510(k) Data Aggregation

    K Number
    K222387

    Validate with FDA (Live)

    Date Cleared
    2022-08-31

    (23 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

    MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

    • Proton density (PD) (also called hydrogen density)
    • Spin-lattice relaxation time (T1)
    • Spin-spin relaxation time (T2)
    • Flow dynamics
    • Chemical Shift

    Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

    Device Description

    The Vantage Galan (Model MRT-3020) is a 3 Tesla Magnetic Resonance Imaging (MRI) System, previously cleared under K220192. This system is based upon the technology and materials of previously marketed Canon Medical Systems and is intended to acquire and display crosssectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body.

    AI/ML Overview

    This document (K222387) is a Special 510(k) Premarket Notification for a modification to an already cleared device: the Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MR.

    This means that the device itself was previously cleared (under K220192) and this submission is for a minor modification: an updated Iterative Motion Correction (IMC) for brain imaging. Therefore, the information provided focuses primarily on the safety and effectiveness of this specific software change relative to the predicate device.

    Given this context, the document explicitly states:

    • Imaging Performance Parameters: "No change from the previous predicate submission, K220192."
    • Indications For Use: "No change from the previous predicate submission, K220192."

    This indicates that a comprehensive de novo study proving the device meets all acceptance criteria for a new device was not performed for this specific submission. Instead, the focus is on demonstrating that the modification does not negatively impact existing performance and safety.

    However, the document does mention "image quality testing was completed which demonstrated that the subject device meets predetermined acceptance criteria" related to the IMC update. Unfortunately, the specific acceptance criteria and detailed performance from this testing are not explicitly provided in the given text.

    Therefore, I cannot fully complete all sections of your request based solely on the provided text, as the detailed information on the original acceptance criteria and validating study for the core device (K220192) is not present here, and only limited information is given for the specific modification in K222387.

    Based on the provided text for K222387, here is what can be extracted:


    Acceptance Criteria and Study for K222387 (Modification to AiCE IMC)

    The provided document (K222387) is a Special 510(k) for a modification to a previously cleared device. Therefore, the "acceptance criteria" and the "study" described here pertain to demonstrating that the modification (updated Iterative Motion Correction for brain imaging) does not compromise the safety and effectiveness of the device as previously cleared. A full, de novo study proving the entire device meets acceptance criteria is not presented, as that would have been part of the original clearance (K220192).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (for IMC Modification)Reported Device Performance (Summary from K222387)
    No degradation in image quality after IMC update."image quality testing was completed which demonstrated that the subject device meets predetermined acceptance criteria." Additionally, the purpose of the update was to "provide more consistent results in reducing motion artifacts."
    Continued compliance with safety parameters (Static field strength, Operational Modes, SAR, dB/dt, Emergency Shutdown).All safety parameters are "Same" as the predicate device and meet IEC standards (e.g., 4W/kg for whole body SAR).
    Continued functionality of the device as a diagnostic imaging modality.No change to the Indications for Use or intended use of the device.
    Software validation requirements met."successful completion of software validation"
    Risk management activities conducted."Risk analysis and verification/validation testing conducted through bench testing are included in this submission." "Risk management activities for this modification are included in this submission."

