    K Number
    K202818

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-12-16

    (448 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Voyant Fine Fusion device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open procedures, including head and neck procedures, where the ligation and division of vessels and tissue bundles is desired.

    The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

    The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

    Device Description

    The Voyant Fine Fusion device is an advanced bipolar instrument that uses RF energy, provided by the Voyant Electrosurgical Generator (K182244), to seal vessels up to and including 7mm in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The device features a mechanical, user-actuated blade for the division of sealed tissue.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Voyant Fine Fusion Device (EB230), based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance CriteriaReported Device Performance
    EMC, Electrical Safety, and MechanicalIn accordance with relevant IEC 60601 series standards; continued performance over multiple device activations.Met all acceptance criteria. Simulated repeated-use testing on predicate sufficient.
    System Testing (Burst Pressure)Met predetermined acceptance criteria.Subject device met predetermined acceptance criteria and substantially equivalent to predicate.
    System Testing (Thermal Spread)Met predetermined acceptance criteria.Subject device met predetermined acceptance criteria and substantially equivalent to predicate.
    Animal Testing (Acute)Demonstrated safety and efficacy for thermal damage and cellular viability.Confirmed substantially equivalent performance to the predicate device.
    Animal Testing (Chronic Stability)Demonstrated long-term seal quality, device performance, and no adverse effects on adjacent structures.Met predetermined acceptance criteria. Confirmed substantially equivalent performance to the predicate device.
    Software VerificationEvaluated design, implementation, and performance of device software script.All levels of software testing conducted (unit, integration, system).
    BiocompatibilityMet all acceptance criteria per ISO 10993-1, 10993-5, 10993-10, 10993-11.Met all acceptance criteria for Cytotoxicity, Intracutaneous Reactivity, Sensitization, and Acute Systemic Toxicity.

    2. Sample Size Used for the Test Set and Data Provenance

    • System Testing (Burst Pressure): Not explicitly stated, but "vessels representative of the devices' indications" were used.
    • System Testing (Thermal Spread): Not explicitly stated.
    • Acute Animal Study: "Large porcine animal models" were used. The number of animals is not specified.
    • Chronic Survival Study: "Large porcine animal models" were used. The number of animals is not specified.
    • Data Provenance: The animal studies were conducted using porcine models, indicating prospective animal study data. The location of these studies is not specified, but the submission is to the U.S. FDA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Acute Animal Study: "Functional staining techniques" were used, and "Samples were sent to a qualified histopathology laboratory for evaluation." The number and specific qualifications of the histopathologists are not provided.
    • Chronic Survival Study: "Vessels representative of the devices' indications were sealed and evaluated by a qualified histopathologist for hemostasis and signs of hematoma." The number and specific qualifications of the histopathologists are not provided.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method for conflicting expert opinions. The evaluations were performed by "qualified histopathologists," implying a certain level of expertise, but no mention of multiple reviewers or a consensus process.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study was mentioned. The studies focus on device performance characteristics rather than human reader improvement with AI assistance. This device is an electrosurgical tool, not an AI-assisted diagnostic or imaging device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable as the Voyant Fine Fusion device is an electrosurgical instrument for direct surgical use, not an AI algorithm. The software mentioned is for the device's internal control, not for diagnostic or analytical tasks that would typically involve standalone algorithm performance evaluation in the context of AI.

    7. Type of Ground Truth Used

    • System Testing (Burst Pressure & Thermal Spread): Objective measurements (burst pressure values, thermal spread measurements) of physical properties served as the ground truth.
    • Animal Studies (Acute & Chronic): Histopathological evaluation of tissue (cellular viability, hemostasis, signs of hematoma) by qualified histopathologists served as the ground truth for biological effects.

    8. Sample Size for the Training Set

    No training set is mentioned in the context of this device. The software verification refers to the internal software script of the device, not an AI model that undergoes training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for an AI model is mentioned for this electrosurgical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153017

    Validate with FDA (Live)

    Date Cleared
    2016-07-06

    (265 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Voyant Fine Fusion device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open procedures, including head and neck procedures, where ligation of vessels and tissue bundles is desired.

