    Intended Use

    The Universal Segmented Cylinder Applicator Set and Universal Stump Applicator Set are intended for cancer treatment of the vagina, vaginal stump and rectum using HDR or PDR brachytherapy.

    The Universal Cervix Probe Sets in combination with the Universal Segmented Cylinder Applicator Set and the Universal Stump Applicator Set are intended for cancer treatment of the vaginal stump, cervix, uterus and endometrium using HDR or PDR brachytherapy.

    The Universal Titanium Cervix Probe Sets in combination with the Universal Segmented Cylinder Applicator Set and the Universal Stump Applicator Set are intended for cancer treatment of the vaginal stump, cervix, uterus and endometrium using HDR or PDR brachytherapy.

    The Universal Multi-Channel Cylinder Set is indicated for use for cancer treatment of the vaginal vaginal stump, cervix, uterus, endometrium and rectum using HDR or PDR brachytherapy.

    The Universal Cervix Probe Sets in combination with the Universal Multi-Channel Cylinder Set are indicated for use for cancer treatment of the vagina, cervix, uterus and endometrium using HDR or PDR brachytherapy.

    The Universal Titanium Cervix Probe Sets in combination with the Universal Multi-Channel Cylinder Set are indicated for use for cancer treatment of the vagina, cervix, uterus and endometrium using HDR or PDR brachytherapy.

    The Universal Interstitial Cylinder Set is indicated for use for cancer treatment of the vaginal vaginal stump, cervix, uterus and endometrium using intracavitary and interstitial HDR or PDR brachytherapy. The Universal Cervix Probe Sets in combination with the Universal Interstitial Cylinder Set are indicated for use for cancer treatment of the vagina, cervix, uterus and endometrium using HDR or PDR brachytherapy. The Universal Titanium Cervix Probe Sets in combination with the Universal Interstitial Cylinder Set are indicated for use for cancer treatment of the vagina, cervix, uterus and endometrium using HDR or PDR brachytherapy.

    Device Description

    The Universal Cylinder Applicator Family (UCAF) is designed to be used with Varian afterloaders (GammaMedplus, VariSource, and Bravos) to deliver high dose rate (HDR) and pulsed-dose-rate (PDR) intracavity and interstitial brachytherapy treatments for gynaecological and rectal applications.

