510(k) Data Aggregation

    K Number
    K210347

    Validate with FDA (Live)

    Device Name
    VSP System
    Manufacturer
    Date Cleared
    2021-04-26

    (77 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VSP® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the VSP® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, and surgical guides for use in maxillofacial surgery. The VSP® System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.

    Device Description

    The VSP® System utilizes a combination of Commercial Off-The-Shelf (COTS) and custom software to manipulate 3D medical images (CT based systems) to create virtual and physical anatomical models, templates, surgical guides, and surgical plans for reconstructive surgical procedures.

    AI/ML Overview

    The provided text is a 510(k) Summary for the VSP® System and primarily focuses on demonstrating substantial equivalence to a predicate device, rather than proving that the device meets specific acceptance criteria through a study with a test set, expert ground truth, and statistical analysis as would be done for an AI/ML-driven device.

    The VSP® System is described as a software system and image segmentation system for maxillofacial surgery, which processes imaging information (CT-based) to create digital models, physical outputs (anatomical models, surgical guides), and aids in pre-operative simulation. The primary change in the current submission is the addition of a new material (Accura® ClearVue™) for anatomical models, not a change or new feature related to an AI/ML algorithm's diagnostic or predictive performance.

    Therefore, the document does not contain the information required to answer the questions regarding acceptance criteria for an AI/ML device, such as performance metrics (e.g., sensitivity, specificity), sample sizes for test sets, expert ground truth establishment, or MRMC studies. The "performance testing" described (Operational Qualification, Performance Qualification, Cleaning Validation, Sterilization Validation, Biocompatibility Validation) are related to manufacturing processes and material safety, not the diagnostic or prognostic performance of a software algorithm on a test dataset.

    The only "performance testing" mentioned is:

    • Operational Qualification: Assessed manufacturing process over allowable parameters. All test method acceptance criteria were met.
    • Performance Qualification: Assessed manufacturing process repeatability at nominal parameters. All test method acceptance criteria were met.
    • Cleaning Validation: Performed in accordance with AAMI TIR 30. All test method acceptance criteria were met.
    • Sterilization Validation: Performed in accordance with ISO 17665-1:2006 to a SAL of 10^-6. All test method acceptance criteria were met.
    • Biocompatibility Validation: Evaluated in accordance with ISO 10993-1. Sensitization, irritation, and acute systemic toxicity testing conducted on the subject device manufactured from Accura ClearVue are within the pre-defined acceptance criteria.

    The document explicitly states: "Clinical Performance Data - Clinical testing was not necessary for the determination of substantial equivalence." This further confirms that no study of the software's performance on a diagnostic/predictive task, using clinical data and ground truth, was performed or presented in this 510(k) summary.

    In summary, based on the provided text, it is not possible to fill out the requested information as it pertains to an AI/ML driven device's performance study because such a study was not conducted or reported for this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192192

    Validate with FDA (Live)

    Device Name
    VSP System
    Manufacturer
    Date Cleared
    2020-02-26

    (197 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VSP® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The is processed by the VSP® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, templates, and surgical guides for use in maxillofacial surgery. The VSP® System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.

    Device Description

    The VSP System utilizes a combination of Commercial Off-The-Shelf (COTS) and custom software to manipulate 3D medical images (CT based systems) to create virtual and physical anatomical models, templates, surgical guides, and surgical plans for reconstructive surgical procedures.

    AI/ML Overview

    The provided text describes the VSP® System, a software and image segmentation system for maxillofacial surgery, and its 510(k) submission (K192192) to the FDA. The submission focuses on demonstrating substantial equivalence to a predicate device (VSP® System, K133907), with modifications primarily related to the expansion of materials used for guides (Polyamide and Titanium Alloy).

    However, the provided document does not contain the specific details required to fully address all parts of your request regarding acceptance criteria and the study that proves the device meets them, especially in the context of AI/algorithm performance.

    The document primarily discusses design validation and process performance qualification, which are standard engineering and manufacturing tests, rather than clinical performance studies of an AI algorithm's diagnostic or predictive capabilities. It explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence." This indicates that the regulatory pathway for this device did not require a human-in-the-loop or standalone AI performance study against a clinical ground truth as might be expected for an AI-driven diagnostic device.

