510(k) Data Aggregation

    K Number
    K152175

    Validate with FDA (Live)

    Date Cleared
    2015-12-04

    (122 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    "Dental Zirconia Blank for Aesthetic Restoration" is mainly used in prosthetic treatment. When the flexural strength of Dental Zirconia Blank for Aesthetic Restoration after sintering is between 300-500 MPa, the product can be used for veneering, inlay, single crown, and substructure ceramic for three-unit prostheses not involving molar restoration; when the flexural strength of Dental Zirconia Blank for Aesthetic Restoration after sintering is over 500 MPa, it can be used for veneering, inlay, single crown, and substrate ceramic for three-unit prostheses. All blanks are processed through dental laboratories or by dental professionals.

    Device Description

    "Dental Zirconia Blank for Aesthetic Restoration" is derived from zirconia powder that has been processed through various molding and sintering techniques – into their final net shapes. These blanks are then further fabricated into various prosthetic dental devices intended for use in the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers. The zirconia powder is composed of ZrO2 + Y2 O3 + HfO2+ Fe2O3+ Er2O3+ Al2O3. The performance of formed zirconia dental blanks conforms to ISO 6872, Dentistry, Ceramic Materials.

    "Dental Zirconia Blank for Aesthetic Restoration" is ceramic dental blanks designed for the manufacture of ceramic dental prosthetic devices. The dental prosthetic devices are fabricated by CAD/CAM machining processes. All prosthetic dental devices are intended for single use applications. At the dental lab, the blanks are held to the CAD/CAM machine which is used to machine to the final dental restoration. At the completion of the machining steps, the dental restoration is fired (i.e., sintered) in the oven to harden the ZrO2 so that its final properties can be achieved.

    "Dental Zirconia Blank for Aesthetic Restoration" is supplied in different shapes, such as blocks, discs, rods, or customer ordered shapes. It is also supplied in the combinations of forty-six different colors, three different translucencies, and two different aesthetic effects (single and multilayer).

    The different colors are originated from the different constituent of color additives (such as Fe203, Er202); the different translucencies are originated from small difference in the amount of Y2 O3, and the different aesthetic effects are originated from the different padding method used in the process of dry pressing.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the "Dental Zirconia Blank for Aesthetic Restoration":

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than defining novel acceptance criteria for a new type of device. The acceptance criteria are largely based on adherence to existing ISO standards for dental ceramic materials and biocompatibility.

    Acceptance Criteria CategorySpecific CriterionReported Device Performance
    Mechanical PropertiesConformance to ISO 6872:2008 (Dentistry, Ceramic Materials)"All tests were verified to meet acceptance criteria."
    "Radioactivity" comparability"very similar to the predicate device."
    "Pre-sintered density" comparability"very similar to the predicate device."
    "Sintered density" comparability"very similar to the predicate device."
    "Flexural strength" comparability"very similar to the predicate device."
    BiocompatibilityConformance to ISO 10993 (various parts)"Results met all relevant requirements in the test standards, and are comparable to the predicate device."
    Cytotoxicity (ISO 10993-5:2009)No cytotoxicity effect"No cytotoxicity effect"
    Irritation Oral Mucosa (ISO 10993-10:2010)Not a primary oral mucosa irritant"Not a primary oral mucosa irritant under the conditions of the study"
    Sensitization (ISO 10993-10:2010)Not a sensitizer"Not a sensitizer under the conditions of the study"
    Subacute and Subchronic Toxicity (ISO 10993-11:2006)No subchronic toxic effects observed"No subchronic toxic effects observed"
    Genotoxicity (ISO 10993-3:2003)No genotoxic effects observed"No genotoxic effects observed"
    Overall PerformanceMeets requirements per ISO 6872 and ISO 10993-1"Its performance meets the requirements of its pre-defined acceptance criteria and intended uses."
    Comparable to predicate device"The test results are also comparable to the predicate device."

