510(k) Data Aggregation

    K Number
    K163524

    Validate with FDA (Live)

    Device Name
    Vessel Dilator
    Date Cleared
    2017-06-29

    (196 days)

    Product Code
    Regulation Number
    870.4475
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fetzer Medical vessel dilators are devices used to enlarge or calibrate a vessel during cardiovascular surgery.

    Device Description

    Fetzer Medical Vessel Dilators are reusable manual surgical instruments. They are rod-like devices with a long, slender, solid shaft and a distal tip of various shapes and sizes. They are sold unsterile and can be reused (cleaned and sterilized) according the instructions for use.

    There are several types of dilators made from stainless steel or aluminum:

    • Cooley Coronary Dilators
    • Cooley Vessel Dilators
    • Garrett Vascular Dilators
    • Hiebert Vascular Dilators
    • DeBakey Vascular Dilators
    • Sarns-Style Dilators
    • Myocardial Dilators

    These dilators are available in a range of different length and diameters.

    The handle of the DeBakey Vascular Dilator, the Cooley Vessel Dilator, the Cooley Coronary Dilator, the Hiebert Vascular Dilator and the Garrett Vascular Dilator is textured with one site flat. The handles of the Sarns and Myocardial Dilators are smooth. The Myocardial Dilator has a flat distal end for holding it and the Sarns Dilator has a hollow handle at the distal end.

    Apart from the Hiebert Vascular Dilator and the Myocardial Dilator, the handle is connected to a thin shaft. The diameter of the shaft of the different dilator types varies. The shaft of the Cooley Coronary Dilator, the Cooley Vessel Dilator, The Garrett Vascular Dilator and the DeBakey Vascular Dilator can be thin so that the shaft is flexible. The handle of the Myocardial Dilator has no shaft while the Hiebert Vascular Dilator is double-ended; it possesses two flexible shafts, with the handle located in between the two thin shafts.

    The tips of all dilators have the aim to enlarge a vessel. The design of the Sarns and Myocardial Dilator is similar to an elephant's tusk. The diameter increases steadily, moving towards the handle's direction. The tips of the other Dilators are olive-shaped. The Cooley Coronary Dilator possesses no formed tip. For all dilators, the penetrating diameter corresponds to the diameter of the shaft. The Hiebert Vascular Dilators working elements are each malleable wires with an olive tip in different diameters.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: Fetzer Medical Vessel Dilators. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the safety and effectiveness from scratch through extensive clinical trials for new medical technologies like AI/ML devices. As such, the information you're asking for, particularly regarding acceptance criteria, study design for proving performance, and expert-driven ground truth establishment (which is typical for AI/ML diagnostic devices), is not present in this document.

    The document focuses on non-clinical performance testing (reprocessing, sterilization, biocompatibility, mechanical tests) to demonstrate that the Fetzer Medical Vessel Dilators are as safe and effective as existing, legally marketed predicate devices.

    Here's a breakdown of why your specific questions cannot be answered by this document and what is provided:

    Why the document does not contain the requested information (related to AI/ML device performance studies):

    • Device Type: The Fetzer Medical Vessel Dilators are manual surgical instruments, specifically vessel dilators. They are not an AI/ML diagnostic or therapeutic device.
    • Regulatory Pathway: A 510(k) submission primarily relies on demonstrating substantial equivalence to a predicate device. This involves comparing indications for use, technological characteristics (design, materials, performance), and demonstrating that any differences do not raise new questions of safety or effectiveness. It does not typically involve defining acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, AUC) or conducting large-scale clinical studies with human readers, ground truth consensus, or advanced statistical analyses like MRMC studies, which are common for AI/ML-driven diagnostic devices.
    • Performance Data Scope: The "performance data" mentioned in this document refers to physical testing of the device (e.g., cleanability, sterilizability, mechanical strength), not diagnostic accuracy or clinical outcomes in patients.

    What the document does provide regarding "Testing":

    The document details performance testing conducted to show the device's characteristics are similar to the predicate devices and that it meets relevant standards for a surgical instrument.

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (Implicit): For this type of device, the "acceptance criteria" are broad and relate to meeting industry standards and demonstrating comparable performance to the predicate device in specific functional aspects.

      • Automatic Reprocessing Validation according to ANSI/AAMI ST81: The implicit acceptance criterion is that "contamination...could be removed...by the given cleaning and disinfection instructions."
      • Sterilization testing: The implicit acceptance criterion is that "the device will be adequately sterilized under the validated sterilization parameters" to achieve a SAL (Sterility Assurance Level) of 10^-6.
      • Biocompatibility testing (ISO 10993-5): The implicit acceptance criterion is that materials are "non cytotoxic for the indicated contact duration."
      • Mechanical tests: The implicit acceptance criterion is that devices "were able to withstand the required pulling force without deformation or failure."
    • Reported Device Performance:

    TestReported Device Performance
    Automatic Reprocessing Validation"This validation provides evidence that contamination of the 'Vessel Dilators' could be removed in health care facilities by the given cleaning and disinfection instructions."
    Sterilization testing"A reduction of test bacteria was observed and provided assurance that the device will be adequately sterilized under the validated sterilization parameters."
    Biocompatibility testing (Cytotoxicity)"All patient contacting materials were from the same material as the equivalent predicate devices. They were considered non cytotoxic for the indicated contact duration."
    Mechanical tests"All devices were able to withstand the required pulling force without deformation or failure."

