    K Number
    K231500

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-08-07

    (75 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

    · medical devices in a single or double pouch configuration*

    · trays containing medical devices in a single or double pouch configuration

    · small items requiring surface sterilization in a single pouch configuration within a tray

    *3-D printed items should not be double pouched.

    to be sterilized in the:

    · Lumen, Non Lumen, Flexible, Fast Non Lumen, Fast and Specialty Cycles of the V-PRO® Low Temperature Sterilization Systems

    • · Default Cycle of the STERRAD** 100S Sterilizer
    • · Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear** Technology Sterilizers

    · Express, Standard, Flex Scope and DUO Cycles of the STERRAD 100NX with ALLClear Technology Sterilizers

    **STERRAD and ALLClear are trademarks of Advanced Sterilization Products

    NOTE: Trays must be legally marketed for use in the V PRO Low Temperature or STERRAD Sterilization Systems and contain a vent surface area to tray volume ratio ≥ 0.135 in-1 with the maximum number of instrument organizers installed.

    The pouches maintain the sterility of the enclosed devices until used.

    When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

    V-PRO 60 & s2 Lumen Cycle

    · Non-lumened instruments and instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

    · Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

    · Medical devices, including single, dual and triple channeled rigid endoscopes, with the following configurations:

    o single or dual lumen devices

    0.77 mm internal diameter (ID) and < 410 mm in length

    ≥ 1.8 mm ID x ≤ 542 mm in length

    o triple lumen devices

    ≥1.2 mm ID and ≤ 275 mm in length

    ≥1.8 mm ID and ≤ 310 mm in length

    or

    ≥2.8 mm ID and ≤ 317 mm in length

    V-PRO 60 & s2 Non Lumen Cycle

    •Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

    V-PRO 60 & s2 Flexible Cycle

    •Load 1: One flexible surgical endoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a:

    o single or dual lumen device with lumens that are ≥ 1 mm ID and ≤ 990 mm in length

    •Load 2: Non-lumened instruments including non-lumened rigid, and flexible endoscopes and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps or scissors. Medical devices, including rigid and semi-rigid endoscopes, with the following configurations:

    ○ ≥ 0.76 mm ID and ≤ 233 mm in length

    o ≥ 1.0 mm ID and ≤ 254 mm in length

    o ≥ 1.8 mm ID and ≤ 542 mm in length

    V-PRO s2 Fast Cycle

    o Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes, and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors

    o Medical devices (including single, dual and triple channeled rigid endoscopes) with the following configurations:

    o Single or dual lumen devices

    ≥ 0.77 mm ID and ≤ 410 mm in length

    ≥ 1.8 mm ID and ≤ 542 mm in length

    o triple lumen devices

    ≥1.2 mm ID and ≤ 275 mm in length

    ≥1.8 mm ID and ≤ 310 mm in length

    Or

    ≥2.8 mm ID and ≤ 317 mm in length

    V-PRO 1, 1 Plus, maX & maX 2 Lumen Cycle

    • Non-lumened instruments and instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

    · Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

    · Medical devices, including single, dual or triple channeled stainless steel lumens that are:

    ≥ 0.77 mm ID and ≤ 527 mm in length

    ≥ 0.8 mm ID and ≤ 542 mm in length

    ≥ 0.48 mm ID and ≤ 100 mm in length

    • Medical devices with Dead end stainless steel lumens that are ≥ 1.3 mm ID and ≤ 73 mm in length

    · Instruments with rigid non-metallic lumens (such as those used in endoscope sheaths, take-apart forceps) that are:

    ≥ 3 mm ID and ≤ 298 mm in length

    ≥ 4 mm ID and ≤ 424 mm in length

    V-PRO 1, 1 Plus, maX & maX2 Non Lumen Cycle

    · Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

    · Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes.

    V-PRO maX and maX 2 Flexible Cycle

    · Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

    · Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two load configurations:

    o Load 1: Two flexible endoscopes with single or dual channel lumens that are ≥ 1 mm ID and ≤ 1050 mm in length with a light cord (if not integral to endoscope) and mat with no additional load

    o Load 2: One flexible endoscope with single or dual channel lumens that are ≥ 1 mm ID and ≤ 1050 mm in length with a light cord (if not integral to endoscope accessories, mat, and additional instruments that may include nonlumened or lumened medical devices with the following configurations:

    o Single, dual or triple channel stainless steel lumen that is ≥ 0.48 mm ID and ≤ 100 mm in length.

    V-PRO maX 2 Fast Non Lumen Cycle

    · Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

    · Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.

