510(k) Data Aggregation

    K Number
    K040126

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-04-19

    (90 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Viatronix V3D Colon is a system for the display and visualization of 3D and 2D medical image data of the colon derived from DICOM 3.0 compliant CT and MR scans, for the purpose of patient screening for detection of colon cancers, polyps, masses, and other lesions. It provides functionality for display, measurement and electronic cleansing to assure complete visualization of the colon from rectum to cecum and vice versa for both prone and supine views. It generates a centerline for guided interactive navigation and fly through of the entire colon, and also includes a reporting facility to enhance workflow for patient screening. It is intended for use by Radiologists, Clinicians and referring Physicians to process, render, review, archive print and distribute colon image studies utilizing PC hardware.

    Device Description

    The Viatronix V3D Colon contains all of the required hardware and software components to provide interactive 3D and 2D views of diagnostic CT and MR scan images of the colon. The views include both inner and outer surface 3D volume rendered images as well as orthogonal and multiplaner reformatted 2D images. This ability to view the dataset in different perspectives from which it was acquired is performed by first transferring DICOM images from the MR or CT scanners to the Viatronix V3D Colon, which automatically identifies regions of interest and displays thesc regions to the user in the above mentioned views. The user can then navigate freely within the dataset/region of interest or follow automatically computed paths to fly through the colon or around the outside of the colon structure. Measurements of the size of colon polyps, masses or lesion can be made for patient screening and for planning treatment.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Viatronix V3D Colon, revision 1.3, based on the provided document:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Preamble Goals
    Accurate 3D model of the entire colon is generatedVerified
    Fly-through in the 3D model for the entire colon is possibleVerified
    Viewing 2D and MPR images of the entire colon is possibleVerified
    Measurements are accurate (using phantom)Verified
    Ability to fly through and visualize the entire colonDemonstrated
    Ability to visualize and measure polyps, masses and lesions for patient screening to detect colon cancersDemonstrated
    Clinical Study Performance (Colorectal Neoplasia Screening in Asymptomatic Adults)
    Virtual Colonoscopy (VC) Sensitivity for 10mm adenomatous polyps93.8%
    Virtual Colonoscopy (VC) Sensitivity for 8mm adenomatous polyps93.9%
    Virtual Colonoscopy (VC) Sensitivity for 6mm adenomatous polyps88.7%
    Virtual Colonoscopy (VC) Specificity for 10mm adenomatous polyps96.0%
    Virtual Colonoscopy (VC) Specificity for 8mm adenomatous polyps92.2%
    Virtual Colonoscopy (VC) Specificity for 6mm adenomatous polyps79.6%
    Detection of malignant polyps (2 total)Both detected on VC

    Note on Acceptance Criteria: The document primarily lists "goals of clinical testing" rather than explicit numerical acceptance criteria for all aspects. For sensitivity and specificity, the reported values can be considered the demonstrated performance against assumed implicit acceptance thresholds for a diagnostic tool in this context.

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: A total of 1233 asymptomatic adults.
    • Data Provenance: US Government's Department of Defense Medical Centers (United States). The study was prospective, as it involved subjects undergoing both virtual colonoscopy and optical colonoscopy on the same day.

    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • The document implies that the "Results" section of the major clinical study (8.1) references data where Optical Colonoscopy (OC) served as a comparator for sensitivity and specificity. However, it does not explicitly state the number of experts who established the ground truth from OC, or their qualifications.
    • For the "Initial Clinical tests" (8.2), a single radiologist reviewed the clinical data.
      • Qualifications of Radiologist (8.2): Experienced with predicate devices.

    3. Adjudication Method for the Test Set:

    • For the main clinical study (8.1), the document mentions a comparison with optical colonoscopy. One paragraph states: "Two polyps were malignant; both were detected on VC and one was missed on OC before the results on VC were revealed." This implies a potential sequential or blinded review process where OC findings were known post-VC review, but a formal adjudication method (like 2+1) is not explicitly detailed.
    • For the "Initial Clinical tests" (8.2), the review was performed by a single radiologist, suggesting no formal adjudication.