    Here's an attempt to answer your questions based on the available information, with clear indications where the information is not provided in the text:


    1. A table of acceptance criteria and the reported device performance

    The document mentions that acceptance criteria were established and met for various tests. However, it does not provide a specific table or quantified acceptance criteria and reported performance values for the device's accuracy in image segmentation or surgical planning. Instead, it describes general conclusions for each test:

    Acceptance Criteria CategoryReported Device Performance
    Design Validation"All acceptance criteria were met." specifically for "accuracy of cutting location/trajectory, drilling location/trajectory, and bony segment length utilizing subject device guides on bone models."
    Process Performance Qualification"All test method acceptance criteria were met," verifying "Both digital and physical outputs from all manufacturing processes... against design specifications."
    Cleaning Validation"All test method acceptance criteria were met."
    Sterilization Validation"All test method acceptance criteria were met."
    Biocompatibility Validation"Within the pre-defined acceptance criteria," adequately addressing biocompatibility.
    Mechanical Performance Validation"Subject devices perform equivalent or better than the predicate device" in flexural deformation testing.
    Packaging Validation"All test method acceptance criteria were met."

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: The document does not specify the sample size for any of the validation tests.
    • Data Provenance: The document generally refers to "bone models" for design validation cases and "representative of orthognathic and reconstruction procedures" for process performance qualification. It does not specify the country of origin of the data or whether it was retrospective or prospective. Given the nature of the device (surgical planning and guide creation from medical scans), the "data" would likely be patient CT scans, but the details of their origin are not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts used to establish ground truth for the test set. It mentions "user needs and intended use of supporting maxillofacial surgeries" for design validation, implying involvement of surgeons, but no specific details on their role in ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not mention any adjudication method used for establishing ground truth or evaluating the test set results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence." Therefore, no MRMC comparative effectiveness study was conducted to evaluate human reader improvement with or without AI assistance. The regulatory pathway for this device, a software system for surgical planning and guide production, did not require such a clinical study to demonstrate safety and effectiveness for substantial equivalence.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A formal standalone performance study of the algorithm's diagnostic or predictive capabilities (e.g., measuring segmentation accuracy against ground truth in a clinical context) is not specifically described in the document as a separate study. The "Design Validation" and "Process Performance Qualification" likely involved evaluating the algorithm's output (digital models, physical guides) against design specifications, which could be considered a form of standalone testing of the system's output creation accuracy, but not a clinical diagnostic performance study.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The document primarily refers to "design specifications" as the basis for validation. For "Design Validation," it assessed "accuracy of cutting location/trajectory, drilling location/trajectory, and bony segment length utilizing subject device guides on bone models." This implies the ground truth for these tests was pre-defined ideal anatomical measurements or surgical plans on synthetic or cadaveric models, rather than clinical pathology or outcomes data from human patients.

    8. The sample size for the training set

    The document does not mention a training set or any details about machine learning model development. The description of the VSP System implies it uses "custom software" and "image transfer and manipulation," but does not explicitly state it employs machine learning or AI that requires a distinct "training set." It's possible the system uses rule-based algorithms or traditional image processing techniques rather than data-driven machine learning models.

    9. How the ground truth for the training set was established

    Since a training set is not mentioned, the method for establishing its ground truth is not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133907

    Validate with FDA (Live)

    Device Name
    VSP SYSTEM
    Date Cleared
    2014-01-22

    (30 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Modeling VSP® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as CT based system. The is processed by the VSP® System and the result is an output data file that may then be provided as digital models or used as ingut to a rapid prototyping portion of the system that produces physical outputs including and surgical guides for use in maxillofacial surgery. The VSP® System is also intended as a pre-operative software tool for simulating surgical reatment options.

    Device Description

    The Medical Modeling VSP® System is a collection of software and associated additive manufacturing (rapid prototyping) equipment intended to provide a variety of outputs to support orthognathic or reconstructive surgery. The system uses electronic medical images of the patient's anatomy with input from the physician, to manipulate original patient images for planning and executing surgery. The system manipulate enginal patient specific outputs including, anatomical models (physical and digital), surgical templates / guides, splints, and patient specific case reports.

    Following the Medical Modeling Quality System and specific Work Instructions, trained employees utilize a combination of Commercial Off-The-Shelf (COTS) and trantom software to manipulate 3D medical scan images which can include Computed Tomography (CT), Cone Beam CT (CBCT), 3D scan images from patient physical models (stone models of the patient's teeth), and / or standard (non-bent) or priyered model implant files to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during planning and process telivery of the final outputs. While the process and data-flow can vary and phor to activery of the particular requirements of a given patient and physician, the following description outlines the functions of key sub-components of the system, following doben pour can be the defined system outputs. It should be noted that the system is operated only by trained Medical Modeling employees, and the the oyotan to open area the information. The physician only provides input for priysioian doos not anothy mactive feedback through viewing of digital models of model mempulation are modified by the engineer during the planning session.

