Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K212750

    Validate with FDA (Live)

    Date Cleared
    2021-09-27

    (28 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XD880B Ultrasonic Surgical Aspirator System is an ultrasonic surgical system consisting of handpieces and associated tips. The product is intended for the fragmentation, emulsification of both soft and hard (i.e. bone and bone approximations) tissue, the intended use is as follows with different configurations:

    • · Neurosurgery
    • · Gastrointestinal and Affiliated Organ Surgery
    • · Urological surgery
    • · Plastic and Reconstructive surgery
    • · General Surgery
    • · Orthopedic Surgery
    • · Gynecological Surgery
    • · Thoracic Surgery
    • · Wound Care
    • · Laparoscopic Surgery
    • · Thoracoscopic Surgery
    Device Description

    The Ultrasonic Surgical Aspirator System (Model: XD880B) is an ultrasonic powered surgical tool. It can be used for precise breakage, aspiration and debridement of soft tissues, and can also be used for precise cutting and shaping of bone tissue, using with the liguid-flow sleeve and liquid-flow tube. The product can be applied to a number of departments according to the suitable parameters and accessories, including:

    • . Department of Neurosurgery, general surgery and other departments for soft tissue breakage and aspiration.
    • Department of Orthopedic surgery, Plastic and Reconstructive surgery, Department of Neurosurgery (only for bone) for bone cutting.
    • . As well as the Department of Dermatology, Department of Trauma Orthopedics and other departments used for debridement.
      The Ultrasonic Surgical Aspirator System consists of a console, application software, foot switch, and accessories. The accessories include handpieces, tips, wrenches, liquid-flow sleeves, liquid-flow tubes, sterilization tray, suction canister and mobile cart. The handpiece, tip, liquid-flow sleeve and liquid-flow tube assemble to ultrasonic tool part to complete the cutting and fragment of soft tissue and bone tissue.
    AI/ML Overview

    This looks like a 510(k) summary for an Ultrasonic Surgical Aspirator System (Model: XD880B) seeking to demonstrate substantial equivalence to a predicate device (K202299).

    Here's an analysis based on your request, focusing on acceptance criteria and supporting studies. It's important to note that for a Class II device like this, the focus is often on demonstrating equivalence to an existing device rather than establishing novel clinical efficacy against specific, quantitative acceptance criteria in the same way a new drug or high-risk device might. The "acceptance criteria" here are implicitly related to demonstrating that the new device performs at least as well as the predicate device, especially for any design changes.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of quantitative acceptance criteria with corresponding performance metrics for the entire device. Instead, it describes changes from the predicate and then cites testing to show these changes do not negatively impact safety or effectiveness.

    However, based on the non-clinical performance data section, we can infer some "acceptance criteria" related to the modifications:

    Acceptance Criteria (Inferred)Reported Device Performance
    Soft Tissue Protection in Bone Cutting Mode: Demonstrated effective protection of soft tissue."Safety and Performance Testing for Soft Tissue Protection in Bone Cutting Mode." (Successful performance implied)
    Acoustic Performance for Modified Tips: Acoustic characteristics meet specifications."Acoustic Performance Testing for modified tips." (Performance meets specifications implied)
    Cutting Efficiency for Modified Tips: Cutting efficiency is maintained or improved."Cutting Efficiency and Thermal testing Report for Modified Tips." (Efficiency maintained/improved implied)
    Thermal Performance for Modified Tips: Thermal output remains within safe limits."Cutting Efficiency and Thermal testing Report for Modified Tips." (Safe thermal performance implied)
    Overall Substantial Equivalence for Modified Tips: Modified tips are substantially equivalent."Comparison of Substantial Equivalence for Modified Tips." (Substantial equivalence demonstrated implied)

