Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K071976

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-09-06

    (51 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VECTEC Disposable Trocar and Laparoscopic Accessories are single-use, sterile devices intended to provide access to and visualization of body cavities, organs, and canals and to cut, dissect, manipulate, and/or cauterize various tissues during endoscopic/laparoscopic, general, vascular, gynecological and thoracic surgical procedures. Accessories include scissors, forceps, dissectors and a hook.

    Device Description

    The VECTEC Disposable Trocars and Laparoscopic Accessories can be used in a variety of procedures (general, vascular, gynecological and thoracic surgical procedures) to cut, dissect, manipulate, and/or cauterize various tissues. Trocars are sterile, single-use devices that allow visualization of body cavities and use of the laparoscopic accessories. The accessory devices are disposable, single use, individually packaged devices that are composed of biocompatible materials. Scissors, forceps, and dissectors have a handle attached to an insulated shaft with different tips, which allows the shaft and tip to rotate. They include a male cautery connector when attached to standard monopolar cautery cables and their generators. All devices are sterilized using a traditional, validated ethylene oxide procedure per EN 550:1994 to a SAL of 106 and with acceptable residual EO levels per ISO 0993-7:1995.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the VECTEC Disposable Trocars and Laparoscopic Accessories, based on the provided document:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the VECTEC device were based on demonstrating substantial equivalence to predicate devices, particularly regarding safety and performance. The primary performance criteria centered on the dielectric withstand testing requirements for electrosurgical devices.

    Acceptance CriteriaReported Device Performance
    Dielectric withstand testing requirementsVECTEC laparoscopic accessories met the dielectric withstand testing
    of ANSI/AAMI HF-18:2001requirements of ANSI/AAMI HF-18:2001.
    Substantial equivalence in Intended UseDemonstrated by similarity in Intended Use to predicate devices.
    Substantial equivalence in characteristicsDevices are similar in characteristics to predicate devices.
    Biocompatible materialsComposed of biocompatible materials; same as predicate devices.
    Energy source compatibilityMonopolar electrosurgical energy is the same as used for predicate devices.
    SterilizationSterilized using traditional, validated ethylene oxide procedure per EN
    550:1994 to a SAL of 10-6 with acceptable residual EO levels per ISO
    0993-7:1995.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a "test set" in the context of clinical data or patient samples. The performance testing described is bench electrosurgical device performance testing. Therefore, there is no information on sample sizes related to human or animal subjects, nor on data provenance (e.g., country of origin, retrospective/prospective). The testing appears to be laboratory-based on the device itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable to this submission. The "ground truth" for the performance testing was adherence to the ANSI/AAMI HF-18:2001 standard for dielectric withstand testing and comparison to the characteristics of predicate devices. This is a technical standard rather than a clinical ground truth established by experts.

    4. Adjudication Method for the Test Set

    This information is not applicable. Since there's no clinical test set requiring human interpretation or consensus, there's no adjudication method mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission relies on substantial equivalence to predicate devices and bench testing.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The device is not an algorithm or AI system. It is a physical medical device (trocars and laparoscopic accessories) designed for direct surgical use.

    7. The Type of Ground Truth Used

    The ground truth used for demonstrating the device's acceptable performance was primarily:

    • Engineering Standards: Meeting the dielectric withstand testing requirements of ANSI/AAMI HF-18:2001.
    • Predicate Device Equivalence: Direct comparison of technological characteristics, intended use, materials, and energy source compatibility with already legally marketed predicate devices.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1