Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K251728

    Validate with FDA (Live)

    Date Cleared
    2025-06-26

    (21 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Velacur is intended to provide estimates of tissue stiffness generated from shear wave speed measurements (40-70 Hz), ultrasound attenuation and Velacur Determined Fat Fraction (VDFF). The Velacur Determined Fat Fraction combines ultrasound attenuation and backscatter coefficient measurements. The device is indicated to non-invasively determine liver tissue stiffness, attenuation, and Velacur Determined Fat Fraction. VDFF is not intended to be used in pediatric patients. These are meant to be used in conjunction with other clinical indicators in order to aid in clinical management of patients with liver diseases, including hepatic steatosis.

    The device is intended to be used in a clinical setting and by trained medical professionals.

    Device Description

    The device that is the subject of this submission is the substantially equivalent to that cleared under K233977, Velacur.

    Velacur ONE is a portable device intended to non-invasively measure the stiffness and attenuation of the liver via measurement of liver tissue shear modulus and ultrasound attenuation. This is done by measuring the wavelength or wave speed of mechanically created shear waves within the organ of the patient. Attenuation is measured directly via the loss in power of the ultrasound beam and Velacur Determined Fat Fraction is a combination of measured ultrasound attenuation and backscatter.

    The device is designed to be used at the point of care, in clinics and hospitals. The device is used by a medical profession, an employee of the clinic/hospital. The activation unit is placed under the patient, while lying supine on an exam bed. The activation unit vibrates at frequencies 40, 50, and 60 Hz causing shear waves within the liver of the patient. The ultrasound transducer is placed on the patient's skin, over the intercostal space, and is used to take volumetric scans of the liver while shear waves are occurring. The device includes two algorithms designed to help users detect good quality shear waves and identify liver tissue. From the scan data, the device calculates tissue stiffness and attenuation.

    Software and hardware changes were made to the device. The intended use of the device is unchanged. The user interface was updated to a more modern platform with the same workflow and outputs. The hardware components have been consolidated, combining the functionality of the computing unit and control unit. The activation unit has been designed to include more voice coils which are distributed throughout the unit. The power output and performance of the activation unit is unchanged.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for Velacur ONE (LI-1100) focuses on substantiating equivalence to a predicate device (Velacur K233977) through bench testing and human factors testing. It explicitly states that "No clinical or animal testing was performed" for this specific submission, meaning there is no mention of a study involving human or animal subjects to prove the device's performance against clinical acceptance criteria.

    Therefore, many of the requested details related to clinical study design, such as sample size for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth types, cannot be extracted from this document as these types of studies were not conducted for the Velacur ONE (LI-1100) submission.

    The information provided primarily relates to the bench testing validation of the algorithms, which is a different type of validation than a clinical study.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance (Bench Testing)

    The document refers to "acceptance criteria" within the context of bench testing against phantoms and comparison to the predicate device.

    MetricAcceptance Criteria (Implicit from "Less than" or "Greater than" statements)Reported Device Performance (Velacur ONE)
    Elasticity - Bias vs. CIRS valuesNot explicitly stated an absolute numerical acceptance criteria, but "overall bias to the CIRS values was used as a comparison point for both systems to determine the substantial equivalence." The average difference between the bias of Velacur ONE and the predicate was 2.45%.Average difference between bias of Velacur ONE and predicate: 2.45%
    Elasticity - Precision (min/max/average)Not explicitly stated an absolute numerical acceptance criteria, but "met all criteria necessary to support clinical suitability" and compared to predicate. Predicate range: 1.1% to 2.6% (average 2.1%).Average Precision: 1.7% (min: 0.3%, max: 2.9%)
    Elasticity - Bland-Altman Correlation"No value falls outside the 1.96*STD lines."Excellent correlation; no values outside 1.96*STD lines.
    Elasticity - Substantial Equivalence"Results of the two systems both passed the defined acceptance criteria when being compared to each other and are considered substantially equivalent."Both systems passed and are considered substantially equivalent.
    Attenuation - Overall BiasLess than 10%9.1%
    Attenuation - Precision (maximum)Less than 10%0.9% (mean); Less than 10% (maximum) (For Velacur ONE)
    VDFF - Overall BiasNot explicitly stated numerical acceptance criteria for overall bias, but compared to predicate (9%) and "clinically meaningful cutoff value for any steatosis of 5%".8% (compared to predicate bias of 9%)
    VDFF - Precision (maximum)Less than 10%5.4%
    VDFF - Pearson Correlation Coefficient (r) vs. Expected VDFFGreater than 0.80.95

    2. Sample Size and Data Provenance

    • Test Set Sample Size:
      • Elasticity: 6 Homogeneous phantoms.
      • Attenuation: 6 attenuation phantoms.
      • VDFF: 6 different phantoms.
      • Note: These are phantoms, not patient data.
    • Data Provenance: The data is from bench testing using phantoms, not clinical data from patients. Therefore, there is no country of origin or retrospective/prospective classification.