    Important Note: The specific quantitative metrics for image quality acceptance (e.g., specific SNR values, resolution tests, or qualitative scores) are not detailed in this summary. It states "predetermined acceptance criteria" were met, but doesn't list them.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated for the image quality testing related to the IMC update. The document mentions "bench testing" and "image quality testing." It does not specify human subject data for this particular modification.
    • Data Provenance: Not explicitly stated. Given it's a modification to an existing product by a Japanese company (Canon Medical Systems Corporation, Japan), the testing would likely have been done internally or with partners. It's unclear if this involved clinical data from specific countries or if it was purely technical/phantom-based testing for the IMC update. The submission type (Special 510(k)) often relies on substantial equivalence to the predicate, meaning extensive new clinical data is often not required for minor modifications.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.
      This information would be crucial for a de novo or comparative effectiveness study, but is not detailed for this type of special 510(k) where the primary focus is on the impact of a software change.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not specified.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • MRMC Study: No, an MRMC comparative effectiveness study is not described in this document. The submission is for a modification to a reconstruction processing unit, not an AI-assisted diagnostic tool that directly aids human readers in interpretation, but rather a component of the image acquisition/processing chain aiming to improve image quality for diagnostic interpretation. Therefore, a study to measure human reader improvement with/without this AI feature is not applicable in the typical sense for this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: The document states "image quality testing was completed which demonstrated that the subject device meets predetermined acceptance criteria." This implies some form of standalone performance measurement (e.g., phantom studies, technical image quality metrics) was performed. However, the exact methodology and results are not detailed. The "AiCE Reconstruction Processing Unit" is an algorithm (Artificial intelligence Cleared Engine) that processes images. The IMC update is a part of this algorithm.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Not explicitly stated. For "image quality testing" for a reconstruction algorithm, ground truth likely involved:
      • Phantoms: Known physical targets with specific properties.
      • Technical Metrics: Measurements like Signal-to-Noise Ratio (SNR), Contrast-to-Noise Ratio (CNR), spatial resolution, artifact suppression measured against theoretical ideals or established benchmarks.
      • Qualitative Assessment: Visual assessment by experts based on predefined criteria, though a formal adjudication process is not described.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable/Not provided in this document. The AiCE (Artificial intelligence Cleared Engine) is a deep learning reconstruction technology. Any training data for the neural network would have been used during the development phase of the original AiCE algorithm (K220192), not specifically detailed for this software update in K222387. This special 510(k) pertains to a modification to an existing algorithm, not the training of a new one.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable/Not provided in this document. As above, this pertains to the initial development of the AiCE algorithm which would have been covered in the K220192 submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220192

    Validate with FDA (Live)

    Date Cleared
    2022-04-08

    (74 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

    MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

    ·Proton density (PD) (also called hydrogen density)

    ·Spin-lattice relaxation time (T1)

    ·Spin-spin relaxation time (T2)

    ·Flow dynamics

    ·Chemical Shift

    Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

    Device Description

    The Vantage Galan (Model MRT-3020) is a 3 Tesla Magnetic Resonance Imaging (MRI) System, previously cleared under K212056. This system is based upon the technology and materials of previously marketed Canon Medical Systems and is intended to acquire and display crosssectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MR. This submission primarily focuses on software and accessory changes to an already cleared device. Due to the nature of a 510(k) summary, detailed acceptance criteria and in-depth study results are not fully elaborated in the provided document, as it emphasizes substantial equivalence to a predicate device.

    However, I can extract the information that is present and highlight what is not explicitly detailed, based on the structure of your request.

    Here's a breakdown of the acceptance criteria and study information provided in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document broadly states that "the test results met predetermined acceptance criteria" for various features. However, it does not provide a specific table or quantified acceptance criteria values, nor does it give specific numerical performance metrics for these criteria. The performance is described qualitatively.

