    K Number
    K193269

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-20

    (24 days)

    Product Code
    Regulation Number
    876.5530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.

    Device Description

    The UroLift System is designed to access the prostatic urethra and deliver one UroLift Implant through a lobe of the prostate. The UroLift System is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. The UroLift Implant is then deployed transversely through the prostatic tissue. Multiple implants are deployed in the UroLift System procedure. The implants secure the retracted position of the urethra, thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant. The procedure typically requires 2-6 implants to retract the obstruction. The UL400 (most recently cleared in K173087), consists of two main components, the UroLift Delivery Device (single use), and the UroLift Implants (one implant per delivery device). Each Delivery Device comes pre-loaded with one UroLift Implant.

    AI/ML Overview

    This document is a 510(k) summary for the UroLift System (UL400) by NeoTract, Inc. It describes a submission concerning a modification to a contraindication, specifically increasing the maximum prostate volume from >80 cc to >100 cc for patients indicated for treatment with the UroLift System.

    Based on the information provided in the document, here's a description of the acceptance criteria and the study that proves the device meets them, specifically regarding the change in contraindication:

    Acceptance Criteria and Study for Modified Contraindication (Increased Prostate Volume)

    The core of this submission is not about proving a new device's performance from scratch, but rather demonstrating that a modification to a contraindication (increasing the prostate volume limit) does not negatively impact the device's established safety and effectiveness. Therefore, the "acceptance criteria" and "study" described herein are focused on demonstrating non-inferiority or equivalence of outcomes in the expanded patient population.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (for expanded prostate volume)Reported Device Performance (as demonstrated by clinical literature review)
    Device Performance Equivalence (Safety & Effectiveness):Outcomes (from sponsored and independent clinical studies):
    - Symptom response in patients with prostate volumes >80cc is equivalent to those with prostate volumes <80cc."These studies show that the symptom response... are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    - Quality of life in patients with prostate volumes >80cc is equivalent to those with prostate volumes <80cc."These studies show that the... quality of life... are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    - Uroflowmetry results in patients with prostate volumes >80cc are equivalent to those with prostate volumes <80cc."These studies show that the... uroflowmetry... are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    - Adverse events in patients with prostate volumes >80cc are equivalent to those with prostate volumes <80cc."These studies show that the... adverse events... are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    - Catheterization rates in patients with prostate volumes >80cc are equivalent to those with prostate volumes <80cc."These studies show that the... catheterization rates are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    Biocompatibility Equivalence:"The modification of the contraindication does not impact the biocompatibility of the existing UroLift System. As such, the data provided with the predicate device (K173087) is applicable for the proposed device with a modified contraindication." (page 5)
    Sterilization & Shelf-Life Equivalence:"The modification of the contraindication does not impact the sterilization or shelf life of the existing UroLift System. As such, the data provided with the predicate device (K173087) is applicable for the proposed device with a modified contraindication." (page 5)
    Technological Characteristics & Principles of Operation Equivalence:"The UroLift System (UL400) described in this submission is substantially equivalent to the previously cleared generations of the device. The UL400 was previously cleared in K173087. This submission concerns the modification of a contraindication of the device and does not impact the device itself." (page 4) "The data provided demonstrated the NeoTract UroLift System with a modified contraindication is as safe and effective, has the same intended use, technological characteristics and principles of operation as the predicate device." (page 5)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Description: The "test set" for this submission is not a newly collected dataset for the modified contraindication. Instead, it relies on a "clinical literature review" of existing studies.
    • Sample Size: The document states "clinical literature review of both sponsored and independent clinical studies that included men with prostate volumes greater than 80cc." (page 5). The specific number of patients or studies reviewed is not provided in this summary.
    • Data Provenance: The document does not specify the country of origin. It indicates the data comes from "sponsored and independent clinical studies," implying a mix of retrospective and prospective data collection, but does not explicitly state the design. For example, the predicate device (K173087) would have included prospective study data to support its original clearance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • This submission relies on a clinical literature review to support the change in contraindication. Therefore, the "ground truth" for the equivalence of outcomes for patients with prostate volumes >80cc is established by the aggregate findings and conclusions of the respective clinical studies included in that review.
    • The document does not specify the number of experts, their qualifications, or their role in reviewing the literature for this specific submission. The "review" itself would have been conducted by the sponsor's clinical/regulatory team, and ultimately evaluated by the FDA.

    4. Adjudication Method for the Test Set

    • Given that the evidence is based on a "clinical literature review" rather than a de novo test set with expert reads, there is no explicit adjudication method described for this submission. The "adjudication" effectively comes from the scientific rigor and peer review (if published) of the underlying studies themselves, and the FDA's assessment of the presented literature review.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done for this specific 510(k) submission. This submission is focused on a change to a contraindication based on existing clinical evidence, not on evaluating the performance of an AI model or the improvement of human readers with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. The UroLift System is a medical device (implantable system), not an AI algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for the claim that outcomes (symptom response, quality of life, uroflowmetry, adverse events, catheterization rates) are equivalent for patients with prostate volumes >80cc compared to those <80cc is based on clinical study outcomes data (clinical endpoints) as reported in sponsored and independent clinical literature.

    8. The Sample Size for the Training Set

    • Not applicable. This submission is for a medical device with a modified contraindication, not an AI/ML algorithm that requires a training set. The device's original clearance (K173087) would have involved clinical trials, which could be considered the "training data" in a broader sense for establishing the device's original safety and efficacy.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. See point 8. The ground truth for the predicate device's efficacy and safety would have been established through well-defined clinical trial endpoints and data collection, usually with a panel of clinical investigators and potentially independent data monitoring committees.