Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K063243

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-12-14

    (414 days)

    Product Code
    Regulation Number
    862.1215
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VIDAS® Troponin I Ultra is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin I in human serum or plasma (lithium heparin) using the ELFA (Enzymc-Linked Fluorescent Assay) technique. VIDAS Troponin I Ultra is intended to be used as an aid in the diagnosis of myocardial infarction.

    Device Description

    The VIDAS Troponin I Ultra (TNIU) Assay is an enzyme-linked fluorescent immunoassay (ELFA) performed in an automated VIDAS® instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as the solid phase as well as a pipettor for the assay. Reagents for the assay are in the sealed TNIU Reagent Strips. The sample is transferred into the wells containing anti-cardiac troponin i antibodies labeled with alkaline phosphatase (conjugate). The sample/conjugate mixture is cycled in and out of the SPR for a specified length of time. Troponin I present in the specimen will bind to the anticardiac troponin i immunoglobulin coating the interior of the SPR. Unbound sample components are washed away. A fluorescent substrate, 4-methylumbellifery| phosphate, is introduced into the SPR. Enzyme remaining on the SPR wall will catalyze the conversion of the substrate to the fluorescent product 4-methylumbelliferone. The optical scanner in the instrument measures the intensity of fluorescence. When the VIDAS TNIU assay is completed, the results are analyzed automatically by the computer, a test value is generated, and a report is printed for each sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the VIDAS® Troponin I Ultra (TNIU) Assay, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in a separate section with specific numerical targets. Instead, it presents performance data for both the VIDAS® TNIU and the predicate device (Dimension RxL® CTNI) for comparison, implying that similar or improved performance relative to the predicate device is the goal for most parameters, and certain established clinical cut-offs are considered.

    Given this, I've constructed a table focusing on the performance characteristics that were measured and compared, as these implicitly serve as the basis for demonstrating substantial equivalence.

    Acceptance Criteria (Implied / Comparison Target)Reported Device Performance (VIDAS® TNIU)Predicate Device Performance (Dimension RxL® CTNI)
    Expected Values (Healthy Patients)99% of 747 patients with no cardiac symptoms had values of <0.01 µg/l97.5% of 101 apparently healthy patients had values of 0.00 -0.05 µg/L
    Cut-off0.11 µg/l0.6-1.5 ng/mL (Note: unit difference)
    Specificity (Cardiac Troponin T) (at 1000 µg/L)0.2%0.34%
    Specificity (Cardiac Troponin C) (at 1000 µg/L)<0.0010.00%
    Specificity (Skeletal Troponin I) (at 1000 µg/L)<0.0010.04%
    Analytical Detection Limit< 0.01 µg/l0.04 µg/L
    Dilution - Recovery Test80-120%98.6-106.4%
    Interference (Bilirubin)No significant interference (up to 510 µM)No significant interference (up to 20 mg/dL)
    Interference (Hemoglobin)No significant interference (up to 332 µM)No significant interference (up to 1000 mg/dL)
    Interference (Lipemia/Triglycerides)No significant interference (up to 30 mg/ml)No significant interference (up to 3000 mg/dL)
    Hook Effect (up to stated concentration)No hook effect found up to 1000 µg/LNo hook effect found up to 1800 ng/mL
    Clinical Method Comparison (Correlation Coefficient)0.97(N/A - This is a comparison between devices)
    Clinical Method Comparison (Slope)0.42 (Confidence Interval: 0.38-0.44)(N/A - This is a comparison between devices)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Method Comparison: 534 samples
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). It is described as a "sample comparison study," implying patient samples were collected and tested.
    • Sample Size for Expected Values (Healthy Patients):
      • VIDAS® TNIU: 747 patients with no cardiac symptoms
      • Predicate Device: 101 apparently healthy patients

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable (N/A) / Not explicitly stated. This device is an immunoassay for a biomarker (Troponin I). The "ground truth" for analytical performance studies would typically be established by reference measurements (e.g., gold standard analytical methods, spiked samples) or by clinical diagnosis for the "expected values" data, not by expert interpretation of images or symptoms in the same way as, for example, an imaging AI device.
    • For the "expected values in healthy patients," the ground truth is simply the clinical status of the patient (e.g., "no cardiac symptoms" or "apparently healthy"). The document doesn't mention expert review of these patient status determinations beyond general clinical criteria.

    4. Adjudication Method for the Test Set

    • Not Applicable (N/A) / Not mentioned. Adjudication methods like 2+1 or 3+1 are typically used for studies where multiple human readers interpret data (e.g., medical images) and their interpretations need to be reconciled to establish a consensus ground truth. For an immunoassay, the measurement itself is the primary focus, and while clinical diagnosis (the outcome) might involve expert consensus, the assay performance itself is not adjudicated in this manner.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. An MRMC study is not relevant for this type of in vitro diagnostic device (immunoassay). MRMC studies compare the diagnostic performance of different readers (humans, AI, or human+AI) on a set of cases. This submission is for an automated immunoassay for measuring a biomarker.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Yes, by nature of the device. The VIDAS® TNIU Assay is a standalone automated quantitative test. The results are generated automatically by the instrument, and then a report is printed. There is no explicit "human-in-the-loop" component in the generation of the quantitative result itself, though a clinician would interpret the result in the context of a patient's overall clinical picture. The performance data presented (e.g., correlation, detection limits) directly reflects the standalone performance of the assay.

    7. The Type of Ground Truth Used

    • Analytical Measurement Comparisons and Clinical Status:
      • For Analytical Detection Limit, Dilution-Recovery, Interference, Hook Effect, Specificity: Ground truth is established by the known concentrations of analytes, spiked samples, or absence/presence of interfering substances under laboratory conditions.
      • For Expected Values (Healthy Patients): Clinical documentation of the patient's health status (e.g., "no cardiac symptoms" or "apparently healthy") serves as the ground truth for that specific study component.
      • For the Clinical Testing (Method Comparison): The "ground truth" is one device's measurement (Dimension RxL® CTNI Assay, the predicate) against which the new device (VIDAS® TNIU Assay) is compared to demonstrate correlation and substantial equivalence.

    8. The Sample Size for the Training Set

    • Not explicitly stated in relation to algorithm training. This device is an immunoassay, not an AI/machine learning algorithm that requires a "training set" in the conventional sense of computational models. The assay is based on chemical and biological reactions, not statistical pattern recognition trained on a large dataset.
    • However, the development of such assays involves extensive research and development with numerous samples to optimize reagents, protocols, and establish performance characteristics, which could be conceptually considered akin to "training" in a broader sense, but not with a distinct "training set" as understood in AI.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable (N/A). As explained above, this is an immunoassay, not an AI/machine learning device that uses a "training set" with ground truth in the typical computational sense. The "ground truth" for optimizing the assay would involve using reference materials, known concentrations, and clinical samples whose true values (or clinical diagnoses) were independently established.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1