    Note: The document states, "the test results confirm that the performance specifications meets the modification inputs." This is the general statement of "acceptance," indicating compliance with internal specifications derived from the predicate's performance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "non-clinical tests" and "bench testing." It does not specify sample sizes for these tests, nor does it provide details on data provenance like country of origin or whether a "retrospective" or "prospective" design applies to bench testing (these terms are more relevant for human data). These are laboratory-based tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is typically not relevant or included for non-clinical, bench testing of an ultrasonic surgical aspirator system. The "ground truth" for such tests would be measurable physical parameters (e.g., thermal output, acoustic frequency, cutting depth/speed) against established engineering specifications or comparisons to the predicate device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods (like 2+1 or 3+1 consensus) are typically used in clinical studies, especially those involving subjective interpretations (e.g., image reading, clinical outcomes). Since the testing described is non-clinical bench testing, this concept and information are not applicable or provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not mentioned as this device is not an AI-powered diagnostic imaging tool requiring human interpretation. It is a surgical tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of an algorithm or AI without human involvement. The Ultrasonic Surgical Aspirator System is a physical surgical device, not an AI algorithm. Therefore, "standalone performance" in this context is not applicable. The device's performance is inherently linked to its physical operation properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests mentioned:

    • Soft Tissue Protection in Bone Cutting Mode: The ground truth would likely be based on objective measurements of tissue damage or protection, adhering to predetermined safety thresholds.
    • Acoustic Performance Testing for modified tips: The ground truth would be quantitative measurements of acoustic parameters (e.g., frequency spectrum, amplitude) compared to design specifications or predicate measurements.
    • Cutting Efficiency and Thermal testing Report for Modified Tips: The ground truth would be objective measurements of cutting speed/volume and temperature generated during operation, compared against predicate performance or safety limits.

    The "ground truth" is thus based on objective physical measurements and engineering specifications, rather than subjective human interpretation, pathology, or clinical outcomes data for these non-clinical tests.

    8. The sample size for the training set

    This question is applicable to machine learning or AI models. Since the device is a physical surgical tool and not an AI or machine learning system, there is no "training set."

    9. How the ground truth for the training set was established

    As there is no training set for an AI/ML model, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202299

    Validate with FDA (Live)

    Date Cleared
    2021-05-21

    (281 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The XD880B Ultrasonic Surgical Aspirator System is an ultrasonic surgical system consisting of handpieces and associated tips. The product is intended for the fragmentation, emulsification of both soft and hard (i.e. bone and bone approximations) tissue, the intended use is as follows with different configurations:

    • · Neurosurgery
    • · Gastrointestinal and Affiliated Organ Surgery
    • · Urological surgery
    • · Plastic and Reconstructive surgery
    • · General Surgery
    • · Orthopedic Surgery
    • · Gynecological Surgery
    • · Thoracic Surgery
    • · Wound Care
    • · Laparoscopic Surgery
    • · Thoracoscopic Surgery
    Device Description

    The XD880B Ultrasonic Surgical Aspirator System is an ultrasonic powered surgical tool. It can be used for precise breakage, aspiration and debridement of soft tissue, and can also be used for precise cutting, crushing and shaping of bone tissue. The product can be applied to a number of departments by using the suitable parameters and accessories, including:

    • . Department of Neurosurgery, general surgery and other departments for soft tissue cutting, breakage and aspiration.
    • . Department of Orthopedic surgery, Plastic and Reconstructive surgery, Department of Neurosurgery (only for bone) for bone cutting.
    • . As well as the Department of Dermatology, Department of Trauma Orthopedics and other departments used for debridement.
      The XD880B Ultrasonic Surgical Aspirator System consists of a console, foot switch, and accessories. The accessories include handpiece, tips, wrench, liquid-flow tube, sterilization tray, suction bag, suction canister and mobile cart. The handpiece, tips, liquid-flow tube are assembled into ultrasonic tool part to complete the cutting and fragment of bone tissue and soft tissues.
    AI/ML Overview

    This document describes the FDA 510(k) clearance for the Ultrasonic Surgical Aspirator System (XD880B), focusing on its substantial equivalence to predicate devices rather than a direct clinical performance study with acceptance criteria.

    Therefore, the specific information requested, such as a table of acceptance criteria, sample sizes for test/training sets, expert qualifications, ground truth establishment, MRMC studies, and standalone algorithm performance, is not present or applicable in the provided document.

