Found 4 results

510(k) Data Aggregation

    K Number
    K160810

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-05-20

    (57 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion DR + Wireless allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.

    Device Description

    The ViZion DR + Wireless system represents the straightforward integration of a modified digital x-ray receptor panel and our previously cleared software. This is a MODIFICATION of our clearance K152279 wherein we have changed the panel scintillator to Cesium Iodide in addition to having the GOS scintillator available. Changing the scintillator is the only modification.

    ViZion DR + Wireless is a Digital Radiography system, featuring an integrated flat panel digital detector (FPD). ViZion DR + Wireless is designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the Viztek PACS and the new receptor panel were not changed.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Vizion DR + Wireless device, presented in the requested format:

    1. A table of acceptance criteria and the reported device performance

    The document describes a modification to an existing device (K152279) where the scintillator material was changed from Gadolinium Oxysulfide (GOS) to Cesium Iodide (CSI). The primary acceptance criteria for equivalence appear to be related to imaging performance metrics (MTF and DQE) and clinical image quality.

    Acceptance Criteria / CharacteristicPredicate Device (Viztek ViZion DR + Wireless K152279 - GOS Scintillator)Modified Device (ViZion DR + Wireless, CSI Scintillator)Performance vs. Predicate
    Intended UseUnchangedUnchangedEquivalent
    ConfigurationUnchangedUnchangedEquivalent
    Digital Panel (size, pixel size)iRay Technology Mars1417V-PSI 14" x 17", 150 µmiRay Technology Mars1417V-TSI 14" x 17", 150 µmNot a meaningful difference
    Internal image storage200 full size imagesUnchangedEquivalent
    Image acquisition time< 5 SecondsUnchangedEquivalent
    DICOM 3YesYesEquivalent
    A/D Conversion14 bitUnchangedEquivalent
    ScintillatorDRZ-Plus Gd2O2S:Tb (GOS)CSI (Cesium Iodide)Changed
    MTF (0.5 (1/mm))0.750.83Better (0.83 > 0.75)
    DQE (0.5 (1/mm))0.270.55Better (0.55 > 0.27)
    InterfaceUnchangedUnchangedEquivalent
    Power sourceUnchangedUnchangedEquivalent
    Electrical safety and EMCComplies with IEC 60601-1, IEC 60601-1-2, FCC Part 15SameEquivalent
    Clinical Image QualityPredicate panel imagesModified panel imagesAs good as or better

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated for either the quantitative (MTF/DQE) or clinical image evaluation. For the clinical images, it states "Clinical images were acquired," implying a set of images, but no number is given.
    • Data Provenance: Not explicitly stated. The document is from the US FDA, and the company address is in North Carolina, USA, suggesting the study was likely conducted in the USA, but this is not confirmed. It is implied to be prospective or specifically acquired for this evaluation, as it mentions "Clinical images were acquired and evaluated..."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: One
    • Qualifications of Experts: "a board certified radiologist" (no mention of years of experience).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: None. A single board-certified radiologist performed the evaluation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. The study described is a comparison of two imaging technologies (different scintillators), not an AI-assisted reading study. The device is a digital X-ray receptor, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. This device is an imaging acquisition system, not an AI algorithm. The performance evaluation focuses on the image quality produced by the system itself before interpretation. The quantitative metrics (MTF, DQE) are "standalone" in the sense that they measure the physical performance of the sensor, but this is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth:
      • For MTF and DQE: These are objective physical measurements of image quality, not dependent on a "ground truth" derived from patient data in the same way clinical diagnostics are. They are measured against known physical standards.
      • For Clinical Image Quality: The "ground truth" for clinical image quality comparison was the subjective opinion of a board-certified radiologist, who evaluated whether images from the modified panel were "as good as or better than" images from the predicate panel. This is a form of expert opinion/subjective evaluation rather than an objective clinical endpoint like pathology.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable. The Vizion DR + Wireless is a hardware device (digital X-ray detector) and associated software, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The "software" was previously cleared and confirmation testing was performed after the hardware modification.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable, as there is no "training set" for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152279

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-09-10

    (29 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion DR + Wireless allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.