    Feature/FunctionalityAcceptance Criteria (Not Quantified)Reported Device Performance (Qualitative)
    sureVOI Liver, LiverLine+, ProstateLine+, W-SpineLine+"worked as intended", "images were of diagnostic quality", "test results met predetermined acceptance criteria"Confirmed that these features worked as intended, the images were of diagnostic quality.
    Exsper in FSE2D"reduces artifacts caused by unfolding errors, compared to traditional SPEEDER"Confirmed that Exsper reduces artifacts caused by unfolding errors, compared to traditional SPEEDER. Functionality testing in FE2D and SE2D confirmed scanning without problems.
    2D-RMC in FSE2D and FASE2D"scanning... can be conducted without problems"Confirmed that the scanning in those sequences can be conducted without problems.
    mART EXP"reduce distortion artifacts"Testing verified mART EXP can reduce distortion artifacts in the readout direction like mART+ and can also reduce distortion artifacts in the slice direction more than mART+.
    IMC (Iterative Motion Correction)"effective in reducing motion artifacts"Testing confirmed that IMC is effective in reducing motion artifacts for head (rigid motion) and C-spine (rigid and non-rigid motion).
    RDC DWI"distortion... reduced"Confirmed that the distortion in phase encoding direction was reduced by RDC DWI as compared to conventional images without RDC DWI in SEEPI2D sequence.
    pCASL (pseudo-continuous ASL)"CBF values... met predetermined acceptance criteria"Testing confirmed CBF values via pCASL met predetermined acceptance criteria.
    Ceiling Camera"percentage of successful patient orientation detection and cases requiring no correction for successful patient anatomy position detection met predetermined acceptance criteria"Confirmed that percentage of successful patient orientation detection and cases requiring no correction for successful patient anatomy position detection met predetermined acceptance criteria. These percentages are not specified.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Sizes: The document states that "sureVOI Liver, LiverLine+, ProstateLine+, and W-SpineLine+ were evaluated using volunteer images," "IMC (Iterative Motion Correction) for head and C-spine was evaluated using volunteer images," and "pCASL (pseudo-continuous ASL) was evaluated utilizing volunteer images." The Ceiling Camera function was evaluated using volunteers. No specific number of volunteers or images is provided. Phantom images were used for Exsper and RDC DWI.
    • Data Provenance: The document implies these studies were conducted by Canon Medical Systems Corporation in Japan, as it lists their manufacturing site and official correspondent there. The patient data origin (country) is not specified. It can be inferred that the data is prospective given the use of "volunteer images" for testing certain features.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth for the test sets. It mentions that images are "interpreted by a trained physician" as part of the overall indications for use, but this is a general statement, not specific to the validation study.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method (such as 2+1, 3+1, or none) for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance vs. without AI assistance, nor does it specify an effect size for such a study. The functionalities described are primarily about improving image quality, reducing artifacts, or automating scan planning, rather than direct diagnostic AI assistance for human readers in a comparative effectiveness study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The document describes several software features that perform automated tasks (e.g., sureVOI Liver, LiverLine+, ProstateLine+, W-SpineLine+ for auto-detection and scan planning; IMC, RDC DWI, Exsper for image reconstruction/artifact reduction). These functions effectively demonstrate "algorithm-only" performance in their specific tasks without explicit human-in-the-loop evaluation being described in the validation section. For instance, the acceptance of automatic detection and planning functions implies a standalone evaluation of the algorithm's ability to identify relevant anatomical regions.

    7. The Type of Ground Truth Used

    The document implies the following types of ground truth:

    • For automated scan planning features (sureVOI Liver, LiverLine+, ProstateLine+, W-SpineLine+): The ground truth seems to be derived from expert-confirmed correct anatomic localization and scan planning, as the outcome measure is whether the features "worked as intended" and "images were of diagnostic quality." This implicitly means agreement with human-expert derived optimal scan planes.
    • For image quality/artifact reduction features (Exsper, mART EXP, IMC, RDC DWI): The ground truth relates to the absence or reduction of artifacts and the diagnostic quality of the resulting images. This would likely be assessed by expert review and comparison to images without the feature, or a recognized "ground truth" image.
    • For pCASL: "CBF values via pCASL met predetermined acceptance criteria." This suggests a quantitative ground truth for Cerebral Blood Flow (CBF) values, possibly established through established methods or phantom measurements.
    • For Ceiling Camera: "percentage of successful patient orientation detection and cases requiring no correction for successful patient anatomy position detection met predetermined acceptance criteria." This suggests a ground truth for correct patient orientation and anatomy position, likely established by human observation or measurement.

    8. The Sample Size for the Training Set

    The document does not provide specific details about the sample size used for the training set (if any) for the machine learning-based features (LiverLine+, ProstateLine+, W-SpineLine+). As this is a 510(k) for a modification, the focus is on validation of the changes rather than a full de novo submission of an AI algorithm, so training data details are often omitted if not directly relevant to the new functionality's validation.

    9. How the Ground Truth for the Training Set was Established

    The document does not provide details on how ground truth was established for any training data used for the machine learning-based features.