Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240940

    Validate with FDA (Live)

    Date Cleared
    2024-05-03

    (28 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vial2Bag Advanced® 20mm Admixture Device is indicated to serve as a connection between a 50, 100 or 250mL IV bag, vial with 20mm closure, and an external IV administration set. The integrated Vial Adapter makes it possible to reconstitute and/or admix drugs prior to administration to the patient. Indicated for adolescent and adult patients only.

    Device Description

    The Vial2Bag Advanced® 20mm Admixture Device is a single-use, sterile, needle-less, nonpyrogenic, fluid transfer device that allows for reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution, through the IV bag administration port. The device consists of the body, Protector, IV Port, and an integrated vial adapter. The device is intended to be used with standard drug vials with a 20mm closure and an elastomeric stopper. The Vial2Bag Advanced® 20mm device is designed to work with a standard 50, 100, or 250mL IV bag and an external IV infusion set.

    AI/ML Overview

    The provided text describes a medical device, the Vial2Bag Advanced® 20mm Admixture Device, and its comparison to a predicate device for substantial equivalence. It is not an AI/ML device, therefore, many of the requested fields are not applicable.

    Here's an analysis based on the information provided:

    1. A table of acceptance criteria and the reported device performance:

    The document outlines performance testing, but the explicit acceptance criteria are not tabulated with specific performance results. Instead, it states that testing was conducted to ensure the device "met the applicable design and performance requirements." For some tests, it mentions the basis for acceptance criteria or sample size, but not the actual criteria set.

    TestAcceptance Criteria (Explicitly Stated)Reported Device Performance (Explicitly Stated)
    Vial Adaptor Tensile Detachment ForceNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    Vial Adaptor Torque TestNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    Detachment Force of Vial Adapter from VialNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    Attachment Force of Vial Adapter to the VialNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    Leakage TestingNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    Internal Diameter of the Upper Skirt for Vial AdapterNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    1m Drop DurabilityNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    FragmentationBased on EN ISO 8536-2, section 6.2.2Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    Mass TransferNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    Residual VolumeNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    Dose ConcentrationNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
    BiocompatibilityIn accordance with ISO 10993-1, classified as externally communicating device with prolonged contact duration (>24 hours to 30 days)Test reports from the reference device (K201415), which uses an identical colorant, were leveraged and deemed applicable to demonstrate biological safety.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Fragmentation Test: "sample size based on EN ISO 7864, Annex B, Section B.4." (Specific number not provided, but the standard is cited.)
    • Sample Size for other tests: Not explicitly stated.
    • Data Provenance: The tests are in-house, suggesting they were conducted by the manufacturer, West Pharma. Services IL, Ltd., in Ra'anana, Israel. There is no mention of retrospective or prospective data in the context of performance testing on the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the device is a physical medical device (fluid transfer device) and not an AI/ML diagnostic or predictive tool that requires expert-established ground truth on a test set. The performance testing is based on engineering and material standards.

    4. Adjudication method for the test set:

    Not applicable for a physical device's performance testing. The outcomes of the tests are objective measurements against established standards or in-house criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI/ML device, and no MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used:

    The performance criteria are established through:

    • In-house test methods (presumably based on engineering principles and intended function).
    • International standards (e.g., EN ISO 8536-2 for Fragmentation, ISO 10993-1 for Biocompatibility).
    • Comparison to a predicate device to demonstrate "substantial equivalence."

    8. The sample size for the training set:

    Not applicable. This is a physical device, not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1