Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K241125

    Validate with FDA (Live)

    Device Name
    VIVIX-S 1751S
    Manufacturer
    Date Cleared
    2024-11-15

    (206 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VIVIX-S 1751S series is used for the general-purpose diagnostic procedures, and as well as intended to replace radiographic film/ screen systems. The VIVIX-S 1751S series is not intended for mammography applications.

    Device Description

    The VIVIX-S 1751S, available in models FXRD-1751SA and FXRD-1751SB, features a 17x51 inch imaging area. This device intercepts x-ray photons and uses a scintillator to emit visible spectrum photons. The FXRD-1751SA model uses a Csl:Tl (Thallium doped Caesium lodide) scintillator, while the FXRD-1751SB model uses a Gadox (Gadolinium Oxysulfide) scintillator. These photons illuminate an array of photo (a-SI) detectors, creating electrical signals are then converted to digital values, which are processed by software to produce digital images displayed on monitors. The VIVIX-S 1751S must be integrated with an operating PC and an X-ray generator, and it can communicate with the generator via cable. It is designed for capturing and transferring digital x-ray images for radiography diagnosis. Note that the X-ray generator and imaging software are not included with the VIVIX-S 1751S.

    AI/ML Overview

    The document describes a 510(k) submission for the VIVIX-S 1751S digital X-ray detector. The acceptance criteria and the study proving the device meets these criteria are primarily demonstrated through a comparison to a predicate device (K190611) and performance testing based on established standards.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by demonstrating substantial equivalence to the predicate device in terms of technological characteristics and performance metrics, as well as compliance with recognized standards. The "performance" column shows the subject device's reported values relative to the predicate device.

    ParameterAcceptance Criteria (Predicate Device K190611, FXRD-1751SB)Reported Device Performance (Subject Device K241125, FXRD-1751SA)Equivalence
    Technological CharacteristicsSame as Predicate deviceSame as Predicate deviceSubstantially Equivalent
    Intended UseVIVIX-S 1751S series is used for general-purpose diagnostic procedures, and to replace radiographic film/screen systems. Not for mammography.VIVIX-S 1751S series is used for general-purpose diagnostic procedures, and to replace radiographic film/screen systems. Not for mammography.Equivalent
    Operating PrincipleSame as Predicate deviceSame as Predicate deviceEquivalent
    Design FeaturesSame as Predicate deviceSame as Predicate deviceEquivalent
    Communication MethodSame as Predicate deviceSame as Predicate deviceEquivalent
    ResolutionSame as Predicate deviceSame as Predicate deviceEquivalent
    Scintillator TypeGadoxCsI (FXRD-1751SA), Gadox (FXRD-1751SB)Different scintillator for FXRD-1751SA model, but performance shown to be comparable. FXRD-1751SB is identical.
    Performance (Optical / Imaging)
    MTF (0.5 lp/mm)≥ 81≥ 83Similar
    MTF (1 lp/mm)≥ 56≥ 63Similar
    MTF (2 lp/mm)≥ 22≥ 30Similar
    MTF (3 lp/mm)≥ 9≥ 14Similar
    DQE (0.5 lp/mm)≥ 29≥ 38Similar
    DQE (1 lp/mm)≥ 22≥ 33Similar
    DQE (2 lp/mm)≥ 11≥ 23Similar
    DQE (3 lp/mm)≥ 4≥ 14Similar
    Compliance with StandardsCompliance with 21CFR1020.30, 21CFR1020.31, IEC 60601-1, CAN/CSA-C22.2 No. 60601-1, ANSI/AAMI ES60601-1, IEC 60601-1-2Complies with all listed standardsCompliant
    Diagnostic CapabilityEquivalent to predicate deviceDemonstrated equivalent diagnostic capabilityEquivalent

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states, "A-Qualified Expert Evaluation study according to CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted..." However, it does not specify the sample size of cases/images used in this clinical evaluation study.
    The data provenance (country of origin, retrospective/prospective) is not explicitly stated in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document mentions an "Expert Evaluation study" but does not specify the number of experts or their exact qualifications (e.g., "radiologist with 10 years of experience"). It only indicates that it was a "Qualified Expert Evaluation study."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe the adjudication method used for the expert evaluation study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) study was conducted as part of the "Qualified Expert Evaluation study." The primary goal of this study was to confirm that the subject device (VIVIX-S 1751S - FXRD-1751SA) provides images of equivalent diagnostic capability to the predicate device (VIVIX-S 1751S - FXRD-1751SB).