    The VSP® System is made up of 14 individual pieces of software and additional manufacturing equipment integrated to provide a range of anatomical models (physical and digital), surgical templates / guides and patient-specific case reports for (priyalour and digital), bargham, burgical region. Surgical guides for graft bone harvesting from patient bone such as, but not limited to, the fibula and hip, are also produced by the system.

    The VSP® System requires an input 3D image file from medical imaging systems (i.e. CT) and / or a standard or patient contoured implant file. This input is then used, (i.e. ST) and from the prescribing physician to provide the following categories of with ourport reconstructive surgery. Each system output is designed with outpute to oupper roosiewed by the physician prior to finalization and distribution. priysician input, and foviotity direct physician involvement to reduce the criticality of the outputs.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Medical Modeling Inc. VSP® System:

    This document describes a "Special 510(k) for Modified VSP® System," which focuses on verifying that a modified version of an already cleared device (the predicate device) is substantially equivalent to the original. Therefore, the study details are primarily about demonstrating equivalence rather than establishing entirely new performance criteria or efficacy for a novel device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of specific, quantitative "acceptance criteria" for the VSP® System's performance in the way one might expect for diagnostic accuracy (e.g., sensitivity, specificity thresholds). Instead, the acceptance criteria are framed around demonstrating equivalence to the predicate device and conformity to specifications for the modifications introduced.

    Acceptance Criteria CategoryReported Device Performance / Assessment Method
    Substantial Equivalence"The modified VSP® System conforms to specifications... and that its performance is substantially equivalent to the unmodified VSP® System." (Section 9.1) "All Design, Process, and other Verification and Validation testing... showed conformity to pre-established specifications and acceptance criteria. The tests were performed in order to demonstrate device performance and substantial equivalence of the system to the unmodified version of the device." (Section 9.5) "Based on a comparison of the intended use and technological characteristics, the modified VSP® System is substantially equivalent to the identified predicate device." (Section 10.1)
    Patient Contoured Implant Files"The final output .STL files the cases designed using a patient contoured implant .STL file as input demonstrated clinical and / or dimensional equivalency to .STL files created using a standard implant .STL file as input." (Section 9.3)
    Clinical Equivalence (for Implant Files)"The reconstructive surgeon approved both design, which indicated that there is no clinically significant difference between the implant template .STL file created from a patient contoured implant .STL file and a standard implant .STL file." (Section 9.3)
    Conformity to Specifications (Software Updates)"Software installation qualification" and "Software validation" tests were executed for each software version upgrade. (Section 9.4.1.1.1, 9.4.1.1.2)
    Conformity to Specifications (Additional SLA Hardware)"Installation Qualification (IQ)", "Operational Qualification (OQ)", "Performance Qualification (PQ)", and "Software Validation" were performed for the additional Stereolithography Apparatus (SLA) model and associated wash tanks. (Section 9.4.1.2.1-9.4.1.2.4)
    ASTM Standards (Accessories)"The modified and unmodified VSP® System Accessories are manufactured to ASTM F899-11". (Section 8.1.6)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" with a quantifiable sample size in terms of number of patients or cases. Instead, the testing appears to be focused on specific modifications and their impact.

    • For Patient Contoured Implant Files: "The final output .STL files the cases designed using a patient contoured implant .STL file as input..." (Section 9.3). The text refers to "cases" (plural) but does not provide a specific number.
    • For Software Updates: "Four (4) software components were upgraded". Tests were executed "for each software version upgrade." (Section 9.4.1.1). The direct sample size of test cases for these software validations is not detailed.
    • For Additional SLA: "An additional SLA model and associated wash tanks have been added..." (Section 9.4.1.2). The testing here (IQ, OQ, PQ, Software Validation) is for the hardware and its integration, not a patient-case-based test set.

    Data Provenance: Not explicitly stated, though the context implies internal testing by Medical Modeling Inc. The document does not mention external data, country of origin of data, or if it was retrospective or prospective in the sense of a clinical trial. It describes a "process validation" (Section 9.2) focused on a "worst-case patient contoured implant file" and subsequent digital file outputs that met requirements.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • For Patient Contoured Implant Files: "The reconstructive surgeon approved both design..." (Section 9.3). This indicates one expert, identified as "the reconstructive surgeon." The specific qualifications (e.g., years of experience) are not provided beyond the title "reconstructive surgeon."