    Device Description

    The ViZion DR + Wireless system represents the straightforward integration of a modified digital x-ray receptor panel and our previously cleared software. This is a MODIFICATION of our clearance K123644 wherein we have changed panel to a wireless version of the previously cleared panel/software combination. Going wireless (with batterv operation possible) are the ONLY modifications.

    ViZion DR + Wireless is a Digital Radiography system, featuring an integrated flat panel digital detector (FPD). ViZion DR + Wireless is designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the Viztek PACS and the new receptor panel were not changed. Our predicate is ViZion DR +, K123644. We do not supply the router, but any router meeting the IEEE 802.11 a/b/g/n specifications will work. With the advent of AED (automatic exposure detection) by the panel, integration with or connection to specific generators is no longer required. If the user decides NOT to use AED, wired synchronization with Sedecal Series SHF generators is known to work.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Viztek ViZion DR + Wireless device based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provided does not explicitly state formal "acceptance criteria" in a quantitative, measurable sense for the device, beyond equivalence to a predicate device. Instead, it focuses on demonstrating substantial equivalence through a comparison of technical specifications and performance characteristics, some of which could be considered de-facto acceptance criteria derived from the predicate's performance.

    Criterion (Implied/Predicate Based)Predicate Device Performance (Viztek ViZion + DR K123644)Vizion DR + Wireless Performance
    Intended UseGeneral radiographic examinations (skull, chest, shoulders, spine, abdomen, pelvis, extremities), excluding fluoroscopy, angiography, mammographyUNCHANGED (Same)
    Digital Panel - Pixel size150 µm150 µm
    Digital Panel - Pixels2288x2800 (6.4 million pixels)2304x2800 (6.4 million pixels)
    Software OutputDICOM imageSAME as K123644
    DICOM 3YesYES
    A/D Conversion14 bitSAME
    MTF (0.5 (1/mm)0.750.75 (Essentially the same)
    DQE (0.5 (1/mm)0.230.27 (Essentially the same)
    ScintillatorGOS scintillation screenUNCHANGED
    InterfaceGigabit EthernetWired: Gigabit Ethernet (1000BASE-T); Wireless: IEEE802.11a/b/g/n
    Power SourceAC LineAC Line and/or Rechargeable Lithium Battery (3 hr run time)
    Electrical SafetyIEC 60601-1SAME
    EMCIEC 60601-1-2SAME, plus complies with FCC Part 15 Rules and Regulations
    Image QualityImplied to be acceptable based on predicate clearance"as good as or better than the images acquired with the predicate panel"

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document simply states "Clinical images were acquired" and "bench, and test laboratory results." It does not specify the number of clinical images or cases used for evaluation.
    • Data Provenance: The document does not explicitly state the country of origin. Given the submission is to the U.S. FDA, it is likely that the clinical image acquisition was performed in the U.S. The study appears to be prospective in the sense that new images were acquired specifically for this evaluation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: One board-certified radiologist.
    • Qualifications: "board certified radiologist." No specific tenure or experience level is mentioned beyond board certification.

    4. Adjudication Method for the Test Set

    • Adjudication Method: None mentioned. A single radiologist evaluated the images.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done. The evaluation was based on a single radiologist's assessment of clinical images from the new panel compared to the predicate, and bench testing.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable / not a software-only algorithm. The device is a digital X-ray receptor panel (hardware) with associated software. The "performance" being evaluated is of the integrated system and its ability to produce diagnostic images. The comparison focuses on the image quality produced by the device, which is inherently designed for human interpretation.

    7. Type of Ground Truth Used

    • Expert Consensus / Expert Interpretation: The ground truth for image quality was established by the interpretation of a single board-certified radiologist. There is no mention of pathology, outcomes data, or other objective measures for ground truth.

