Found 2 results

510(k) Data Aggregation

    K Number
    K123072

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-02-14

    (136 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vashe® Wound Therapy Solution is intended for cleansing, irrigating, moistening, and debriding acute and chronic dermal lesions, such as Stage I-IV pressure ulcers, stasis ulcers, diabetic ulcers, post-surgical wounds, first and second degree burns, abrasions and minor irritations of the skin in addition to moistening and lubricating absorbent wound dressings.

    Device Description

    Vashe® Wound Therapy Solution is a wound cleanser solution that contains hypochlorous acid generated from sodium chloride solution through the proprietary electrochemical process. Hypochlorous acid acts as a preservative that inhibits microbial contamination within the solution. The device is presented as a prescription product that requires the practitioner to diagnose the disease state and prescribe the product.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Vashe® Wound Therapy Solution, based on the provided 510(k) summary:

    The document does not detail a clinical study with human subjects, but rather focuses on non-clinical equivalency testing related to product specifications and stability. This is a common approach for 510(k) submissions where a device is substantially equivalent to a predicate device and the changes made do not significantly alter its fundamental scientific technology or intended use.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the modified device (Vashe® Wound Therapy Solution manufactured centrally) are primarily based on the chemical specifications of the solution, which were slightly expanded compared to the predicate device.

    Acceptance Criteria (Modified Device)Reported Device Performance
    Available Free Chlorine (AFC)150 to 330 ppm
    pH3.5 to 6.75 pH

    Note: The document states "Lower specification expanded pH at 3.5 to 6.75ppm", which appears to be a typo and should likely read "pH at 3.5 to 6.75". The predicate device had a pH specification of "5.3 to 6.75ppm".

    Study Details

    The provided 510(k) summary indicates non-clinical equivalency testing was conducted.

    1. Sample size used for the test set and the data provenance: Not specified in terms of number of samples, but the testing was for Shelf Life and Chemical Stability of the modified product. The data provenance is internal to PuriCore Inc. as the manufacturing process was moved from on-site generation to central manufacturing at PuriCore in Malvern, PA.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For chemical stability testing, the "ground truth" is established by standard analytical chemistry methods and adherence to the defined chemical specifications.

    3. Adjudication method for the test set: Not applicable. This was chemical and stability testing against predefined specifications.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a wound therapy solution, not an AI-powered diagnostic device.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a chemical solution, not an algorithm.

    6. The type of ground truth used: For the non-clinical testing, the ground truth was chemical specifications and stability over time, assessed through analytical methods.

    7. The sample size for the training set: Not applicable. This is a non-clinical submission for a chemical product, not a machine learning model.

    8. How the ground truth for the training set was established: Not applicable.

    Summary of Substantial Equivalence Justification

    The core of the submission for Vashe® Wound Therapy Solution relies on demonstrating substantial equivalence to its predicate device (Vashe® Wound Therapy System, K100918) through non-clinical testing. The key changes are:

    • Place of Production: Moving from on-site generation at the customer's location to central manufacturing at PuriCore Inc.
    • Expanded Specifications: The acceptable range for Available Free Chlorine (AFC) and pH was slightly expanded.
    • Shelf Life and Stability: Non-clinical testing was performed to ensure that the modified device, with its expanded specifications and centralized production, maintains chemical stability and preservative effectiveness (due to hypochlorous acid) throughout its shelf life, including under worst-case scenarios for initial and final concentrations and pH.

    The conclusion drawn by PuriCore Inc. (and accepted by the FDA for 510(k) clearance) is that these modifications have not changed the Intended Use or altered the Fundamental Scientific Technology of the device, thereby demonstrating substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093697

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-04-13

    (134 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vashe® Wound Therapy Solution is intended for OTC use for management of minor skin abrasions, minor lacerations, minor irritations, minor cuts, and intact skin.

    Device Description

    The subject device is a wound cleansing solution that is intended for cleansing, irrigating, debriding dermal wounds in addition to moistening and lubricating absorbent wound dressings. The mechanical action of fluid moving across the wound provides for the mechanism of action and aids in the removal of foreign objects such as dirt and debris.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study proving that the device meets those criteria. The document is a 510(k) summary for the Vashe® Wound Therapy Solution, a wound cleansing solution, and a letter from the FDA confirming its substantial equivalence to predicate devices.

    The information primarily focuses on:

    • Device Identification: Trade Name, Common Name, Classification Name, Product Code.
    • Submitter and Contact Information.
    • Predicate Devices: A list of previously cleared devices to which Vashe® Wound Therapy Solution is compared for substantial equivalence.
    • Device Description: What the device is and its mechanical action.
    • Indications for Use: What the device is intended for (management of minor skin abrasions, minor lacerations, minor irritations, minor cuts, and intact skin).
    • Substantial Equivalence Argument: How it is similar in function and intended use to the predicate devices.
    • Non-clinical Performance: A brief mention that pre-clinical testing demonstrated biocompatibility.
    • FDA Clearance Letter: Confirmation of substantial equivalence and regulatory information.

    There is no mention of:

    • Specific acceptance criteria (e.g., performance metrics, thresholds).
    • A formal study designed to demonstrate the device meets acceptance criteria.
    • Sample sizes for test sets or training sets.
    • Data provenance.
    • Number or qualifications of experts.
    • Adjudication methods.
    • MRMC comparative effectiveness study.
    • Standalone algorithm performance.
    • Type or establishment of ground truth.

    The "Non-clinical Performance" section states: "Pre-clinical testing demonstrated biocompatibility of the Vashe® Wound Therapy Solution." This is the only mention of any testing, but it's limited to biocompatibility and does not involve performance metrics related to acceptance criteria for the indicated uses. The basis for clearance is substantial equivalence to already marketed predicate devices, not meeting specific performance criteria through a new clinical or benchmark study.