    The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

    The device has not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures.

    Device Description

    The Applied Medical Voyant Fine Fusion instrument is designed for use with the Voyant ESG (cleared in K141288). This device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 7mm in diameter and in head and neck procedures. The device may also be used to seal tissue bundles that can be captured in the device jaws. The device features a mechanical, user-actuated blade for the division of sealed tissue.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Voyant Fine Fusion electrosurgical device. It details the device’s indications for use, its substantial equivalence to predicate devices, and the performance testing conducted.

    Here’s a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" and "reported device performance" in a quantitative format for clinical effectiveness. Instead, it describes various tests and states whether the device "passed" or demonstrated "substantial equivalence" to a predicate device.

    However, based on the Discussion of Performance Testing section, we can infer some criteria and outcomes:

    Test CategoryAcceptance Criteria (Inferred)Reported Device Performance
    BiocompatibilityCompliance with ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Intracutaneous Reactivity, Sensitization), ASTM F2148 (Sensitization)"The device passed all biocompatibility testing according to the acceptance criteria contained in the respective standards."
    SterilizationCompliance with ISO 11137-2 and AAMI TIR 33"Validation of the sterilization process for this device was performed according the recognized consensus standards ISO 11137-2 and AAMI TIR 33." The packaging, sterilization method, and shelf life are "identical to that of the previously cleared Voyant Fine Fusion device (K143536)."
    EMCCompliance with applicable sections of IEC 60601-1-2 and IEC 60601-2-2"There have been no changes to the subject device that impact its electromagnetic compatibility (EMC) as presented in K143536. The device was designed and previously tested according to the applicable sections of IEC 60601-1-2 and IEC 60601-2-2."
    Electrical SafetyCompliance with applicable sections of IEC 60601-1 and IEC 60601-2-2"An electrical safety evaluation... was been conducted according to the applicable sections of IEC 60601-1 and IEC 60601-2-2 and presented in K143536." "Relevant tests from these standards were repeated to confirm the safety of the updated button design." (Implies compliance was confirmed).
    Preclinical BenchPerformance substantially equivalent to the LigaSure predicate for vessels up to 7mm and head and neck procedures, including seal evaluation and burst pressures."Preclinical bench (ex vivo porcine vessels/tissue) and animal (in vivo ovine model) studies were performed to evaluate system safety and to demonstrate that the subject device performance is substantially equivalent to the predicate device for the expanded indications including vessels up to 7mm in diameter and head and neck procedures." Specific results included "Seal evaluation" and "Burst pressures of arteries and veins."
    Preclinical AnimalPerformance substantially equivalent to the LigaSure predicate for vessels up to 7mm and head and neck procedures, including seal evaluation, thermal damage, and thermal spread metrics; chronic hemostasis and no hematomas."Preclinical bench (ex vivo porcine vessels/tissue) and animal (in vivo ovine model) studies were performed to evaluate system safety and to demonstrate that the subject device performance is substantially equivalent to the predicate device for the expanded indications including vessels up to 7mm in diameter and head and neck procedures." "Acute animal studies (in vivo) [evaluated] Seal evaluation and Thermal damage and thermal spread (Minimum, maximum, median, mode)." "In addition, chronic studies (in vivo ovine and canine model) were conducted using the subject device to evaluate long term seal quality and resulted in chronic hemostasis and no signs of hematomas."

