Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K183300

    Validate with FDA (Live)

    Date Cleared
    2019-01-03

    (37 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizer cycles:

    · Lumen, Non Lumen, Flexible, Fast and Fast Non Lumen Cycles of the V-PRO® 1, 1 Plus, maX, maX 2, 60 and s 2 Sterilizers

    · STERRAD® 100S Sterilizer (default cycle)

    • STERRAD® 200 Sterilizer (default cycle)

    · Standard and Advanced Cycles of the STERRAD® NX Sterilizers with All Clear Technology

    · Express, Flex Scope and Standard Cycles of the STERRAD® 100NX Sterilizers with All Clear technology

    The VERIFY V24 BI Challenge Pack is intended for qualification testing of the V PRO 1 Plus, VPRO maX, V-PRO maX 2, V-PRO 60 and V PRO s 2 Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs.

    The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.

    The challenge pack is not intended for routine of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.

    Device Description

    The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare facilites to monitor the V-PRO® Low Temperature Sterilization Systems. It is designed to accompany medical devices placed in the sterilizer. The user places the VERIFY V24 SCBI into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation. The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    The VERIFY V24 Biological Indicator Challenge Pack is used by healthcare facilities for qualification testing of the V-PRO Low Temperature Sterilization Systems following installation, relocations, malfunctions or major repairs. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital defined challenge load is not included. The user places the VERIFY V24 Biological Indicator Challenge Pack into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the VERIFY HPU Chemical Indicator (CI) and the VERIFY V24 Self-Contained Biological Indicator (SCBI) contained in the challenge pack are retrieved. The CI is assessed for a passing color change immediately and the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation. The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the VERIFY V24 Self-Contained Biological Indicator and VERIFY V24 Biological Indicator Challenge Pack, based on the provided document.

    1. Table of Acceptance Criteria and Reported Device Performance

    It's important to note that the document presents separate acceptance criteria and testing for the "Self-Contained Biological Indicator (SCBI)" and the "Challenge Pack."

    For the VERIFY V24 Self-Contained Biological Indicator (SCBI):

    TestAcceptance CriteriaReported Device Performance
    ½ & ¾ Cycle PerformanceAll test SCBIs exposed in either the ½ or ¾ cycle display passing resultsPASS
    Simulated UseAll processed SCBIs exhibit negative growth results All processed SCBI Label PIs exhibit a “pass” result All processed CIs exhibit a “pass” resultPASS
    Verification of Viable Spore Population after exposure to ALLClearThe mean population of V24 SCBI exposed to the ALLClear pre-conditioning was -50% to +300% of the unexposed V24 SCBI populationPASS
    Reduced Incubation Time (RIT) TestingMeets FDA's requirement of > 97% alignment of the 20-minute results with the conventional incubation time of 7 daysPASS
    Viable spore population2.0 - 3.4 x 10^6 spore/SCBIPASS (stated to contain "greater than 10^6 spores/BI" and "Same" as predicate in Table 5-1)
    D-value4-8 secPASS (stated "Same" as predicate in Table 5-1)
    Survival Time4-30 secPASS (stated "Same" as predicate in Table 5-1)
    Kill Time≤ 16 minPASS (stated "Same" as predicate in Table 5-1)
    Hold TimePerformance not affected if incubated within 72 hours of exposure to VHP sterilizationPASS

    For the VERIFY V24 Biological Indicator Challenge Pack:

    TestAcceptance CriteriaReported Device Performance
    Comparative Dose Response to Biological ModelThe challenge pack shall demonstrate equal or greater resistance as compared to the worst case biological modelPASS
    Simulated UseAll processed SCBIs exhibit negative growth results All processed SCBI Label PIs exhibit a “pass” result All processed CIs exhibit a “pass” resultPASS

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size used for the test set in any of the studies. It refers to "All test SCBIs" and "All processed SCBIs" without providing a count.

    The data provenance is not specified beyond being "nonclinical tests" and "performance testing to demonstrate substantial equivalence to the predicate." There is no mention of country of origin or whether the studies were retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The document describes tests related to the biological and chemical indicators themselves and their performance in sterilization cycles, not a diagnostic or interpretive task that would typically involve human expert adjudication for ground truth.

    4. Adjudication Method for the Test Set

    This information is not provided as the tests described are objective measures of biological indicator performance (e.g., growth/no growth, color change, spore population) rather than subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. The device described (Biological Indicator and Challenge Pack) is a sterilization process indicator, not an AI or imaging device that would typically be involved in human-in-the-loop diagnostic tasks or benefit from a reader study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable as the device is a physical biological indicator, not an algorithm. Its performance is evaluated through laboratory and simulated use testing.

    7. The Type of Ground Truth Used

    The ground truth for the performance tests appears to be based on:

    • Biological viability: Growth or lack thereof of Geobacillus stearothermophilus spores after exposure to sterilization cycles, as determined by a color change in the growth medium.
    • Established physical and chemical properties: D-value, survival time, kill time, and viable spore population are laboratory-measured physical characteristics against established standards.
    • Chemical indicator color change: A visible, objective change in color on the indicator.
    • Comparative resistance: The challenge pack's resistance compared to a "worst case biological model."

    Essentially, the "ground truth" is defined by the expected biological or chemical response to a sterilization process.

    8. The Sample Size for the Training Set

    This is not applicable. There is no concept of a "training set" for a biological indicator. The tests described are performance and validation tests, not machine learning model training.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172748

    Validate with FDA (Live)

    Date Cleared
    2018-02-09

    (150 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizers and cycles:

    • · Non Lumen, Flexible, Lumen and Fast Non Lumen Cycles of the V-PRO 1, 1 Plus, maX, 60 and maX 2 Sterilizers
    • · Standard Cycle of the STERRAD® 100S Sterilizer
    • · Standard and Advanced Cycles of the STERRAD® NX Sterilizer
    • · Express, Flex Scope and Standard Cycles of the STERRAD® 100 NX Sterilizer

    The VERIFY V24 Challenge Pack is intended for qualification testing of the V-PRO 1 Plus, V-PRO maX, V-PRO maX 2 and V-PRO 60 Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs.

