Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150778

    Validate with FDA (Live)

    Date Cleared
    2015-05-29

    (65 days)

    Product Code
    Regulation Number
    878.4635
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INTENDED USE: This ultraviolet lamp is intended for use in sunlamp products for the exclusive purpose of cosmetic tanning of the human skin.

    Device Description

    This 510(k) applies to more than one hundred and sixty individual devices that fall into a generic class of ultraviolet lamps that have the same basic technological features and exactly the same intended use. The ultraviolet sunlamps of this 510(k) are classified scientifically as low pressure, mercuryrare gas discharge devices. The general lamp construction and technological principle of operation of all of the devices of this 510(k) is the same as that of a common fluorescent lamp. The main differentiators for the various sunlamps of this 510(k) include: lamp length, lamp diameter, lamp wattage, spectral characteristics and private labeling thereof.

    The ultraviolet lamps named herein comprise a tubular glass envelope, the internal surface of which is coated with a fluorescent phosphor. To each end of the tube, a glass mount is sealed. The main purpose of the mount is to provide 1) for a means of hermetically sealing the internal atmosphere of the tube. 2) a means of supporting an electrode and a means of 3) electrically accessing the internal atmosphere of the device via electrically conducting lead-wires. Once assembled as described, the envelope is evacuated of air, the air is replaced by an inert gas at a pressure significantly lower than that of the normal atmosphere, and a drop of mercury is injected into the enclosed envelope. Finally, the envelope is sealed.

    In operation, the device, now called a "lamp", is connected to an electrical source typically in series with a ballast that provides for 1) ignition of a discharge through the inert-gas-mercury vapor mixture and 2) its subsequent stabilization and sustainment of the electrical behavior of the lamp. Once connected to the applied electric field, free electrons are accelerated through the mercury vapor wherein collisions take place with the mercury atoms. These collisions cause excitation of the electrons in the mercury (Hg) atom and the production of photons at various wavelengths. Certain of these photons travel to the bulb wall and excite the phosphor powder. The phosphor, in turn, emits radiation that is transmitted through the transparent glass envelope. The radiation that is emitted from a standard fluorescent general lighting lamp is primarily in the visible region of the spectrum (380nm-760nm). The radiation that is emitted from an ultraviolet sunlamp is primarily in the UV range of the electromagnetic spectrum (300nm-400nm). Modern day, low pressure ultraviolet sunlamps typically emit most of their energy in the wavelength range that is closest to the visible; that is, in the UVA range (320nm-400nm) range. The emission spectrum of sunlamps usually contains a relatively small proportion of UVB radiation. (260nm-320nm)

    AI/ML Overview

    This submission concerns Cosmedico Light Inc.'s Ultraviolet lamps, specifically for tanning purposes. The document states that the device's performance is not being proven by a study, but rather by demonstrating substantial equivalence to predicate devices that were legally marketed prior to September 2, 2014. The claim is that the contemporary ultraviolet lamps are identical to these predicate devices in terms of intended use, technological design characteristics, and conformity to special controls required by reclassification. Therefore, the "acceptance criteria" and "device performance" in the table refer to meeting the characteristics of these predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Performance Requirements of Predicate Devices)Reported Device Performance
    Functionality / Light upYes
    Electrical CharacteristicsYes
    Burn-in BehaviorYes
    Dimensional CharacteristicsYes
    Mechanical SafetyYes
    Electrical SafetyYes
    Physical Design (e.g., reflector angle)Yes
    UV PerformanceYes
    UVA IrradianceYes
    Erythemal Effective IrradianceYes
    UVC/UVB Ratio (per 21 CFR 1040.20)Yes

    2. Sample Size Used for the Test Set and Data Provenance:

    The document explicitly states: "As noted, the subject lamps of this 510(k) are identical to the predicate devices. Therefore, the subject lamps and the predicate devices have the same performance requirements."

    This indicates that a separate "test set" in the traditional sense, involving a distinct validation study with a specified sample size of the new device, was not conducted because the claim is that the devices are fundamentally the same as already-marketed devices. Therefore, there is no specific sample size for a test set or data provenance mentioned for a new study. The performance is assessed against the established characteristics of the predicate devices.

    The document mentions that all lamps are "100% tested for functionality at the end of the manufacturing process" and that a "Quality Sampling Plan" is used for in-process and post-production testing. However, this is described as routine quality control, not a specific "test set" for a device performance study as typically understood in regulatory submissions for new or modified devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    No experts were used to establish ground truth for a test set related to this submission because the device's performance is asserted to be identical to predicate devices. The "ground truth" here is essentially the established performance and safety characteristics of the legally marketed predicate devices, which are assumed to be known and accepted.

    4. Adjudication Method for the Test Set:

    Not applicable, as no separate test set requiring adjudication was conducted.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. This submission focuses on demonstrating substantial equivalence based on identical technological characteristics and intended use, not on comparing performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    Not applicable. This device is an ultraviolet lamp, not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    The "ground truth" implicitly used is the established performance and safety profiles of the legally marketed predicate ultraviolet sunlamps. The submission argues that the new devices are essentially the same as these predicate devices.

    8. The Sample Size for the Training Set:

    Not applicable. This device is an ultraviolet lamp, not an AI/ML algorithm that requires a training set. The submission does not refer to any training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1