    V-PRO maX 2 Specialty Cycle:

    Patient-specific surgical guides (e.g. osteotomy, shoulder, hip, knee, spine) or anatomical models fabricated via additive manufacturing (3D printing) processes and intended for single-use during operative procedures .* or

    Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes. **

    • The validation studies were conducted using a validation load consisting of pouched guide(s) (with or without tray) for a total weight of 5 lbs (2.3 kg) 3D printed material.

    ** The validation studies were conducted using a validation load consisting of one pouched instrument tray or one pouched instrument tray or one pouch with guide(s) (with or without tray) for a total weight of 11 lbs (5 kg).

    MaterialManufacturerSpecialty CycleLumens
    Surgical Guide ResinFormlabsF≥3 mm ID x ≤30 mm L
    BioMed Amber ResinFormlabsF≥3 mm ID x ≤30 mm L
    Dental LT Clear V2 ResinFormlabsD≥3 mm ID x ≤30 mm L
    BioMed Clear ResinFormlabsD≥3 mm ID x ≤30 mm L
    Biocompatible Clear MED610StratasysE≥3 mm ID x ≤20 mm L
    Biocompatible Opaque MED615RGDStratasysE≥3 mm ID x ≤20 mm L
    VeroGlaze™ MED620StratasysE≥3 mm ID x ≤20 mm L

    STERRAD 100S Default Cycle

    Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

    • Metal and nonmetal lumened instruments with:
    • ≥ 6 mm ID and < 310 mm in length
    • · Medical devices with a single stainless steel lumen with:
    • ≥ 1 mm ID and ≤ 125 mm in length
    • ≥ 2 mm ID and ≤ 250 mm in length
    • ≥ 3 mm ID and ≤ 400 mm in length

    STERRAD NX and NX with ALLClear Technology Standard Cycle

    · Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

    · Medical devices with a single stainless steel lumen with:

    ≥ 2 mm ID and ≤ 400 mm in length

    STERRAD NX and NX with ALLClear Technology Advanced Cycle

    · Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

    · Medical Devices, including most flexible endoscopes, with:

    a single stainless steel lumen with:

    ≥ 1 mm ID and < 500 mm in length

    Single channel polyethylene and Teflon (polytetrafluoroethylene)

    ≥1 mm ID and < 850 mm in length

    STERRAD 100NX and 100NX with ALLClear Technology Standard Cycle

    · Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

    · Medical devices with a single stainless steel lumen with:

    ≥ 0.7 mm ID and ≤ 500 mm in length

    STERRAD 100NX and 100NX with ALLClear Technology Flex Scope Cycle

    Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

    Medical Devices, including most flexible endoscopes, with:

    · Single channel polyethylene and Teflon (polytetrafluoroethylene)

    ≥ 1 mm ID and ≤ 850 mm in length

    STERRAD 100NX and 100NX with ALLClear Technology Express Cycle

    • Metal and nonmetal medical devices (surfaces sterilization only) and instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

    STERRAD 100NX and 100NX with ALLClear Technology Duo Cycle

    Medical devices including:

    o most flexible endoscopes with a single channel of polyethylene and Teflon (polytetrafluoroethylene) with ≥ 1 mm ID and ≤ 875 mm in length

    o accessory devices that are normally connected to a flexible endoscope during use

    o flexible endoscopes without lumens

    • Trays must be legally marketed for use in the V-PRO Low Temperature Sterilization Systems and contain a vent surface area to tray volume ratio ≥0.135 in-1 with the maximum number of instrument organizers installed.

    Ethylene Oxide Sterilization

    The Vis-U-All Low Temperature Sterilization Pouch has been qualified by STERIS as suitable for use by health care providers to enclose and seal other medical devices to be sterilized by ethylene oxide (ETO). The Vis-U-All Low Temperature Sterilization Pouch for ethylene oxide is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.

    The following are the validated test conditions:

    • · 1 hour exposure, at 130(±5) °F, * >30% RH using 100% ETO (750-790 mg/L)
    • · 4.5 hour exposure at 100(±5) °F, * >30% RH using 100% ETO (750-790 mg/L)
    • *±5 °F is used during sterilization phase following an equilibrium period of 10% of exposure time.

    The ethylene oxide process indicator is intended to be used by health care providers with the Vis- U-All Low Temperature Sterilization Pouches to distinguish between processed and unprocessed units.

    Type of Use (Select one or both, as applicable)

    | Prescription Use (Part 21 CFR 801 Subpart D)

    X Over-The-Counter Use (21 CFR 801 Subpart C)

    Device Description

    The proposed Vis-U-All Low Temperature Sterilization Pouches/Tubing is identical to the predicate and is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in V-PRO or STERRAD Low Temperature Sterilization Systems. As is the predicate device, the proposed device is available as a self seal pouch, a heat seal pouch, or heat seal tubing. Available sizes and configurations are shown in Table 5-1.

