates "practically no interference by coexistent serum and plasma substances."
    • Convenience and Stability: It is also "convenient ready-to-use liquid type reagent."

    Table of Acceptance Criteria and Reported Device Performance (Inferred):

    Acceptance Criteria Category (Inferred)Specific Criteria (Inferred/Implicit)Reported Device Performance
    Accuracy/ComparabilitySubstantial equivalency to predicate device (510(K) # 970985) for serum samples; good correlation with conventional methods.Meets: Demonstrated substantial equivalency to predicate device. Shows good correlation with conventional methods (diazo coupling, bilirubin oxidase enzymatic).
    InterferenceMinimal interference from coexistent serum and plasma substances.Meets: "practically no interference by coexistent serum and plasma substances."
    Stability/Usability"Convenient ready-to-use liquid type reagent." (This is more a design feature than a performance criterion, but contributes to overall assessment.)Meets: Described as "convenient ready-to-use liquid type reagent." This implies stability and ease of use, addressing a disadvantage of older methods (unsatisfactory stability of reagents after preparation).
    New Sample Type PerformancePerformance for plasma samples is equivalent to its established performance for serum samples and to the predicate device for serum.Meets: The submission "adds the use of plasma as a sample" with "no changes to performance claims already established in 510(k) # 970985" for serum, implying equivalent performance for plasma.

    Study Details

    Given the nature of the submission (510(k) for an in vitro diagnostic reagent, primarily adding a sample type), the study described is an analytical validation comparing the new device to a predicate and conventional methods. It is not a clinical study involving human readers or extensive ground truth establishment in the way a diagnostic imaging device would be studied.

    1. Sample Size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated in the provided text.
      • Data Provenance: Not explicitly stated, but typical for such submissions would be laboratory-based analytical studies using human serum and plasma samples. It does not specify country of origin or whether it's retrospective or prospective, but these are generally prospective analytical studies for IVD reagents.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the sense of expert human interpretation of images/data. The "ground truth" (or reference values) would be established by the "conventional methods" mentioned (diazo coupling, bilirubin oxidase enzymatic method) which are laboratory-based chemical analyses, often performed by trained medical technologists or clinical chemists. No specific number or qualifications of individual "experts" are provided as this is a chemical assay validation.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This is an analytical performance study of a chemical reagent, not a diagnostic interpretation study requiring adjudication of expert opinions.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI-assisted diagnostic device, nor is it a study involving human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is inherently a "standalone" device in the sense that it is a chemical reagent that performs its function without a human-in-the-loop interpretative step once the sample is added. However, the term "standalone" usually refers to AI algorithms. In the context of IVDs, this refers to the assay's analytical performance on its own. The description implies such a standalone analytical validation was performed.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for bilirubin measurement is established through reference laboratory methods, specifically the "conventional methods" like the diazo coupling method and the bilirubin oxidase enzymatic method, which are well-established chemical assays for bilirubin quantification.
    7. The sample size for the training set:

      • Not applicable. This is a chemical reagent, not a machine learning algorithm that requires a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable. (See #7).
    Ask a Question

    Ask a specific question about this device

    K Number
    K970985

    Validate with FDA (Live)

    Date Cleared
    1997-04-21

    (34 days)

    Product Code
    Regulation Number
    862.1110
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Not Found

    Device Description

    When a sample is mixed with the reagent containing the detergent and the vanadate, at around pH 3, total bilirubin in the sample is oxidized to biliverdin. This causes the absorbance of yellow, specific to bilirubin, to decrease. Therefore, the total bilirubin concentration in the sample can be obtained by measuring the absorbance before and after the vanadate oxidation.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Wako Total Bilirubin V assay, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit)Reported Device Performance
    Correlation with predicate deviceCorrelation coefficient of 0.997; regression equation y = 1.003x + 0.09
    Precision (day-to-day)Precision studies indicate acceptable values
    Minimum detectable level0.03 mg/dL
    LinearityLinear to 40 mg/dL
    Absence of interference by coexistent substancesPractically no interference by coexistent substances (claimed)

    Study Details

    1. Sample size used for the test set and the data provenance:

      • The document states that "serum samples" were used for comparison studies against the predicate assay. The exact number of samples is not specified.
      • The data provenance (country of origin, retrospective or prospective) is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not applicable as the study compares a new device to a predicate device, not against a clinical expert's interpretation. The "ground truth" for the test set is established by the results from the predicate device (Wako's previous Total Bilirubin assay, K912024/A).
    3. Adjudication method for the test set:

      • Not applicable. This is a comparison of quantitative measurements between two devices, not a qualitative assessment requiring adjudication of human interpretations.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a study comparing two laboratory diagnostic devices for bilirubin measurement, not a study involving human readers or AI assistance.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes. This study describes the performance of a standalone laboratory assay (Wako Total Bilirubin V) in comparison to another standalone laboratory assay (the predicate device). There is no "human-in-the-loop" component in the direct measurement process being evaluated.
    6. The type of ground truth used:

      • The "ground truth" for the comparative study was the results obtained from the predicate device (Wako's previous Total Bilirubin assay, 510(k)#K912024/A).
    7. The sample size for the training set:

      • This concept is not applicable to this type of device and study. The Wako Total Bilirubin V assay is a chemical measurement method; it does not involve machine learning algorithms that require a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this device.