    Key takeaway: The primary "acceptance criteria" for this 510(k) submission is showing substantial equivalence to existing, legally marketed predicate devices, particularly the LigaSure Curved, Small Jaw, Open Sealer/Divider (K113572) and the previous Voyant Fine Fusion (K143536). Performance is demonstrated by comparing the subject device against these predicates in a laboratory and animal setting.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set:
      • Bench (ex vivo): "porcine vessels/tissue" were used, but a specific number is not provided.
      • Acute Animal (in vivo): "ovine model" was used, but a specific number of animals or vessels is not provided.
      • Chronic Animal (in vivo): "ovine and canine model" were used, but a specific number of animals is not provided.
    • Data Provenance: The studies were preclinical (ex vivo and in vivo animal models). They are performed in a laboratory setting, not from human patients. The specific country of origin for the data is not provided in this summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This type of information is typically not applicable to preclinical bench or animal studies where objective measurements (e.g., burst pressure, thermal spread measurements, histological examination by veterinary pathologists) establish "truth" rather than expert consensus on imaging or clinical findings. The document does not mention the use of human experts to establish ground truth for the test set in the context of diagnostic performance.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a human clinical trial involving subjective assessments that would require adjudication. Preclinical studies rely on objective measurements and established scientific methodologies.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "Clinical: This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an electrosurgical instrument, not an AI or imaging algorithm. Therefore, the concept of "standalone performance" in the context of an algorithm does not apply. The device's performance is intrinsically linked to its operation by a human surgeon.

    7. The Type of Ground Truth Used

    • For Biocompatibility, Sterilization, EMC, Electrical Safety: Ground truth is established by adherence to recognized consensus standards (e.g., ISO 10993 series, ISO 11137-2, AAMI TIR 33, IEC 60601 series).
    • For Preclinical Bench (ex vivo): Ground truth is established through objective measurements such as seal evaluation methods and quantitative burst pressure measurements of vessels.
    • For Preclinical Animal (in vivo): Ground truth is established through objective measurements and observations including seal evaluation, thermal damage and spread metrics (minimum, maximum, median, mode), and histological assessment for chronic hemostasis and lack of hematomas. Comparison against the predicate device's performance also serves as a benchmark.

    8. The Sample Size for the Training Set

    Not applicable. The Voyant Fine Fusion is a physical electrosurgical device, not a machine learning model. Therefore, there is no "training set" in the AI/ML sense. "Training" would refer to the engineering design and iteration process, not data-driven model training.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no "training set" for an AI/ML model for this device. Ground truth in the context of product development for this type of device would relate to engineering specifications, material properties, and functional requirements, all established through standard engineering and scientific principles.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143536

    Validate with FDA (Live)

    Date Cleared
    2015-04-16

    (122 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Voyant Fine Fusion is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open procedures where the ligation and division of vessels and tissue bundles is desired.

    The device can seal and divide vessels up to and including 3mm in diameter and tissue bundles that can be captured in the jaws of the device.

    The device has not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures.

    Device Description

    The Applied Medical Voyant Fine Fusion instrument is designed for use with the Voyant ESG (cleared in K141288). This device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 3mm in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The device features a mechanical, useractuated blade for the division of sealed tissue.