    8. Sample Size for the Training Set

    • Not applicable / Not explicitly a machine learning model requiring a training set. The device is primarily a hardware modification (wireless panel) to an existing cleared system. While there's "software," the document states "Since no software modifications were required, confirmation testing was performed." This suggests the software itself was not re-trained or developed with a new dataset for this specific submission, but rather integrated with the new wireless panel.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As stated above, a distinct "training set" for a machine learning model isn't indicated as part of this device's submission or the testing described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K112661

    Validate with FDA (Live)

    Device Name
    VIZION DR
    Manufacturer
    Date Cleared
    2011-10-20

    (37 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ViZion DR is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.

    Device Description

    The ViZion DR system represents the straightforward integration of two cleared devices: ViZion DR, K102123 and K102587, the Samsung Digital Flat Panel. ViZion DR is a Digital Radiography system, featuring an integrated flat panel digital detector (FPD) (K102587, Samsung Flat-Panel X-Ray Detector), made by Samsung Mobile Display Co., Ltd. (this is the 510(k) for the flat panel detector) and Viztek's proprietary OPAL-RAD PACS image viewing and acquire interface software technology, (K063337) which incorporates state of the art object-oriented software and connectivity. ViZion is designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a film less. environment and improves cost effectiveness. The major functions and principle of operation of the Viztek PACS and Samsung FPD were not changed.

    AI/ML Overview

    This is a 510(k) premarket notification for a new version of the Viztek ViZion DR, a Digital Radiography (DR) system. The submission focuses on replacing the digital flat panel detector (FPD) with a new model while maintaining the existing software and overall functionality. As such, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are framed in terms of demonstrating substantial equivalence to a predicate device, rather than proving a specific diagnostic accuracy against a clinical ground truth.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the demonstration of substantial equivalence to the predicate device, specifically regarding safety and effectiveness, and the absence of new indications for use or technological differences that would raise new questions of safety or effectiveness. The reported performance is a comparison to the predicate.

    CharacteristicAcceptance Criteria (Implicitly, as per Predicate)Reported Device Performance (New Device)
    Intended UseSame as Viztek ViZion DR K102123: Digital image capture in general radiographic examinations (excluding fluoroscopy, angiography, mammography), imaging skull, chest, shoulders, spine, abdomen, pelvis, and extremities.SAME (Matches predicate exactly)
    Digital PanelSamsung LTX240AA01-A (K090742) with Pixel size 143 µm, 3072 x 3072 pixels, 9 million pixels.Samsung LLX240AB01 (K102587) with Pixel size 143 µm, 3072 x 3072 pixels, 9 million pixels.
    SoftwareEmploys OPAL-RAD PACS image viewing and acquire interface software technology, K063337.SAME (Matches predicate exactly)
    Electrical SafetyElectrical Safety per IEC-60601. UL listed.SAME (Matches predicate exactly)
    Safety and EffectivenessAs safe and effective as predicate devices."The results of clinical image inspection, bench, and test laboratory indicates that the new device is as safe and effective as the predicate devices. Clinical images collected demonstrate equal or better image quality as compared to our predicates."
    Technological DifferencesNo new technological differences that raise new questions of safety or effectiveness."have few technological differences" (only change is the FPD) and "no new indications for use, thus rendering it substantially equivalent".

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Clinical images collected demonstrate equal or better image quality as compared to our predicates." However, it does not specify the sample size for this clinical image inspection, nor does it provide details about the data provenance (e.g., country of origin, retrospective or prospective nature). Given the nature of a 510(k) for a component change (the FPD), this might have been a limited comparative study rather than a large-scale clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document mentions "clinical image inspection" but does not provide any information on the number of experts involved in this inspection or their qualifications.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method. It only mentions "clinical image inspection."