    The VERIFY V24 Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.

    The VERIFY V24 Challenge Pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing the Sterilizers.

    Device Description

    The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare facilites to monitor the V-PRO® Low Temperature Sterilization Systems. It is designed to accompany medical devices placed in the sterilizer.

    The user places the VERIFY V24 Self-Contained Biological Indicator into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the VERIFY V24 Self-Contained Biological Indicator can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.

    The VERIFY V24 Self-Contained Biological Indicator is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the VERIFY V24 Self-Contained Biological Indicator vial and releases the growth media.

    The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The VERIFY V24 Self-Contained Biological Indicator indicates a pass if the media remains orange and non-turbid. The VERIFY V24 Self-Contained Biological Indicator indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    The VERIFY V24 Challenge Pack is used by healthcare facilities for qualification testing of the V-PRO Low Temperature Sterilization Systems following installation, relocations, malfunctions or major repairs. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital defined challenge load is not included.

    The user places the VERIFY V24 Challenge Pack into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the VERIFY HPU Chemical Indicator (CI) and the VERIFY V24 Self-Contained Biological Indicator (SCBI) contained in the challenge pack are retrieved. The CI is assessed for a passing color change immediately and the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.

    The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.

    The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.

    AI/ML Overview

    This document describes the acceptance criteria and study proving the VERIFY® V24 Self-Contained Biological Indicator (SCBI) and the VERIFY® V24 Challenge Pack meet these criteria.

    VERIFY® V24 Self-Contained Biological Indicator

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance/Conclusion
    1/4, 1/2 & 3/4 Cycle PerformancePartial positive results obtained in 1/4 and 1/2 cycles; all negative results obtained in 3/4 cycle evaluations.PASS
    Growth InhibitionUninoculated VERIFY V24 SCBIs processed in a Fast Non Lumen Cycle and inoculated with low numbers (<100 CFU) of Geobacillus stearothermophilus should all exhibit growth.All tested SCBIs exhibited growth. PASS
    Simulated UseVERIFY V24 SCBIs processed in the Fast Non Lumen Cycle under simulated use conditions should result in: All processed SCBIs exhibit negative growth results; All processed SCBI Label PIs exhibit a “pass” result; All processed CIs exhibit a “pass” result.PASS
    Reduced Incubation Time (RIT) Testing30 – 80% survival of exposed BI with a ≥ 97% correspondence between the 24-hour results and the conventional incubation time of 7 days.PASS
    Viable Spore Population2.0 – 3.4 x 10^6 spore/SCBIPASS (met the range)
    D-value4-8 seconds (at 2.7 mg/L H₂O₂)PASS (met the range)
    Survival Time4-30 seconds (at 2.7 mg/L H₂O₂)PASS (met the range)
    Kill Time≤ 16 minutes (at 2.7 mg/L H₂O₂)PASS (met the criteria)
    Hold TimePerformance of VERIFY V24 SCBIs held for 72 hours after exposure to VHP was evaluated.PASS

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated in the provided text for individual tests, but implied to be sufficient for each test type.
    • Data Provenance: The studies were conducted by STERIS Corporation, the manufacturer. The information does not specify country of origin for specific data sets, but the company is located in the USA. The studies appear to be prospective, designed specifically for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The "ground truth" for a biological indicator and its performance characteristics (like D-value, survival time, kill time, and growth/no-growth) are established through microbiological laboratory methods and adherence to recognized standards (e.g., FDA guidance on BI submissions, ISO 11138-1), not through expert consensus in the same way an AI diagnostic tool might be. The interpretation of results (growth/no growth, color change) is objective.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is a scientific and objective assessment of a biological and chemical indicator's performance against defined quantifiable metrics, not a subjective human interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a biological indicator that provides a visual readout.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth is established through microbiological testing and adherence to established industry-specific standards and guidelines. This includes:
      • Direct enumeration of viable spores.
      • Controlled exposure to the sterilant (H₂O₂ at specific concentrations and times) to determine D-value, survival time, and kill time.
      • Incubation under controlled conditions to observe microbial growth (turbidity, color change) as an indicator of sterilization failure.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.

    VERIFY® V24 Challenge Pack

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance/Conclusion
    Comparative Dose Response to Biological ModelThe VERIFY V24 Challenge Pack shall demonstrate equal or greater resistance as compared to the worst-case biological model in the Fast Non Lumen Cycle.PASS
    Simulated UseVERIFY V24 Challenge Packs processed in the Fast Non Lumen Cycle under simulated use conditions should result in: All processed SCBIs exhibit negative growth results; All processed SCBI Label PIs exhibit a “pass” result; All processed CIs exhibit a “pass” result.PASS

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated in the provided text for individual tests, but implied to be sufficient for each test type.
    • Data Provenance: The studies were conducted by STERIS Corporation, the manufacturer, in the USA. The studies appear to be prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Similar to the SCBI, the ground truth for the challenge pack's performance is based on objective microbiological and chemical indicator principles, not expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth is established through microbiological testing and adherence to established industry-specific standards and guidelines for process challenge devices. This includes:
      • Comparison to a "worst-case biological model" to ensure adequate challenge.
      • Observation of growth/no-growth for the contained SCBI and color change for chemical indicators after exposure to sterilization cycles.

    8. The sample size for the training set:

    • Not applicable.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1