Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN150022

    Validate with FDA (Live)

    Date Cleared
    2016-12-20

    (574 days)

    Product Code
    Regulation Number
    878.5050
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultravision™ Visual Field Clearing System is indicated for the clearance of smoke and other particulate matter that is created during laparoscopic surgery.

    Device Description

    The Ultravision™ Visual Field Clearing System is a device that precipitates surgical smoke generated during laparoscopic procedures to clear the visual field. The device components are summarized briefly in Table 1, below.

    AI/ML Overview

    Acceptance Criteria and Device Performance for ULTRAVISION™ Visual Field Clearing System

    The ULTRAVISION™ Visual Field Clearing System is indicated for the clearance of smoke and other particulate matter created during laparoscopic surgery. The acceptance criteria address various aspects of the device's safety, performance, and usability.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of the device as a "surgical smoke precipitator," the acceptance criteria were primarily qualitative and focused on safety, usability, and the ability to maintain a clear visual field.

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceStudy Supporting Performance
    BiocompatibilityAbsence of cytotoxic responses from patient-contacting components.Grade 1 (mild reactivity) in cytotoxicity, Scores of 0 for sensitization, Pass for intracutaneous reactivity and acute systemic toxicity. Endotoxin levels within limits.Biocompatibility Testing (Table 2 in source)
    Chemical SafetyNo measurable chemical modifications or newly generated chemical species in treated surgical smoke.No measurable chemical modifications to surgical smoke observed.Risk assessment concluded no new patient risk beyond standard of care.Chemical Characterization of Treated Smoke/Risk Assessment
    Sterility & Shelf LifeSterility of patient-contacting components (Ionwand™, catheter, trocar).Sterility assurance level of 1x10^-6 achieved (ethylene oxide sterilization).Shelf Life/Sterility section
    Sterility & Shelf LifeEthylene oxide and ethylene chlorohydrin residuals below limits.Residuals below limits for limited exposure devices.Shelf Life/Sterility section
    Sterility & Shelf LifePackaging integrity and device functionality maintained over identified shelf life.Packaging and functional testing support a three-year shelf life.Shelf Life/Sterility section
    Electrical SafetyCompliance with IEC 60601-1 General requirements and specific tests.All applicable requirements for AAMI/ANSI ES60601-1 passed. All IEC 60601-1 tests passed.Electromagnetic Compatibility and Electrical Safety - IEC 60601-1 section
    Electrical SafetyCompatibility with third-party active electrodes.All IEC 60601-2-2 tests passed.Electromagnetic Compatibility and Electrical Safety - IEC 60601-2-2 section
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2 for electromagnetic disturbances.All tests passed for both generator unit and battery charging station. No modifications made to achieve compliance.Electromagnetic Compatibility and Electrical Safety - IEC 60601-1-2 section
    Software PerformanceSoftware functions according to specification for audible alerts and fault lights.Software functioned according to specification for alarms under fault and normal conditions. Hazards properly mitigated.SOFTWARE section
    Animal Study - Functionality & UsabilityFunctionality of Ionwand™ trocar/introducer assembly.Demonstrated functionality without usability concerns.Performance Testing - Animal
    Animal Study - Visual Field ClearanceAbility to maintain a smoke-free laparoscopic surgical field.Demonstrated ability to maintain a smoke-free field.Performance Testing - Animal
    Animal Study - SafetyAbsence of clinical chemistry or histology concerns (e.g., tissue damage).No identified clinical chemistry or histology concerns. Device demonstrated to be non-damaging.Performance Testing - Animal
    Clinical Study - Primary Endpoint: Visual FieldMaintenance of a clear visual field during surgery.Treatment group had a higher mean proportion of procedures with effective visibility than the control group.Summary of Clinical Information - Primary Endpoint
    Clinical Study - Secondary Endpoint: SafetyNo adverse events attributable to the device.No adverse events attributable to the device. No detectable difference in CO or MetHb levels.Summary of Clinical Information - Secondary Endpoints
    Clinical Study - Secondary Endpoint: Procedure InterruptionsReduction in interruptions due to visual field impairment.Treatment group had 8 procedures with no interruptions; control group had no cases without interruptions.Summary of Clinical Information - Secondary Endpoints
    Clinical Study - Secondary Endpoint: Camera CleaningReduction in frequency of camera cleaning.85% of active Ultravision™ procedures required no camera cleaning vs. 35% in control group.Summary of Clinical Information - Secondary Endpoints

