Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151523

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-07-16

    (41 days)

    Product Code
    Regulation Number
    870.4210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass surgical procedures.

    Device Description

    Medtronic uncoated Tubing and Connector Components are intended for use in the extracorporeal circuit during cardiopulmonary bypass surgery. These components are primarily used to interconnect the primary devices of the bypass circuit. The lumen of the uncoated Tubing and Connector Components consist of smooth blood-contacting surfaces which provide a continuous pathway for the flow of blood and/or other fluids during cardiopulmonary bypass surgical procedures. The uncoated Tubing Components are comprised of polyvinyl chloride (PVC) and the uncoated Connector Components are comprised of polycarbonate and Plastisol. The uncoated Connectors are provided in Y-type, straight, and reducer configurations with ½ inch to ½ inch connection sites.

    AI/ML Overview

    This document is a 510(k) summary for the Medtronic Uncoated Tubing and Connector Components. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving efficacy or a specific level of performance through extensive clinical studies like those typically associated with AI/ML devices or novel therapies.

    Therefore, the information requested in your prompt regarding acceptance criteria, study details, expert involvement, and ground truth is generally not applicable in the context of this 510(k) submission for a non-active medical device like tubing and connectors. The approval is based on demonstrating that the new device is as safe and effective as a legally marketed predicate device, primarily through performance and biological testing.

    However, I can extract the relevant information that is present in the document.


    1. A table of acceptance criteria and the reported device performance

    The document states: "Performance and biological tests confirm that the uncoated Tubing and Connector Components met pre-determined acceptance criteria". However, the specific quantitative acceptance criteria and the detailed reported performance values for each test are not provided in this 510(k) summary. The summary only lists the types of performance tests conducted.

    Test TypeAcceptance Criteria (Not Detailed)Reported Performance (Not Detailed)
    Shelf lifeMet pre-determined criteriaMet pre-determined criteria
    Pressure IntegrityMet pre-determined criteriaMet pre-determined criteria
    Kink ResistanceMet pre-determined criteriaMet pre-determined criteria
    Pressure DecayMet pre-determined criteriaMet pre-determined criteria
    SpallationMet pre-determined criteriaMet pre-determined criteria
    Pull ForceMet pre-determined criteriaMet pre-determined criteria
    Method of sterilization and sterility assurance levelMet pre-determined criteriaMet pre-determined criteria
    Blood TraumaMet pre-determined criteriaMet pre-determined criteria
    BiocompatibilityMet pre-determined criteria (EN ISO 10993-1:2009)Met pre-determined criteria

    2. Sample sizes used for the test set and the data provenance

    The document does not specify sample sizes for any of the performance or biological tests. All testing is implied to be prospective and conducted by Medtronic, likely in a lab setting, given the nature of the device (tubing and connectors). There is no mention of country of origin for data as it's not a clinical data study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable. The device is an inert medical component. "Ground truth" in the context of clinical outcomes or expert consensus for AI/ML devices is not relevant here. Performance is evaluated against engineering specifications and biological safety standards.

    4. Adjudication method for the test set

    This is not applicable. There is no "adjudication" in the sense of resolving discrepancies in expert opinions for a diagnostic task. Performance is measured against physical and chemical standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is not an AI-powered diagnostic tool, nor does it have human "readers" in the context of image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    "Ground truth" for this type of device refers to objectively measurable physical, chemical, and biological properties. This includes:

    • Engineering specifications: For properties like pressure integrity, kink resistance, pull force, and pressure decay.
    • Material standards: For aspects like spallation.
    • Sterilization standards: For sterility assurance level.
    • Biocompatibility standards: Specifically EN ISO 10993-1:2009 for biological evaluation.
    • Chemical analysis/testing: For blood trauma (e.g., hemolysis).

    8. The sample size for the training set

    This is not applicable. This is not an AI/ML device that requires a "training set."

    9. How the ground truth for the training set was established

    This is not applicable. As above, there is no training set. The "ground truth" for the device's characteristics is established through recognized industry standards, test methods, and scientific principles applied during the performance and biological testing.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1