    AI/ML Overview

    The provided text describes the acceptance criteria and study for the STERIS Vis-U-All Low Temperature Sterilization Pouches/Tubing (K231500).

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Worst-case test articles shall be reproducibly sterilized under worst-case ½ cycle conditions for the 136L V-PRO Sterilizer Specialty CyclePASS

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific "sample size used for the test set" in terms of the number of individual pouches or sterilization cycles. However, it mentions "worst-case test articles" were used in the testing.

    The data provenance is not explicitly stated in terms of country of origin. The study appears to be retrospective as it's a validation study for an existing product to extend its claims (qualify for use in the V-PRO maX 2 Specialty Cycle).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The study's ground truth for sterilization effectiveness is based on the ability of the worst-case test articles to be reproducibly sterilized under specific cycle conditions, which implies a direct measurement of sterility rather than expert consensus on an image or clinical assessment.

    4. Adjudication Method for the Test Set

    This information is not applicable as the ground truth is based on physical sterilization validation, not expert review or consensus.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic or AI-assisted interpretation devices to assess human performance with and without AI assistance. The described device is a sterilization pouch, and its performance is evaluated through physical tests of sterilization effectiveness.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation was done. The study assessed the device's ability to achieve sterilization in the specified V-PRO cycles without human intervention in the sterilization process itself, beyond loading and operating the sterilizer. The performance criteria directly relate to the pouch's material and design allowing sterilant penetration.

    7. The Type of Ground Truth Used

    The ground truth used is based on physical sterilization effectiveness, specifically demonstrating that "Worst case test articles shall be reproducibly sterilized under worst case ½ cycle conditions." This implies microbiological testing to confirm the absence of viable microorganisms after sterilization.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided. The device is a physical product (a sterilization pouch), not an AI algorithm that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. As mentioned above, the device is a physical product and does not involve an AI algorithm with a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222440

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-09-09

    (28 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

    • medical devices in a single or double pouch configuration
    • trays* containing medical devices in a single or double pouch configuration
    • small items requiring surface sterilization in a single pouch configuration within a tray*
      NOTE: Trays must be legally marketed for use in the V PRO Low Temperature or STERRAD Sterilization Systems and contain a vent surface area to tray volume ratio ≥ 0.135 in-1 with the maximum number of instrument organizers installed.

    to be sterilized in the:

    • Lumen. Non Lumen. Flexible. Fast Non Lumen and Fast Cvcles of the V-PRO® Low Temperature Sterilization Systems
    • Default Cycle of the STERRAD 100S Sterilizer
    • Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers
    • Express, Standard, Flex Scope and DUO Cycles of the STERRAD 100NX with ALLClear Technology Sterilizers

    *STERRAD and ALLClear are trademarks of Advanced Sterilization Products

    The pouches maintain the sterility of the enclosed devices until used.

    When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

    Intended Sterilization Cycles and Intended Pouch Loads when Medical Devices are:

    • Directly pouched
    • Placed inside of a tray and the tray pouched

    The PRO-LITE Sterilization Trays are used to contain medical devices for sterilization in the following Cycles:

    • Lumen, Non Lumen, Flexible, Fast Non Lumen and Fast Cycles of the V-PRO Low Temperature Sterilization Systems
    • Default Cycle of the STERRAD®* 100S Sterilizer
    • Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers
    • Standard, Flex Scope, Express and DUO Cycles of the STERRAD 100NX with ALLClear Technology Sterilizers

    *STERRAD and ALLClear are trademarks of Advanced Sterilization Products

    Prior to placing in the Sterilizer, the trays must either be:

    • wrapped with a legally marketed sterilization wrap for use in the Sterilizers listed above
      Or
    • placed inside a legally marketed pouch for enclosing trays in the Sterilizers listed above.

    The PRO-LITE Sterilization Trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated, FDA-cleared sterilization wrap or pouch in order to maintain sterility of the enclosed medical instruments.

    Device Description

    The proposed Vis-U-All Low Temperature Sterilization Pouches/Tubing is identical to the predicate and is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in V-PRO or STERRAD Low Temperature Sterilization Systems. As is the predicate device, the proposed device is available as a self-seal pouch, a heat-seal pouch, or heat-seal tubing.

    The PRO-LITE Sterilization Trays contain medical devices for sterilization in the V-PRO Low Temperature Sterilization Systems and STERRAD Sterilizers. The trays are available in various sizes to accommodate the loads to be processed. The proposed trays are identical in design to the predicate Sterilization Tray (K183402) and are composed of a base and a lid. The lid includes clamping mechanisms designed to secure the lid onto the base. There are numerous holes in the base and lid for sterilant penetration. The tray is categorized as a cassette and requires complete enclosure in a legally-marketed sterilization wrap or pouch to maintain sterility of the enclosed devices. Both the base and the lid for the proposed tray are made of a mineral-filled polypropylene material. Optional instrument organizers and sterilization mats are provided as accessories.

