510(k) Data Aggregation

    K Number
    K153352

    Validate with FDA (Live)

    Date Cleared
    2016-08-01

    (255 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vertu® & Vertu® Ti-Bond devices are stand-alone interbody fusion devices intended for spinal fusion procedures at one level from the C2/C3 disc space to the C7/T disc space in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radies) of the cervical spine. Implanted via an open, anterior approach and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

    Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.

    The implant is designed to accommodate two screws. Two screws should be used to ensure adequate fixation of the implant.

    Crystal® & Crystal® Ti-Bond devices are intended for spinal fusion at one or two contiguous levels in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implanted via an open, anterior approach from the C2-C3 disc space to the C7-T1 disc space and packed with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

    The device is intended to be used with supplemental fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).

    Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.

    Device Description

    The Vertu and Vertu Ti-Bond Cervical Interbody System is composed of an implant body made from titanium alloy (Ti-6Al-4V) conforming to ASTM F136 or ISO 5832-3 or polyetheretherketone (PEEK-Optima®) conforming to ASTM F2026 or PEEK-Optima® coated with commercially pure titanium per ASTM F1580 and fixation screws made from Ti-6Al-4V per ASTM F136 or ISO 5832-3 with a Nitinol clip conforming to ASTM F2063. All implant bodies have a titanium insert lining the internal surface of the screw holes conforming to ASTM F136 or ISO 5832-3. The implant body is generally boxshaped device with holes through its body for the placement of graft material. Additionally, it has teeth located on its superior and inferior external surfaces to help keep the device from migrating once placed in its desired location. Each screw hole is lined in its internal surface with a titanium ring insert. Screws pass through screw holes of the implant body and affix to bone to help prevent implant migration. When fully seated, the screw head rests on the titanium insert of the screw hole.

    The Crystal and Crystal Ti-Bond Cervical Interbody devices are generally a box-shaped device with various holes located throughout its geometry to allow for packing of bone graft material. The body of the implant is made from titanium alloy (Ti-6A1-4V) conforming to ASTM F136 or ISO 5832-3 or polyetheretherketone (PEEK-Optima®) conforming to ASTM F2026 or PEEK-Optima® coated with commercially pure titanium per ASTM F1580. Superior and inferior surfaces of the device have teeth that help keep the device from migrating once placed in its desired location.

    AI/ML Overview

    This FDA 510(k) Premarket Notification document (K153352) describes the Vertu® & Vertu® Ti-Bond and Crystal® & Crystal® Ti-Bond Cervical Interbody Systems. Per the document, "No changes were made to the existing Vertu. Vertu Ti-Bond. Crystal Ti-Bond devices nor were any new components added to the system. Therefore, no additional testing was required or performed." This implies that new acceptance criteria and a study to prove these devices meet those criteria were not established for this specific submission. Instead, the document relies on the substantial equivalence to previously cleared predicate devices.

    Therefore, the following information cannot be directly extracted from the provided text for this specific K153352 submission:

    1. A table of acceptance criteria and the reported device performance: Not explicitly stated as new criteria or performance data were generated for this submission. The device is determined to be substantially equivalent based on existing data and characteristics of predicate devices.
    2. Sample size used for the test set and the data provenance: Not applicable as a new test set was not created for this submission. The submission references previous clearances (K122771, K133218, K073351) for the predicate devices, which would have had their own testing.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. These are interbody fusion devices, not AI-assisted diagnostic tools.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used: Not applicable.
    8. The sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.

    Summary based on the provided document:

    The provided document (K153352) is a 510(k) premarket notification for expanded indications for the Vertu® & Vertu® Ti-Bond and Crystal® & Crystal® Ti-Bond Cervical Interbody Systems. The FDA determined the device is substantially equivalent to legally marketed predicate devices. The basis for this determination is that no changes were made to the existing devices, and no new components were added, thus no additional testing was required or performed for this specific submission. The substantial equivalence is argued based on a comparison of indications for use, technological characteristics, surgical technique, design features, and instrumentation to established predicate devices.

    The acceptance criteria and performance data for the original clearances of these devices and their predicates would be found in the associated 510(k) submissions (K122771, K133218, K073351, K150053, K142264, K133653), which are not included in this document.