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Study): 30 patients were enrolled. Follow-up data was available for 25/30 enrolled patients.
    • Data Provenance: The clinical study was a randomized, double-blinded, controlled, prospective trial conducted at a "single site." The location of this single site (e.g., country of origin) is not explicitly stated in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: The text refers to "The surgeons and reviewing panel" rating the treatment group's visibility. The specific number of surgeons or members of the reviewing panel is not provided.
    • Qualifications of Experts: The qualifications of the "surgeons" are implied by their role in performing and assessing laparoscopic cholecystectomy. The qualifications of the "reviewing panel" are not specified beyond their role in judging visual field effectiveness.

    4. Adjudication Method

    The text indicates that "The surgeons and reviewing panel rated the treatment group to have a higher mean proportion of procedures with effective visibility than the control group." This suggests some form of consensus or independent rating was used, but a specific adjudication method (e.g., 2+1, 3+1) is not explicitly detailed. The blinding of the surgeon to the device's activation status suggests an effort to reduce bias in assessment.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A formal MRMC comparative effectiveness study, as typically described in imaging, was not conducted. However, a controlled clinical trial was performed where surgeons served as "readers" of the visual field.

    • Effect size of human readers with AI vs. without AI assistance: The device is not an AI-assisted diagnostic tool but a physical device that clears the visual field. Therefore, the "effect size" isn't about human reader improvement with AI, but rather the improvement in the surgical environment (visual field clarity) provided by the device.
      • Visual Field Clarity: The treatment group (with device active) had a "higher mean proportion of procedures with effective visibility" compared to the control group (device deactivated).
      • Procedure Interruptions: In the treatment group, 8 procedures had no interruptions, while in the control group, no cases were completed without interruptions due to visual field impairment.
      • Camera Cleaning: 85% of procedures with the active device required no camera cleaning, compared to only 35% in the control group.

    These results indicate a substantial positive impact of the device on the surgeon's ability to maintain a clear visual field and proceed without interruptions.

    6. Standalone Performance Study

    Yes, a standalone performance (algorithm only without human-in-the-loop performance) was implicitly done for several aspects, though not for visual field assessment.

    • Biocompatibility Testing (Table 2): These are laboratory tests evaluating material properties directly, independent of human interaction.
    • Chemical Characterization: This study assessed the effect of the device on surgical smoke chemistry, independent of human observers.
    • Sterilization and Shelf Life Validation: These studies verify the effectiveness of the sterilization process and integrity of the packaging over time, independent of human application.
    • Electrical Safety and EMC Testing: These are objective engineering tests demonstrating compliance with safety standards, not requiring human interpretation of the device's primary function.
    • Software Verification and Validation: These analyze the software's functionality independently.
    • Animal Simulated-Use Testing: While involving an animal subject, the assessment of functionality, visual field clearance, and non-damaging effects were objectively observed and recorded, serving as a standalone performance validation in a controlled environment prior to human clinical use.

    7. Type of Ground Truth Used

    • Clinical Study: The primary endpoint, "maintenance of a clear visual field," was established by the subjective assessment and consensus of the surgeons performing the procedures and a "reviewing panel." Secondary endpoints like adverse events and physiological measurements (CO, MetHb) were objective data (lab results) but visual field was assessed by human observation.
    • Animal Study: The "acceptance criteria for the study were qualitative verification of performance and usability, through observed clearance of the visual field and an absence of observed clinical chemistry or histology concerns." This indicates a combination of expert observation/assessment (for visual field and usability) and objective laboratory results (clinical chemistry and histology).
    • Other Studies (Biocompatibility, Chemical Characterization, Electrical Safety, etc.): These rely on objective test results against established standards and scientific methodologies (e.g., ISO standards, spectrographic analysis, electrical measurements).

    8. Sample Size for the Training Set

    The provided document describes a medical device, not an AI/ML algorithm that requires a "training set" in the conventional sense for model development. Therefore, there is no explicit training set sample size mentioned for an AI/ML model for this device. The clinical trial and animal studies serve as validation, not training.

    9. How the Ground Truth for the Training Set Was Established

    As stated above, this device is not an AI/ML algorithm that uses a "training set." Therefore, this question is not applicable in the context of the provided information.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1