510(k) Data Aggregation

    K Number
    K952095

    Validate with FDA (Live)

    Date Cleared
    1996-03-12

    (313 days)

    Product Code
    Regulation Number
    866.2660
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    is intended for use in conjunction with the Vitek® System for the automated identification of clinically significant streptococci, staphylococci, and a selected group of Gram positive bacilli.

    Device Description

    The VITEK® Gram Positive Identification Card containing new claims for the identification of Enterococcus casseliflavus Enterococcus gallinarum Enterococcus hirae

    AI/ML Overview

    Here's the analysis of the provided text regarding the acceptance criteria and study for the Vitek® GPI Card:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Correct genus call percentage (after repeat and supplemental testing for enterococcal isolates)96.7%
    Correct species call percentage (for correctly identified genus)91.1%
    Mis-identification rate for enterococcal isolates3.3%
    Equivalency of identification between new and current software (overall agreement)99.8%
    Reproducibility of identificationThe new software reproducibly identifies specified gram-positive organisms.
    No adverse effect on performance of GPI Card for non-enterococcal isolatesPerformance remains the same as current software.
    No adverse effect on performance of GPI Card for existing species with new softwarePerformance remains the same as current software.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the total sample size used for the test set of enterococcal isolates. It mentions "clinical testing was performed" and "Correlation Studies on enterococcal isolates."
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be retrospective as they involve analyzing existing clinical isolates or comparing current software with new software.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not mentioned in the provided text.

    4. Adjudication Method for the Test Set

    Not mentioned in the provided text.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The study described focuses on the performance of the automated Vitek® system and its software, not on human readers' performance with or without AI assistance.

    6. Standalone Performance Study (Algorithm Only)

    Yes, a standalone study was performed. The entire context describes the performance of the Vitek® GPI Card and its software in identifying bacteria, which is an automated algorithm-only performance. The "Correlation Studies on enterococcal isolates" and "Equivalency Study" directly describe the algorithm's performance.

    7. Type of Ground Truth Used

    The type of ground truth is not explicitly specified as "pathology" or "outcomes data." However, based on the context of bacterial identification in a clinical laboratory setting, the ground truth would likely be established through:

    • Reference laboratory methods: Often involving biochemical tests, molecular methods (e.g., PCR, sequencing), or advanced phenotypic characterization, considered the gold standard for bacterial identification.
    • Expert consensus: Among microbiologists using a combination of methods.

    8. Sample Size for the Training Set

    Not mentioned in the provided text. The document describes validating the addition of new claims to existing software, implying the core identification algorithms were already established and likely trained on a separate dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not mentioned in the provided text.