Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230968

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-07-06

    (92 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VYBE RF Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis and the percutaneous treatment of facial wrinkles for use with Fitzpatrick Skin Type I to Skin Type V when using the 1MHz setting. The 2MHz setting has not been evaluated for use in the percutaneous treatment of facial wrinkles and is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The VYBE RF Electrosurgical System is comprised of the following components:

    • . The system main body, consisting of:
      • o LCD touch screen control panel
      • High-frequency generating output section (main P.C.B board or RF Generator) O
      • Power supply component Switch Mode Power Supply (SMPS) O
    • The accessories to the device include:
      • Handpiece with Handpiece Connector O
      • Disposable micro-needle cartridge (electrode) for insertion into Handpiece o
      • Foot switch O
        Radio frequency current (RF energy) is delivered from the RF Generator, through the handpiece and electrode tip into the target tissue. As RF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating (causing minor dermal damage) the tissue. The bi-polar RF energy is delivered between independent adjacent electrode pairs (total 36 needle electrode, 6 x 6 array insertion). The RF generator and hand piece are not disposable. Each disposable micro-needle cartridge (electrode) is supplied sterile and is for single patient use only and cannot be re-sterilized.
    AI/ML Overview

    The document provided describes the VYBE RF Electrosurgical System and its substantial equivalence to a predicate device, the VIVACE Electrosurgical System. It includes performance data, but does not describe an acceptance criteria and study proving the device meets the acceptance criteria in the manner requested by the prompt for a typical AI/ML medical device submission.

    Specifically, the document focuses on regulatory compliance through electrical, biocompatibility, sterility, software verification and validation, and thermal testing. These are standard tests for medical devices to ensure safety and fundamental functionality, not necessarily performance benchmarks against clinical criteria in the same way an AI/ML diagnostic or prognostic tool would be evaluated.

    Therefore, many of the requested categories in the prompt regarding acceptance criteria, human reader studies, and ground truth establishment are not applicable or not explicitly detailed in the provided text.

    Here's a breakdown of what can and cannot be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document lists various tests and states that "All acceptance criteria were met." However, it does not provide a table with specific acceptance criteria values (e.g., a specific thermal threshold) or the quantitative reported performance for each of these tests. It only lists the standards followed.

    Acceptance Criteria CategoryReported Device Performance
    Electrical TestingMet IEC 60601-1, IEC 60601-1-2, IEC 60101-2-2, IEC 60601-1-6, IEC 62366 standards.
    Biocompatibility TestingMet ISO 10993-1, ISO 10993-5, ISO 10993-10 standards.
    Sterility Testing (micro-needle cartridge)Conforms to a Sterility Assurance Level (SAL) of 10^-6, meeting ISO 11135, ISO 11138-1, ISO 11138-2, ISO 11737-1, ISO 11737-2, ISO 10993-7 standards.
    Software Verification and Validation TestingConducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued May 11, 2005.
    Thermal TestingCompleted in accordance with FDA's Guidance for Industry and FDA Staff "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", issued March 9, 2020.

    2. Sample size used for the test set and the data provenance

    Not applicable for the types of tests conducted. These are primarily engineering, laboratory, and compliance tests, not clinical studies with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth as typically understood in AI/ML clinical studies (e.g., expert consensus on image interpretation) is not relevant to these engineering and compliance tests.

    4. Adjudication method for the test set

    Not applicable for the types of tests conducted.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an electrosurgical system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is an electrosurgical system.

    7. The type of ground truth used

    For the sterility testing, the ground truth is the established Sterility Assurance Level (SAL) of 10^-6 and adherence to the listed ISO standards.
    For other tests, the "ground truth" is adherence to the specified engineering and biocompatibility standards.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1