    4. Adjudication Method for the Test Set

    • For Patient Contoured Implant Files: The document states, "The reconstructive surgeon approved both design." This suggests a single expert's approval was sufficient, implying no formal adjudication method (e.g., 2+1 or 3+1 consensus) among multiple experts was necessary for this specific part of the testing. The "clinical input and review from the physician during planning and processing" (Section 7.2) and "reviewed by the physician prior to finalization and distribution" (Section 7.4) of system outputs also points to individual physician review.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This submission is for a special 510(k) for a modified Virtual Surgical Planning (VSP) system, where the modifications primarily relate to input file types (patient contoured implant files) and internal manufacturing/software processes. The device itself is a software system for surgical planning and creating physical guides/models, not a diagnostic AI system meant to assist human readers in interpretation. The goal was substantial equivalence, not improved human reader performance with AI.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The VSP® System is itself a software component, so in a sense, its underlying algorithms operate in a "standalone" fashion to process inputs and generate outputs. However, the system is explicitly designed with "human-in-the-loop" aspects:

    • "The process requires clinical input and review from the physician during planning and delivery of the final outputs." (Section 7.2)
    • "The system is operated only by trained Medical Modeling employees, and the physician only provides input for interactive feedback through viewing of digital models... models are modified by the engineer during the planning session." (Section 7.2)
    • "Each system output is designed with physician input, and direct physician involvement to reduce the criticality of the outputs." (Section 7.4)

    So, while the software performs its functions without direct human intervention at every step of its internal computations, the overall device workflow is inherently human-("clinician and trained employee")-in-the-loop, with outputs subject to expert review and approval. The testing described (e.g., "clinical and/or dimensional equivalency," "reconstructive surgeon approved") assesses the output of the system in the context of its intended use, which includes human oversight.


    7. The Type of Ground Truth Used

    The ground truth for the "patient contoured implant file" modification was based on:

    • Expert Consensus/Approval: "The reconstructive surgeon approved both design, which indicated that there is no clinically significant difference..." (Section 9.3). This relies on the expert judgment of a reconstructive surgeon.
    • Dimensional Equivalency: The comparison also sought "dimensional equivalency" (Section 9.3), implying quantitative measurements were involved, compared against established specifications or direct comparison of the designs.

    For other internal modifications (software, hardware), the ground truth was conformity to "pre-established specifications and acceptance criteria" (Section 9.5) as demonstrated through standard qualification processes (IQ, OQ, PQ, software validation).


    8. The Sample Size for the Training Set

    The document does not specify a training set sample size. The VSP® System is not described as a machine learning or AI system that undergoes "training" in the conventional sense (i.e., learning from large datasets to develop an algorithm). It's a system that utilizes a combination of Commercial Off-The-Shelf (COTS) and custom software to manipulate 3D medical scan images based on predefined algorithms and engineering principles (Section 7.2). The modifications described are related to inputs, software updates, and manufacturing hardware, not algorithm retraining.


    9. How the Ground Truth for the Training Set Was Established

    As no training set is described (see point 8), the concept of establishing ground truth for a training set is not applicable to this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120956

    Validate with FDA (Live)

    Device Name
    VSP SYSTEM
    Date Cleared
    2012-12-12

    (257 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Modeling VSP® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the VSP® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, templates, and surgical guides for use in maxillofacial surgery. The VSP® System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.

    Device Description

    The Medical Modeling VSP® System is a collection of software and associated additive manufacturing (rapid prototyping) equipment intended to provide a variety of outputs to support reconstructive surgery. The system uses electronic medical images of the patients' anatomy with input from the physician, to manipulate original patient images for planning and executing surgery. The system produces a variety of patient specific outputs including; anatomical models (physical and digital), surgical templates / guides, splints, and patient specific case reports.

    Following the Medical Modeling Quality System and specific Work Instructions, trained employees utilize a combination of Commercial Off-The-Shelf (COTS) and custom software to manipulate 3-D medical scan images which can include Computed Tomography (CT), Cone Beam CT (CBCT), and/or 3-D scan images from patient physical models (stone models of the patient's teeth) to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during planning and prior to delivery of the final outputs. While the process and dataflow vary somewhat based on the particular requirements of a given patient and physician, the following description outlines the functions of key sub-components of the system, and how they interact to produce the defined system outputs. It should be noted that the system is operated only by trained Medical Modeling employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session.

