Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182897

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-26

    (161 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pressure Rated: The Q2® Extension Set with needleless connector is for single use only. The extension set may be used for direct injection, intermittent infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.

    Non Pressure Rated: The Q2® Extension Set with needleless connector is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration.

    Device Description

    Sterile, single use non-pyrogenic intravenous administration extension set comprised of 5.75 inch tubing bonded to a male luer on one end and a T-connector on the other end with a clamp in between. The T-connector has a female luer on one side and a swabable, needle-free port on the other side. This device is not made with the plasticizer Diethylhexylphthalate (DEHP). The Velano Vascular Q2® Extension Set is compatible with PIVO™ devices.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device (Velano Vascular Q2 Extension Set), which focuses on demonstrating substantial equivalence to a predicate device, rather than proving that an AI/ML device meets specific performance criteria through a rigorous study. Therefore, the information typically required to answer your prompt about AI acceptance criteria and study details (such as sample size for training/test sets, expert adjudication, MRMC studies, or standalone algorithm performance) for an AI/ML device is not present in this document.

    This document describes a traditional medical device (an intravascular administration set) and performs mechanical and functional testing to demonstrate its safety and effectiveness relative to a legally marketed predicate device.

    However, I can extract the relevant information regarding the device's functional and safety testing, despite it not being an AI-driven device.

    Here's a breakdown of the acceptance criteria and study information provided in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly present a table of "acceptance criteria" against "reported performance" in a quantitative manner for each test. Instead, it lists the types of functional and safety tests performed and states that these tests ensured the device design meets its functional and performance requirements. The conclusion explicitly states that "Velano Vascular considers the Velano Vascular Q2® Extension Set device to be substantially equivalent to the predicate device listed above. This conclusion is based upon the intended use, design specifications, patient contacting materials, manufacturing and sterilization processes." This implies that the device met the implicit acceptance criteria for each test to achieve substantial equivalence.

    Implicit Acceptance Criteria and Demonstrated Performance (Inferred from the document):

    Test PerformedImplicit Acceptance CriterionReported Device Performance (Implied)
    Visual inspectionDevice must be free from visible defects.Met requirements (ensured device design meets functional and performance requirements).
    Simulated shippingDevice must withstand shipping stresses without damage or compromise to function.Met requirements.
    Priming volumePriming volume must be consistent with specifications and safe for intended use.< 1 mL (consistent with predicate and acceptable).
    Backpressure leak (normal use & power injection)No leakage under normal operating pressures and specified power injection pressures (max 325 psi).Met requirements (demonstrates the addition of the T-connector and female luer does not alter the safety or effectiveness of the device compared to the predicate, which also withstands 325 psi).
    Flow rate (normal use & power injection)Flow rate must meet specified values for normal and power injection (10ml/sec at 325 psi).Met requirements (demonstrates the addition of the T-connector and female luer does not alter the safety or effectiveness of the device, and the shorter length of the new device still results in acceptable flow rate and kink resistance).
    Simulated useDevice must perform as intended during simulated clinical use.Met requirements.
    Tubing bond strengthBonds must withstand specified forces without separation.Met requirements.
    Multiple EngagementNeedleless connector must function accurately after multiple engagements.Met requirements.
    Continuous EngagementNeedleless connector must maintain integrity during continuous engagement.Met requirements.
    Activation ForceForce required to activate the device must be within acceptable limits.Met requirements (demonstrates the addition of the T-connector and female luer does not alter the safety or effectiveness of the device).
    Tubing kink resistanceTubing must resist kinking under normal manipulation.Met requirements (demonstrates the change in length does not risk safety or effectiveness).
    Multiple clampingClamp must function effectively after multiple closures.Met requirements.
    Prolonged clampingClamp must maintain occlusion over prolonged periods.Met requirements.
    PIVO compatibility testingDevice must be compatible with PIVO™ devices as claimed.Met requirements (demonstrates compatibility with PIVO™ devices does not affect safety or effectiveness).
    Sterilization (ISO 11135:2014)Sterilization process (EtO) achieves a sterility assurance level (SAL) of 1 x 10^-6.All patient contacting materials, sterilization, and design specifications are identical and unchanged from the predicate device. Additional sterilization assessments and testing were performed in accordance with ISO 11135:2014, achieving 1 x 10^-6 SAL.

    2. Sample sized used for the test set and the data provenance:

    • Sample Size: The document repeatedly refers to "representative samples" being used for mechanical testing (Page 7). However, it does not specify the exact number of samples (sample size) used for each test.
    • Data Provenance: Not applicable in the context of "data" for an AI/ML device. For a physical device, the data provenance is the location where the manufacturing and testing occurred. This information is not detailed beyond the manufacturer's address (San Francisco, CA). The tests are presumably performed in a controlled laboratory environment. The testing is prospective in the sense that it's performed as part of the submission process for the new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as this is not an AI/ML device requiring human expert ground truth for classification or interpretation. The performance is assessed through objective physical and functional tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable as this is not an AI/ML device requiring human expert adjudication. The test results are based on objective measurements and established standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This is not applicable as this is not an AI/ML device. The device's "standalone" performance is its physical and functional performance during the mechanical tests.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For physical/mechanical devices, "ground truth" is established by engineering specifications, industry standards (e.g., ISO), and physical measurements against known values and benchmarks (like the predicate device's performance). For example, a leak test's "ground truth" is "no leak detected under specified pressure." This is not based on expert consensus in the diagnostic sense, but rather adherence to predefined engineering limits and physical laws.

    8. The sample size for the training set:

    • This is not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • This is not applicable as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1