Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162424

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-12

    (43 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The System is indicated for fixation of fractures and non-unions of the ulna, particularly in osteopenic bone.

    Device Description

    The Ulna Plating System is comprised of two new low profile plates anatomically contoured for the ulna, existing screws and both system specific and general instrumentation. The plates, locking screws and non-locking screws are manufactured from titanium alloy Ti6Al4V ELI per ASTM F136. All the plates are Type II anodized. The plates incorporate both locking and nonlocking holes that allow the surgeon to stabilize the fracture by the use of compression plating techniques and then obtain neutralization of the construct with the locking screws. The locking screws construct can protect the compression screws while providing relative stability across extensive comminution of in osteopenic bone. The plate material and the inclusion of both locking and non-locking holes are design features also included in the predicate device cleared in K083843. The plates are designed to accept existing 2.7mm locking screws, 2.7mm non-locking screws, and 2.7mm multidirectional screws. The non-locking and locking screws are manufactured from Titanium Alloy per ASTM F-136 and are color anodized. The multidirectional screws are manufactured from Cobalt Chromium Alloy per ASTM 1537. The predicate also includes Cobalt Chromium (ASTM F1537) multidirectional screws. The Ulna Plating System will offer the surgeon several system specific instruments as well as various general instruments to facilitate the installation of the implants.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Ulna Plating System." It describes the device, its intended use, and argues for its substantial equivalence to previously cleared predicate devices.

    Here's an analysis of the acceptance criteria and study information provided, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    Mechanical Performance (Construct Testing per ASTM F382)Results indicate substantial equivalence to legally marketed devices.
    Mechanical Performance (4-point bend testing per ASTM F382)Results indicate substantial equivalence to legally marketed devices.
    Biocompatibility (Galvanic Corrosion engineering evaluation)Found to meet the acceptance criteria.
    Substantial Equivalence (Overall Safety & Effectiveness)The Ulna Plating System has shown to be substantially equivalent to the predicate devices. No new issues of safety and effectiveness have been raised.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical tests are provided for basis of substantial equivalence." This means there was no test set in the context of clinical data. The performance was assessed through non-clinical (mechanical and engineering) tests.

    • Test Set Sample Size: Not applicable for clinical data. For non-clinical tests, specific sample sizes (e.g., number of plates, screws, test specimens) are not detailed in this summary.
    • Data Provenance: Not applicable for clinical data. For non-clinical testing, the data would have originated from laboratory testing conducted by the manufacturer or a contracted lab. The document does not specify the country of origin for this testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. As no clinical tests were performed, there was no clinical "ground truth" to be established by experts.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a bone fixation system, not an AI-powered diagnostic tool. Therefore, an MRMC study related to readers and AI assistance would not be relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical implant, not an algorithm or AI.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance data, the "ground truth" was established by engineering standards and specifications (e.g., ASTM F382 for mechanical properties, and internal criteria for galvanic corrosion). The comparison was primarily against the performance of legally marketed predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. There is no mention of a "training set" as this is not an AI/ML device requiring such. The design of the device would have been informed by engineering principles, predicate device designs, and potentially cadaver studies or biomechanical modeling, but not a "training set" in the context of AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set in the context of AI/ML.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The study proving the device meets acceptance criteria was a non-clinical performance evaluation. This included:

    • Construct testing per ASTM F382: This standard specifies requirements for metallic bone plates and tests their mechanical integrity. The results demonstrated substantial equivalence to predicate devices.
    • 4-point bend testing per ASTM F382: This is another mechanical test under the ASTM F382 standard, specifically assessing the bending strength and stiffness of the plates. Results also indicated substantial equivalence to predicate devices.
    • Galvanic Corrosion engineering evaluation: This evaluated the potential for corrosion when different metals are in contact, a crucial safety aspect for implants. The evaluation found that the device met acceptance criteria.

    The conclusion drawn from these non-clinical tests was that the Ulna Plating System is substantially equivalent to legally marketed predicate devices (specifically K083843 - Locking Anatomic and Composite Plate System, and K112345 - Distal Volar Radius Plating System) in terms of its intended use, indications for use, materials, design features, and sterilization method. No clinical tests were performed, nor were they deemed necessary by the FDA for this 510(k) submission based on the non-clinical data and comparison to predicates.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1