    The VSP® System is made up of 13 individual pieces of software and 9 pieces of manufacturing equipment integrated to provide a range of anatomical models (physical and digital), surgical guides, and patient specific planning reports for reconstructive surgery in the maxillofacial region. Surgical guides for graft bone harvesting which can include the fibula and hip are also produced by the system.

    The VSP® System requires an input 3-D image file from medical imaging systems (i.e. CT). This input is then used, with support from the prescribing physician to provide the following potential outputs to support reconstructive surgery. Each system output is designed with physician input, and reviewed by the physician prior to finalization. All outputs are used only with direct physician involvement reducing the criticality of the outputs.

    System Outputs:
    Anatomical Models
    Surgical Positioning Templates / Guides
    Osteotomy Templates / Guides
    Plate Bending Templates/Guides
    Patient Specific Case Reports

    The VSP® System also contains a set of Stainless Steel Cutting and Drill Inserts (VSP® System Accessories) which are intended to be used by the physician to guide cutting and drilling activities during the surgical procedure. The inserts fit into a standard slot/hole in the cutting/drill guides, and can be used across all the VSP® System guides and templates.

    AI/ML Overview

    The Medical Modeling VSP® System is a collection of software and associated additive manufacturing (rapid prototyping) equipment intended to provide a variety of outputs to support reconstructive surgery. The system uses electronic medical images of the patients' anatomy with input from the physician, to manipulate original patient images for planning and executing surgery.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The provided document describes a 510(k) submission for a medical device and outlines various verification and validation testing performed. However, it does not explicitly state a table of specific acceptance criteria with quantifiable metrics (e.g., "accuracy must be > 95%") alongside reported device performance for the VSP® System.

    Instead, the document details a comprehensive set of engineering and process validation activities. The general acceptance criterion is implicit: "All, Design, Process, and Other Verification and Validation Testing, which was required as a result of risk analysis and design impact assessments, showed conformity with pre-established specifications and acceptance criteria." This implies that individual test plans for each category would have contained the specific acceptance criteria.

    The study proves the device meets the acceptance criteria through a series of "Verification and Validation testing" categories. The overarching goal was to demonstrate that "the manufacturing process could correctly translate 3 dimensional electronic models of patient anatomies, templates and guides into matching physical models within the allowed tolerance ranges from the process specifications."

    Given the lack of a direct table in the provided text, I will infer the categories of performance testing as "acceptance criteria" through successful completion, and "reported device performance" as the successful demonstration against these criteria.

    Acceptance Criteria Category (from 9.3)Reported Device Performance (Implied from 9.4)
    Installation Qualification (IQ)Showed conformity with pre-established specifications and acceptance criteria.
    Operational Qualification (OQ)Showed conformity with pre-established specifications and acceptance criteria.
    Performance Qualification (PQ)Showed conformity with pre-established specifications and acceptance criteria.
    Software ValidationShowed conformity with pre-established specifications and acceptance criteria; Verified independent subsystems, interfaces, and validated the fully integrated system.
    Simulated Use (Planning Session)Showed conformity with pre-established specifications and acceptance criteria.
    Labeling ValidationShowed conformity with pre-established specifications and acceptance criteria.
    Cleaning ValidationShowed conformity with pre-established specifications and acceptance criteria.
    Sterilization ValidationShowed conformity with pre-established specifications and acceptance criteria.
    BiocompatibilityShowed conformity with pre-established specifications and acceptance criteria.
    Packaging ValidationShowed conformity with pre-established specifications and acceptance criteria.

    Additional Information on the Study:

    1. Sample size used for the test set and the data provenance:
      The document does not specify a distinct "test set" sample size in terms of patient cases or imaging data. The testing described is focused on the manufacturing process and the system's ability to translate 3D electronic models into physical models within tolerance. Therefore, the "sample" for testing would likely refer to the number of physical models, guides, or templates produced and evaluated during OQ, PQ, and Simulated Use. These details are not provided in the summary.
      Data provenance (e.g., country of origin, retrospective/prospective) is not mentioned, as the focus is on system functionality rather than clinical performance on a patient cohort for this type of device (a system for creating models and guides).

