510(k) Data Aggregation

    K Number
    K973659

    Validate with FDA (Live)

    Device Name
    VASOCOR PVR 100
    Manufacturer
    Date Cleared
    1997-12-17

    (83 days)

    Product Code
    Regulation Number
    870.2780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vasocor PVR-100 (Pulse Volume Recorder) is a non-invasive medical device used in conjunction with other devices such as; continuous-wave Doppler ultrasound, treadmill testing, and ultrasonic imaging techniques for vascular studies of limbs and digits. Use of the PVR in this manner allows the physician to non-invasively diagnosis extremity arterial and venous disorders.

    Device Description

    The Vasocor PVR-100 provides Pulse Volume Recording (PVR) made possible through air plethysmography. The PVR-100 device, which is based on a PC platform, includes various standard blood pressure cuff sizes that allow use of the PVR measurement on various limb and digit extremities, a pneumatic circuit that allows for internal calibration as well as inflation cuff pressure, and a user interface that includes various menus and screens for inputting patient demographic data, performing tests, displaying and entering results, and printing reports. The PVR tracing is accomplished by attaching a standard blood pressure cuff on various areas of the extremities. Once atmospheric air is injected into the cuff to provide skin contact, a pressure transducer records the pressure change in the PVR cuff secondary to change in volume of the extremity segment under the cuff during repeated cardiac cycles. PVR measurement is accomplished by calculating the volume change in the extremity segment under evaluation, where the PVR tracing shows the change in pressure of the same segment during the cardiac cycle.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Vasocor PVR-100 device:

    Vasocor PVR-100: Acceptance Criteria and Study Details

    The provided document describes the Vasocor PVR-100, a Pulse Volume Recorder (PVR) that utilizes air plethysmography for non-invasive vascular studies. The submission aims to establish substantial equivalence to a predicate device, the Life Sciences Pulse Volume Recorder.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" in the traditional sense (e.g., target specificity, sensitivity, or amplitude correlation coefficient). Instead, the non-clinical and clinical testing sections describe a comparison methodology to demonstrate equivalence to the predicate device. The primary "acceptance criteria" can be inferred as: producing PVR amplitude measurements and contours that are comparable to those produced by the predicate device.

    Acceptance Criteria (Inferred)Reported Device Performance (Vasocor PVR-100)
    PVR amplitude measurements are comparable to the predicate device.Amplitudes were measured in millimeters to the nearest 0.5mm. (Comparison to predicate was made, implying similar performance and measurement precision).
    PVR contour analysis is comparable to the predicate device.Contour analysis was made by visual analysis (method used clinically). (Comparison to predicate was made, implying similar visual characteristics).
    Ability to mimic human circulation in lower extremities under varying hemodynamic conditions (non-clinical).Hemodynamic model produced Pressure and Volume contours mimicking normal human circulation and was responsive to changes in flowrate, system compliance, and resistance.
    Functions equivalently for both normal subjects and subjects with known peripheral vascular disease (clinical).Successfully tested on 3 normal subjects and 3 subjects with known peripheral vascular disease on various limb and digit extremities.
    Overall, device is "substantially equivalent" to predicate within the stated intended use.Conclusion: The Vasocor PVR is substantially equivalent to the Life Sciences Pulse Volume Recorder.

    2. Sample Size and Data Provenance

    • Test Set Sample Size:
      • Non-clinical (In-vitro): Not explicitly stated how many distinct "tests" or parameter variations were run on the hemodynamic model, but the model itself served as the synthetic "test set."
      • Clinical: Six subjects (3 normal, 3 with known peripheral vascular disease).
    • Data Provenance: The document does not specify the country of origin for the clinical subjects. Given the submitter's address (Miami, Florida), it's highly probable the clinical data was collected in the USA.
    • Retrospective or Prospective:
      • Non-clinical: The hemodynamic model was "developed specifically for the Vasocor PVR," suggesting a controlled, prospective experimental setup.
      • Clinical: The six subjects were "evaluated using the new Vasocor PVR-100 subject to this 510(k) and the predicate device," indicating a prospective comparative study.

    3. Number of Experts and Qualifications for Ground Truth

    • Non-clinical: The ground truth for the in-vitro model was based on its design to "produce Pressure and Volume contours that mimic the normal human circulation in lower extremities." This effectively means the model itself is the ground truth based on established hemodynamic principles. No external experts are mentioned for this part.
    • Clinical: The document mentions "known peripheral vascular disease" for three subjects, implying a pre-existing clinical diagnosis by qualified medical professionals (e.g., vascular specialists, radiologists). The "visual analysis" of PVR contours is stated as a "method used clinically," implying interpretation by clinical experts. However, the exact number and qualifications of experts involved in establishing the ground truth for the study itself (e.g., confirming the disease status of the patients, or interpreting the PVR tracings for comparison) are not explicitly stated.

    4. Adjudication Method

    The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The comparison of PVR amplitude and contour was made between the Vasocor PVR-100 and the predicate device. "Amplitudes were measured in millimeters to the nearest 0.5mm." "Contour analysis were made by visual analysis (method used clinically)." This suggests a direct comparison, likely by the study investigators or a designated clinical expert, rather than a multi-reviewer adjudication process.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was NOT done. The study's focus was on the substantial equivalence of the device's output (PVR tracings) compared to a predicate device, not on how human readers' diagnostic performance might improve with or without AI assistance. This device is an output-generating tool, not an AI-assisted diagnostic aid for human interpretation.

    6. Standalone Performance Study

    Yes, a standalone performance essentially was done for the device itself. The PVR-100's ability to generate PVR tracings and measurements was evaluated directly, both in a hemodynamic model and on human subjects, in comparison to the predicate device. The performance being measured is the device's output, independent of human interpretation for diagnostic purposes, though the output is intended for human interpretation.

    7. Type of Ground Truth Used

    • Non-clinical: Hemodynamic model (designed to mimic normal human circulation) and its controlled variations.
    • Clinical:
      • For subject selection: Pre-existing clinical diagnosis (e.g., "known peripheral vascular disease").
      • For performance comparison: The output of the predicate device (Life Sciences Pulse Volume Recorder) was used as the reference against which the Vasocor PVR-100's output was compared. This implies the predicate device's output is considered the "ground truth" for what a functional PVR tracing should look like.

    8. Sample Size for the Training Set

    Not applicable. The document describes a 510(k) submission for a medical device that produces physiological measurements (PVR tracings), not a machine learning or AI model. Therefore, there is no "training set" in the context of AI model development. The device's operation is based on pneumatic and pressure transducer technology, not learned patterns from data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used/described.