    AI/ML Overview

    The provided text describes the Universal Cylinder Applicator Family (UCAF), a medical device used for brachytherapy. The document is a 510(k) premarket notification to the FDA, asserting substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (What the device must do)Reported Device Performance (How the device was shown to meet the criteria)
    Functional Performance (with specified afterloaders)Test results demonstrate conformance to applicable requirements and specifications, indicating the device functions correctly with compatible afterloaders (GammaMedplus, VariSource, Bravos).
    Durability (number of use cycles)Testing has been performed to demonstrate the device can withstand the number of use cycles it will experience in its specified lifetime.
    Radioactive Source Positional AccuracyTesting has been performed to demonstrate the device enables the radioactive source to be located to the required accuracy and that the positional accuracy of the source in the devices is adequate.
    Material Resistance to RadiationTesting has been performed to demonstrate the device is constructed of materials that are not significantly affected by the radiation to which it is exposed during its specified lifetime.
    Sterilization EffectivenessTesting has been performed to demonstrate the device may be sterilized effectively and that it can be used and sterilized for the specified cycle limits. Conforms to ISO 14161:2009, ANSI AAMI ST79:2017, ISO 17665-1:2006, ISO TS 17665-2:2009, ISO 17664:2017.
    BiocompatibilityThe biocompatibility evaluation was conducted in accordance with ISO 10993-1, including Chemical characterization, Cytotoxicity, Sensitization, Irritation (intracutaneous reactivity), and Material mediated pyrogenicity. Test results demonstrate conformance. The device is categorized as a surface device in contact with mucosal membrane and prolonged contact duration. Conforms to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-6:2016, ISO 10993-10:2010, ISO 10993-11:2017, ISO 10993-12:2012, ISO 10993-17:2009, ISO 10993-18:2005, ANSI AAMI BE83:2006.
    CT Compatibility and MRI SafetyCT compatibility was evaluated. MRI safety testing was conducted in accordance with FDA guidance and included Magnetically-induced displacement force, Magnetically-induced torque, RF-induced heating, and MRI image artifacts. Testing demonstrated the UMC and UIC sets are safe to use in MRI environments under specified conditions. Conforms to ASTM F2503-13, ASTM F2052-15, ASTM F2213-11, ASTM F2182-11a, ASTM F2119-07.
    Conformance to Quality Management, Risk Management, and Usability StandardsVerification and validation testing were conducted in accordance with FDA Quality System Regulation 21 CFR 820, ISO 13485 – Quality Management System standard, ISO 14971 – Risk Management standard, and IEC 62366-1:2015 for usability. Conforms to EN ISO 13485:2016, EN ISO 14971:2012, IEC 62366-1:2015.
    Electrical Safety (if applicable as part of the afterloader system interaction)Conforms to IEC 60601-1:2005, MOD and IEC 60601-2-17:2013. (Implied for the system with which it interacts.)
    Packaging IntegrityConforms to EN ISO 11607-1:2017, ISTA 3A:2008, ASTM D4332-13, ASTM D4169-09, ASTM D4728-06, ASTM D5276-98. (Packaging tests are listed as part of standards conformance, implying the device meets these for safe delivery and preservation.)
    No remaining critical discrepancy reports (DRs) classified as Safety or Customer Intolerable (Implied)"There were no remaining discrepancy reports (DRs) which could be classified as Safety or Customer Intolerable." (This is a final conclusion based on the testing performed, suggesting that all identified issues were resolved or deemed non-critical).

    2. Sample Size Used for the Test Set and the Data Provenance

    The submission explicitly states: "No animal studies or clinical tests have been included in this pre-market submission." Therefore, there is no human test set sample size and no data provenance in terms of country of origin or retrospective/prospective human data for this device. All testing was non-clinical bench testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Since no human studies or clinical trials were conducted, there was no "ground truth" established by human experts for a test set in the context of clinical performance. The ground truth for the engineering and safety performance was established by adherence to recognized national and international standards and internal verification and validation protocols by the manufacturer's qualified personnel.

    4. Adjudication Method for the Test Set

    Not applicable, as no human test set or clinical adjudication processes were involved. Performance was assessed via engineering and bench testing against predefined specifications and recognized standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly states: "No animal studies or clinical tests have been included in this pre-market submission." An MRMC study involves human readers, which was not performed. Therefore, there is no effect size given for human readers improving with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This device is not an AI algorithm. It is a physical medical device (applicators for brachytherapy). Therefore, the concept of a "standalone algorithm" performance is not applicable. The device's performance was evaluated as a physical entity through bench testing.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical bench testing, the "ground truth" was based on:

    • Engineering specifications and design requirements: The device had to perform within specified tolerances (e.g., source positional accuracy, durability).
    • Recognized national and international standards: Adherence to standards like ISO 10993 (biocompatibility), ASTM (MRI safety), ISO 13485 (quality management), etc., served as the benchmark or "ground truth" for demonstrating safety and effectiveness in specific aspects.
    • Manufacturer's internal verification and validation protocols.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Universal Segmented Cylinder Applicator Set and Universal Stump Applicator Set are intended for cancer treatment of the vagina, vaginal stump and rectum using HDR or PDR brachytherapy.

    The Universal Cervix Probe Sets in combination with the Universal Segmented Cylinder Applicator Set and the Universal Stump Applicator Set are intended for cancer treatment of the vaginal stump, cervix, uterus and endometrium using HDR or PDR brachytherapy.

    The Universal Titanium Cervix Probe Sets in combination with the Universal Segmented Cylinder Applicator Set and the Universal Stump Applicator Set are intended for cancer treatment of the vaginal stump, cervix, uterus and endometrium using HDR or PDR brachytherapy.