    3. Number of Experts and Qualifications for Ground Truth

    • Not applicable for this submission. Ground truth for bench testing is established by the known properties of the phantoms (e.g., CIRS values, labeled backscatter and attenuation). There is no mention of experts involved in establishing ground truth for the bench tests.

    4. Adjudication Method for the Test Set

    • Not applicable for this submission. As there's no clinical test set requiring human interpretation, no adjudication method is mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. The document explicitly states "No clinical or animal testing was performed."

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop)

    • The performance data provided (bias, precision, correlation coefficients for elasticity, attenuation, and VDFF) appears to be standalone algorithm performance as measured on phantoms. The stated purpose of these tests is to validate the algorithms. While "User holds the probe against the skin, and software... is shown on the device display to indicate when the location is acceptable," the actual measurement and calculation of stiffness, attenuation, and VDFF are algorithmic. The performance metrics listed directly reflect the algorithm's output against the known phantom values.

    7. Type of Ground Truth Used

    • Phantom Properties: The ground truth for the bench testing was derived from the known and labeled properties of the reference phantoms (e.g., CIRS values for stiffness, labeled backscatter and attenuation for VDFF, and expected attenuation values).

    8. Sample Size for the Training Set

    • Not mentioned or applicable for this submission. This submission is a 510(k) for a modified device (Velacur ONE) claiming substantial equivalence to a predicate (Velacur). The focus is on verifying that the changes to the device (hardware consolidation, updated UI, new activation unit) do not negatively impact performance compared to the predicate, and that the algorithms maintain their validated performance. Details about the original training of the algorithms (which would have occurred for the predicate device's clearance) are not part of this specific document.

    9. How the Ground Truth for the Training Set Was Established

    • Not mentioned or applicable for this submission. Similar to point 8, this information would pertain to the original development and validation of the predicate device's algorithms, not this specific 510(k) for a modified device.

    The study proves the device meets the acceptance criteria through extensive bench testing on phantoms, demonstrating that the modified Velacur ONE systems performs equivalently to its predicate (Velacur) and within pre-defined numerical tolerances for accuracy and precision when measuring phantom properties. This approach is common for 510(k) submissions of modified devices where clinical performance has already been established for the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233977

    Validate with FDA (Live)

    Device Name
    Velacur
    Manufacturer
    Date Cleared
    2024-09-04

    (261 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Velacur is intended to provide estimates of tissue stiffness generated from shear wave speed measurements (40-70 Hz), ultrasound attenuation and Velacur Determined Fat Fraction (VDFF). The Velacur Determined Fat Fraction combines ultrasound attenuation and backscatter coefficient measurements. The device is indicated to non-invasively determine liver tissue stiffness, attenuation, and Velacur Determined Fat Fraction. VDFF is not intended to be used in pediatric patients. These are meant to be used in conjunction with other clinical indicators in order to aid in clinical management of patients with liver diseases, including hepatic steatosis.

    The device is intended to be used in a clinical setting and by trained medical professionals.

    Device Description

    Velacur is a portable device intended to non-invasively measure the stiffness and attenuation of the liver via measurement of liver tissue shear modulus and ultrasound attenuation. This is done by measuring the wavelength or wave speed of mechanically created shear waves within the orqan of the patient. Attenuation is measured directly via the loss in power of the ultrasound beam.

    The device is designed to be used at the point of care, in clinics and hospitals. The device is used by a medical profession, an employee of the clinic/hospital. The activation unit is placed under the patient, while lying supine on an exam bed. The activation unit vibrates at frequencies 40, 50, and 60 Hz causing shear waves within the liver of the patient. The ultrasound transducer is placed on the patient's skin, over the intercostal space, and is used to take volumetric scans of the liver while shear waves are occurring. The device includes two algorithms designed to help users detect good quality shear waves and identify liver tissue. From the scan data, the device calculates tissue stiffness and attenuation.

    Minor hardware and software changes were made to the organ guide (cleared in K223287) was also extended to add more optional overlays on top of the liver overlay to help with optimizing the scan and training users to obtain adequate images. The significant change is the addition of a new output measure for Velacur, an ultrasound derived fat fraction (VDFF).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Velacur device, as described in the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes two separate machine learning algorithms: the Velacur Determined Fat Fraction (VDFF) algorithm, the Organ Guide Extension, and the Wave Quality Guide. Each has its own acceptance criteria and performance metrics.

