510(k) Data Aggregation

    K Number
    K993468

    Validate with FDA (Live)

    Date Cleared
    1999-12-16

    (64 days)

    Product Code
    Regulation Number
    862.1100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AST/GOT measurements are used in the diagnosis and treatment of certain types of liver and heart disease.

    Device Description

    The Wako L-Type GOT.J2 test is an in vitro assay for the quantitative determination of glutamic-0xaloacetic transaminase (GOT) activity in serum.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device PerformanceStudy Type
    Correlation0.9996 correlation coefficient against predicateComparative Study
    Regression Equationy = 1.055x + 2.38 against predicateComparative Study
    PrecisionAcceptable values obtained on a day-to-day basisPrecision Studies
    Minimum Detectable Level4 IU/LAnalytical Performance

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document refers to "comparison studies," but the number of samples or patients included in these studies is not provided.
    • Data Provenance: Not explicitly stated. There is no mention of the country of origin or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable/not stated. The ground truth for this diagnostic device (Wako L-Type GOT.J2 test) is established by comparing its performance against a predicate device (Wako UA 30R GOT product) and through analytical precision studies, not through expert consensus on interpretation.
    • Qualifications of Experts: Not applicable/not stated.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. As the comparison is analytical (device vs. predicate device), an adjudication method for human interpretation is not relevant here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) for quantitative determination of GOT activity, not an imaging device or a device requiring human interpretation of complex outputs in the same way an AI-assisted diagnostic tool might. The comparison is between two analytical assays.
    • Effect Size: Not applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: Yes, the described performance is a standalone performance of the L-Type GOT.J2 test. As an in vitro diagnostic assay, its performance (correlation, regression, precision, minimum detectable level) is inherent to the assay itself and does not involve a human-in-the-loop for its primary function of measuring GOT activity.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The ground truth is indirectly established by comparison to a legally marketed predicate device (Wako UA 30R GOT product) for which safety and effectiveness have already been demonstrated. The predicate device's results serve as the reference standard for the "correlation" and "regression equation" studies. For "precision," the ground truth is the inherent reproducibility of the assay itself.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable/not stated. This document describes a traditional in vitro diagnostic assay, not a machine learning or AI-based device that would typically involve a "training set" in the common sense. The development of such assays involves reagent formulation and analytical validation, not data-driven model training.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not applicable. As mentioned above, this document does not describe an AI/ML development process with a training set and corresponding ground truth.