    2. Sample Sizes used for the test set and the data provenance:

    • Sample Size for Test Set: The document does not explicitly state the specific sample sizes used for each individual test (e.g., how many specimens for flexural strength, how many animals for biocompatibility tests). It refers to "Bench testing" performed "per ISO 6872:2008 and internal procedures" and "Biocompatibility testing" performed "following the recommended procedures outlined in the FDA recognized consensus standard of ISO 10993." These ISO standards typically specify minimum sample sizes for material testing.
    • Data Provenance: The studies were conducted by Liaoning Upcera Co., Ltd. for their "Dental Zirconia Blank for Aesthetic Restoration." The country of origin for the studies is implicitly China, as the submitter is based in Benxi, Liaoning, China. The studies are prospective in the sense that they were performed on the subject device to demonstrate compliance with standards and equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable and not provided in the document. The tests performed are objective, laboratory-based material characterization and biocompatibility assessments, not subjective evaluations requiring expert consensus for ground truth. The "ground truth" here is adherence to established international standards (ISO).

    4. Adjudication method for the test set:

    Not applicable. The tests are scientific measurements and biological evaluations, not requiring human adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a material science and biocompatibility study for a dental restorative material, not an AI-assisted diagnostic or imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a material science and biocompatibility study, not an algorithm.

    7. The type of ground truth used:

    The "ground truth" used for this device's acceptance is based on:

    • International Standards: Primarily ISO 6872:2008 for ceramic materials and ISO 10993 series for biological evaluation of medical devices.
    • Predicate Device Performance: The device's performance is explicitly compared against a legally marketed predicate device (K093560, "Upcera Zirconia Blanks") to establish substantial equivalence.

    8. The sample size for the training set:

    Not applicable. There is no machine learning or AI component described in this device submission, so the concept of a "training set" is irrelevant.

    9. How the ground truth for the training set was established:

    Not applicable, as no training set is involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141724

    Validate with FDA (Live)

    Date Cleared
    2014-10-22

    (118 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed thought dental laboratories or by dental professionals.

    Device Description

    Upcera Dental Zirconia Blanks are derived from Zirconia powder that has been processed into their final net shapes. These blanks are then being further fabricated into all-ceramic restorations such as crowns, bridges, veneers, inlay/ onlay. The zirconia powder is composed of ZrO2 + Y2O3 + HfO2 + Al2O3 with its composition conforms to ISO 13356, Implants for Surgery -- Ceramic Materials Based on Yttria-Stabilized Tetragonal Zirconia (Y-TZP). The performance of the dental blanks conforms to ISO 6872, Dentistry: Ceramic Materials.

    Upcera Dental Zirconia Pre-Shaded Blanks are derived from the same Zirconia powder as the regular Upcera Dental Zirconia Blanks, with the addition of very small amount of inorganic pigments, before the composite material is processed into their final net shapes. These blanks are then being further fabricated into all-ceramic restorations such as crowns, bridges, veneers, inlay/ onlay. The purpose of the inorganic pigments is to generate the color on the prosthetic dental devices, after sintering at dental labs, that matches natural color of patient's teeth. The performance of the pre-shaded dental zirconia blanks conforms to ISO 6872: 2008, Dentistry: Ceramic Materials.

    AI/ML Overview

    The document describes the submission of a 510(k) premarket notification for "Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank." The primary goal of the submission is to demonstrate substantial equivalence to a legally marketed predicate device (K093560, "Upcera Zirconia Blanks"). The acceptance criteria and the study used to demonstrate that the device meets these criteria are primarily based on international standards for dental materials.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from the specified ISO standards, and the reported performance indicates compliance with these standards.