    The document does not mention the involvement of AI in this study, nor does it quantify any improvement of human readers with AI assistance. The study described is a comparison of two different X-ray detector technologies, not an AI-assisted reading study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device is an X-ray detector, not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" performance without a human-in-the-loop is not applicable in the same way it would be for an AI diagnostic software. The performance metrics reported (MTF, DQE) are physical image quality parameters of the detector itself, which could be considered "standalone" in this context as they characterize the device's inherent imaging capability.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the "clinical test" (Expert Evaluation study) was established by "Qualified Expert Evaluation" to assess "equivalent diagnostic capability" of the images. This suggests a form of expert consensus or individual expert readings to determine the diagnostic quality of the images produced by the subject device compared to the predicate. It does not mention pathology or outcomes data as the primary ground truth.

    8. The sample size for the training set

    The document does not mention a training set in the context of this device because it is a hardware device (X-ray detector), not an AI algorithm that requires a training phase.

    9. How the ground truth for the training set was established

    Since there is no mention of a training set for an AI algorithm, the question of how its ground truth was established is not applicable to this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190611

    Validate with FDA (Live)

    Device Name
    VIVIX-S 1751S
    Manufacturer
    Date Cleared
    2019-04-08

    (28 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VIVIX-S 1751S series is used for the general-purpose diagnostic procedures, and as well as intended to replace radiographic film/ screen systems. The VIVIX-S 1751S series is not intended for mammography applications.

    Device Description

    VIVIX-S 1751S, whose model name is VIVIX-S 1751S with 17X51 inches of imaging area. The device intercepts x-ray photons, and its Gadox (Gadolinium Oxysulfide) scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create electrical signals. After the electrical signals are generated, it is converted to digital value, and the software acquires and processes the data values from the detector. The resulting digital images will be displayed on monitors. These devices should be integrated with an operating PC and an X-Ray generator. The detector can communicate with the generator by cable. It is utilized for capturing and transfer of digital x-ray images for radiography diagnosis. The X-ray generator and the imaging software are not part of the VIVIX-S 1751S.

    AI/ML Overview

    Acceptance Criteria and Study for VIVIX-S 1751S Digital X-ray Detector

    The VIVIX-S 1751S Digital X-ray detector was evaluated for substantial equivalence to its predicate device (K181003, VIVIX-S 1717V) through non-clinical performance testing and a clinical image concurrence study.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the VIVIX-S 1751S are based on demonstrating performance equivalent to or better than the predicate device (VIVIX-S 1717V). Specifically, the device was tested against recognized performance metrics for X-ray detectors.

