Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K090426

    Validate with FDA (Live)

    Date Cleared
    2009-06-12

    (113 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VAKU-8™ Blood Collection Needles are designed for routine blood collection by a qualified practitioner

    Device Description

    VAKU-8™ Single use Multi Sample blood collection needles are composed of a stainless steel needle tube that is ground at both ends and affixed to a plastic hub. A rubber sleeve covers the backend needle tube and hub tip. The front end needle is used to withdraw blood once placed in the patient. The needle tube is protected by a plastic needle cover on either end. The "front end needle" is designed to withdraw blood from the patient. The "back end needle" is covered by a rubber sleeve and is designed with a threaded hub that can be attached to a blood collection tube holder. The tube holder facilitates attachment to a evacuated blood collection tube. Each needle is individually packaged sterile utilizing two plastic end covers. A perforated label serves to simplify identification of needle size and also acts as a seal of integrity.

    AI/ML Overview

    The provided text describes the 510(k) submission for the VAKU-8™ Blood Collection Needle, establishing its substantial equivalence to a predicate device. The information details the device, its intended use, a comparison with the predicate device, and various tests and certifications.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on the device conforming to British Standard 4843 for measurements and tolerances, color coding according to ISO 6009, and comparable physical and performance characteristics to the predicate device. The "Bench Mark Study of Blood Collection Needle" provides the reported performance.

    Element of ComparisonAcceptance Criteria (based on predicate or standard)Reported Device Performance (VAKU-8™)
    Intended UseSterile, Single Use Needle for Blood CollectionSterile, Single Use Needle for Blood Collection
    Cannula MaterialSUS304 Stainless SteelSUS304 Stainless Steel
    Needle Gauges21G, 22G21G, 22G, 23G
    Needle Lengths1" and 1.5"1" and 1.5"
    LumenSingleSingle
    Bottom Cover MaterialPPPP
    Bottom Cover TransparencyTransparentTransparent
    Rubber Sleeve ColorGrayGray
    Rubber Sleeve TransparencyOpaqueOpaque
    Rubber Sleeve MaterialLatex free IsopreneLatex free Isoprene
    Hub DesignThreadedThreaded
    Front End Needle Effective Length22.60-23.00mm25.00-25.50mm
    Back End Needle Effective Length14.60-14.80mm15.0-15.50mm
    Front End Bevel Length3.69-3.80mm3.55-3.65mm
    Front Section Bevel Length1.55-1.60mm1.60-1.66mm
    Front Angle of Bevel9° 30' - 10° 00'10° 30' - 11° 30'
    Removal of CoversComfortable (Predicate device performance)Comfortable
    Fitment with HolderComfortable (Predicate device performance)Comfortable
    Maximum Load0.83-1.00N (Predicate device performance)0.67-0.88N
    Shaft Friction0.13-1.5N (Predicate device performance)0.10-0.15N
    Retraction of tube over cannulaOK (Predicate device performance)OK
    Cannula Penetration over rubber sleeveOK (Predicate device performance)OK
    BiocompatibilityUSP Class VI Plastic requirements (for PP)Meets USP Class VI Plastic
    SterilitySterilized by ETO Gas, SAL requirements per guidance document, USP sterility testsSterilized by ETO Gas, validated per EN 550, USP sterility tests
    Freedom from PyrogenUSP test for "freedom from Pyrogenic materials"Meets USP requirements

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample size for the "Bench Mark Study of Blood Collection Needle" or for any performance tests that directly compared the VAKU-8™ to its predicate. It mentions:

    • "Samples are drawn from each batch produced as per defined sampling plan, performance tests are conducted as per the standards cited above and defined criteria for acceptance are applied." (Section 13)
    • "Samples are drawn from each sterilization load and subjected to the tests for sterility as per USP." (Section 15)
    • "Freedom from Pyrogen is determined by taking samples from each sterilizer load and subjected to the test 'freedom from Pyrogenic materials' as per USP" (Section 15)

    This indicates that testing is done on batches and sterilization loads, but the exact number of samples per test is not specified.

    The data provenance is not explicitly mentioned in terms of country of origin for the test data, nor is it specified if it's retrospective or prospective. However, given that it's a 510(k) submission for a new device, the performance data would be prospective, obtained during the development and validation of the VAKU-8™.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide information about the number of experts or their qualifications used to establish ground truth for the test set. The tests appear to be a mix of objective measurements against standards (BS 4843, ISO 6009) and functional assessments (e.g., "Comfortable," "OK") that would likely be evaluated by trained personnel, but specific expert involvement is not detailed.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1, none) is mentioned or implied for the performance tests. The assessments seem to be based on direct measurement or functional observation against a standard or predicate.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study was done. This type of study is typically associated with diagnostic imaging devices or algorithms where human interpretation is a key component, and the device aims to assist or replace human readers. The VAKU-8™ is a physical medical device (blood collection needle) where human interaction is primarily mechanical (e.g., proper technique by a qualified practitioner), not interpretative.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The concept of a "standalone" performance for an algorithm is not applicable to this device, as it is a physical medical instrument, not an algorithm. Performance tests are conducted on the physical device itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the VAKU-8™ is established through:

    • Objective Measurements/Standards: Conformance to British Standard 4843 for measurements and tolerances, and ISO 6009 for color coding.
    • Predicate Device Performance: Direct comparison of physical attributes and functional characteristics (e.g., "Removal of Covers," "Fitment with holder," "Maximum load," "Shaft friction") against the established performance of the legally marketed predicate device (K992699).
    • Biocompatibility Certifications: Testing of materials (e.g., Topilene PP) to meet USP Class VI Plastic requirements.
    • Sterilization and Pyrogenicity Standards: Conformance to EN 550 for sterilization validation and USP tests for sterility and freedom from pyrogenic materials.

    8. The Sample Size for the Training Set

    This information is not applicable. The VAKU-8™ Blood Collection Needle is a physical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as the device does not involve a "training set."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1