    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size:

    • No, a formal MRMC comparative effectiveness study aiming to measure the effect size of human readers with vs. without AI assistance is not described.
    • The main clinical study (8.1) compares the device's performance (VC) directly against Optical Colonoscopy (OC) in terms of sensitivity and specificity, effectively comparing the AI-assisted diagnostic pipeline (VC reviewed by an unnamed number of readers) to the gold standard (OC). It does not isolate the improvement from AI assistance to a human reader.

    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, the studies described do not appear to be standalone algorithm-only performance evaluations.
    • The device is a "system for the display and visualization" intended for use by "Radiologists, Clinicians and referring Physicians." The clinical study evaluated "Virtual Colonoscopy (VC) ... by patient," which implicitly involves human interpretation of the V3D Colon output. The "Initial Clinical tests" section explicitly states that a radiologist reviewed the rendering.

    6. The Type of Ground Truth Used:

    • For the main clinical study (8.1): Optical Colonoscopy (OC) findings were used as the primary ground truth. This is a form of clinical/procedural ground truth.
    • For the non-clinical tests (7): Phantoms with structures of known size and distance were used. This is a form of physical phantom ground truth.

    7. The Sample Size for the Training Set:

    • The document does not specify a sample size for the training set. It discusses the development process ("developed in a manner consistent with accepted standards for software development, including test protocols") and testing, but not the data used to train the AI/algorithm component of the system.

    8. How the Ground Truth for the Training Set was Established:

    • The document does not provide information on how ground truth was established for any training set, as a training set is not explicitly discussed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K020658

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-03-07

    (6 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Viatronix V3D Colon is a system for the display and visualization of 3D and 2D medical image data of the colon derived from DICOM 3.0 compliant CT and MR scans, for the purpose of screening a colon to detect polyps, masses, cancers and other lesions. It provides functionality for display, measurement and electronic cleansing to assure complete visualization of the colon from rectum to cecum and vice versa for both prone and supine views. It generates a centerline for guided interactive navigation and fly through of the entire colon, and also includes a reporting facility to enhance workflow for screening a colon. It is intended for use by Radiologists, Clinicians and referring Physicians to process, render, review, archive, print and distribute colon image studies utilizing PC hardware.

    Device Description

    The Viatronix V3D Colon contains all of the required hardware and software components to provide interactive 3D and 2D views of diagnostic CT and MR scan images of the colon. The views include both inner and outer surface 3D volume rendered images as well as orthogonal and multiplaner reformatted 2D images. This ability to view the dataset in different perspectives from which it was acquired is performed by first transferring DICOM images from the MR or CT scanners to the Viatronix V3D Colon, which automatically identifies regions of interest and displays these regions to the user in the above mentioned views. The user can then navigate freely within the dataset/region of interest or follow automatically computed paths to fly through the colon or around the outside of the colon structure. Measurements of the size of colon polyps, masses or lesions can be made for patient screening and for planning treatment.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Viatronix V3D Colon, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria with specific numerical targets. Instead, the acceptance criteria are implicitly defined by the goals of the clinical testing, focusing on functionality, accuracy, and substantial equivalence to predicate devices.