    Feature/AlgorithmAcceptance CriteriaReported Device Performance
    Velacur Determined Fat Fraction (VDFF)
    Correlation (VDFF vs. MRI-PDFF)Not explicitly stated an "acceptance criterion" value, but implied to be strong based on predicate device testing.0.85 [0.77-0.91] (correlation coefficient [95% CI] in validation cohort)
    AUC for 5% Steatosis DetectionNot explicitly stated an "acceptance criterion" value.0.97 [0.89-0.99] (AUC [95% CI] for detection of 5% steatosis)
    Organ Guide Extension
    Dice Coefficient> 0.7Not explicitly stated the exact achieved value, but implies met acceptance criteria.
    Pixel Accuracy> 80%Not explicitly stated the exact achieved value, but implies met acceptance criteria.
    Wave Quality Guide
    Dice Coefficient> 0.7Not explicitly stated the exact achieved value, but implies met acceptance criteria.
    Sensitivity> 80%Not explicitly stated the exact achieved value, but implies met acceptance criteria.
    Specificity> 80%Not explicitly stated the exact achieved value, but implies met acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Velacur Determined Fat Fraction (VDFF):

      • Test Set Sample Size: 70 new patients
      • Data Provenance: From 3 separate sites (with different Velacur operators). Implied to be prospective, as it's separate from the training data. The document states "Data was collected from sites across the US and Canada" for training, and "Evaluation data was collected... from separate patients, sites and collected by different users than the data used for training in order to ensure data independence," suggesting a similar geographical distribution for the test set. Retrospective/prospective not explicitly stated for this particular validation cohort, but the nature of MRI scans for ground truth implies it would be collected alongside the Velacur scans.
    • Organ Guide Extension:

      • Test Set Sample Size: More than 800 images from 21 patients.
      • Data Provenance: "Evaluation data was collected from volunteers and patients with non-alcoholic fatty liver disease and non-alcoholic steatohepatitis. All patients were recruited from hepatology clinics." Data was collected from sites across the US and Canada (implied to be similar for evaluation as for training). Data for evaluation was explicitly stated to be from "separate patients, sites and collected by different users than the data used for training," indicating an independent, possibly prospective collection for evaluation.
    • Wave Quality Guide:

      • Test Set Sample Size: More than 4,000 images from 36 patients.
      • Data Provenance: "Evaluation data was collected from volunteers and patients with non-alcoholic fatty liver disease and non-alcoholic steatohepatitis. All patients were recruited from hepatology clinics." Data was collected from sites across the US and Canada (implied to be similar for evaluation as for training). Data for evaluation was explicitly stated to be from "separate patients, sites and collected by different users than the data used for training," indicating an independent, possibly prospective collection for evaluation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Velacur Determined Fat Fraction (VDFF):

      • Number of Experts: Not specified.
      • Qualifications of Experts: Not specified. The MRI scans were "assessed to create the final measurement," implying expert interpretation without explicitly stating the number or qualifications.
    • Organ Guide Extension:

      • Number of Experts: Not specified.
      • Qualifications of Experts: "experts in the field of sonography and/or ultrasound elastography." No specific experience level (e.g., years) is provided.
    • Wave Quality Guide:

      • Number of Experts: Not specified.
      • Qualifications of Experts: "experts in the field of ultrasound elastography." No specific experience level (e.g., years) is provided.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (such as 2+1 or 3+1) for establishing ground truth for any of the algorithms. It generally states that ground truth was established by experts or through MRI-PDFF assessment, implying a single assessment per case or a consensus without detailing the process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not reported in this document. The studies focus on the standalone performance of the AI algorithms.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, standalone performance studies were done for all three components:

    • Velacur Determined Fat Fraction (VDFF): The reported correlation coefficient and AUC are measures of the algorithm's direct performance against a ground truth (MRI-PDFF).
    • Organ Guide Extension: The Dice Coefficient and pixel accuracy measure the algorithm's ability to segment organs directly.
    • Wave Quality Guide: The Dice Coefficient, sensitivity, and specificity measure the algorithm's direct ability to identify good quality waves.

    7. The Type of Ground Truth Used

    • Velacur Determined Fat Fraction (VDFF): MRI Proton Density Fat Fraction (MRI-PDFF). This is considered a highly accurate quantitative measure for liver fat.
    • Organ Guide Extension: Manual image segmentation by experts.
    • Wave Quality Guide: Manual image segmentation by experts.