    Device Description

    The applicator sets in this submission are designed to be used with the GammaMed Plus afterloader series, VariSource 200 and iX afterloaders to deliver high dose rate (HDR) and pulsed-dose-rate (PDR) brachytherapy treatment for gynaecological and rectal applications. The intracavitary applicators will be used for the treatment of cancerous tumors, and are designed to be inserted into a body cavity.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the Universal Cylinder Applicator Family (K172611). This is a medical device, not a diagnostic AI/ML algorithm. Therefore, the information requested regarding acceptance criteria and studies that prove a device meets those criteria (especially sections related to AI/ML such as test set sample size, data provenance, ground truth, MRMC studies, standalone performance, and training set details) are not applicable to this document.

    Instead, this document describes a traditional medical device (brachytherapy applicators) and its substantial equivalence to a predicate device based on technological characteristics and functional testing.

    Here's an analysis based on the document's content, focusing on what is provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of formal acceptance criteria with numerical performance targets alongside reported device performance in the way one might expect for an AI/ML algorithm (e.g., sensitivity, specificity, AUC). Instead, it describes various bench tests conducted to demonstrate conformance to applicable requirements and specifications. The "reported device performance" is a general statement of compliance.

    Acceptance Criteria (Implied from Bench Tests)Reported Device Performance
    Correct functioning with specified afterloadersConformance to requirements
    Withstanding specified number of use cyclesConformance to requirements
    Accurate radioactive source localizationConformance to requirements
    Material resistance to radiation exposureConformance to requirements
    Effective sterilizationConformance to requirements
    Usability and sterilizability for cycle limitsConformance to requirements
    Adequate positional accuracy of the sourceConformance to requirements
    Biocompatibility (per ISO 10933)Conformance to requirements
    Safe and effective use in CT environmentsConformance to requirements

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is a hardware medical device; testing was primarily bench testing, not involving clinical data sets in the AI/ML sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth, in the context of expert consensus for AI/ML, is not relevant here. Device functionality was assessed against design specifications and relevant standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No expert adjudication method was described as this is not an AI/ML diagnostic device requiring clinical image interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an applicator for brachytherapy, not an AI system. No MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable for AI/ML ground truth. The "ground truth" for this device's performance is its ability to meet engineering specifications, functional requirements, material properties, and safety standards verified through bench testing.

    8. The sample size for the training set

    Not applicable. This is a hardware medical device; it does not utilize a training set in the AI/ML context.

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no AI/ML training set.


    Summary regarding the device and its studies:

    The document states:

    • Non-Clinical Tests: "Bench Testing has been performed to demonstrate that... [list of functional and safety attributes]."
    • Clinical Tests: "No clinical tests have been included in this pre-market submission."
    • Conclusions: "All the tests that were performed met the applied pass criteria. Varian considers the devices to be safe and effective and to perform as well or better than the predicate."

    This 510(k) submission relies on a comparison to a predicate device (K160516 - Universal Cylinder Applicator Family) and extensive bench testing to demonstrate substantial equivalence, rather than clinical studies or AI/ML performance evaluation.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Universal Segmented Cylinder Set and Universal Stump Applicator Set are indicated for use for cancer treatment of the vaginal stump and rectum using HDR or PDR brachytherapy.

    The Universal Cervix Probe Sets in combination with the Universal Segmented Cylinder Applicator Set and the Universal Stump Applicator Set are indicated for use for cancer treatment of the vaginal stump, cervix, uterus and endometrium using HDR or PDR brachytherapy.

    Device Description

    The applicator sets in this submission are designed to be used with the GammaMed Plus afterloader series, Varisource 200 and IX afterloaders to deliver high dose rate (HDR) and pulsed-dose-rate (PDR) brachytherapy treatment for gynecological and rectal applications. The intracavitary applicators will be used for the treatment of cancerous tumors, and are designed to be inserted into a body cavity.

    AI/ML Overview

    The provided document pertains to a 510(k) premarket notification for the "Universal Cylinder Applicator Family" by Varian Medical Systems. This family of devices is used for brachytherapy and its primary purpose is to deliver radioactive sources for cancer treatment.

