Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173881

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-07-11

    (202 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unifine SafeControl™ range of pen needles are intended for use with multi-dose injection devices for the subcutaneous injection of FDA approved drugs, including insulin.

    Device Description

    The Unifine SafeControl™ safety pen needle is a sterile, single-use, disposable device intended for use with multi-dose Injector devices for the subcutaneous injection of FDA approved drugs, including insulin. The device is designed for prescription and over-the-counter use and to be used by self-administering patients, care-givers and healthcare professionals. The safety pen needle is a 30-gauge needle available in sizes between 5mm and 8mm lengths. The pen needle assembly consists of a cannula attached to a needle carrier assembled into a plastic moulded needle hub and safety guard, a primary container that houses the entire assembly and a sterility seal that covers the assembly inside the primary container. The entire device is packaged and labeled as a sterile single-use device.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Unifine SafeControl™ safety pen needle. It describes the device's characteristics, intended use, and performance data to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the provided text to extract the requested information regarding acceptance criteria and the study proving the device meets them:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally "Meets standard" for each test listed, implying compliance with the specified ISO or ASTM standards. The reported device performance uniformly states "Meets standard" for all listed tests.

    TestStandard/RequirementAcceptance Criteria (Implicit)Reported Device Performance
    Visual InspectionISO 11608-2:2012 Section 12.2.2Meets standardMeets standard
    Torque to attach to penISO 11608-2:2012 Section 11.4.1.5Meets standardMeets standard
    Torque to remove from penISO 11608-2:2012 Section 11.4.3Meets standardMeets standard
    Dose accuracyISO 11608-2:2012 Section 11.4.2Meets standardMeets standard
    Flow RateISO 11608-2:2012 Section 4.3Meets standardMeets standard
    Force to over-ride safety featureISO 11608-5:2012 Section 5.1.11.2Meets standardMeets standard
    Needle RetentionISO 11608-2:2012 Section 9Meets standardMeets standard
    Needle DislocationISO 11608-2:2012 Section 4.8Meets standardMeets standard
    Sterility - Seal integrityISO 11607-1:2007 Section 6.3Meets standardMeets standard
    Internal pressure test (Burst Testing)ASTM-F-1140-07:2007, ASTM 4169 2016 Section 16.2Meets standardMeets standard
    NIS CompatibilityISO 11608-2:2012 Section 4.9Meets standardMeets standard
    Simulated Clinical use testing (sharps safety testing)ISO 23908:2011 & Guidance for Industry and FDA Staff-Medical Devices with Sharps Injury Prevention FeaturesMeets standardMeets standard
    Biocompatibility (Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Hemocompatibility, Pyrogenicity)ISO 10993-1, ISO 10993-11 Annex FMaterials are biocompatibleMaterials are biocompatible
    Sterilization (Dose mapping)ISO 11137: "Sterilization of Healthcare Products - Radiation - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices."Achieves SAL 10-6Achieves SAL 10-6
    Shelf lifeISO 11608-2, ISO 23908, and ASTM F1980-162-year initial, 5-year upon studies completed2-year initial, studies for 5-year ongoing

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each of the bench tests. It mentions that "Written protocols specified the scope, objectives, specifications, test equipment, test methods and acceptance criteria," which would typically include sample sizes, but these details are not provided in the summary.

    The data provenance is not explicitly stated as retrospective or prospective clinical data. Given that it's a 510(k) summary focusing on bench testing, the data is primarily from non-clinical performance testing conducted by the manufacturer, Owen Mumford Ltd., which is based in the United Kingdom.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable to this submission. The "ground truth" for these tests relates to meeting objective engineering and material standards (e.g., specific force measurements, flow rates, sterility assurance levels), not a clinical diagnosis or interpretation by human experts.

    For biocompatibility, it states that "Full device evaluation includes a review of all the materials, consumables, manufacturing processes and environments by an expert toxicologist." The number of toxicologists and their specific qualifications are not provided, but the expert's role here is to review and ensure compliance with biocompatibility standards based on material properties and testing results, not to establish a "ground truth" in the diagnostic sense.

    4. Adjudication Method for the Test Set

    This is not applicable as the tests are objective bench tests based on measurable physical properties and adherence to established standards, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI/ML-driven diagnostic devices where human readers' performance is compared with and without AI assistance. The Unifine SafeControl™ is a physical medical device (pen needle) and its effectiveness is determined by its physical and material properties, not through AI-assisted human interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone algorithm-only performance study was not done. This concept is relevant for AI/ML-driven diagnostic software. The Unifine SafeControl™ is a physical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance evaluation is established through compliance with recognized industry standards and specifications. This includes:

    • Engineering specifications: For physical properties like torque, force, flow rate, needle retention, etc. (e.g., ISO 11608-2, ISO 11608-5).
    • Material and safety standards: For sterility assurance and biocompatibility (e.g., ISO 11607-1, ASTM F1980-16, ISO 10993 series, ISO 11137).
    • Sharps safety guidelines: For the safety feature (e.g., ISO 23908:2011 & FDA Guidance).

    Essentially, the ground truth is that the device performs as expected and safely according to these predefined, objective standards.

    8. The Sample Size for the Training Set

    There is no training set as this is a physical medical device (a pen needle) and not an AI/ML algorithm that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no training set mentioned, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1