    The document details the device's technical specifications and non-clinical performance data to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's an analysis of the provided information in the context of your request:

    1. A table of acceptance criteria and the reported device performance:

    • Not applicable in the context of this document. This document focuses on demonstrating substantial equivalence based on technical characteristics and non-clinical testing, not on achieving specific clinical performance metrics against predefined acceptance criteria from a clinical trial.
    • The "performance" described is in terms of passing various non-clinical standard tests (e.g., electrical safety, EMC, software verification, sterilization validation, shelf-life, lifetime validation, acoustic performance, efficiency and temperature characteristics).

    2. Sample sizes used for the test set and the data provenance:

    • Not applicable. This submission relies on non-clinical testing and comparison to predicate devices, not on a clinical test set in the way a diagnostic AI would.
    • The document implies that various units of the device and its components were tested to meet the specified non-clinical standards, but explicit "sample sizes" for a clinical test set are not mentioned as no clinical performance study was conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No clinical ground truth establishment involving human experts is described. The "truth" for this submission is based on engineering standards and technical specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No clinical test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. A MRMC study was not conducted as this is an ultrasonic surgical aspirator system, not a diagnostic AI system designed to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This device is a surgical tool, not a standalone diagnostic algorithm. Its function inherently involves human operation.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • Technical Specifications and Standard Compliance: The "ground truth" for this submission is adherence to engineering specifications and compliance with international and national standards for medical devices (e.g., IEC 60601-1 for electrical safety, IEC 60601-1-2 for EMC, IEC 62304 for software, ISO 17665-1 and ISO 11135 for sterilization). The device's performance is measured against these established technical benchmarks.

    8. The sample size for the training set:

    • Not applicable. This is a hardware medical device, not an AI/ML model that requires a "training set" in the computational sense.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.

    Summary of Device Acceptance (from the document's perspective):

    The acceptance mechanism for this device's FDA clearance is "Substantial Equivalence" to legally marketed predicate devices. This is achieved by demonstrating that the new device:

    • Has the same intended use.
    • Has the same or similar indications for use.
    • Has the same or similar technological characteristics.
    • Does not raise new or different questions of safety or effectiveness.

    Studies/Data that Proved Device Meets "Acceptance Criteria" (i.e., Substantial Equivalence):

    The document states that the following non-clinical tests were conducted and passed to support substantial equivalence:

    • Electrical safety testing per IEC 60601-1
    • EMC testing per IEC 60601-1-2
    • Software verification and validation per IEC 62304 / FDA Guidance
    • Moist heat sterilization validation per ISO 17665-1
    • EO sterilization validation per ISO 11135 (demonstrates SAL of 10^-6)
    • Shelf-life Testing
    • Lifetime validation test for handpieces
    • Acoustic Performance Test per IEC 61847
    • Efficiency and Temperature Characteristics Test

    These tests confirm that the device's design inputs and performance specifications are met, allowing the FDA to determine that it is as safe and effective as the predicate devices, thereby meeting the "acceptance criteria" for 510(k) clearance by way of substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062471

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-10-26

    (63 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Misonix Inc. FS 1000 RF Ultrasonic Aspirator System is indicated for use in the fragmentation, emulsification and aspiration of both soft and hard (i.e. bone) tissue in the following surgical specialties:
    Neurosurgery Gastrointestinal and Affiliated Organ Surgery Urological Surgery Plastic and Reconstructive Surgery General Surgery Orthopedic Surgery Gynecological Surgery Thoracic Surgery Laparoscopic Surgery Thoracoscopic Surgery
    The system may also be combined with electrosurgery using optional RF Surgery interface components.

