510(k) Data Aggregation

    K Number
    K060327

    Validate with FDA (Live)

    Date Cleared
    2006-04-12

    (62 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Twin-Pass catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the coronary and peripheral arterial vasculature, to facilitate placement and exchange of guidewires and other interventional devices, for use during two guidewire procedures and to subselectively infuse/deliver diagnostic or therapeutic agents.

    The Skyway support catheters are intended to be used in conjunction with steerable guidewires in order to access discreet regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The Skyway OTW also may be used to subselectively infuse/deliver therapeutic agents.

    Device Description

    The Twin-Pass™ Dual Access Catheter is a hydrophilically coated, dual lumen catheter designed for use in the arterial vasculature. The catheter provides support for 0.014"/0.36mm guidewires, and the dual lumen design allows for the delivery of a second guidewire into distal vasculature while leaving the initial guidewire in place. The Twin-Pass catheter comes with a stiffening mandrel to provide support and pushability during catheter insertion.

    The SKYWAY catheters are single lumen, hydrophilically coated catheters designed for use in the arterial vasculature. The catheters provide support for 0.014"/0.36mm guidewires during interventional procedures, and allow for the exchange of one distally located guidewire for another one while maintaining access to distal vasculature.

    AI/ML Overview

    This document is a 510(k) summary for the Twin-Pass and Skyway Catheters, which are intravascular catheters. It presents a "Summary of Non-Clinical Testing" and explicitly states "No clinical evaluations of this product were conducted."

    Therefore, the requested information regarding acceptance criteria and a study proving the device meets these criteria cannot be provided from the given document, as no such study was conducted or reported. The 510(k) relies on substantial equivalence to predicate devices rather than direct demonstration of performance against specific acceptance criteria through a study.

    The document indicates:

    1. Summary of Non-Clinical Testing: Flow-rate testing was conducted using saline and a 76% contrast medium. This constitutes non-clinical testing, not a study to prove acceptance criteria in the context of device performance as one might expect for a diagnostic or AI-based device.
    2. Summary of Clinical Testing: "No clinical evaluations of this product were conducted."

    Therefore, the sections of your request regarding acceptance criteria, device performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, and ground truth establishment for a study are not applicable to the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052257

    Validate with FDA (Live)

    Date Cleared
    2005-11-23

    (97 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Twin-Pass Dual Access Catheter is to be used in conjunction with steerable quidewires in order to access discrete regions of the coronary and peripheral arterial vasculature, to facilitate placement of guidewires and other interventional devices, and for use during two guidewire procedures.

    Device Description

    The Twin-Pass Dual Access Catheter is a 3F O.D. catheter that has two lumens-a short distal lumen and a second full length lumen-each of which are compatible with a 0.014" standard quide wire. The Twin-Pass catheter has a working length of 135cm and contains positioning markers at 95 and 105cm which provide a visual indication of the relative positions of Twin Pass and the end of a standard 105cm guide catheter. Two radiopaque marker bands at the end of each wire lumen provide for a radiographic means of locating the position of each lumen. The softer, distal end of the catheter is coated with a hydrophilic coating to assist passage through the guide catheter and vessels while the proximal end of the catheter contains a strain relief and a standard luer hub. A 126cm stiffening mandrel is included which provides support and pushability to the Twin-Pass.

    AI/ML Overview

    This document does not contain an acceptance criteria table or information about a study proving device performance in the context of device outputs (e.g., diagnostic accuracy metrics like sensitivity, specificity, or AUC) against a set of predetermined criteria. This is primarily because the submitted document is a 510(k) premarket notification for a medical catheter (Twin-Pass Dual Access Catheter), not a diagnostic algorithm or AI-powered device.

    For a medical device like the Twin-Pass Dual Access Catheter, performance is generally established through design verification testing (non-clinical testing) and substantial equivalence to predicate devices, not through studies that involve human readers, ground truth consensus, or metrics like sensitivity/specificity for disease detection.

    Here's why the requested information is not applicable or cannot be extracted from the provided text:

    • Acceptance Criteria & Reported Device Performance (Table): Not present. The document summarizes non-clinical testing for design verification and states the device is suitable for its intended use based on these tests. It does not provide specific performance metrics in the way a diagnostic algorithm would (e.g., "Sensitivity > X%").
    • Sample Size for Test Set & Data Provenance: No clinical test set as such was used for performance evaluation that would require this. The non-clinical tests would have involved samples of the device itself.
    • Number of Experts & Qualifications for Ground Truth: Not applicable, as there was no clinical study involving human assessment of device outputs against a ground truth.
    • Adjudication method for Test Set: Not applicable.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. This type of study is for evaluating diagnostic algorithms, not passive medical devices like catheters.
    • Standalone (algorithm only) Performance: Not applicable. This device is a catheter, not an algorithm.
    • Type of Ground Truth Used: Not applicable. The "ground truth" for a catheter is its ability to perform its mechanical function as intended during design verification.
    • Sample Size for Training Set: Not applicable. There is no algorithm to train.
    • How Ground Truth for Training Set Was Established: Not applicable.

    What the document does state regarding device evaluation:

    • Non-Clinical Testing: "Testing conducted included assessments of the design verification of the Twin-Pass Dual Access Catheter along with biocompatibility assessments. The results of this battery of tests confirmed the suitability of the Twin-Pass Dual Access Catheter for its intended use." This indicates engineering and materials testing, not accuracy testing against a clinical "ground truth."
    • Clinical Testing: "No clinical evaluations of this product have been conducted." This explicitly states that clinical studies were not performed to evaluate this product's performance in humans.
    • Predicate Device: The device demonstrates "substantial equivalence" to existing legally marketed predicate devices (Lumend Percutaneous Catheter, Quick-Cross Catheter, and Dual Lumen Catheter). This is the primary means of clearance for this type of device.

    In summary, the provided document describes a 510(k) clearance for a medical catheter based on non-clinical testing and substantial equivalence to predicate devices, not on a clinical performance study with human readers or AI algorithms.