Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240176

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-13

    (234 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    V-VITFREEZE is intended for use in the vitrification of oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.

    V- VITWARM is intended for use in the thawing of vitrified of oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.

    Device Description

    V-VITFREEZE AND V-VITWARM include a set of five media intended for use in the vitrification and thawing of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos as part of InVitro Fertilization (IVF) procedures.

    V-VITFREEZE (2mL vials) includes two media components, Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol, trehalose (only in VS), and dimethyl sulfoxide. During the vitrification process, embryos are first exposed to ES and then to VS.

    Using this methodology, the permeating cryoprotectants can replace water in the oocyte or PN through blastocyst stage embryos prior to vitrification and storage in liquid nitrogen.

    V-VITWARM (4mL vials) is composed of three media used stepwise for thawing and removing cryoprotectants from vitrified oocytes and PN through blastocyst stage embryos. It is composed of Thawing Solution (TS), Diluent Solution (DS) and Washing Solution (WS).

    The base medium for the vitrification and thawing media includes water, salts, buffering agents, minerals, chelating agents, energy substrates, amino acids, hydroxypropyl cellulose, polymeric substance, gentamicin, and phenol red. Cryoprotectants in the media include ethylene glycol, (7.5% in ES and 15% in VS), DMSO (7.5% in ES and 15% in VS), and trehalose in VS, TS, and DS.

    The five solutions in the V-VITFREZE and V-VITWARM kit are aseptically filtered (storage vials sterilized by radiation) and provided in polypropylene vials with high density polyethylene screw caps. They have a shelf-life of 12 months when stored at 2-8℃. Media in the vials can be used for up to seven days after vial opening.

    AI/ML Overview

    This is an FDA 510(k) clearance letter for the V-VITFREEZE and V-VITWARM devices, which are reproductive media and supplements. The document describes the device, its intended use, and a comparison to a predicate device. It also summarizes non-clinical performance testing.

    However, the provided text does not contain detailed acceptance criteria and the results of a specific study in the format requested (e.g., a table of acceptance criteria and reported device performance with statistical metrics like sensitivity, specificity, accuracy). It focuses on physical and biological characteristics of the media rather than performance metrics of an AI or diagnostic device.

    Therefore, I cannot fulfill all parts of your request exactly as specified because the document does not describe the acceptance criteria and a study in the context of an AI or diagnostic device's performance in the way you've outlined.

    The document does list certain specifications for the media and the tests conducted to ensure the media meets those specifications, which are a form of acceptance criteria for manufacturing and quality control. I will extract these details.

    Here's the information that can be extracted relevant to acceptance criteria and testing, adapted to the context of the provided document:

    1. Table of Acceptance Criteria and the Reported Device Performance (as per non-clinical testing of the media):

    Test/CharacteristicAcceptance CriteriaReported Device Performance (Implied by "met" without specific data)
    AppearanceAll solutions pink to rose color without precipitatesMet (demonstrated through shelf-life testing)
    pH (WS, DS, TS)7.20 - 7.45Met (demonstrated through shelf-life testing)
    pH (VS)7.50 - 7.70Met (demonstrated through shelf-life testing)
    pH (ES)7.30 - 7.60Met (demonstrated through shelf-life testing)
    Osmolality (ES)2300-3100 mOsm/kg (1:10 dilution)Met (demonstrated through shelf-life testing)
    Osmolality (VS)4600-6150 mOsm/kg (1:20 dilution)Met (demonstrated through shelf-life testing)
    Osmolality (TS)1250-1600 mOsm/kg (1:4 dilution)Met (demonstrated through shelf-life testing)
    Osmolality (DS)730-930 mOsm/kg (1:2 dilution)Met (demonstrated through shelf-life testing)
    Osmolality (WS)240-300 mOsm/kg (No dilution)Met (demonstrated through shelf-life testing)
    SterilityNo microbial growthMet (demonstrated through shelf-life testing and aseptic filtration/filling validation)
    Bacterial Endotoxin< 0.25 EU/mLMet (demonstrated through shelf-life testing)
    Mouse Embryo Assay≥ 80% embryos developed to expanded blastocyst at 96 hours (1-Cell MEA)Met (demonstrated through shelf-life testing)
    Shelf-Life12 months (product specifications met at time 0 and after accelerated aging)Demonstrated (product specifications were met for 12 months)

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify exact sample sizes for each test listed (e.g., how many vials were tested for pH, how many mouse embryos were used). It refers to "shelf-life testing," "aseptic filtration and aseptic filling validation," etc., which are general statements of testing performed.
    • Data Provenance: Not specified in terms of country of origin or whether prospective/retrospective for these non-clinical tests. These are typically internal quality control and validation tests conducted by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. These are laboratory-based, non-clinical tests for reproductive media, not an AI or diagnostic device requiring expert interpretation of results. The "ground truth" is defined by established laboratory testing standards (e.g., USP monographs, ISO standards, FDA guidance).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is for expert consensus in diagnostic or AI studies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document describes reproductive media, not an AI-assisted diagnostic device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This document describes reproductive media.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the non-clinical tests are established by industry standards and regulatory requirements. For example:
      • Chemical/Physical tests (pH, Osmolality, Appearance): Defined by analytical standard methods (e.g., USP monographs).
      • Sterility: Defined by microbiological testing methods (e.g., USP <71>).
      • Bacterial Endotoxin: Defined by Limulus Amebocyte Lysate (LAL) test standards (e.g., USP <85>).
      • Mouse Embryo Assay (MEA): Defined by the specified FDA guidance for ART devices, where successful development of a certain percentage of embryos to expanded blastocyst within 96 hours is the benchmark.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device, so there is no training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1