    The document indicates that this is a 510(k) submission, which means it aims to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness through clinical trials. Therefore, the information provided focuses on engineering and bench testing rather than clinical study results demonstrating performance against acceptance criteria in a clinical context.

    Here's an analysis of the requested information based on the document's content:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a formal table of acceptance criteria with corresponding performance metrics in a typical clinical study format. Instead, it lists several non-clinical tests and states that the "Results of Bench Testing showed conformance to applicable requirements and specifications" and "All the tests that were performed met the applied pass criteria."

    Here's a summary of the stated performance based on the non-clinical tests:

    Acceptance Criteria (Implied from Test Purpose)Reported Device Performance
    Correct function with specified afterloaders.Devices function correctly with the specified afterloaders (GammaMed Plus, VariSource 200, IX).
    Withstand specified number of use cycles over lifetime.Devices can withstand the number of cycles of use they will experience in their lifetime.
    Accurate radioactive source location.Devices enable the radioactive source to be located to the accuracy required.
    Material integrity under radiation exposure over lifetime.Devices are constructed of materials that are not significantly affected by the radiation to which they are exposed in the lifetime of the product.
    Effective sterilization.Devices may be sterilized effectively.
    Maintain usage and sterilization integrity for specified number of times.Devices can be used and sterilized for the specified number of times.
    Adequate positional accuracy of the source within the devices.The positional accuracy of the source within the devices is adequate.
    Biocompatibility according to ISO 10933 standards.Devices are biocompatible as per ISO10933 standards.
    Safe and effective use in CT environments.Devices can be used safely and effectively in CT environments (CT compatible).
    Safe use in MRI environments under specified conditions.Testing in MRI environments has demonstrated they are safe to use under the conditions specified in the labeling (MR conditional).
    Conformance to usability requirements (IEC 62366:2007).Usability has been assessed to the requirements of IEC 62366:2007, and results showed conformance.

    2. Sample Size for the Test Set and Data Provenance

    The document describes non-clinical bench testing. Therefore, there isn't a "test set" in the sense of patient data. The "samples" would be a certain number of the Universal Cylinder Applicator Family devices that underwent various physical and functional tests. The document does not specify the exact number of devices tested for each non-clinical test.

    • Data Provenance: The tests are "Bench Testing," implying they were conducted in a laboratory or engineering environment. There is no mention of country of origin for this data or if it's retrospective or prospective, as these terms are generally applicable to human subject studies.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable as the document describes non-clinical bench testing of a physical medical device (an applicator for brachytherapy), not an AI or diagnostic device that requires expert-established ground truth from medical images or patient cases.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reason as point 3. Adjudication is relevant for studies involving human interpretation or clinical outcomes, not for engineering bench tests.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical tests have been included in this pre-market submission." This type of study would compare human readers' performance with and without an AI-assisted device, which is not relevant for this physical brachytherapy applicator.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    A standalone study was not done. This device is a physical applicator system used in conjunction with "afterloaders" and operated by medical professionals. It is not an algorithm or a device that functions independently without human intervention in a diagnostic or interpretive capacity.

    7. Type of Ground Truth Used

    The concept of "ground truth" as pathology, outcomes data, or expert consensus is not applicable in the context of this 510(k) submission. The "truth" for the bench tests would be defined by engineering specifications, material properties, and functional requirements (e.g., source position accuracy within manufacturing tolerances, sterilization efficacy according to validated protocols, biocompatibility meeting ISO standards).

    8. Sample Size for the Training Set

    This information is not applicable. There is no "training set" as this device is a physical medical device, not a machine learning or AI model that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as point 8. There is no training set for this type of device.

    In summary, the provided document is a 510(k) premarket notification for a physical medical device, not a diagnostic or AI-based system. Therefore, many of the requested items related to clinical studies, AI performance, and ground truth establishment in a diagnostic context are not present or applicable. The submission relies solely on non-clinical bench testing to demonstrate substantial equivalence to a predicate device.