    Device Name
    VAR MONITOR
    Date Cleared
    2008-05-23

    (185 days)

    Product Code
    Regulation Number
    868.5925
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is to be used by properly trained personnel to deliver emergency, short term, constant flow - pressure cycled ventilatory support.

    Device Description

    The VAR-Monitor is a battery (9 VDC) operated device that is mounted onto the VAR to monitor cycling conditions during operation.

    AI/ML Overview

    The provided text describes a 510(k) summary for the "VAR with VAR-Monitor™" device, which is a powered emergency ventilator.

    Here's an analysis based on the given information:

    1. Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for the VAR with VAR-Monitor is its compliance with ASTM F 920-93 (Reapproved 1999), "Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans."

    Acceptance Criteria (Standard)Reported Device Performance
    Mechanical Performance: Meets or exceeds ASTM Designation: F 920-93Passed: "The VAR with VAR-Monitor meets and exceeds 'Standard
    (Reapproved 1999) "Standard Specification for Minimum Performance and SafetySpecification for Minimum Performance and Safety Requirements for Resuscitators
    Requirements for Resuscitators Intended for Use with Humans."Intended for Use with Humans,' ASTM Designation: F 920-93 (Reapproved 1999)."
    Alarm Functionality: Detection of non-cycling condition and activation ofFunctionality Confirmed: "VAR-Monitor detects a non-cycling condition of
    visual and audible alarms within a preset time of eight (8) seconds.the VAR modulator Indicating that no pressure changes have occurred. When the
    VAR-Monitor is powered on and ready, the device will check for the 'pressure
    signal.' If the pressure stays unchanged for more than the preset time of eight
    (8) seconds, both visual and audible alarms will be activated."
    Clinical Efficacy (Indirect through equivalence): Substantially equivalentSubstantially Equivalent: "In conclusion, the modified VAR with VAR-Monitor
    to legally marketed predicate devices in terms of intended use,is substantially equivalent (SE) to legally marketed predicate devices."
    operational characteristics, and safety.
    Absence of Adverse S&E Information: No new safety or effectiveness concerns.None: "Adverse S & E Information: None"
    Overall Conclusion: Passed all required tests and met all predeterminedConclusion: "The VAR with VAR-Monitor passed all required tests and
    acceptance criteria for applicable standards.demonstrated that it meets all predetermined acceptance criteria for applicable
    standards."

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "Clinical Tests: None" and does not refer to a "test set" in the context of clinical data. The evaluation appears to be based on engineering tests against the ASTM standard and comparison to predicate devices, rather than a clinical study with patients or a specific dataset. Therefore, there is no mention of sample size or data provenance in this context.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    As "Clinical Tests: None" is stated, there is no indication of a test set requiring ground truth established by experts. The evaluation primarily relies on compliance with the ASTM standard and comparison to predicate devices.

    4. Adjudication Method for the Test Set

    Given that no clinical test set is described, there is no mention of an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly states "Clinical Tests: None." This type of study would fall under clinical testing.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    The device is a physical medical device (ventilator) with a monitor, not an algorithm in the typical AI sense. The performance evaluation focuses on the mechanical and alarm functions of the device itself, which are inherently "standalone" in their operation rather than involving a human-in-the-loop for interpreting algorithm outputs.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's acceptable performance is primarily derived from:

    • Engineering Standards: Compliance with ASTM F 920-93 (Reapproved 1999) for performance and safety.
    • Operational Characteristics: Whether the device successfully detects a non-cycling condition and triggers alarms within the specified time.
    • Predicate Device Comparison: The established safety and effectiveness of the existing predicate devices (RespirTech PRO, RespirTech PRO O₂C, VAR-Plus Model PCM).

    There is no mention of expert consensus, pathology, or outcomes data as "ground truth" for a specific clinical dataset.

    8. The Sample Size for the Training Set

    The document does not describe the development of an AI algorithm or a "training set." The device is a ventilator with an alarm monitor, evaluated against engineering standards and substantial equivalence to predicate devices, not through machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set or AI algorithm.