Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110083

    Validate with FDA (Live)

    Date Cleared
    2012-01-13

    (367 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Esprit and V200 Ventilators are microprocessor controlled, electrically powered, mechanical ventilators. They are intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult, pediatric, and neonatal patients as prescribed by a physician. The Esprit and V200 Ventilators are intended for use in either invasive or non-invasive applications.

    The Esprit and V200 Ventilators with APRV Mode are intended for use for invasively ventilated adult and pediatric patients as prescribed by a physician

    The intended use is the same as that of the predicate devices, except that the APRV Mode is for use on a subset of the patient population the original devices are cleared for (e.g. neonatal patients and non-invasive applications are excluded).

    Device Description

    The Esprit and V200 Ventilators are microprocessor controlled, electrically powered, mechanical ventilators. This modification to the currently marketed Esprit Ventilator and V200 Ventilators is the addition of the APRV Mode.

    The APRV Mode is an optional software upgrade. It is both a breath type and ventilation mode intended for invasively ventilated adult and pediatric patient populations. APRV enables the ventilator to deliver gas via an endotracheal tube or tracheostomy tube at two levels of pressure (Press High and Press Low), and allows for spontaneous or supported breathing at both levels.

    The APRV Mode is activated via a software download through an I-button and is integrated into the Esprit and V200 Ventilators in the same way as other currently released software options. It can either be installed in the factory or in the field as an upgrade to existing Esprit and V200 ventilators. Downloading this option will add a "button" to the Graphical User Interface (GUI), which is used to turn APRV on and off.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Esprit Ventilator and V200 Ventilator with APRV Option.

    It's important to note that the provided text is a 510(k) summary and FDA clearance letter, which typically summarizes the validation rather than detailing the full study protocols and results. As such, some specific details like exact acceptance criteria or raw performance data might not be explicitly stated in quantitative terms.


    Acceptance Criteria and Reported Device Performance

    The document states that "performance testing and a clinical simulation were conducted and support the assertion that the APRV Mode does not raise any new questions regarding safety and effectiveness."

    Since this is a 510(k) for an addition of a mode (APRV) to existing cleared ventilators, the acceptance criteria are generally focused on demonstrating that this new mode performs as intended and does not negatively impact the overall safety and effectiveness of the existing device. The performance is assessed against the established specifications and safety profile of the predicate devices.

    Acceptance Criteria CategoryReported Device Performance (as inferred from the document)
    Software Verification & ValidationSuccessfully completed in accordance with Respironics California, Inc. policies and procedures and the FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices dated May 11, 2005.
    Performance Testing (APRV Mode)Conducted, supports that the APRV Mode performs as intended for invasively ventilated adult and pediatric patients and does not raise new questions regarding safety and effectiveness.
    Clinical Simulation (APRV Mode)Conducted, supports that the APRV Mode performs as intended for invasively ventilated adult and pediatric patients and does not raise new questions regarding safety and effectiveness.
    Technological Characteristics (Ventilator Base)Unchanged (control mechanism, operating principle, energy type, ergonomics of patient interface, firmware, environmental specifications, performance specifications).
    Non-APRV Ventilation Modes FunctionalityIdentical to those in the currently marketed Esprit and V200 Ventilators.
    Safety & EffectivenessNo new questions regarding safety and effectiveness raised by the addition of the APRV Mode.
    Intended UseThe APRV Mode is intended for use for invasively ventilated adult and pediatric patients, a subset of the patient population for the original cleared devices, aligning with the "same intended use" principle for substantial equivalence.

    Study Details

    Due to the nature of the provided document (510(k) summary), many of the specific details for a full study report are not present. Information below is extracted or inferred from the text.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Size: Not explicitly stated. The document mentions "performance testing and a clinical simulation." For a ventilator, performance testing typically involves bench testing on a ventilator, and a clinical simulation usually involves scenarios or mock patients rather than a large cohort of real patients.
      • Data Provenance: Not specified. Given the context of performance testing and clinical simulation, it would likely be laboratory or in-house data rather than patient data from specific countries. It's prospective in the sense that the testing was conducted on the modified device.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Not specified. For performance and clinical simulations of a ventilator, the "ground truth" would be the expected performance according to engineering specifications, physiological models, and clinical guidelines. This would typically be established by internal engineering, clinical, and regulatory experts within Respironics.
    3. Adjudication Method for the Test Set:

      • Not specified. Given the nature of performance testing and clinical simulation for a medical device's functional mode, adjudication methods (like 2+1 reads) typical for diagnostic image analysis are not directly applicable. Performance is usually assessed against predefined technical and clinical thresholds.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not explicitly mentioned or conducted as described. This type of study is more common for diagnostic imaging AI devices where human reader performance (with and without AI assistance) is being evaluated against ground truth. The current device is a ventilator with an added mode, not a diagnostic tool requiring reader interpretation in the same way. The evaluation focused on the device's functional performance and safety.
    5. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance):

      • Yes, in essence. The "performance testing" and "clinical simulation" mentioned for the APRV Mode would constitute a standalone evaluation of the algorithm's (software's) performance within the ventilator system, independent of human interpretation or assistance during operation. The software's outputs (e.g., pressure, flow, volume delivery as per the APRV settings) are directly measured and compared against specifications.
    6. Type of Ground Truth Used:

      • For performance testing: Engineering specifications, physiological models, and established clinical parameters for ventilation. These would define the expected output and behavior of the APRV mode under various simulated patient conditions.
      • For clinical simulation: Clinically acceptable ranges and responses as determined by medical professionals or established medical guidelines for ventilation.
    7. Sample Size for the Training Set:

      • Not applicable / Not specified. This device is a software-controlled mechanical ventilator, not a machine learning or AI algorithm that 'learns' from a training dataset in the typical sense (e.g., image recognition). The "training" for such a system involves software development, coding, and internal validation against design specifications, not a dataset of examples.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable. As above, there isn't a "training set" in the sense of data used to train a machine learning model. The software's logic and behavior are designed based on established medical science, engineering principles for mechanical ventilation, regulatory requirements, and internal development methodologies. The "ground truth" during development and testing refers to what the device is designed to do and what it should achieve safely and effectively.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1