Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121675

    Validate with FDA (Live)

    Device Name
    VERSAVIT
    Manufacturer
    Date Cleared
    2012-06-21

    (15 days)

    Product Code
    Regulation Number
    886.4150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synergetics VersaVit is an ophthalmic microsurgical system that is indicated for posterior segment (i.e. vitreoretinal) ophthalmic surgery. The integrated light source is intended to illuminate the eye during vitreoretinal procedures.

    Device Description

    The VersaVit is a compact, stand-alone, portable device with a pneumatic vitrector drive, aspiration, and illumination through an imbedded solid state LED light source.

    AI/ML Overview

    The provided document is a 510(k) summary for the Synergetics VersaVit, a vitreous aspiration and cutting instrument. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than providing extensive clinical study data with acceptance criteria and device performance metrics.

    Based on the information provided in the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in the format typically seen for clinical trials. Instead, it states that the device "has undergone testing and is in compliance with the applicable requirements of safety standards." and was "found to perform equivalently to the predicate device in a series of bench tests."

    ElementSynergetics VersaVit (Subject Device Performance)Predicate Device (Alcon Accurus, K911808)Acceptance Criteria (Implied Equivalence)
    Intended UsePosterior Segment VitrectomiesAnterior and Posterior Segment Ophthalmic SurgerySubstantially Equivalent (for posterior segment)
    Microprocessor basedYesYesEquivalent
    Aspiration Pump TypeDiaphragmVenturiSufficient for intended use, functionally equivalent
    Operating Pressure72.5 - 120 PSI70 - 120 PSIWithin acceptable/equivalent range
    Pneumatic SourceCompressed Air, CO2 cartridgesCompressed AirFunctionally equivalent methods
    Unit Height x Weight x Depth10" x 13" x 11"20" x 19" x 20.5"Differences allowed if not impacting safety/effectiveness
    Unit weight25 pounds90 poundsDifferences allowed if not impacting safety/effectiveness
    Electrical power specifications100-120V, 220-240V, 50/60 Hz100-120V, 220-240V, 50/60 HzEquivalent
    Consumable Packs provided sterileYesYesEquivalent
    Consumable Packs method of sterilizationETOETOEquivalent
    Consumable Packs Sterility Assurance Level10-610-6Equivalent

    The "acceptance criteria" here are largely implied to be "substantially equivalent" or "functionally equivalent" to the predicate device for parameters where direct identical values are not present, or "compliant with applicable safety standards" for overall performance.

    2. Sample size used for the test set and the data provenance:

    The document mentions "a series of bench tests" and "internal comparative bench testing" but does not specify sample sizes or data provenance (e.g., country of origin, retrospective/prospective) for these tests. This is typical for a 510(k) application relying on bench testing for equivalence, rather than clinical studies with human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided because the submission primarily relies on bench testing and comparison to an existing predicate device rather than human-read image interpretation or diagnostic performance that would require expert-established ground truth.

    4. Adjudication method for the test set:

    Not applicable, as the document describes bench testing and not a study requiring adjudication of human interpretations or diagnoses.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The document describes a medical device (vitreous aspiration and cutting instrument) and not an AI/CAD system for image interpretation or diagnosis. Therefore, an MRMC study related to human reader improvement with AI assistance would not be relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI algorithm. The device is a surgical instrument.

    7. The type of ground truth used:

    For bench testing, the "ground truth" would typically refer to established engineering specifications, performance standards, or known performance characteristics of the predicate device against which the subject device's performance is measured. The document states compliance with "applicable requirements of safety standards" and equivalence to the predicate device's performance.

    8. The sample size for the training set:

    Not applicable, as this is a medical device and not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as this is a medical device and not an AI/machine learning model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1