    Acceptance Criteria (Inferred from Clinical Testing Goals)Reported Device Performance
    Generate an accurate 3D model of the entire colonVerified that an accurate 3D model of the entire colon is generated.
    Enable fly-through in the 3D model of the entire colonVerified that fly-through in the 3D model of the entire colon is possible. Demonstrated the ability to fly through and visualize the entire colon.
    Enable viewing of 2D and MPR images of the entire colonVerified that viewing 2D and MPR images of the entire colon is possible.
    Accurate measurements (e.g., of polyps, masses) using a phantomUsing a phantom, verified that measurements are accurate. Demonstrated the ability to visualize and measure polyps, masses, and lesions for patient screening. Accuracy was assessed by comparing test results with actual phantom structures.
    Ability to visualize structures (compared to optical endoscopy)In comparison to optical endoscopy, the Viatronix V3D Colon is able to visualize structures of similar size and shape. Qualitative judgment by radiologists indicated accurate and medically useful rendering.
    Substantial equivalence in essential features to predicate devices (e.g., quality, effectiveness, rendering accuracy)Clinical data reviewed by a radiologist experienced with predicate devices found the Viatronix V3D Colon "substantially equivalent in essential features," with improvements in speed of rendering and ease of segmentation. Non-clinical phantom testing determined its level of accuracy to be "substantially equivalent to that of its predicate devices." Concluded to be substantially equivalent.
    Perform as intended with a broad sampling of input data (core functionality: fly-through and visualization)Each patient was assessed as to whether the core functionality permitted fly-through and visualization.
    Reliability and ease of useThe product has shown itself to be reliable, easy to use, and capable of rendering useful 3D medical images.

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-clinical (Phantom) Test Set:

      • Sample Size: Not explicitly stated, but implies multiple "structures of a known size and distance" inserted into the phantom.
      • Data Provenance: N/A (phantom data)
    • Clinical (Human) Test Set:

      • Sample Size: Not explicitly stated, described as "broad sampling of input data" and "each patient was assessed."
      • Data Provenance: Prospective (clinical studies were performed under IRB overview, which included Patient Consent Forms). Country of origin is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Non-clinical (Phantom) Test Set:

      • Number of Experts: One independent reviewer.
      • Qualifications: "An independent reviewer." Specific qualifications (e.g., years of experience, medical specialty) are not detailed.
    • Clinical (Human) Test Set:

      • Number of Experts: One radiologist.
      • Qualifications: "A radiologist who determined that the rendering was accurate and medically useful. The radiologist was experienced with the predicate devices." Specific years of experience or subspecialty are not detailed.

    4. Adjudication Method for the Test Set

    • For the non-clinical phantom study, the "independent reviewer" compared test results with the actual phantom to assess accuracy. This implies direct comparison rather than a formal adjudication process involving multiple experts for disagreement resolution.
    • For the clinical study, a single radiologist made the determination of accuracy and medical usefulness. There is no mention of a multi-reader adjudication method (e.g., 2+1, 3+1).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus human readers without AI assistance was not explicitly described. The clinical evaluation involved a single radiologist assessing the device's output and comparing it qualitatively to optical endoscopy and predicate devices.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the testing described appears to be a standalone evaluation of the device. The system automatically identifies regions of interest, displays them, generates centerlines for navigation, and provides measurement tools. The radiologist then reviews and uses this output. The evaluation focuses on the device's ability to render images, permit navigation, and make accurate measurements, which are functions of the algorithm without continuous human-in-the-loop intervention for its primary output generation. The radiologist's role was to review and determine accuracy and usefulness, implying the device produced results first.

    7. The Type of Ground Truth Used

    • Non-clinical (Phantom): Physical ground truth – "structures of a known size and distance" inserted into the phantom, against which the device's measurements were compared.
    • Clinical (Human): Relative ground truth –
      • Qualitative judgment by a radiologist.
      • Comparison with "optical endoscopy" for visualization capabilities.
      • Comparison to the performance of predicate devices.
      • The "System Clinical Test Results (Visualization of Polyps)" suggests comparison to pathology or clinical findings might have been part of the evaluation, but the direct method for establishing truth related to specific lesions is not fully detailed beyond "qualitative judgment."

    8. The Sample Size for the Training Set

    The document does not provide any information about a training set or its sample size. The focus is on the evaluation of the finalized device.

    9. How the Ground Truth for the Training Set was Established

    Since no training set information is provided, how its ground truth was established is also not present in the document.