    AI/ML Overview

    Based on the provided text, the acceptance criteria and study details for the Voyant Fine Fusion device can be summarized as follows:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Electrical SafetyCompliance with IEC 60601-1 and IEC 60601-2-2.Demonstrated compliance.
    Mechanical StrengthCompliance with IEC 60601-1.Tested and demonstrated.
    DurabilityNot explicitly defined, but implied through durability testing.Tested and demonstrated.
    Destructive TestingCompliance with IEC 60601-2-2.Tested and demonstrated.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2.Tested and demonstrated.
    Safety SystemsNot explicitly defined, but implied through safety systems testing.Tested and demonstrated.
    Seal Evaluation (Bench/Ex Vivo)To show substantial equivalence to the predicate device for vessels up to 3mm. Specific metrics (e.g., burst pressure threshold) are not detailed but are implied.Performed a side-by-side comparison against the predicate device to evaluate seal quality. Conclusion: Substantially equivalent seal performance to the predicate device for vessels up to and including 3mm.
    Thermal Damage (Bench/Ex Vivo)To show substantial equivalence to the predicate device. Specific metrics (e.g., thermal spread limits) are not detailed but are implied.Performed a side-by-side comparison against the predicate device to evaluate thermal damage. Conclusion: Substantially equivalent local tissue effects to the predicate device.
    Burst Pressures (Bench/Ex Vivo)To show substantial equivalence to the predicate device. Specific thresholds are not detailed.Tested comparing the subject device to the predicate device. Conclusion: Substantially equivalent vessel sealing performance (implied by burst pressure testing) to the predicate device.
    Seal Evaluation (Acute Animal Study)To show substantial equivalence to the predicate device.Performed a side-by-side comparison against the predicate device in an in vivo canine model to evaluate seal quality. Conclusion: Substantially equivalent seal performance to the predicate device.
    Thermal Damage (Acute Animal Study)To show substantial equivalence to the predicate device.Performed a side-by-side comparison against the predicate device in an in vivo canine model to evaluate thermal damage. Conclusion: Substantially equivalent local tissue effects to the predicate device.
    Seal Quality & Chronic Hemostasis (Chronic Animal Study)To demonstrate efficacy and absence of hematomas over time.Conducted a chronic study in an in vivo canine model to evaluate seal quality and chronic hemostasis, with no signs of hematomas.
    BiocompatibilityCompliance with ISO 10993, FDA Blue Book Memorandum #G95-1, and 2013 FDA Draft Guidance.Device passed all biocompatibility testing according to acceptance criteria in respective standards.
    SterilizationValidated to an SAL (Sterility Assurance Level) of 10⁻⁶ in accordance with ISO 11137-2 and AAMI TIR 33.The Fine Fusion Device is validated to an SAL of 10⁻⁶.
    Shelf LifeMaintenance of sterility and protection from damage throughout shelf life (via ship testing, environmental conditioning).Testing conducted on aged and conditioned units met all acceptance criteria, ensuring sterility and protection from damage.

    2. Sample Sizes Used for the Test Set and the Data Provenance

    • Test Set Sample Sizes:
      • Bench (Ex Vivo): "ex vivo porcine vessels/tissue" - specific number of vessels/tissue samples not provided.
      • Acute Animal Study (In Vivo): "in vivo canine model" - specific number of animals not provided.
      • Chronic Animal Study (In Vivo): "in vivo canine model" - specific number of animals not provided.
    • Data Provenance:
      • The bench testing used "porcine vessels/tissue," indicating animal tissue.
      • The animal studies (acute and chronic) used an "in vivo canine model," indicating animal data.
      • The studies appear to be prospective as they were specifically performed to evaluate the new device and compare it against the predicate.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not provide information on the number of experts, their qualifications, or their involvement in establishing ground truth for the performance testing. The "ground truth" for these tests seems to be defined by the direct measurements and observations from the experimental procedures (e.g., burst pressure, thermal spread, visual assessment of seals and hematomas).

    4. Adjudication Method for the Test Set

    The document does not describe any expert adjudication methods (e.g., 2+1, 3+1). The evaluation criteria suggest direct measurement and comparison rather than subjective human assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an electrosurgical tool, not an AI-assisted diagnostic or imaging device, so such a study would not be applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The Voyant Fine Fusion is a physical electrosurgical device, not an algorithm. Its performance is evaluated by its physical characteristics and its effect on tissue, not by an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for the performance testing was primarily based on:

    • Direct physical measurements: Such as mechanical strength, electrical safety parameters, burst pressures.
    • Histological/Pathological observations: For thermal damage, seal quality, and absence of hematomas in animal tissue.
    • Biological responses: Observed in chronic animal studies for hemostasis.

    8. The Sample Size for the Training Set

    There is no mention of a training set. This device is not an AI/ML-based device; therefore, the concept of a "training set" for an algorithm does not apply.

    9. How the Ground Truth for the Training Set was Established

    As there is no training set for an algorithm, this question is not applicable.