Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K141288

    Validate with FDA (Live)

    Date Cleared
    2014-08-22

    (98 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Voyant electrosurgical generator is indicated for use with Voyant devices in open and laparoscopic surgical procedures where the electrosurgical ligation of vessels or tissue bundles is desired.

    The Voyant 5mm Sealer-Divider device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in laparoscopic procedures where the ligation and division of vessels and tissue bundles is desired.

    The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

    The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

    Device Description

    The Applied Medical Voyant Electrosurgical Generator (ESG) is a tabletop radiofrequency (RF) power supply designed for use in electrosurgery. It operates outside the sterile field and is equipped with receptacles for Applied Medical bipolar devices. The ESG enclosure is constructed of various metals and polymers and houses all electrical hardware and software components. The front panel features an LCD and backlit buttons for the navigation, adjustment, and selection of ESG and device settings.

    The Applied Medical Voyant 5mm Laparoscopic Tissue Sealer-Divider instrument is designed for use with the Voyant ESG. This device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 7mm in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The device features a mechanical, useractuated blade for the division of sealed tissue.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Voyant® Electrosurgical Generator and Voyant® 5mm Sealer-Divider System. It outlines the device's indications for use, technological characteristics, and performance testing conducted to demonstrate substantial equivalence to predicate devices. However, the document does not contain a table of acceptance criteria and reported device performance in a way that directly answers the request. It states that testing was performed but does not list specific numerical criteria or outcomes in a clear tabular format for direct comparison.

    Therefore, for the information requested in your prompt, I can only extract what is available and highlight what is missing.

    Here's an attempt to answer based on the provided text, noting the limitations:

    1. A table of acceptance criteria and the reported device performance

    The document states its Indications for Use which implicitly sets a performance expectation: "The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device."

    However, specific numerical acceptance criteria (e.g., minimum burst pressure, maximum thermal spread) and the precise reported performance values against these criteria are not explicitly presented in a table. The document generally concludes that the system is "substantially equivalent in performance to the predicate system with respect to vessel sealing performance and local tissue effects."

    Acceptance Criteria (Implied from Indications for Use and Testing Goals)Reported Device Performance (General Statement in Document)
    Ability to seal and divide vessels up to 7mm in diameter."substantially equivalent in performance to the predicate system with respect to vessel sealing performance"
    Ability to seal and divide tissue bundles that can be captured in jaws."substantially equivalent in performance to the predicate system with respect to vessel sealing performance"
    Acceptable local tissue effects (thermal damage)."substantially equivalent in performance to the predicate system with respect to... local tissue effects."
    Electrical, mechanical, and functional capabilities.Tested in accordance with IEC 60601-1-2 and IEC 60601-2-2. Basic mechanical and functional capabilities were tested.
    System safety and efficacy.Evaluated in preclinical bench and animal studies; chronic hemostasis evaluated.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: The document mentions "preclinical bench (ex vivo porcine vessels/tissue)" and "acute animal study (in vivo canine model)," and "a chronic study (in vivo canine model)." No specific sample sizes (e.g., number of porcine vessels, number of canine subjects) are provided.
    • Data Provenance:
      • Country of Origin: Not specified.
      • Retrospective or Prospective: The preclinical and animal studies are prospective in nature, as they involve testing the device under controlled conditions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the studies are device performance studies on animal/ex vivo models, not studies requiring expert interpretation of diagnostic images or clinical outcomes in humans to establish ground truth. The "ground truth" for these types of studies would be objective measurements (e.g., burst pressure measurements, histological assessment of thermal damage) performed by trained technicians or researchers. The document does not specify the number or qualifications of individuals performing these measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable for the type of device performance studies described. Adjudication methods are typically used in clinical or diagnostic studies to resolve discrepancies in expert interpretation of data (e.g., radiology reads).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The document describes the performance of an electrosurgical device, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is an electrosurgical instrument used by a surgeon, not a standalone algorithm. Its performance inherently involves human-in-the-loop operation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance studies appears to be based on:

    • Objective physical measurements: Energy output, mechanical strength, durability, electrical safety, burst pressures of sealed vessels.
    • Histopathological assessment: Evaluation of thermal damage in tissue, and chronic hemostasis in the animal studies.
    • Comparative performance: Against a legally marketed predicate device to establish "substantial equivalence."

    8. The sample size for the training set

    This is not applicable. The document describes a physical medical device (electrosurgical generator and sealer-divider), not a machine learning or AI model that requires a "training set."

    9. How the ground truth for the training set was established

    This is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1