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This device is a digital X-ray detector system, not an AI-assisted diagnostic tool. The comparison is between the new detector and a previous detector system, with the focus on image quality and equivalence, not reader performance improvement with AI.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to an algorithm's performance without human intervention. Since the device is a digital X-ray detector system (hardware), not a diagnostic algorithm, this concept does not apply. The "standalone performance" for this device would relate to its hardware specifications and image acquisition capabilities, which are covered by bench and laboratory testing.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The available information suggests that the "ground truth" for the comparison was based on "clinical image inspection" and possibly "bench, and test laboratory" results, comparing the image quality of the new device to that of the predicate device. This implies a subjective assessment of image quality by potentially experts, but the specifics are not detailed. It is unlikely to involve pathology or outcomes data for this type of 510(k) submission.

    8. The Sample Size for the Training Set

    This submission is for a digital X-ray detector system, not an AI algorithm that requires a training set. Therefore, there is no training set in the context of this device.

    9. How the Ground Truth for the Training Set was Established

    As there is no training set for this hardware device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102123

    Validate with FDA (Live)

    Device Name
    VIZION DR
    Manufacturer
    Date Cleared
    2011-01-24

    (179 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ViZion DR is intended for digital image capture use in general radiographic examinations for adult and pediattic, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. ViZion allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis and extremities.

    Device Description

    The ViZion DR system represents the straightforward intearation of two cleared devices. ViZion DR is a Digital Radiography system, featuring an integrated flat panel digital detector (FPD) K090742 (Samsung Flat-Panel X-Ray Detector), Samsung Mobile Display Co., Ltd. (this is the 510(k) for the flat panel detector) and Viztek's proprietary OPAL-RAD PACS image viewing and acquire interface software technology, (K063337) which incorporates state of the art object-oriented software and connectivity. ViZion is designed to perform digital radiographic examinations in replacement for conventional film. This integrated platform provides the benefits of PACS with the with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the Viztek PACS and Samsung FPD were not changed.

    AI/ML Overview

    The provided text is a 510(k) summary for the Viztek ViZion DR, a digital flat panel X-Ray detector system. It describes the device, its intended use, and states that bench and test laboratory results indicate the new device is as safe and effective as predicate devices, with clinical images demonstrating equal or better image quality.

    However, the document does not contain specific acceptance criteria, detailed study results, sample sizes for test or training sets, ground truth establishment methods, information about expert involvement, adjudication methods, or MRMC study details. The 510(k) summary is a high-level comparison to predicate devices, focusing on substantial equivalence rather than a detailed performance study with quantifiable acceptance criteria.

    Therefore, most of the requested information cannot be extracted from the provided text.

    Here is what can be inferred or stated based on the text provided:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified in documentClinical images demonstrate equal or better image quality as compared to predicate devices.
    Not specified in documentDevice is as safe and effective as predicate devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size (Test Set): Not specified. The document only states "Clinical images collected."
    • Data Provenance: Not specified. It's unclear if the data was retrospective or prospective, or from which country.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not specified. The document does not mention the involvement of experts in establishing ground truth or evaluating the clinical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not specified. No adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not specified. The document does not mention an MRMC study or any AI assistance. The device is a digital X-ray detector system, not an AI-powered analysis tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/Not specified. The device is a digital X-ray detector system, not an algorithm, so the concept of "standalone" performance for an algorithm isn't directly relevant in the context of this document. The document focuses on the image quality produced by the detector.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not specified. The document only states "Clinical images collected demonstrate equal or better image quality." It does not elaborate on how "image quality" was objectively assessed or against what ground truth.

    8. The sample size for the training set

    • Not applicable/Not specified. This device is an imaging hardware system. It does not appear to involve machine learning models that would require a "training set" in the conventional sense. The "training" here would refer to engineering and calibration, not data-driven model training.

    9. How the ground truth for the training set was established

    • Not applicable/Not specified. (See point 8).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1