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      The document states that "The process requires clinical input and review from the physician during planning and prior to delivery of the final outputs." and "All outputs are used only with direct physician involvement reducing the criticality of the outputs." This indicates that physicians are key domain experts providing input and review. However, the specific number of experts used for establishing "ground truth" during validation testing (e.g., for evaluating the accuracy of physical models against digital plans) and their qualifications are not detailed. It can be inferred that "trained Medical Modeling employees" (engineers/CAD specialists) operate the system, and physicians provide clinical feedback.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
      The document does not specify any formal adjudication method like "2+1" or "3+1" for establishing ground truth or evaluating the test set results. The process involves physician input and review, but a formal adjudication process for test results is not detailed in this summary.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No MRMC study was conducted or reported. The VSP® System is a tool for planning and manufacturing patient-specific devices, not an AI-assisted diagnostic or interpretation system that would involve "human readers" in the typical sense of an MRMC study. There is no mention of "AI assistance" in this context.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      The system is explicitly described as having "human-in-the-loop" performance. "The system is operated only by trained Medical Modeling employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session." Therefore, no standalone (algorithm only) performance was presented or is applicable to the intended use.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For mechanical and dimensional testing (OQ, PQ), the ground truth for physical models would be established by comparing them to the digital models/plans generated by the system, referencing "allowed tolerance ranges from the process specifications." This essentially uses the digitally designed model as the 'truth' against which the manufactured physical output is measured. For "Simulated Use (Planning Session)," the ground truth would likely be based on expert (physician) review and acceptance of the surgical plan and resulting outputs as clinically appropriate and accurate based on the patient's anatomy.

    7. The sample size for the training set:
      The document does not describe a "training set" in the context of machine learning or AI, as this system is not presented as an AI-driven device. It is a software system for image transfer, segmentation, and pre-operative planning, utilizing COTS and custom software operated by trained personnel. Therefore, the concept of a "training set" for an algorithm is not applicable here.

    8. How the ground truth for the training set was established:
      As there is no mention of an algorithm training set, this question is not applicable. The software within the VSP® system undergoes standard software validation (9.3.4), where its performance against defined requirements and interfaces (likely functional specifications) would be the "ground truth."

    Ask a Question

    Ask a specific question about this device

    K Number
    K984350

    Validate with FDA (Live)

    Device Name
    VSP SYSTEM
    Date Cleared
    1999-01-20

    (47 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VSP System is indicated for degenerative spondylolisthesis, in skeletally mature patients, with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). Levels of fixation are for the thoracic, lumbar and sacral spine.

    The VSP System is also indicated for pedicle screw fixation for severe spondylolisthesis (Grades 3 and 4) at L5-S1, in skeletally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is attained. Levels of fixation are from L3-S1.

    Device Description

    The primary purpose of this premarket notification is to add indications to the marketing clearance for the pedicle screws which may be used as a spinal anchor in the VSP System. All implant components are manufactured of either ASTM F-138 or F-1314 stainless steel or ASTM F-136 titanium alloy.

    AI/ML Overview

    The provided text describes a 510(k) submission for the ISOLA SPINAL SYSTEM Pedicle Screw Indications (VSP System), which is an orthopedic device. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving clinical efficacy or safety through new clinical studies with defined acceptance criteria and performance metrics.

    Therefore, the input does not contain information about:

    1. Acceptance criteria and reported device performance
    2. Sample size and data provenance for a test set
    3. Number and qualifications of experts for ground truth
    4. Adjudication method
    5. Multi-Reader Multi-Case (MRMC) comparative effectiveness study
    6. Standalone (algorithm only) performance
    7. Type of ground truth used
    8. Sample size for the training set
    9. How ground truth for the training set was established

    The document is a 510(k) summary and an FDA clearance letter, which means that the device's acceptable performance is demonstrated by its substantial equivalence to an existing, legally marketed device. The core of this submission is to add new indications for use for the pedicle screws within the VSP System, not to present new performance data from a clinical study.

    Therefore, I cannot populate the requested table and answer the questions based on the provided text, as the information is not present.

    The document states: "The VSP System manufactured from either stainless steel or titanium alloy is substantially equivalent, for purposes of this 510(k) adding indications, to itself. This 510(k) seeks to add labeled indications pursuant to a reclassification order. The device design remains the same." This explicitly indicates that no new performance data or studies defining acceptance criteria are being presented; rather, the focus is on regulatory reclassification and expanded indications for an existing, already cleared device.