    8. The Sample Size for the Training Set

    • Velacur Determined Fat Fraction (VDFF): 112 patients from 4 sites (used for parameter fitting/training).
    • Organ Guide Extension: More than 5,000 patient images.
    • Wave Quality Guide: More than 15,000 patient images from 100+ patients.

    9. How the Ground Truth for the Training Set Was Established

    • Velacur Determined Fat Fraction (VDFF): MRI-PDFF scans were assessed to create the final measurement.
    • Organ Guide Extension: Manual image segmentation by experts in the field of sonography and/or ultrasound elastography.
    • Wave Quality Guide: Manual image segmentation by experts in the field of ultrasound elastography.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232459

    Validate with FDA (Live)

    Device Name
    Velacur
    Manufacturer
    Date Cleared
    2023-09-12

    (28 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Velacur is intended to provide estimates of tissue stiffness generated from shear wave speed measurements (40-70 Hz) and coefficient of attenuation. The device is indicated to non-invasively determine liver tissue stiffness and attenuation. These are meant to be used in conjunction with other clinical indicators in order to assist in clinical management of patients with liver disease. The device is intended to be used in a clinical setting and by appropriately trained medical professionals.

    Device Description

    Velacur is a portable device intended to non-invasively measure the stiffness and attenuation of the liver via measurement of liver tissue shear modulus and ultrasound attenuation. This is done by measuring the wavelength or wave speed of mechanically created shear waves within the organ of the patient. Attenuation is measured directly via the loss in power of the ultrasound beam. The device is designed to be used at the point of care, in clinics and hospitals. The device is used by a medical profession, an employee of the clinic/hospital. The activation unit is placed under the patient, while lying supine on an exam bed. The activation unit vibrates at frequencies 40, 50, and 60 Hz causing shear waves within the liver of the patient. The ultrasound transducer is placed on the patient's skin, over the intercostal space, and is used to take volumetric scans of the liver while shear waves are occurring. The device includes two algorithms designed to help users detect good quality shear waves and identify liver tissue. From the scan data, the device calculates tissue stiffness and attenuation.

    AI/ML Overview

    The provided document is a 510(k) premarket notification summary for the Velacur device, which is an ultrasound elastography system. The document focuses on demonstrating substantial equivalence to a predicate device, specifically regarding algorithmic changes for elasticity and attenuation calculations.

    Here's an analysis based on the provided text, addressing your questions:

    1. A table of acceptance criteria and the reported device performance

    ParameterAcceptance Criteria (Maximum)Reported Device Performance
    Elasticity (Homogeneous Phantoms)
    Bias between MRE & Velacur< 10%< 10% (for 4 phantoms)
    Precision (Elasticity)< 2%< 2% (for 4 phantoms)
    Elasticity (Boundary-less Homogeneous Phantom)
    Bias< 10%< 10%
    Precision< 10%0.8%
    Attenuation
    Maximum Bias (vs. phantom specification)< 10%5.21%
    Maximum Precision< 10%3.22%
    Mean PrecisionN/A (but implicitly <10%)2.02%

    Note: The document states for elasticity that "Bland-Altman plots show that no value falls outside the 1.96STD lines." This is a common method for assessing agreement and implies good performance, but specific numerical criteria for the 1.96STD ranges are not provided as "acceptance criteria." The explicit acceptance criteria listed are for bias and precision.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The testing described is non-clinical bench testing performed on phantoms (materials with known physical properties simulating tissue).

    • Sample Size:
      • Elasticity: "4 phantoms of various elasticities" were used for homogeneous elasticity testing. For the "boundary-less" phantom, it implies a single phantom of that type.
      • Attenuation: "three attenuation phantoms" were used.
    • Data Provenance: The data is from bench testing in a lab setting, not from human subjects. The location of the testing is not specified beyond "Sonic Incytes" (which is based in Vancouver, Canada), but this is not human data with geographical or retrospective/prospective distinctions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable as the ground truth was established using physical phantoms with known properties. The "Magnetic Resonance Elasticity (MRE)" measurements are referred to for elasticity validation, but this refers to a different measurement technique for phantom properties, not human expert consensus.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable as the ground truth was established by phantom specifications and MRE measurements, not by human expert review that would require adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not conducted. The document explicitly states: "No animal or clinical performance was performed." This submission focuses on validating algorithmic changes through bench testing against phantoms, not through human-in-the-loop clinical studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in a sense. The "performance data" presented is the algorithm's performance when processing data from phantoms, validated against the known physical properties of those phantoms or readings from alternative methods (like MRE). While the device still has human operation (e.g., placing the transducer), the validation itself pertains to the accuracy of the algorithm's output (stiffness and attenuation values) from the acquired phantom data.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used was phantom specifications and measurements from other established physical measurement techniques (Magnetic Resonance Elasticity - MRE).