    2. Sample size used for the test set and the data provenance:

    • Test Sets: The document refers to "worst case devices" for reprocessing validation and "test specimens" for sterilization testing. The exact number of units or biological samples tested is not specified.
    • Data Provenance: This is not applicable in the context of clinical patient data. The tests are laboratory-based. No country of origin for clinical data is mentioned as no clinical studies were performed. The tests are retrospective in the sense they were done before submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications:

    • Not applicable. The "ground truth" for these tests relates to measurable physical/biological parameters (e.g., residual protein levels, bacterial reduction, material cytotoxicity, mechanical force applied), not expert interpretation of medical images or patient outcomes.

    4. Adjudication method for the test set:

    • Not applicable. This refers to consensus building among experts for subjective interpretations, which is not relevant for the objective laboratory tests performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No. This type of study is for evaluating observer performance, typically in diagnostic imaging. It was not conducted for this surgical instrument.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This relates to AI/ML algorithm performance. The device is a manual surgical tool.

    7. The type of ground truth used:

    • Laboratory-based measurements against established standards:
      • Reprocessing: Measured contamination levels (red blood cells, residual protein, organisms, total organic carbon) against AAMI TIR 30 guidelines.
      • Sterilization: Reduction of bioindicators/challenge suspension to achieve SAL 10^-6.
      • Biocompatibility: Cytotoxicity results per ISO 10993-5.
      • Mechanical: Resistance to pulling forces.

    8. The sample size for the training set:

    • Not applicable. This is typically for machine learning models. No training set for an algorithm was used.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K963388

    Validate with FDA (Live)

    Device Name
    VESSEL DILATOR
    Manufacturer
    Date Cleared
    1997-03-11

    (195 days)

    Product Code
    Regulation Number
    870.1310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To enlarge the opening in a vessel in preparation of percutaneous entry of a catheter.

    Device Description

    The Vessel Dilator is a device used over a quide wire to dilate the opening in a vessel in preparation of percutaneous entry of a catheter. The device may be used separately or in conjunction with a sheath introducer. The device consists of a plastic hub attached to a plastic cannula. The distal end of the cannula is thermally tapered to assist in dilation. The hub proximal end has a female luer lock configuration An optional hub collar may be used to attach the dilator to an introducer. The device is available in sizes 4F through 15F and is supplied as a sterile, disposable single put up in sealed pouches, in non-sterile bulk quantities, or included as a component in catheter introducer kits,trays, or packs.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called a "Vessel Dilator." It describes the device, its intended use, and compares it to a predicate device. However, it does not contain any information about acceptance criteria or a study proving the device meets those criteria in the context of clinical performance or diagnostic accuracy.

    The information provided focuses on:

    • Biocompatibility: Stating that "Biocompatibility of the device has been established by cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity and hemocompatibility tests."
    • Material Testing: "The device meets the requirements of the USP XXIII physicochemical tests for plastics."
    • Tensile Strength: "Results of performance testing indicate that the Vessel Dilator has improved tensile strength characteristics compared to the predicate device."

    These are engineering and material compliance statements, not clinical performance acceptance criteria or studies in the way you've outlined with terms like "test set," "ground truth," "MRMC study," or "standalone performance."

    Therefore, I cannot populate the table or answer most of your specific questions based on the provided text.

    Here's how I can address what is available:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    BiocompatibilityEstablished by cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, and hemocompatibility tests.Biocompatibility has been established by the listed tests. (Implies positive outcome without specific performance metrics).
    Material ComplianceMeets requirements of the USP XXIII physicochemical tests for plastics.Device meets the requirements of the USP XXIII physicochemical tests for plastics. (Implies positive outcome without specific performance metrics).
    Mechanical Performance(Implicit: Tensile strength at least equivalent to predicate device; explicitly stated improvement)"Results of performance testing indicate that the Vessel Dilator has improved tensile strength characteristics compared to the predicate device." (Specific metrics not provided).
    Equivalence to PredicateIdentical in design, function, intended use, and similar in composition to predicate device, and performs at least as well as the legally marketed predicate.Device is identical in design, function, and intended use, similar in composition. Performance testing shows improved tensile strength, leading to the conclusion it "performs at least as well."

    The remaining questions cannot be answered from the provided text as they relate to clinical study design and performance, which is not present in this 510(k) summary.

    1. Sample size used for the test set and the data provenance: Not applicable; no clinical test set described.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable; no clinical ground truth described.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable; no clinical test set described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable; this device is a physical medical instrument, not an AI diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable; this device is a physical medical instrument, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable; no clinical ground truth described.
    7. The sample size for the training set: Not applicable; no training set described.
    8. How the ground truth for the training set was established: Not applicable; no training set described.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1