    Test / Performance MetricAcceptance Criteria (from Standards)Reported Device Performance (Upcera Dental Zirconia Blank & Pre-Shaded Blank)
    Mechanical Properties
    Flexural StrengthConforms to ISO 6872:2008, Dentistry: Ceramic Materials (Specific values not provided in document, but compliance stated)"Very similar to the predicate device," met all relevant requirements in test standard.
    RadioactivityConforms to ISO 6872:2008, Dentistry: Ceramic Materials (Specific limits not provided, but compliance stated)"Very similar to the predicate device," met all relevant requirements in test standard.
    Pre-sintered DensityConforms to ISO 6872:2008, Dentistry: Ceramic Materials (Specific values not provided, but compliance stated)"Very similar to the predicate device," met all relevant requirements in test standard.
    Sintered DensityConforms to ISO 6872:2008, Dentistry: Ceramic Materials (Specific values not provided, but compliance stated)"Very similar to the predicate device," met all relevant requirements in test standard.
    Biocompatibility
    CytotoxicityConforms to ISO 10993-5:2009 (No cytotoxicity effect)No cytotoxicity effect
    Irritation (Oral Mucosa)Conforms to ISO 10993-10:2010 (Not a primary oral mucosa irritant)Not a primary oral mucosa irritant under the conditions of the study
    SensitizationConforms to ISO 10993-10:2010 (Not a sensitizer)Not a sensitizer under the conditions of the study
    Subacute/Subchronic ToxicityConforms to ISO 10993-11:2006 (No subacute and subchronic toxic effects)No subacute and subchronic toxic effects observed
    GenotoxicityConforms to ISO 10993-3:2003 (No genotoxic effects)No genotoxic effects observed
    CompositionConforms to ISO 13356, Implants for Surgery -- Ceramic Materials Based on Yttria-Stabilized Tetragonal Zirconia (Y-TZP) (Specific composition requirements for ZrO2, Y2O3, HfO2, Al2O3)Regular: ZrO2 + Y2O3 + HfO2 + Al2O3 ≥ 99.0%Pre-Shaded: ZrO2 + Y2O3 + HfO2 + Al2O3 ≥ 98.0% + Inorganic pigments (Fe2O3, Pr2O3, and Er2O3, <2.0%)

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: Not explicitly stated in the provided text. The document mentions "Bench testing was performed per ISO 6872:2008 and internal procedures" and "Biocompatibility testing was performed to verify the equivalent safety of the materials that are used," as well as "Tests were conducted following the recommended procedures outlined in the FDA recognized consensus standard of ISO 10993." These standards typically specify sample sizes for their respective tests, but the specific number of samples tested for this submission is not detailed.
    • Data Provenance: The studies were conducted by the manufacturer, Liaoning Upcera Company Limited, based in China (Liaoning Upcera Co., Ltd, No.122 Xianghuai Road, Economic Development Zone, Benxi, Liaoning, China). The data appears to be prospective as it was generated specifically for this 510(k) submission to demonstrate the device's properties.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable and not provided. This submission concerns material properties and biocompatibility as per international standards, not the interpretation of medical images or diagnostic outputs where expert consensus would typically be required for ground truth. The "ground truth" here is defined by meeting predefined technical specifications and safety profiles established by ISO standards.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. As explained above, the "ground truth" is based on objective measurements against ISO standards. There is no mention of human-based adjudication for these types of material tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not provided. An MRMC study is relevant for diagnostic imaging devices where human readers interpret results, often with AI assistance. This submission is for a dental material (zirconia blanks) and does not involve AI or human interpretation in that context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable and not provided. This device is a material, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for both mechanical properties and biocompatibility is defined by international consensus standards:

    • For mechanical properties (flexural strength, radioactivity, density), the ground truth is established by the requirements and test methods outlined in ISO 6872:2008, Dentistry: Ceramic Materials.
    • For biocompatibility, the ground truth is established by the requirements and test methods outlined in ISO 10993 series (specifically ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-11:2006, ISO 10993-3:2003).
    • For material composition, the ground truth is established by ISO 13356, Implants for Surgery -- Ceramic Materials Based on Yttria-Stabilized Tetragonal Zirconia (Y-TZP).

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. "Training set" is a concept typically associated with machine learning or AI models. This device is a physical material, and its performance is evaluated through direct testing against established material standards, not through training a model.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and not provided for the same reasons as #8.