    Device Description

    The FS 1000 RF Ultrasonic Surgical Aspirator System is comprised of a generator, which feeds a 23 kHz electrical signal to a piezoelectric crystals mounted in a hand-held handpiece; the crystals then vibrate at the same frequency. The titanium tip attached to the handpiece amplifies the vibration. An irrigation unit is provided to introduce irrigation solution to the operative site. An aspirator system removes fragmented material and waste liquids from the area. Accessories include various probe tips. wrenches, tube sets and cleaning brushes.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (Misonix Inc. FS 1000 RF Ultrasonic Surgical Aspirator System). This type of submission focuses on demonstrating substantial equivalence to a predicate device, primarily through non-clinical testing and comparison of design, materials, and operating parameters. It does not present a study with acceptance criteria and device performance in the way that would be expected for a diagnostic AI or machine learning device.

    Therefore, I cannot provide the requested information for an "AI or machine learning device" as the document describes a traditional surgical aspirator system. The questions about AI, reader studies, ground truth establishment, training sets, and sample sizes for diagnostic performance are not applicable to the content provided.

    However, I can extract the information related to the device's performance based on the provided document, interpreting "acceptance criteria" as the voluntary standards it was tested against and "reported device performance" as the assertion of substantial equivalence based on meeting those standards and direct comparison to predicate devices.


    Non-AI/ML Device Acceptance Criteria and Performance (Based on K062471)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Voluntary Standards)Reported Device Performance
    UL 2601-1 Medical Electrical Equipment, Part 1: General Requirements for SafetyDesigned and tested to pass this standard.
    EN 60601-1 Medical Electrical Equipment, Part 1: General Requirements for SafetyDesigned and tested to pass this standard.
    EN 60601-2-2 Medical Electrical Equipment, Part 2: Particular Requirements for Safety of High Frequency Surgical EquipmentDesigned and tested to pass this standard.
    EN 60601-1-2:2001 Electromagnetic CompatibilityDesigned and tested to pass this standard.
    FCC Part 18 EMC Requirement (Electromagnetic Compatibility)Designed and tested to pass this standard.
    Other Non-Clinical Tests (to demonstrate Substantial Equivalence)Performed and deemed acceptable to support substantial equivalence.
    Output Frequency MeasurementsPerformed.
    Output Power Measurements (No Load to Maximum Load)Performed.
    Tip Displacement MeasurementsPerformed.
    Irrigation Flowrate Measurements (Ultrasound On and Flush Mode)Performed.
    Life TestsPerformed.
    Acoustic Output TestPerformed.
    Vacuum Flowrate and Pressure MeasurementsPerformed.
    Input Power MeasurementsPerformed.
    EMI TestsPerformed.
    Dielectric Tests on Mains CircuitsPerformed.
    Patient Current Leakage and Patient Sink Current MeasurementsPerformed.
    Power Line Ground Leakage MeasurementsPerformed.
    Dielectric Tests on Patient CircuitsPerformed.
    RF Cautery Life TestsPerformed.
    Dielectric Tests with RF Cautery Unit AttachedPerformed.
    RF Cautery Unit Output Power TestsPerformed.
    Software ValidationSoftware validations reviewed and cleared under 510(k) K032690; not affected by change of indication for use.
    Sterilization ValidationsValidation statements contained in Exhibit J.

    Since this is a filing for a traditional surgical device and not an AI/ML diagnostic, the following questions are not directly applicable to the content provided in the 510(k) summary. I will state why for each:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable to this document. This document describes non-clinical engineering and performance testing against standards, and a comparison of physical device characteristics to predicate devices. There is no "test set" of patient data in the context of diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable to this document. There is no "test set" requiring expert ground truth in the context of diagnostic assessment. The "ground truth" for non-clinical engineering tests would be the established scientific and engineering principles and the specifications of the device/standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable to this document. No diagnostic test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable to this document. This is a surgical device, not a diagnostic imaging AI. No MRMC study or AI assistance is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable to this document. This is not an algorithm-only device. It is a manually operated surgical tool.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable to this document in the diagnostic sense. For the engineering tests, the ground truth would be the validated measurements, specifications, and performance against established engineering standards. For substantial equivalence, the "ground truth" is the performance and safety profile of the predicate devices.

    8. The sample size for the training set

    • Not applicable to this document. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not applicable to this document. There is no "training set" as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1