    AI/ML Overview

    This document, a 510(k) Premarket Notification from the FDA, describes the acceptance criteria and the study proving the device meets these criteria for two products: "Vis-U-All Low Temperature Sterilization Pouches/Tubing" and "PRO-LITE Sterilization Trays."

    Since the request is about "acceptance criteria and the study that proves the device meets the acceptance criteria" specifically for a device (implying a medical device with an associated study), and the document describes the performance testing for these sterilization products, I will extract and synthesize the information relevant to these performance tests as evidence of meeting acceptance criteria.

    Note: The provided document is a 510(k) summary for medical devices (sterilization pouches and trays), not an AI/ML powered device. Therefore, typical AI/ML study components like "number of experts used to establish ground truth," "adjudication method," "MRMC study," "effect size of human readers improving with AI," "standalone performance," "training set size," and "ground truth for training set" are not applicable to this type of submission. The document focuses on demonstrating substantial equivalence through non-clinical performance testing (e.g., sterilization efficacy).

    Here's an analysis of the acceptance criteria and proof, as presented for these sterilization devices:

    1. A table of acceptance criteria and the reported device performance

    For both devices, the acceptance criteria implicitly align with demonstrating "Sterile efficacy" and the "ability to sterilize medical devices" under various conditions and cycles. The performance is reported as meeting these implicit criteria, indicated by "PASS" results for all tests.

    For "Vis-U-All Low Temperature Sterilization Pouches/Tubing" (Table 3, Page 18):

    Acceptance Criteria (Implicit)Reported Device Performance (Result)Conclusion
    Sterile Efficacy for Mated Surfaces under Worst-Case Conditions in V-PRO Sterilizer CyclesSterile efficacy was demonstrated for mated surfaces under worst case conditions in the V-PRO Sterilizer cycles.PASS
    Sterile Efficacy with Standard Injection WeightThe standard injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle.PASS
    Ability to Sterilize Medical Devices under Worst-Case Processing Conditions (Simulated Use)Simulated use testing verified the ability of the sterilizer cycles to sterilize medical devices under worst-case processing conditions.PASS
    Ability to Sterilize Patient-Soiled, Clinically-Cleaned Medical Instruments (In Use)The in use investigation demonstrated the ability of the V-PRO Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.PASS

    For "PRO-LITE Sterilization Tray" (Table 8, Page 29-30):

    Acceptance Criteria (Implicit)Reported Device Performance (Result)Conclusion
    Sterile Efficacy for Mated Surfaces under Worst-Case Conditions in V-PRO Sterilizer CyclesSterile efficacy was demonstrated for mated surfaces under worst case conditions in the V-PRO Sterilizer cycles.PASS
    Sterile Efficacy with Standard Injection WeightThe standard injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle.PASS
    Ability to Sterilize Medical Devices under Worst-Case Processing Conditions (Simulated Use)Simulated use testing verified the ability of the sterilizer cycles to sterilize medical devices under worst-case processing conditions.PASS
    Ability to Sterilize Patient-Soiled, Clinically-Cleaned Medical Instruments (In Use)The in use investigation demonstrated the ability of the V-PRO Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.PASS

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes (e.g., number of pouches/trays, number of sterilization cycles) used for each test. It refers to "worst-case conditions" and "validation load." Similarly, data provenance (country of origin, retrospective/prospective) is not specified, but the testing would typically be prospective and conducted in a controlled laboratory or simulated clinical environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable to the type of device and study presented. Ground truth for sterilization efficacy is typically established through scientific methods like sterility testing (e.g., microbial challenge tests, biological indicators) and physical/chemical monitoring, not human expert interpretation of images or data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as this is related to human interpretation of data, which is not the primary method for evaluating sterilization efficacy.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. These are physical devices for sterilization, not AI-powered diagnostic tools.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This concept pertains to AI algorithms, not sterilization instruments.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for sterilization efficacy is established through direct measurements of sterility (e.g., absence of viable microorganisms after sterilization) and functional performance under simulated and actual use conditions as described in the summary:

    • Sterile efficacy: Demonstrated through microbial challenge and physical/chemical indicators.
    • Simulated Use Test: Verifies the ability of the sterilizer cycles to sterilize medical devices under worst-case processing conditions.
    • In Use Test: Demonstrates the ability to sterilize patient-soiled, clinically-cleaned medical instruments.

    This involves direct, objective testing for sterility, rather than interpretation by human experts.

    8. The sample size for the training set

    Not applicable. These devices do not use "training sets" as they are not AI/ML models.

    9. How the ground truth for the training set was established

    Not applicable.