    8. The sample size for the training set

    The document does not provide information on the training set sample size. This submission focuses on the validation of algorithmic changes, implying the algorithms were already developed. Furthermore, as this is a 510(k) for an updated algorithm in an existing device, details of the original training data (if any was used for the initial algorithm development) are not typically required for this type of submission unless new machine learning models are being introduced for new functionalities that would require it.

    9. How the ground truth for the training set was established

    The document does not provide information on how the ground truth for the training set (if any) was established. As noted in point 8, the focus is on the validation of algorithmic changes, not on the developmental process of the original algorithms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223287

    Validate with FDA (Live)

    Device Name
    Velacur
    Manufacturer
    Date Cleared
    2023-04-20

    (177 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Velacur is intended to provide estimates of tissue stiffness generated from shear wave speed measurements (40-70 Hz) and coefficient of attenuation. The device is indicated to non-invasively determine liver tissue stiffness and attenuation. These are meant to be used in conjunction with other clinical indicators in order to assist in clinical management of patients with liver disease.

    Device Description

    Velacur is a portable device intended to non-invasively measure the stiffness and attenuation of the liver via measurement of liver tissue shear modulus and ultrasound attenuation. This is done by measuring the wavelength or wave speed of mechanically created shear waves within the organ of the patient. Attenuation is measured directly via the loss in power of the ultrasound beam.

    The device is designed to be used at the point of care, in clinics and hospitals. The device is used by a medical profession, an employee of the clinic/hospital. The activation unit is placed under the patient, while lying supine on an exam bed. The activation unit vibrates causing shear waves within the liver of the patient. The ultrasound transducer is placed on the patient's skin, over the intercostal space, and is used to take volumetric scans of the liver while shear waves are occurring. From the scan data, the device calculates tissue stiffness and attenuation.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Velacur device, as extracted from the provided text:

    Velacur Device Performance and Study Details

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Organ Segmentation Guide (for liver segmentation):
    Average Dice Score > 0.7Average Dice Score > 0.7
    Overall pixel-based accuracy > 80%Overall pixel-based accuracy > 80%
    Wave Quality Guide (for shear wave detection and segmentation):
    Dice Scores ≥ 0.7Dice Scores ≥ 0.7
    Sensitivity ≥ 80%Sensitivity ≥ 80%
    Specificity ≥ 80%Specificity ≥ 80%

    2. Sample Size for Test Set and Data Provenance

    • Test Set Sample Size: More than 1,500 images, from 35-40 patients.
    • Data Provenance: Patient data from the US and Canada. The data was collected from sites in the US and Canada.
    • Retrospective/Prospective: Not explicitly stated, but the description "Evaluation data was collected from volunteers and patients" suggests it was prospectively collected for evaluation.

    3. Number of Experts for Ground Truth and Qualifications

    • Organ Segmentation Guide:
      • Number of Experts: At least three sonographers.
      • Qualifications: All had more than 20 years of experience in abdominal ultrasound imaging.
    • Wave Quality Guide:
      • Number of Experts: Not specified beyond "all experts."
      • Qualifications: All experts hold a masters or PhD in a relevant field and have at least 10 years of experience with ultrasound imaging and elastography.

    4. Adjudication Method for Test Set

    • Organ Segmentation Guide: Ground truth was established using a pixel-based voting method (implying the consensus of the three sonographers).
    • Wave Quality Guide: Ground truth was established through expert consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No information on a Multi-Reader Multi-Case (MRMC) comparative effectiveness study (i.e., comparing human readers with AI vs. without AI assistance) is provided in the document. The study focuses on the standalone performance of the AI algorithms.

    6. Standalone (Algorithm Only) Performance

    Yes, standalone (algorithm only) performance was done. The reported Dice Scores, pixel accuracy, sensitivity, and specificity are metrics of the algorithm's performance in segmenting organs and detecting/segmenting shear waves independently.

    7. Type of Ground Truth Used

    Expert consensus based on manual image segmentation.

    8. Sample Size for Training Set

    More than 5,000 patient images were used for training each algorithm.

    9. How Ground Truth for Training Set Was Established

    Not explicitly detailed for the training set, but it can be inferred that a similar process of manual image segmentation by experts was used, as described for the validation set. The text states: "Ground truth was established using manual image segmentation by experts in the field of sonography and/or ultrasound elastography," followed by details for both validation algorithms. It is reasonable to assume a similar method was employed for the training data's ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1