    ParameterAcceptance Criteria (Predicate Device K181003)Reported Device Performance (VIVIX-S 1751S)Met Acceptance Criteria?
    Image Quality
    DQE (1lp/mm)FXRD-1717VA: 4060 (typ. 48%)FXRD-1717VB: 2040 (typ. 27%)FXRD-1751SB: 20% (for Gd2O2S:Tb scintillator)Equivalent
    MTF (1lp/mm)FXRD-1717VA: 5070 (typ. 70%)FXRD-1717VB: 4060 (typ. 58%)FXRD-1751SB: 40% (for Gd2O2S:Tb scintillator)Equivalent
    Resolution3.5 lp/mm3.5 lp/mmYes
    Pixel Pitch0.14mm0.14mmYes
    Active Area430.08 x 430.08 (mm)430.08(mm) x 430.08(mm)430.08(mm) x 860.16(mm)430.08(mm) x 1290.24(mm)Yes (offers larger options)
    Pixel Size140 μm140 μmYes
    Grayscale16 bit16 bitYes
    Spatial Resolution3.5lp/mm3.5lp/mmYes
    Functional Equivalence
    Intended UseGeneral diagnostic procedures, replace film/screen systems, not mammography.General diagnostic procedures, replace film/screen systems, not mammography.Yes
    Operating PrincipleX-ray photons strike scintillator, visible light, electric charge, digital value.X-ray photons strike scintillator, visible light, electric charge, digital value.Yes
    Design FeaturesSimilar to predicateSimilar to predicateYes
    Communication MethodWired, Max. 1GbpsWired, Max. 1GbpsYes
    Scintillator MaterialsGd2O2S:Tb, Csl:TIGd2O2S:Tb (Csl:TI type not available for this model)Equivalent
    Active Array3072 x 3072 pixels3072 x 3072 pixels3072 x 6144 pixels3072 x 9216 pixelsYes (offers larger options)
    Method of Generator InterfaceDR Trigger, AED, Passive TriggerDR Trigger, AEDEquivalent (Passive Trigger omitted)
    TechnologyStructured scintillator for efficiency and resolution.Structured scintillator for efficiency and resolution.Yes
    Principle of OperationX-ray conversion to visible light then electric charge.X-ray conversion to visible light then electric charge.Yes

    Note: While the exact DQE and MTF values vary for the specific FXRD-1751SB model compared to some predicate models (FXRD-1717VA vs FXRD-1717VB), the applicant states "Similar performance as follow" and declares the performance as "Equivalent," suggesting the overall performance is within an acceptable range for substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided summary does not explicitly state the specific sample size used for the clinical image concurrence study test set. It mentions "a single-blinded concurrence study." The data provenance (e.g., country of origin, retrospective or prospective) for this clinical study is also not detailed in the provided information.

    3. Number of Experts and Qualifications for Ground Truth

    The number of experts used to establish the ground truth for the clinical test set and their specific qualifications (e.g., "radiologist with 10 years of experience") are not specified in the provided 510(k) summary. It only indicates that "CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted" and that it was a "single-blinded concurrence study."

    4. Adjudication Method

    The adjudication method for the test set is not explicitly stated. The summary mentions "a single-blinded concurrence study," which implies a comparison of images by experts, but the process for resolving disagreements or establishing a final ground truth is not described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no indication that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to assess how much human readers improve with AI vs. without AI assistance. The device is a digital X-ray detector, not an AI-powered diagnostic tool, and the study described is a comparison of image diagnostic capability between two detector models.

    6. Standalone (Algorithm Only) Performance Study

    A standalone performance study of the algorithm only without human-in-the-loop performance is not applicable in the context of this device. The VIVIX-S 1751S is a digital X-ray detector, a hardware component that produces images for human interpretation, not an independent algorithm for diagnosis. The non-clinical performance (DQE, MTF, resolution) can be considered "standalone" in terms of objective image quality metrics of the device itself.

    7. Type of Ground Truth Used

    For the clinical study, the ground truth was established through expert concurrence. The summary states that the study "confirmed that the new x-ray detectors VIVIX-S 1751S provide images of equivalent diagnostic capability to the predicate devices." This implies that expert readers (though unspecified in number or qualifications) evaluated the images and concurred on their diagnostic equivalence. Pathology or outcomes data are not mentioned as the ground truth.

    8. Sample Size for the Training Set

    The provided 510(k) summary describes a digital X-ray detector, which is a hardware device for image acquisition. This device does not have a "training set" in the context of artificial intelligence or machine learning algorithms. Its design and performance are based on physical and technical specifications.

    9. How the Ground Truth for the Training Set was Established

    As clarified in point 8, the concept of a "training set" and associated "ground truth" for a training set is not applicable to this device as it is a digital X-ray detector, not an AI/ML diagnostic software. The performance evaluations are based on engineering specifications and clinical comparison to a predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1